Context.—The Samsung LABGEOHC10 Hematology Analyzer (LABGEOHC10) is a recently developed automated hematology analyzer that uses impedance technologies. The analyzer provides 18 parameters including 3-part differential at a maximum rate of 80 samples per hour.
Objective.—To evaluate the performance of the LABGEOHC10.
Design.—We evaluated precision, linearity, carryover, and relationship for complete blood cell count parameters between the LABGEOHC10 and the LH780 (Beckman Coulter Inc) in a university hospital in Korea according to the Clinical and Laboratory Standards Institute guidelines. Sample stability and differences due to the anticoagulant used (K2EDTA versus K3EDTA) were also evaluated.
Results.—The LABGEOHC10 showed linearity over a wide range and minimal carryover (<1%) for white blood cell, hemoglobin, red blood cell, and platelet parameters. Correlation between the LABGEOHC10 and the LH780 was good for all complete blood cell count parameters (R > 0.92) except for mean corpuscular hemoglobin concentration. The bias estimated was acceptable for all parameters investigated except for monocyte count. Most parameters were stable until 24 hours both at room temperature and at 4°C. The difference by anticoagulant type was statistically insignificant for all parameters except for a few red cell parameters.
Conclusions.—The accurate results achievable and simplicity of operation make the unit recommendable for small to medium-sized laboratories.
Automated blood cell counters have been routinely used as cost-effective and valuable tools for the diagnosis and evaluation of blood disorders. Currrently, point-of-care testing is also becoming an important adjunct to hematology laboratory practice and is becoming widely available at primary health care clinics and in general practice. Although several hematologic tests, especially the measurement of hemoglobin concentration, are appropriate for point-of-care testing, other various hematologic tests are also required within such settings for the delivery of effective point-of-care testing.1–3 Nevertheless, small to medium-sized laboratories usually prefer hematology analyzers that require minimal training for operation and deliver accurate test results in laboratory conditions.
The Samsung LABGEOHC10 Hematology Analyzer (Samsung Electronics Co, Suwon, Korea) is a simple, compact, automated hematology analyzer that had been designed for small to medium-sized laboratories with limited laboratory space. The analyzer is composed of an automatic analysis area and supportive reagents (diluent, lysing reagent, and cleaner) area and uses impedance technology for cell counting and photometry for hemoglobin (Hb) assay. The analyzer provides 18 parameters, including 3-part differential at a maximum rate of 80 samples per hour, with a sample requirement of 25 μl of EDTA-anticoagulated blood.
This study was performed to evaluate the performance of the LABGEOHC10 in terms of precision, linearity, carryover, difference by anticoagulant used (K2EDTA versus K3EDTA), stability during storage, and method comparison with the LH780 (Beckman Coulter Inc, Brea, California). We followed the Clinical and Laboratory Standards Institute (CLSI) guidelines EP5-A2,4 EP6-A,5 EP9-A2,6 and H26-A27 for evaluation of precision, linearity, and method comparison. Importantly, after method comparison, we will be able to interpret results and compare them to internal performance criteria according to the EP9-A2 of the CLSI to determine whether the candidate methods (LABGEOHC10) are suitable replacements for current methods.
MATERIALS AND METHODS
Instrumentation and Parameters
The LABGEOHC10 weighs 13.5 kg, with dimensions of 344 mm (width) × 293 mm (depth) × 390 mm (height). The analyzer features a graphic color LCD module with a touch screen and built-in thermal printer module. Its internal memory is capable of storing 1000 records with full histograms and individual patient data. The analyzer can be connected to a host computer through a USB port and allows stored records to be archived or restored. The operation of the analyzer requires minimal training of laboratory personnel.
The analyzer is capable of determining 18 parameters including 3-part differential; white blood cell count (WBC), hemoglobin concentration (Hb), red blood cell count (RBC), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), mean corpuscular hemoglobin concentration (MCHC), hematocrit (HCT), RBC distribution width (RDW), platelet count (PLT), mean platelet volume (MPV), plateletcrit, platelet distribution width (PDW), granulocyte percentage, lymphocyte percentage, monocyte percentage (MON%), granulocyte count, lymphocyte count, and monocyte count (MON#).
Calibration and Quality Control
The LABGEOHC10 and LH780 were calibrated according to the manufacturers' recommendations before the study started. Manufacturers' quality control material was run on both instruments on a daily basis during the entire study period.
Blood Samples
Blood samples drawn from 244 patients and 26 healthy adults with informed consent were used for this study. Residual K2EDTA-anticoagulated blood samples were taken from the hematology laboratory of Ajou University Hospital, Suwon, Republic of Korea, after routine testing was complete. The leftover specimens were used for precision, linearity, stability, and method comparison. K2EDTA- and K3EDTA-anticoagulated blood samples were taken from participants with their prior agreement, and were used for evaluation of the difference by EDTA tube (K2/K3) usage. We used K2EDTA plastic tubes (Becton Dickinson, Franklin Lakes, New Jersey) and K3EDTA plastic tubes (Greiner Bio-One, Monroe, North Carolina) as containers. All specimens were stored at room temperature and were tested within 6 hours, except in the case of sample stability testing. This study was performed under the approval of the Ajou University Hospital Institutional Review Board.
Precision
Within-run imprecision was established according to the H26-A2 guideline.7 Twelve samples of various range were assessed by analyzing them 31 times consecutively by LABGEOHC10.
Repeatability and device/method precision of LABGEOHC10 was also established according to the EP5-A2 guideline.4 Three levels of commercial quality control material (DiatroCont 3 Hematology Control, Clinical Diagnostic Solutions Inc, Plantation, Florida) were run in pairs, 2 times per day (with at least 5-hour intervals) for 25 operating days.
Linearity
Linearity for WBC, Hb, RBC, and PLT was evaluated according to the EP6-A guideline.5 Briefly, we prepared 5 levels of analytes by mixing, diluting, or concentrating blood samples. The prepared samples were tested 4 times each and the linearity was obtained by polynomial regression analysis for first-, second- and third-order polynomials.
Method Comparison (Correlation)
The LABGEOHC10 was compared with the LH780 for the following parameters: WBC; the numbers and percentages of granulocytes, lymphocytes, and monocytes; RBC; Hb; HCT; MCV; MCH; MCHC; RDW; PLT; and MPV according to the EP9-A2.6 Briefly, each sample was measured in duplicates with both units, and the data of 40 samples were selected to cover clinically relevant ranges for each parameter. Another 24 samples with PLT less than 50 × 103/μL were used for an additional evaluation of low PLT.
Carryover
Carryover for WBC, Hb, RBC, and PLT was assessed by following the recommendation of the International Council for Standardization in Haematology8 and the H26-A2 guideline.7 Briefly, 3 consecutive analyses of a patient sample with high analyte concentration (H1, H2, and H3) were followed by 3 consecutive analyses of a patient sample with low analyte concentration (L1, L2, and L3). Carryover (%) was calculated from the formula: 100 × (L1–L3)/(H3–L3). This process was repeated 9 times and 95%, 97.5%, and 99% confidence intervals were calculated.
Stability During Storage
Samples from 10 healthy adults were used for the evaluation of sample stability. Two sets of samples, which were drawn at the same time, were analyzed 10 times at 0, 6, 12, and 24 hours after blood collection, with 1 set stored at room temperature (RT) and the other at 4°C.
Difference by Type of Anticoagulant (K2EDTA Versus K3EDTA)
We collected blood into 2 different EDTA tubes (K2EDTA and K3EDTA) from 50 patients and analyzed them to evaluate the difference of the results.
Data Analysis
Data analysis was carried out by using Microsoft Excel 2007 (Microsoft, Redmond, Washington) and SPSS v11.0 (SPSS Inc, Chicago, Illinois). We performed polynomial regression analysis for first-, second-, and third-order polynomials for linearity access.5 The effect of EDTA tube (K2 versus K3) was analyzed by paired t test and the sample stability by 1-way analysis of variance test. For the method comparison and bias estimation, we computed 95% confidence interval of predicted bias and compared it to internal performance criteria.6
RESULTS
Precision
Within-run precision for normal and abnormal patient sample results is shown in Table 1; results are shown to satisfy the manufacturer's specification for all of the measured parameters. Repeatability and device/method precision is shown in Table 2; results are shown to satisfy the manufacturer's specification for all of the measured parameters, except for low PLT. No technologic problems with the instrument occurred during the evaluation period of 8 weeks.
Linearity
Linearity analysis for WBC, Hb, RBC, and PLT is shown in Table 3; it showed nonsignificance for all nonlinear coefficients for all 4 parameters measured.
Method Comparison and Bias Estimation
Results for the method comparison of the LABGEOHC10 with the LH780 are presented in the Figure. All parameters showed r values >0.95 with the exception of MON# (r = 0.9309), MCHC (r = 0.3778), and MPV (r = 0.9259). Results for the bias estimation of the LABGEOHC10 with the LH780 are presented in Table 4.
Comparison of the blood count parameters (A through O) between the LABGEOHC10 (Samsung Electronics Co, Suwon, Korea) and the LH780 (Beckman Coulter Inc, Brea, California). Abbreviations: GRA#, granulocyte count; Hb, hemoglobin concentration; HCT, hematocrit; Low PLT, less than 50 × 103/μL of platelet count; LYM#, lymphocyte count; MCH, mean cell hemoglobin; MCHC, mean cell hemoglobin concentration; MCV, mean corpuscular volume; MON#, monocyte count; MPV, mean platelet volume; PCT, plateletcrit; PLT, platelet count; RBC, red blood cell count; RDW, red blood cell distribution width; WBC, white blood cell count.
Comparison of the blood count parameters (A through O) between the LABGEOHC10 (Samsung Electronics Co, Suwon, Korea) and the LH780 (Beckman Coulter Inc, Brea, California). Abbreviations: GRA#, granulocyte count; Hb, hemoglobin concentration; HCT, hematocrit; Low PLT, less than 50 × 103/μL of platelet count; LYM#, lymphocyte count; MCH, mean cell hemoglobin; MCHC, mean cell hemoglobin concentration; MCV, mean corpuscular volume; MON#, monocyte count; MPV, mean platelet volume; PCT, plateletcrit; PLT, platelet count; RBC, red blood cell count; RDW, red blood cell distribution width; WBC, white blood cell count.
Carryover
Minimal carryover (<1%) was observed for all the parameters investigated (Table 3).
Stability During Storage
All parameters, except for the percentage and count of monocytes, showed 24-hour long-term stability for normal samples both at RT and at 4°C. The MON% and MON# showed statistically significant differences even after only 6 hours at 4°C, but were stable up to 6 hours and 12 hours, respectively, at RT.
Difference by EDTA Tube (K2 Versus K3)
All parameters except Hb, RBC, and HCT showed no statistically significant differences by the type of EDTA tubes used. The difference of the 3 red cell parameters was minimal (<1%) and the correlation between the different tubes was excellent for Hb (r = 0.997), RBC (r = 0.992), and HCT (r = 0.993).
COMMENT
Despite the application of previously proven technology to a commercial product, instruments of such critical importance as hematology analyzers should be validated technically before implementation in daily practice. The automated LABGEOHC10 Hematology Analyzer is easy to use and requires minimal training for operation. This study showed that the analyzer provides reliable and comparable results to instruments that are currently used in clinical laboratories.9–12
Overall results of the repeatability and device/method precision satisfied the manufacturer's specifications, except for low PLT. The coefficient of variation at low PLT was slightly higher than the coefficient of variation claimed by the manufacturer, though it was still acceptable according to EP5-A2 specifications.4 The imprecision of automated platelet counts in the thrombocytopenic range is a well-documented problem.11,13–15 The LABGEOHC10 showed an overall good correlation with the LH780, except for MCHC. The poor correlation of MCHC has also been previously reported in the literature.9,11,12,16
We determined if the LABGEOHC10 is a suitable replacement for the LH780, even though the 2 analyzers are very different in terms of their scale. The LABGEOHC10 showed an acceptable bias for most parameters investigated. Nonequivalence for MON# is thought to be due to the difference between LH780 providing a 5-part differential and LABGEOHC10, a 3-part differential. Limitation of LABGEOHC10 that provides 3-part differential and no flags other than numerical flag (high or low) might require an application of broader criteria for slide review.
Our study showed that most parameters were stable for at least 24 hours at RT and at 4°C while the monocyte count was unstable at both RT and 4°C, with an unexpectedly shorter stability of refrigerated samples than for samples at RT, and opposite directional change dependent on the storage temperature, using the LAGBEOHC10. Imeri et al17 reported that the stability of CBC parameters varied not only according to the investigated parameter but also according to storage temperature and the measurement system used. In addition, sample stability is substantially shorter than that stated by the manufacturers of hematology analyzers.17,18 Therefore, laboratory physicians might need to validate stability before the introduction of a new hematology analyzer in their laboratories.
All parameters, except Hb, RBC, and HCT, did not show statistically significant differences according to the different EDTA tubes. In this study, we used spray-dried K2EDTA plastic tubes and K3EDTA plastic tubes and therefore, the previously reported dilution effect of K3EDTA can be excluded.19 The lower MCV value of K3EDTA-anticoagulated blood by RBC shrinkage, reported in the previous studies,19,20 was not observed in this study. Although Hb, RBC, and HCT were statistically significantly lower in samples collected into K2EDTA tubes than in those collected into K3EDTA tubes, they showed excellent correlation between tubes. The differences of these parameters were minimal (<1%) and were not likely to be of any clinical relevance.
In conclusion, the overall performance of the LABGEOHC10 is comparable to that of the LH780. The accurate results and simplicity of operation make it recommendable for small to medium-sized laboratories and applicable as a back-up instrument in larger laboratories.
References
Author notes
The authors have no relevant financial interest in the products or companies described in this article.
Competing Interests
This study was supported by a research fund from Samsung Electronics Company (Suwon, Korea).
Presented in part at the XXVth International Symposium on Technological Innovations in Laboratory Hematology; May 21–24, 2012; Nice, France.