Nickel-titanium (NiTi) wires produced by various manufacturers may have different corrosion resistance in acidic oral environments. The purpose of this study was to investigate the variation during in vitro corrosion resistance of commercial NiTi dental orthodontic wires from different manufacturers using the fast electrochemical technique. The linear polarization test was used to evaluate the corrosion resistance, in terms of polarization resistance (Rp), of as-received commercial NiTi wires in acidic artificial saliva at 37°C. One-way analysis of variance was used to analyze Rp with the wire manufacturer as the variable factor. Atomic force microscopy was used to analyze the three-dimensional surface topography and roughness (Ra). Electron spectroscopy for chemical analysis was used to identify the chemical structure of the passive film on the NiTi wires. The results showed that NiTi wires from different manufacturers had a statistically significant difference in Rp (P < .001). Different surface topography was present among the tested NiTi wires, whereas the same surface chemical structure was observed for the tested NiTi wires. The surface roughness of the commercial NiTi wires with similar surface chemical structure does not correspond with the difference in corrosion resistance.

Nickel-titanium (NiTi) alloy is a prospective material for dental orthodontic wires because of its favorable mechanical properties and corrosion resistance. Although a protective passive film exists on the NiTi alloy, Ni or Ti ions may still be released from the metal surface in the acidic oral environment through the corrosion processes.1 

The potential danger associated with corrosion in the use of NiTi wire comes from the biologically negative effects of the Ni ion.2–5 NiTi wire with good corrosion resistance is crucial to dental prosthesis biocompatibility.

Many studies on NiTi and stainless steel wire corrosion resistance for dental orthodontic applications have been conducted.5–10 NiTi wires produced by various manufacturers may have different corrosion resistances in the acidic oral environment. However, related information in the literature concerning the comprehensive difference in corrosion resistance among commercial NiTi wires from various manufacturers is still limited.

In this investigation, the dissimilarity of in vitro corrosion resistance of commercial NiTi wires, produced by different manufacturers, in acidic artificial saliva was studied. The surface analyses results were compared and correlated with the corrosion resistance.

Four different commercial NiTi dental orthodontic wires (as-received condition) with 0.016 inch diameters were used. The studied wires were designated as RMO (Orthonol® Nickel-Titanium Wire, RMO Inc, Denver, Colo), Ormco (NI-TI®, Ormco, Glendora, Calif), SY (NiTi Wire, Shin-Ya Co, Taipei, Taiwan), and KH (Orthodontics-NiTi Wire, Kuo-Hua Co, Taipei, Taiwan). Chemical compositions (in wt %) of the as-received NiTi wires were analyzed using an energy-dispersive spectrometer (LINK exLII, Oxford, UK) and shown in the following: 57% Ni–43% Ti for RMO; 57% Ni–43% Ti for Ormco; 57% Ni–42% Ti–1% Fe for SY; 56% Ni– 44% Ti for KH.

A scanning electron microscope (SEM) (ABT-150S, Topcon, Tokyo, Japan) was used to observe the surface morphologies of the NiTi wires. Three-dimensional surface roughness (Ra) of the NiTi wires was evaluated using atomic force microscopy (AFM) (Nanoscope III, Digital Instruments Inc, Santa Barbara, Calif). The outermost surface chemical analyses of the passive film on NiTi wires were assessed from electron spectroscopy for chemical analysis (ESCA) (ESCALAB 210, VG Scientific Ltd, East Grinstead, UK).

A potentiostat (AUTOLAB PGSTAT 30, Eco Chemie BV, Utrecht, The Netherlands) was used to perform the linear polarization test. NiTi wires were used as working electrodes. A saturated calomel electrode and platinum sheet were used as the reference electrode and counter electrode, respectively. Modified Fusayama artificial saliva1,11 was used as the corrosion test electrolyte, which consisted of NaCl (400 mg/L), KCl (400 mg/L), CaCl2·2H2O (795 mg/L), NaH2PO4·H2O (690 mg/L), KSCN (300 mg/L), Na2S·9H2O (five mg/ L), and urea (1000 mg/L). The electrolyte was adjusted to a pH of 6.25 using sodium hydroxide and maintained at 37°C. The artificial saliva used in this study has good correlation in corrosion trends with the results obtained from natural saliva12 though it lacks some biological species. The electrolyte was deaerated with argon gas for one hour before the specimen was dipped into the electrolyte for the following corrosion test. The linear polarization curves of the test specimens were measured from −10 to +10 mV (vs corrosion potential) with a scan rate of 0.1 mV/second after dipping the specimen into the test electrolyte for two hours. The polarization resistance (Rp), which is inversely proportional to the corrosion rate, is defined as the slope of the potential vs the current density near corrosion potential in the linear polarization curves.13 The Rp value was statistically analyzed using one-way analysis of variance (ANOVA) for analyzing the wire manufacturer factor (α = 0.05). The sample size for the corrosion test of each NiTi wire was 10.

Figure 1 shows the SEM observations of the as-received NiTi wires. Figure 2 shows the AFM observations and the corresponding surface roughness (Ra: nm) of the as-received NiTi wires. Surface defects produced during the manufacturing processes were noticeably observed on Ormco, SY, and KH wires (Figures 1 and 2). A rougher surface was visible on Ormco (Ra: 1272 nm) and KH wires (Ra: 1078 nm), whereas RMO wire had the lowest surface roughness (Ra: 187 nm) (Figure 2). ESCA surface analysis results, which are not shown here, indicated that the outermost surface of the passive film on all tested NiTi wires was the same and contained mainly TiO2 with trace amounts of NiO.

FIGURE 1.

Scanning electron microscope (SEM) observations of the as-received commercial nickel-titanium (NiTi) wires from different manufacturers

FIGURE 1.

Scanning electron microscope (SEM) observations of the as-received commercial nickel-titanium (NiTi) wires from different manufacturers

Close modal
FIGURE 2.

Atomic force microscope (AFM) observations (left: three-dimension, right: top view) and corresponding surface roughnesses of the as-received commercial nickel-titanium (NiTi) wires from different manufacturers

FIGURE 2.

Atomic force microscope (AFM) observations (left: three-dimension, right: top view) and corresponding surface roughnesses of the as-received commercial nickel-titanium (NiTi) wires from different manufacturers

Close modal

Table 1 shows the Rp values of the as-received NiTi wires from different manufacturers after linear polarization tests in acidic artificial saliva. The ranking of the mean Rp value was KH (1.02 × 105 Ω cm2) > Ormco (9.12 × 104 Ω cm2) > RMO (6.26 × 104 Ω cm2) > SY (4.55 × 104 Ω cm2). Results of one-way ANOVA showed that the manufacturers of NiTi wires had a statistically significant influence on the Rp value (P < .001).

TABLE 1.

Polarization Resistances (Rp [Ω cm2]) of the As-Received Commercial Nickel-Titanium Wires From Different Manufacturers After Linear Polarization Tests in Acidic Artificial Saliva

Polarization Resistances (Rp [Ω cm2]) of the As-Received Commercial Nickel-Titanium Wires From Different Manufacturers After Linear Polarization Tests in Acidic Artificial Saliva
Polarization Resistances (Rp [Ω cm2]) of the As-Received Commercial Nickel-Titanium Wires From Different Manufacturers After Linear Polarization Tests in Acidic Artificial Saliva

Reports concerning the corrosion resistance of orthodontic wires focus mainly on the corrosion parameters, such as corrosion potential, pitting potential, or passive current density obtained from the potentiodynamic polarization curve.5–10,14–17 In this study, from the linear polarization test, taken as a nondestructively fast electrochemical technique, the Rp value could be rapidly calculated and used as a parameter for corrosion resistance evaluation with respect to the above-mentioned corrosion parameters. Although the ESCA analysis results showed that the chemical structure of the outermost surface of the passive film on all NiTi wires was the same, a statistically significant difference in Rp (or corrosion resistance) among the NiTi wires from different manufacturers was found (P < .001).

The significant difference in Rp (corrosion resistance) could be due to the variation in surface. In this study, NiTi wires (such as Ormco and KH wires) with a rougher surface (Figure 2) and preexisting surface defects (Figure 1) did not exhibit a lower Rp (Table 1) or a lower corrosion resistance. This was similar to the results reported by Huang et al,1 who reported that NiTi wires with rougher surfaces do not exhibit higher ion release in acidic artificial saliva. Therefore, the difference in Rp (or corrosion resistance) among the NiTi wires with the same surface-passive film, containing TiO2 with small amounts of NiO, was believed to be related to the different surface residual stress produced during the various manufacturing processes, instead of the surface roughness, surface defects, and chemical structure. Conflicting results have been reported by Widu et al14 who stated that surface roughness should be taken into account as an indicator of the tendency toward corrosion for orthodontic wires. Oshida et al18 noted that surface defects on NiTi orthodontic archwire produced during the manufacturing process can be possible sites for corrosion.

It is known that the biocompatibility of a metal is related mainly to the character of the surface-passive film.17,19–21 The TiO2-based passive film on NiTi wires can provide a good measure of NiTi alloy biocompatibility.5,17,19 Therefore, NiTi wire with long-term good corrosion resistance, namely, with a durably protective TiO2-based passive film, in an acidic oral environment is crucial to biocompatibility. In this study, Ormco and KH wires showed a higher Rp (or a more protective TiO2-based passive film) in the acidic artificial saliva. The passive film on Ormco and KH wires had a higher ion transfer resistance, although no difference in the outermost surface structure of the passive film was present. Therefore, these two NiTi wires were suggested as potential candidates for dental orthodontic applications.

For highly corrosion-resistant surgical implants, the Rp value may even reach 106 Ω cm2.13 Huang and coworkers22,23 reported that in acidic modified Fusayama artificial saliva the Rp value for Ti with a surface structure of TiO2 is around 106 Ω cm2, whereas the Rp value for Ti-6Al-4V with a surface structure of TiO2 and small amounts of Al2O3 and V2O5 is around 4.5 × 104 Ω cm2. In this study, the Rp values of the investigated NiTi wires with the same TiO2 and NiO surface chemical structure ranged between 4.5 × 104 and 105 Ω cm2, which were lower than that of Ti but close to or higher than that of Ti-6Al-4V alloy. Therefore, the corrosion resistance of the investigated NiTi wires in the slightly acidic oral environment was acceptable, although the variance in the corrosion resistance of the NiTi wires was present.

The surface roughness of the as-received commercial NiTi orthodontic wires (RMO, Ormco, KH, and SY) with identical surface chemical structure (TiO2 and small amounts of NiO) does not correspond with the difference in corrosion resistance for the polarization resistance in artificial saliva. From the corrosion resistance point of view, KH and Ormco wires, showing a higher polarization resistance of about 105 Ω cm2, were suggested as suitable candidates for dental orthodontic applications. Surface residual stress produced during the manufacturing processes on the commercial NiTi wires may play an important role in the corrosion resistance instead of the surface roughness and chemical structure.

1
Huang
,
H. H.
,
Y. H.
Chiu
,
T. H.
Lee
,
S. C.
Wu
,
H. Y.
Yang
,
K. H.
Su
, and
C. C.
Hsu
.
Ion release from NiTi orthodontic wires in artificial saliva with various acidities.
Biomaterials
2003
.
24
:
3585
3592
.
2
Guyuron
,
B.
and
C. I.
Lasa
Jr.
.
Reaction to stainless steel wire following orthognathic surgery.
Plast Reconstr Surg
1992
.
89
:
540
542
.
3
Koster
,
R.
,
D.
Vieluf
,
M.
Kiehn
,
M.
Sommerauer
,
J.
Kähler
,
S.
Baldus
,
T.
Meinertz
, and
C. W.
Hamm
.
Nickel and molybdenum contact allergies in patients with coronary in-stent stenosis.
Lancet
2000
.
356
:
1895
1897
.
4
Ryhanen
,
J.
,
E.
Niemi
,
W.
Serlo
,
E.
Niemela
,
P.
Sandvik
,
H.
Pernu
, and
T.
Salo
.
Biocompatibility of nickel-titanium shape memory metal and its corrosion behavior in human cell cultures.
J Biomed Mater Res
1997
.
35
:
451
457
.
5
Veldhuizen
,
A. G.
,
D. J.
Wever
,
J.
De Vries
,
H. J.
Busscher
,
D. R. A.
Uges
, and
J. R.
Van Horn
.
Electrochemical and surface characterization of a nickel-titanium alloy.
Biomaterials
1998
.
19
:
761
769
.
6
Rondelli
,
G.
and
B.
Vicentini
.
Evaluation by electrochemical tests of the passive film stability of equiatomic NiTi alloy also in presence of stress-induced martensite.
J Biomed Mater Res
2000
.
51
:
47
54
.
7
Huang
,
H. H.
Corrosion resistance of stressed NiTi and stainless steel orthodontic wires in acid artificial saliva.
J Biomed Mater Res
2003
.
66A
:
829
839
.
8
Shin
,
J. S.
,
K. T.
Oh
, and
C. J.
Hwang
.
In vitro surface corrosion of stainless steel and NiTi orthodontic appliances.
Aust Orthod J
2003
.
19
:
13
18
.
9
Kim
,
H.
and
J. W.
Johnson
.
Corrosion of stainless steel, nickel-titanium, coated nickel-titanium, and titanium orthodontic wires.
Angle Orthod
1999
.
69
:
39
44
.
10
Hunt
,
N. P.
,
S. J.
Cunningham
,
C. G.
Golden
, and
M.
Sheriff
.
An investigation into the effects of polishing on surface hardness and corrosion of orthodontic archwires.
Angle Orthod
1999
.
69
:
433
440
.
11
Geis-Gerstorfer
,
J.
and
H.
Weber
.
Effect of potassium thiocyanate on corrosion behavior of non-precious metal dental alloys.
Dtsch Zahnärztl Z
1985
.
40
:
87
91
.
12
Mayer
,
J. M.
and
J. N.
Nally
.
Influence of artificial salivas on the corrosion of dental alloys.
J Dent Res
1975
.
54
:
678
.
13
Jones
,
D. A.
Principles and Prevention of Corrosion. 2nd ed.
Upper Saddle River, NJ: Prentice-Hall Inc; 1996:143–167
.
14
Widu
,
F.
,
D.
Drescher
,
R.
Junker
, and
C.
Bourauel
.
Corrosion and biocompatibility of orthodontic wires.
J Mater Sci Mater Med
1999
.
10
:
275
281
.
15
Rondelli
,
G.
and
B.
Vicentini
.
Localized corrosion behavior in simulated human body fluids of commercial Ni-Ti orthodontic wires.
Biomaterials
1999
.
20
:
785
792
.
16
Thierry
,
B.
,
M.
Tabrizian
,
C.
Trepanier
,
O.
Savadogo
, and
L'H.
Yahia
.
Effect of surface treatment and sterilization processes on the corrosion behavior of NiTi shape memory alloy.
J Biomed Mater Res
2000
.
51
:
685
693
.
17
Wever
,
D. J.
,
A. G.
Veldhuizen
,
J.
De Vries
,
H. J.
Busscher
,
D. R. A.
Uges
, and
J. R.
Van Horn
.
Electrochemical and surface characterization of nickel-titanium alloy.
Biomaterials
1998
.
19
:
761
769
.
18
Oshida
,
Y.
,
R. C. L.
Sachdeva
, and
S.
Miyazaki
.
Microanalytical characterization and surface modification of TiNi orthodontic archwires.
Biomed Mater Eng
1992
.
2
:
51
69
.
19
Shabalovskaya
,
S. A.
and
J. M.
Anderegg
.
Surface spectroscopy characterization of NiTi nearly equiatomic shape memory alloy for implants.
J Vac Sci Technol A
1995
.
13
:
2624
2632
.
20
Oshida
,
Y.
and
S.
Miyazaki
.
Corrosion and biocompatibility of shape memory alloys.
Corros Eng
1991
.
40
:
1009
1025
.
21
Nakayama
,
Y.
,
T.
Yamamuro
,
Y.
Kotoura
, and
M.
Oka
.
In vivo measurement of anodic polarization of orthopaedic implant alloys: comparative study of in vivo and in vitro experiments.
Biomaterials
1989
.
10
:
420
424
.
22
Huang
,
H. H.
,
C. H.
Hsu
,
S. J.
Pan
,
J. L.
He
, and
C. C.
Chen
.
Corrosion and cell adhesion behavior of TiN-coated and ion-nitrided titanium for dental applications.
Proceedings of the 12th International Conference on Solid Films and Surfaces (ICSFS). Fukuda Y, Watanabe K, eds. Shizuoka, Japan: Shizuoka University; 2004:225
.
23
Huang
,
H. H.
Effect of fluoride and albumin concentration on the corrosion behavior of Ti-6Al-4V alloy.
Biomaterials
2003
.
24
:
275
282
.

Author notes

Corresponding author: Her-Hsiung Huang, BSc, MSc, PhD, Associate Professor, Institute of Oral Materials Science, Chung Shan Medical University, 110, Sec. 1, Chien-Kuo North Road, Taichung 402, Taiwan ([email protected])