Context.—

The clinical applications of point-of-care testing (POCT) are gradually increasing in many health care systems. Recently, POCT devices using molecular genetic method techniques have been developed. We need to examine clinical pathways to see where POCT can be applied to improve them.

Objective.—

To introduce up-to-date POCT items and equipment and to provide the content that should be prepared for clinical application of POCT.

Data Sources.—

Literature review based on PubMed searches containing the terms point-of-care testing, clinical chemistry, diagnostic hematology, and clinical microbiology.

Conclusions.—

If medical resources are limited, POCT can help clinicians make quick medical decisions. As POCT technology improves and menus expand, areas where POCT can be applied will also increase. We need to understand the limitations of POCT so that it can be optimally used to improve patient management.

Point-of-care testing (POCT) is defined as a laboratory test performed outside a central laboratory, usually at or near a clinical treatment site or by a patient. In most circumstances, medical staff perform POCT, but patients perform some POCT at home. Point-of-care testing is usually performed when quick decision-making is required, such as in an emergency room or when urgent treatment is to be determined. The advantages of POCT compared with central laboratory testing include shorter wait times for results and earlier discharge home.1  The Table shows common POCT and clinical situations. Current commercial point-of-care (POC) platforms can produce results faster than laboratory-based assays, but this improvement generally results in reduced accuracy.

Common Point-of-Care Tests

Common Point-of-Care Tests
Common Point-of-Care Tests

Many analytes can be qualitatively detected or quantitatively measured using POCT; examples of commonly used POCT are tests for glucose, blood gases, cardiac markers, urinalysis, creatinine, prothrombin time/international normalized ratio (INR), infectious diseases (human immunodeficiency virus [HIV], respiratory syncytial virus, influenza, etc), and drug screening. The considerable success of POCT is attributed to increasing clinical demand, heavy industry promotion, short turnaround time, economical and practical factors, and advancements in technology, such as lab-on-a-chip systems that use miniaturization, micromachining, microfluidics, nanotechnology, and wireless communication.2 

This paper aims to describe the different applications of POCT and to review their limitations so that they can be used in the appropriate clinical pathways.

In hematology, POCT is generally limited to hemoglobin and hematocrit testing by blood gas devices and point-of-care coagulation monitoring, such as prothrombin time and INR for patients on warfarin.

Some studies have investigated portable instruments using newer technology, such as single-sample cuvettes and image analysis, microfluidic cartridges, and the lab-on-a-chip/micro total analysis system platform, which provides complete blood cell count tests including 3-part differential white blood cells, red blood cells, platelet count, and hemoglobin for rapid diagnosis in resource-poor environments.35  A rapid complete blood cell count and differential blood panel can help clinicians determine whether a patient is suffering from a viral or bacterial infection and whether he or she requires admission and antibiotics.

The use of POC INR testing has become popular for at-home testing and allows patients to easily use a device to monitor their INR and report their results to a clinician (either in person or via telephone), who can then adjust anticoagulant dose, if necessary.6  The most common symptom of patients with venous thromboembolism is chest pain or dyspnea, but patients may also present with leg problems, arm pain, or chest tightness, or be asymptomatic.7  In a systematic review and meta-analysis of 11 trials with data for 6417 participants and 12 800 person-years of follow-up, there was a significant reduction in thromboembolic events in the self-monitoring INR group (hazard ratio, 0.51; 95% CI, 0.31–0.85) on oral anticoagulation.8  Participants younger than 55 years showed a striking reduction in thrombotic events (hazard ratio, 0.33; 95% CI, 0.17–0.66), as did participants with mechanical heart valves (hazard ratio, 0.52; 95% CI, 0.35–0.77), so the authors8  suggested that patients should be offered the option to self-manage their disease with suitable health care support.

In one investigation,9  negative D-dimer test was valuable in ruling out acute pulmonary embolism in patients with a low pretest probability in hospital outpatient or accident and emergency settings. Combined with pretest probability scores, point-of-care D-dimer tests are a quick and safe way to rule out venous thromboembolism and improve patient experience.10  In addition, the D-dimer POC device was comparable with the laboratory device and was sufficiently accurate for use as a screening tool in the emergency department (ED) setting.11  In a prospective observational study12  of 104 patients who underwent simultaneous D-dimer measurements using the 2 analyzers, the median time for D-dimer results from triage by VIDAS (bioMérieux SA) was 258 minutes (interquartile range, 173–360 minutes) and that by AQT90 FLEX POCT analyzer (Radiometer Medical ApS) was 146 minutes (interquartile range, 55–280.5 minutes). Following implementation of the rapid whole blood D-dimer test in the ED, there was a 13.8% decrease in patients admitted to the hospital, a 7.3% increase in patients discharged from the ED, and a 6.4% increase in patients admitted for observation only.13  General practitioners can safely exclude pulmonary embolism by using the Wells criteria in combination with either a qualitative POC D-dimer or a quantitative D-dimer test.14 

Glucose testing is the most commonly used POCT item in the field of clinical chemistry. The most common POC glucose test is glucose testing strips, which comprise 53.7% of the total global POCT market.15  Blood glucose POCT was first used primarily at home or in nursing homes and was later expanded to inpatient settings, such as intensive care and postoperative inpatient environments, to help achieve strict glycemic control. Intensive glycemic control reduces long-term microvascular complications of diabetes. Previous representative reports include the Diabetes Control and Complication Trial16  in subjects with type 1 diabetes and the United Kingdom Prospective Diabetes Study17  in subjects with type 2 diabetes.

Van den Berghe et al18  followed 1548 patients and reported that intensive blood glucose control to levels of 79–110 mg/dL using insulin decreased mortality from 8% to 4.6% in a small homogenous population of patients in the surgical intensive care unit (ICU). Although POCT glucose meters using whole blood may be ideal in the ICU, risks of POCT include increased variability of results; as such, routine use of capillary blood sampling to monitor glucose level is not recommended in critically ill patients.19,20  The reliability of POCT glucose measurements depends upon a variety of factors, including underlying disease, patient drug regimens, interfering substances, and instrument analytical performance. Therefore, laboratory blood glucose analysis is recommended if a POCT glucose value is in the critical hypoglycemic or hyperglycemic range.21  A newer-generation (StatStrip, Nova Biomedical) POC glucose meter met more stringent accuracy criteria because of reduced bias compared with the previous-generation device,22  and implementation of StatStrip led to better agreement with venous plasma glucose, improved detection of critical low glucose results, and more efficient test utilization.23,24 

New technologies, such as continuous glucose monitoring (CGM) systems, may help alleviate the risks associated with glucose fluctuations in the ICU.2527  Recently, the implantable CGM system (Eversense CGM System, Senseonics) was reported to provide accurate glucose readings through the intended 90-day sensor life with a favorable safety profile in participants with type 1 or type 2 diabetes.28,29  However, one study30  indicated that the use of intermittent retrospective CGM did not reduce the risk of macrosomia in 300 randomized pregnant women with type 1, type 2, or gestational diabetes to either CGM (n = 147) or standard treatment (n = 153). Women who were randomized to CGM had a lower hemoglobin A1c (HbA1c) concentration by the end of the pregnancy and smaller babies (reduced median customized birth weight centile and reduced rate of large-for-gestational-age [>90th percentile] infants) compared with women in the usual-care group.31 

In addition to serum glucose tests, HbA1c measurements are useful for tracking patients with diabetes. Point-of-care testing for HbA1c is increasingly performed in outpatient conditions to monitor glucose control in diabetes mellitus. Laboratory HbA1c is established for both monitoring glycemic control and diagnosing diabetes.32,33  The American Diabetes Association32  recommends an HbA1c cutoff of 6.5% (48 mmol/mol) to diagnose diabetes, which is the threshold at which moderate retinopathy is detected. In a case-control study34  of 607 pregnant women between the 24th and 28th weeks of gestation, screening for gestational diabetes mellitus by measuring HbA1c by G8 (Tosoh Corporation) reduced use of the glucose challenge test. Availability of rapid HbA1c measurements using the DCA 2000 instrument (Bayer) increased the frequency of therapy intensification and lowered HbA1c level in patients with type 2 diabetes in an urban neighborhood health center.35  The Abbott (formerly Alere) Afinion AS100 HbA1c Dx device is the first and only POC HbA1c method cleared by the US Food and Drug Administration to aid health care professionals in the diagnosis of diabetes.

Several reports3540  have cited advantages and disadvantages of the POC HbA1c test in diabetes management, treatment adaptation, and glycemic control. In addition, many studies38,4145  have shown advantages of the POC HbA1c test in terms of patient satisfaction and cost-effectiveness. Several studies4649  of various POC HbA1c systems identified significant differences in analytical performance, including precision and accuracy. Some HbA1c POC devices had excellent analytical performance, whereas some were not acceptable; continuous performance improvements are needed.4649  Imprecision (coefficient of variation) ranged from 1.8% to 4.9% at an HbA1c value of 4.7% to 5.2% for 6 POC instruments (DCA Vantage, Siemens Diagnostics; In2it, Bio-Rad; Afinion, Axis-Shield; Nycocard, Axis-Shield; Clover, Infopia; and InnovaStar, DiaSys).48  According to data of 6 years of accuracy-based proficiency testing for HbA1c, POCT for HbA1c measurement showed the highest “unacceptable” rate and imprecision.50  This indicates that the limitations of POCT equipment on precision and accuracy need to be understood and improved. Intensive educational efforts combined with external quality assessment improve the preanalytical phase in general practitioner offices and nursing homes.51 

Point-of-care testing for creatinine and cardiac markers in the ED is also rapidly changing. Molecularly imprinted polymers have been used in biomimetic sensors as recognition elements. These polymers contain binding sites complementary in shape and function to their target analyte and their use has been critically evaluated in the detection of biomarkers of acute coronary syndrome.52  The most obvious contribution of POCT to cardiovascular disease evaluation is diagnosis of acute coronary syndrome. Currently, diagnosis of acute myocardial infarction is dependent on the change of cardiac troponin (cTn) concentrations, with at least one measurement above the 99th-percentile upper reference limit and evidence of one other listed criterion specified by the European Society of Cardiologists.53  Despite the importance of the analytical sensitivity of cTn measured by POC devices, there are significant differences in limit of detection among equipment using biosensors and biomimetic sensors for cardiac biomarkers.52  Although high-sensitivity cTn assays measure relatively low values and document small increases above the 99th percentile upper reference limit, many contemporary and POC cTn assays may not detect small increases within the reference interval or at slightly above the 99th-percentile upper reference limit. This may lead to substantial differences in frequency of events based solely on the cTn assay used.53  The decision to implement cTn POCT can significantly increase the capacity and efficiency for diagnosing acute myocardial infarction and other related conditions.54,55 

The analytical sensitivity (limit of detection) of cTnI and cTnT in 12 POC assays varies 20-fold, ranging from 0.008 to 0.15 μg/L.56  The 99th percentile of cTn values ranges from 0.01 to 0.2 μg/L.56  In addition, the 99th percentile variability between assays is significant and shows lack of standardization of cTnI and cTnT assays.56  When a single common population of patients with suspected acute coronary syndrome was used for analysis, the Alere Triage Cardio3 TnI assay (Alere) and the PathFast cTnI-II (Mitsubishi Chemical Medience Corporation) assay had similar diagnostic performance to the central laboratory assay Singulex Erenna TnI (Singulex).57  The use of point-of-care cTnI measurement allows early rapid diagnosis or exclusion of myocardial infarction. The use of the additional measurement of myoglobin and CK-MB does not provide further diagnostic information.58  Additional use of ultrasensitive copeptin improves the diagnostic performance of conventional sensitive POCT assays for cTn to overcome lower sensitivities at the cost of decreased clinical specificity.59 

Point-of-care systems could be used to assess natriuretic peptide, B-type natriuretic peptide, or N-terminal pro–B-type natriuretic peptide levels in the ED and community outpatient settings to monitor the risk of acute heart failure; median time to discharge was reduced with versus without POCT (8.8 hours versus 14.2 hours, P < .001).60  Natriuretic peptides are an important objective tool in diagnosis, prognosis, and management of heart failure, and POCT is a valuable tool for obtaining real-time information, particularly in the emergency room.61 

Other rapidly developing areas of POCT in the ED are creatinine measurement and pregnancy tests, both of which are used to determine whether patients can be properly referred for radiologic procedures.

Contrast-induced acute kidney injury occurs rarely in patients exposed to iodinated contrast. Screening patients at risk for postcontrast acute kidney injury may be feasible with creatinine POCT technology, as all at-risk patients were identified by several analyzers: i-STAT (Abbott), StatSensor (Nova Biomedical), epoc (Siemens), ABL90 Flex Plus (Radiometer Benelux), ABL800 FLEX (Radiometer Benelux), and STAT Profile Prime+ (Nova Biomedical).62,63  In another study, i-STAT and epoc were the most well-functioning POC devices but were less user-friendly. StatSensor did not meet any of the error criteria for creatinine or estimated glomerular filtration rate measurements. However, it was more user-friendly than the other POC devices.64 

Stomach pain, cramps, spasms, and obstetric complications are leading primary causes for women between 15 and 64 years of age to visit the ED.65  Qualitative urine human chorionic gonadotropin (hCG) POCT is widely used in the ED to assess patient pregnancy status because presenting symptoms may be indicative of an abnormal pregnancy (abdominal pain, vaginal bleeding, pelvic discomfort, and miscarriage), or exclusion of pregnancy is required prior to performing diagnostic or therapeutic interventions that could harm a developing fetus.55  The claimed lower limit of detection for urine hCG in many POC devices ranges from 6.3 to 50 mIU/mL, with most falling between 20 and 25 mIU/mL.66  Some reports have shown that qualitative hCG POC devices can result in erroneous negative consequences due to excess hCG variants6771  and poor analytical sensitivity.7274  Kamer et al75  reported that 4 commonly used hCG POC devices (Alere hCG Combo Cassette [Alere], ICON 20 hCG [Beckman Coulter, Inc.], OSOM hCG Combo Test [Sekisui Diagnostics, LLC], and Sure-Vue Serum/Urine hCG-STAT [Fisher Scientific]) were susceptible to false-negative results at low concentrations of urine hCG. An evaluation of sensitivity for the hook effect caused by hCG β core fragments showed that susceptibility to inhibition of 11 types of POCT equipment varied greatly; only 2 devices exhibited minimal to no susceptibility to hCG β core fragments.67  Laboratory physicians and clinicians should be aware of the limitations of using urine hCG POC devices to rule out early pregnancy.

Drug screening tests are frequently ordered in patients presenting to the ED with acute psychiatric symptoms such as agitation, ataxia, delirium, altered mental status, or psychosis. There are advantages and disadvantages regarding the usefulness of drug of abuse screening in an emergency room. Tenenbein76  reported that emergency drug screening was unlikely to significantly impact patient management in the ED. Point-of-care testing for drugs of abuse and therapeutic drugs in the ED may be effective, less costly, and more rapid than the laboratory-based screening system.7779  However, immunoassays for drugs of abuse have poor specificity (inaccuracy), and there is no correlation between the presence of drugs or metabolites in the urine and toxic effects or clinical impairment.80  Point-of-care testing is based on immunochromatography, in which a drug in the patient's sample competes with drug and antibody conjugates in the test to allow or block development of a colored line; most POCT is visually interpreted in a few minutes.81  An optimal screening test is not an optimal diagnostic test, and urine drug screens are primarily designed as a screening test.76  Interpretation of negative screening results, whether expected or unexpected, can also be challenging. Several factors must be considered, including assay cutoff, cross-reactivity, time since last dose, low immunoreactivity, and individual metabolism.80  The nature of the screening test allows false-positive results in some cases, but additional gold standard confirmatory testing, such as chromatography and mass spectrometry, can be carried out for confirmation.

Urine and serum (or plasma) are the most frequently used specimens for clinical toxicologic testing, but oral fluid may also be a rapid and accurate way to screen for psychoactive substances.82  Overall estimates of reliability in oral fluid with POCT devices showed high variability for detecting amphetamines, benzodiazepines, cannabinoids, cocaine, and opioids.83  When choosing a POCT device, certain features should be considered, such as forensic utility, reliability of device manufacturer, specimen type, test menu, analytical methodology and performance, result interpretation, and cost.83,84 

Point-of-care testing blood gas analyzers for electrolytes and basic metabolic panels including β-hydroxybutyrate and lactate measurements facilitate management of several scenarios and may be useful in ED and ICU settings.55  There is fair evidence that POCT of arterial blood gas results in the ICU and ED leads to improved clinical outcomes when POCT reduces therapeutic turnaround time compared with central laboratory testing.85  Although arterial blood gas analysis remains the gold standard to assess acid-base, ventilation, and oxygenation status in critically ill patients, venous blood gas analysis has been shown to correlate with arterial blood gas analysis and has been proposed as a safer, less invasive alternative to arterial blood gas analysis for undifferentiated critically ill patients in the ED and ICU.86,87  Calculated arterial blood gases from venous samples and pulse oximetry are comparable to arterial blood gas values. Point-of-care testing could reduce the logistic burden of arterial sampling, facilitate improved screening and follow-up, and reduce patient pain.88  A retrospective study and prospective analytical observational studies at multiple health care sites showed sufficient agreement between sodium, potassium, and ionized calcium results obtained from blood gas and central laboratory analyzers to enable prompt clinical decision-making.8991  With proper training and education of the ED care team, POCT can be used as an effective tool for managing patient flow in the ED.92 

C-reactive protein (CRP) provides diagnostic value for ruling in or ruling out serious bacterial infection in febrile children.93  Preantiretroviral therapy POC CRP testing may reduce mortality by identifying HIV patients at high risk for poor outcomes.94  Brouwer and van Pelt95  compared 2 semiquantitative strips (Actim CRP strips [Medix Biochemica] and Cleartest CRP strips [Servopax]) and 6 quantitative CRP tests (Afinion AS 100 analyzer [Axis Shield], QuikRead go [Orion Diagnostica], Smart analyzer [Eurolyser Diagnostica], iChroma analyzer [Boditech Med Inc.], Microsemi [Horiba Ltd.], and AQT90 FLEX [Radiometer Medical ApS]) to the Synchron CRP method (Synchron analyzer; Beckman Coulter), using the Clinical and Laboratory Standards Institute EP9 protocol. They concluded that the semiquantitative CRP strips could be used to discriminate between normal and increased levels of CRP, and that the Smart and Afinion would be the preferred quantitative analyzers for POCT, taking into account both analytical validation and practical evaluation.95  Immunochromatographic methods were used in semiquantitative strips, and various methods such as immunoturbidimetric assay, solid-phase immunochemical assay, fluorescence sandwich immunoassay, and solid-phase sandwich immunoassay were used in quantitative CRP analyzers.95  In addition, immediate POC CRP testing is reliable in forensic settings such as during autopsy and at postmortem inspection.96 

Point-of-care testing immunodiagnostics in routine clinical laboratories commonly depend on antibodies as biorecognition elements in which the biomolecular interaction is monitored by sophisticated equipment provided by the assay manufacturer. Most widely applied recognition elements are monoclonal or polyclonal antibodies.97  The World Health Organization has called for development of a rapid, biomarker-based, nonsputum test capable of detecting all forms of tuberculosis at the POC to enable immediate treatment initiation. Lipoarabinomannan is the only World Health Organization–endorsed tuberculosis biomarker that can be detected in urine, an easily collected sample.98 

Lateral flow immunoassay is the most popular diagnostic tool that meets the required standards for colorimetric assays.99  Lateral flow immunoassay provides sensitive and precise quantitative determination of target analytes because it carries out immunologic recognition at a variety of concentrations with frequently used labels, such as gold nanoparticles, quantum dots, and up-converting phosphorescent labels.100  However, limitations should be considered according to the characteristics of the lateral flow immunoassay, such as visual interpretation and the possibility of misunderstanding, incorrect timing (read within a specific time), detection sensitivity, and manual input of results.

Quantitative reverse transcription polymerase chain reaction is the reference standard laboratory test for influenza diagnosis. New rapid diagnostic tests for influenza that show sensitivity and specificity comparable to those of real-time polymerase chain reaction assays are available (Figure).101104  Diagnosis of patients with influenza by POCT resulted in significantly higher rates of antiviral prescription in a systematic review of the impact of POCT for influenza on outcomes of patients with acute respiratory tract infection.101  Although rapid molecular POCT for respiratory viruses (FilmArray Respiratory Panel, BioFire) in adults presenting to a hospital with acute respiratory illness was not associated with reduction in overall duration of antibiotics, more patients in the POCT group than in the control group received single doses or brief courses of antibiotics.105  Importantly, a negative POCT test does not confirm the absence of influenza, and ED clinicians should use other confirmatory tests to avoid diagnostic delays or to confidently exclude influenza.106 

Generalized steps in methods of diagnostic testing for influenza (adapted from Egilmezer et al101  with permission). Abbreviations: POC, point-of-care; POCT, point-of-care test; RT-qPCR, quantitative reverse transcription polymerase chain reaction; TAT, turnaround time.

Generalized steps in methods of diagnostic testing for influenza (adapted from Egilmezer et al101  with permission). Abbreviations: POC, point-of-care; POCT, point-of-care test; RT-qPCR, quantitative reverse transcription polymerase chain reaction; TAT, turnaround time.

Close modal

Multiplex detectability is necessary for emergency treatment depending on disease stage or interactional infections. Brendish et al107  reported several molecular platforms with potential for POCT use and evidence for clinical and economic benefits of testing for respiratory viruses in adults: Alere i Influenza A&B (Alere), FilmArray Respiratory Panel (BioFire Diagnostics), and Xpert Flu (Cepheid). Use of the FilmArray Respiratory Panel for detecting influenza A (H1 and H3), influenza B, respiratory syncytial virus, rhinovirus, enterovirus, human metapneumovirus, parainfluenza virus types 1 through 4, coronaviruses (OC43, 229E, HKU1, and NL63), and adenovirus was associated with reduced length of stay, improved influenza detection, and antiviral use, and appeared to be safe.105  Hagen et al108  suggested that introduction of a FilmArray meningitis/encephalitis panel into routine clinical procedures is associated with significantly reduced length and days of therapy for empiric anti-infective treatment in children with suspected meningoencephalitis (4.0 versus 3.0, P = .04, and 8.0 versus 6.0, P = .02, respectively). Patients with exacerbation of airway disease should undergo respiratory virus testing at the POC using a comprehensive syndromic multiplex panel rather than a molecular POCT for influenza alone, which would not detect the majority of viruses associated with early antibiotic discontinuation.109 

An evaluation of 4 POC HIV tests (Chembio DPP HIV-1/2, Determine HIV-1/2 Ag/Ab Combo, INSTI HIV-1/HIV-2 Rapid Antibody Test, and OraQuick ADVANCE Rapid HIV-1/2 Antibody Test) using unprocessed oral fluid and whole blood specimens showed that sensitivity of whole blood POCT ranged from 95.53% to 97.21%, and specificity was high for all tests (range, 99.44%–100.00%).110  Despite relatively high specificity and sensitivity of POC HIV tests, the authors110  suggested that organizations should have a plan to manage false results that occur because of the limitations of POCT (eg, lower sensitivity to identify acute HIV infection, effects of antiretroviral therapy and pre-exposure prophylaxis on oral fluid tests, low positive predictive value among pre-exposure prophylaxis populations). In a systematic review and meta-analysis,111  rapid POC HIV testing was highly accurate compared with conventional tests and offered a clear advantage of timely interventions to reduce mother-to-child transmission of HIV. If POCT results for HIV are indeterminate, the test result should not be considered preliminarily positive or negative, and repeat POC HIV testing or HIV testing at an approved HIV testing laboratory is required.112  Dual POCT screening for HIV and syphilis (SD BIOLINE HIV/Syphilis Duo Test, Standard Diagnostics; MedMira Multiplo Rapid TP/HIV Antibody Test, MedMira; and Chembio Dual Path Platform HIV/Syphilis Assay, Chembio Diagnostic System) was more cost-effective than single rapid tests and prevented more adverse pregnancy outcomes.113  The FACSPresto POC CD4+ T-cell test (Becton Dickinson) is useful and does not have significant variability in reliability when performed by nonlaboratory health care workers. Hence, it may be a valuable instrument to increase access and coverage of CD4 estimations for monitoring HIV-infected individuals in developing countries.114116  Expansion of molecular POC type Mycobacterium tuberculosis testing, including drug resistance testing, is also highly desired and necessary.2 

The coronavirus disease 2019, which began in December 2019, is a severe acute respiratory syndrome coronavirus 2 infection, and real-time reverse transcription polymerase chain reaction is currently the most reliable diagnostic method for coronavirus disease 2019 around the world.117  Serologic testing may be helpful for diagnosis of suspected patients with negative reverse transcription polymerase chain reaction results and for identification of individuals with an adaptive immune response to severe acute respiratory syndrome coronavirus 2, indicating recent or prior infection.118  But the POC immunochromatographic immunoglobulin (Ig) M/IgG antibody assay had low sensitivity during the early phase of severe acute respiratory syndrome coronavirus 2, and immunochromatographic assay alone is not recommended for initial diagnostic testing for coronavirus disease 2019.119 

Point-of-care testing is likely to play an increasing role in health care delivery in the future. It will improve access to health care and increase the efficacy of service provided to patients.49  Although POCT provides laboratory results faster than the traditional central laboratory, process improvement is needed to optimize the accuracy of laboratory results.92 

Molecular POCT for common pathogens in select populations, such as in intensive care or other common illness presentations, needs to be evaluated to further improve patient care and effectively manage health care resources.120  Despite several advantages of POCT, limitations include cost, imprecision and inaccuracy, requirement for an interdisciplinary approach, and human error. Although the use of POCT is expanding in many areas, the limitations must be understood and improvements in analytical performance achieved to properly interpret POCT results.

1.
Kankaanpaa
M,
Raitakari
M,
Muukkonen
L,
et al
Use of point-of-care testing and early assessment model reduces length of stay for ambulatory patients in an emergency department
.
Scand J Trauma Resusc Emerg Med
.
2016
;
24
(1)
:
125
.
2.
Abel
G.
Current status and future prospects of point-of-care testing around the globe
.
Expert Rev Mol Diagn
.
2015
;
15
(7)
:
853
855
.
3.
Abbasi
U,
Chowdhury
P,
Subramaniam
S,
et al
A cartridge based point-of-care device for complete blood count
.
Sci Rep
.
2019
;
9
(1)
:
18583
.
4.
Smith
S,
Madzivhandila
P,
Sewart
R,
et al
Microfluidic cartridges for automated, point-of-care blood cell counting
.
SLAS Technol
.
2017
;
22
(2)
:
176
185
.
5.
Larsson
A,
Carlsson
L,
Karlsson
B,
Lipcsey
M.
Rapid testing of red blood cell parameters in primary care patients using HemoScreen point of care instrument
.
BMC Fam Pract
.
2019
;
20
(1)
:
77
.
6.
Lingervelder
D,
Koffijberg
H,
Kusters
R,
IJzerman
MJ.
Point-of-care testing in primary care: a systematic review on implementation aspects addressed in test evaluations
.
Int J Clin Pract
.
2019
;
73
(10)
:
e13392
.
7.
Konstantinides
SV,
Torbicki
A,
Agnelli
G,
et al
2014 ESC guidelines on the diagnosis and management of acute pulmonary embolism
.
Eur Heart J
.
2014
;
35
(43)
:
3033
3069
,
3069a–3069k.
8.
Heneghan
C,
Ward
A,
Perera
R,
et al
Self-monitoring of oral anticoagulation: systematic review and meta-analysis of individual patient data
.
Lancet
.
2012
;
379
(9813)
:
322
334
.
9.
Crawford
F,
Andras
A,
Welch
K,
Sheares
K,
Keeling
D,
Chappell
FM.
D-dimer test for excluding the diagnosis of pulmonary embolism
.
Cochrane Database Syst Rev.
2016
(8):CD010864.
10.
Marquardt
U,
Apau
D.
Point-of-care D-dimer testing in emergency departments
.
Emerg Nurse
.
2015
;
23
(5)
:
29
35
.
11.
Sen
B,
Kesteven
P,
Avery
P.
Comparison of D-dimer point of care test (POCT) against current laboratory test in patients with suspected venous thromboembolism (VTE) presenting to the emergency department (ED)
.
J Clin Pathol
.
2014
;
67
(5)
:
437
440
.
12.
Perveen
S,
Unwin
D,
Shetty
AL.
Point of care D-dimer testing in the emergency department: a bioequivalence study
.
Ann Lab Med
.
2013
;
33
(1)
:
34
38
.
13.
Lee-Lewandrowski
E,
Nichols
J,
Van Cott
E,
et al
Implementation of a rapid whole blood D-dimer test in the emergency department of an urban academic medical center: impact on ED length of stay and ancillary test utilization
.
Am J Clin Pathol
.
2009
;
132
(3)
:
326
331
.
14.
Lucassen
WA,
Erkens
PM,
Geersing
GJ,
et al
Qualitative point-of-care D-dimer testing compared with quantitative D-dimer testing in excluding pulmonary embolism in primary care
.
J Thromb Haemost
.
2015
;
13
(6)
:
1004
1009
.
15.
Reynolds
TM,
Pillay
TS.
Rumsfeld
Donald
and point-of-care testing (POCT): the problem of unknown unknowns—venous blood and glucose oxidase-based test strips
.
J Clin Pathol
.
2016
;
69
(10)
:
839
840
.
16.
Nathan
DM,
Genuth
S,
Lachin
J,
et al
The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus
.
N Engl J Med
.
1993
;
329
(14)
:
977
986
.
17.
Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33)
.
UK Prospective Diabetes Study (UKPDS) Group.
Lancet
.
1998
;
352
(9131)
:
837
853
.
18.
Van den Berghe
G,
Wouters
P,
Weekers
F,
et al
Intensive insulin therapy in critically ill patients
.
N Engl J Med
.
2001
;
345
(19)
:
1359
1367
.
19.
Petersen
JR,
Graves
DF,
Tacker
DH,
Okorodudu
AO,
Mohammad
AA,
Cardenas
VJ
Jr.
Comparison of POCT and central laboratory blood glucose results using arterial, capillary, and venous samples from MICU patients on a tight glycemic protocol
.
Clin Chim Acta
.
2008
;
396
(1–2)
:
10
13
.
20.
Shearer
A,
Boehmer
M,
Closs
M,
et al
Comparison of glucose point-of-care values with laboratory values in critically ill patients
.
Am J Crit Care
.
2009
;
18
(3)
:
224
230
.
21.
Janssen
K,
Delanghe
J.
Importance of the pre-analytical phase in blood glucose analysis
.
Acta Clin Belg
.
2010
;
65
(5)
:
311
318
.
22.
Karon
BS,
Blanshan
CT,
Deobald
GR,
Wockenfus
AM.
Retrospective evaluation of the accuracy of Roche AccuChek Inform and Nova StatStrip glucose meters when used on critically ill patients
.
Diabetes Technol Ther
.
2014
;
16
(12)
:
828
832
.
23.
Raizman
JE,
Shea
J,
Daly
CH,
et al
Clinical impact of improved point-of-care glucose monitoring in neonatal intensive care using Nova StatStrip: evidence for improved accuracy, better sensitivity, and reduced test utilization
.
Clin Biochem
.
2016
;
49
(12)
:
879
884
.
24.
Lockyer
MG,
Fu
K,
Edwards
RM,
et al
Evaluation of the Nova StatStrip glucometer in a pediatric hospital setting
.
Clin Biochem
.
2014
;
47
(9)
:
840
843
.
25.
Honiden
S,
Inzucchi
SE.
Metabolic management during critical illness: glycemic control in the ICU
.
Semin Respir Crit Care Med
.
2015
;
36
(6)
:
859
869
.
26.
Brady
V.
Management of hyperglycemia in the intensive care unit: when glucose reaches critical levels
.
Crit Care Nurs Clin North Am
.
2013
;
25
(1)
:
7
13
.
27.
Strasma
PJ,
Finfer
S,
Flower
O,
et al
Use of an intravascular fluorescent continuous glucose sensor in ICU patients
.
J Diabetes Sci Technol
.
2015
;
9
(4)
:
762
770
.
28.
Christiansen
MP,
Klaff
LJ,
Brazg
R,
et al
A prospective multicenter evaluation of the accuracy of a novel implanted continuous glucose sensor: PRECISE II
.
Diabetes Technol Ther
.
2018
;
20
(3)
:
197
206
.
29.
Christiansen
MP,
Klaff
LJ,
Bailey
TS,
Brazg
R,
Carlson
G,
Tweden
KS.
A prospective multicenter evaluation of the accuracy and safety of an implanted continuous glucose sensor: the PRECISION study
.
Diabetes Technol Ther
.
2019
;
21
(5)
:
231
237
.
30.
Voormolen
DN,
DeVries
JH,
Sanson
RME,
et al
Continuous glucose monitoring during diabetic pregnancy (GlucoMOMS): a multicentre randomized controlled trial
.
Diabetes Obes Metab
.
2018
;
20
(8)
:
1894
1902
.
31.
Feig
DS,
Murphy
HR.
Continuous glucose monitoring in pregnant women with type 1 diabetes: benefits for mothers, using pumps or pens, and their babies
.
Diabet Med
.
2018
;
35
(4)
:
430
435
.
32.
Diagnosis and classification of diabetes mellitus.
Diabetes Care
.
2010
;
33
(suppl 1)
:
S62
S69
.
33.
Florkowski
C,
Crooke
M.
Standardisation of reporting haemoglobin A1c: adoption of the New Zealand Society for the Study of Diabetes (NZSSD) position statement
.
N Z Med J
.
2010
;
123
(1310)
:
43
49
.
34.
Maesa
JM,
Fernandez-Riejos
P,
Gonzalez-Rodriguez
C,
Sanchez-Margalet
V.
Screening for gestational diabetes mellitus by measuring glycated hemoglobin can reduce the use of the glucose challenge test
.
Ann Lab Med
.
2019
;
39
(6)
:
524
529
.
35.
Miller
CD,
Barnes
CS,
Phillips
LS,
et al
Rapid A1c availability improves clinical decision-making in an urban primary care clinic
.
Diabetes Care
.
2003
;
26
(4)
:
1158
1163
.
36.
Schnell
O,
Crocker
JB,
Weng
J.
Impact of HbA1c testing at point of care on diabetes management
.
J Diabetes Sci Technol
.
2017
;
11
(3)
:
611
617
.
37.
Al-Ansary
L,
Farmer
A,
Hirst
J,
et al
Point-of-care testing for Hb A1c in the management of diabetes: a systematic review and metaanalysis
.
Clin Chem
.
2011
;
57
(4)
:
568
576
.
38.
Thaler
LM,
Ziemer
DC,
Gallina
DL,
et al
Diabetes in urban African-Americans, XVII: availability of rapid HbA1c measurements enhances clinical decision-making
.
Diabetes Care
.
1999
;
22
(9)
:
1415
1421
.
39.
Egbunike
V,
Gerard
S.
The impact of point-of-care A1C testing on provider compliance and A1C levels in a primary setting
.
Diabetes Educ
.
2013
;
39
(1)
:
66
73
.
40.
Rust
G,
Gailor
M,
Daniels
E,
McMillan-Persaud
B,
Strothers
H,
Mayberry
R.
Point of care testing to improve glycemic control
.
Int J Health Care Qual Assur
.
2008
;
21
(3)
:
325
335
.
41.
Laurence
CO,
Gialamas
A,
Bubner
T,
et al
Patient satisfaction with point-of-care testing in general practice
.
Br J Gen Pract
.
2010
;
60
(572)
:
e98
e104
.
42.
Crocker
JB,
Lee-Lewandrowski
E,
Lewandrowski
N,
Baron
J,
Gregory
K,
Lewandrowski
K.
Implementation of point-of-care testing in an ambulatory practice of an academic medical center
.
Am J Clin Pathol
.
2014
;
142
(5)
:
640
646
.
43.
Lee-Lewandrowski
E,
Lewandrowski
K.
Perspectives on cost and outcomes for point-of-care testing
.
Clin Lab Med
.
2009
;
29
(3)
:
479
489
.
44.
Laurence
CO,
Moss
JR,
Briggs
NE,
Beilby
JJ.
The cost-effectiveness of point of care testing in a general practice setting: results from a randomised controlled trial
.
BMC Health Serv Res
.
2010
;
10
:
165
.
45.
Chadee
A,
Blackhouse
G,
Goeree
R.
Point-of-care hemoglobin A1c testing: a budget impact analysis
.
Ont Health Technol Assess Ser
.
2014
;
14
(9)
:
1
23
.
46.
Lenters-Westra
E,
Slingerland
RJ.
Three of 7 hemoglobin A1c point-of-care instruments do not meet generally accepted analytical performance criteria
.
Clin Chem
.
2014
;
60
(8)
:
1062
1072
.
47.
Criel
M,
Jonckheere
S,
Langlois
M.
Evaluation of three hemoglobin A1c point-of-care instruments
.
Clin Lab
.
2016
;
62
(3)
:
285
291
.
48.
Lenters-Westra
E,
Slingerland
RJ.
Six of eight hemoglobin A1c point-of-care instruments do not meet the general accepted analytical performance criteria
.
Clin Chem
.
2010
;
56
(1)
:
44
52
.
49.
Lenters-Westra
E,
English
E.
Evaluation of four HbA1c point-of-care devices using international quality targets: are they fit for the purpose?
J Diabetes Sci Technol
.
2018
;
12
(4)
:
762
770
.
50.
Chung
S,
Jun
SH,
Song
WH,
Song
J.
Six years' experience of accuracy-based proficiency testing for HbA1c in Korea
.
Journal of Laboratory Medicine and Quality Assurance
.
2015
;
37
(2)
:
92
100
.
51.
Solvik
UO,
Bjelkaroy
WI,
Berg
KVD,
Saga
AL,
Hager
HB,
Sandberg
S.
Intensive educational efforts combined with external quality assessment improve the preanalytical phase in general practitioner offices and nursing homes
.
Clin Chem Lab Med
.
2017
;
55
(12)
:
1857
1864
.
52.
Regan
B,
Boyle
F,
O'Kennedy
R,
Collins
D.
Evaluation of molecularly imprinted polymers for point-of-care testing for cardiovascular disease
.
Sensors (Basel)
.
2019
;
19(16).
53.
Thygesen
K,
Alpert
JS,
Jaffe
AS,
et al
Fourth universal definition of myocardial infarction (2018)
.
Circulation
.
2018
;
138
(20)
:
e618
e651
.
54.
Bingisser
R,
Cairns
C,
Christ
M,
et al
Cardiac troponin: a critical review of the case for point-of-care testing in the ED
.
Am J Emerg Med
.
2012
;
30
(8)
:
1639
1649
.
55.
Florkowski
C,
Don-Wauchope
A,
Gimenez
N,
Rodriguez-Capote
K,
Wils
J,
Zemlin
A.
Point-of-care testing (POCT) and evidence-based laboratory medicine (EBLM)—does it leverage any advantage in clinical decision making?
Crit Rev Clin Lab Sci
.
2017
;
54
(7–8)
:
471
494
.
56.
Apple
FS,
Sandoval
Y,
Jaffe
AS,
Ordonez-Llanos
J.
Cardiac troponin assays: guide to understanding analytical characteristics and their impact on clinical care
.
Clin Chem
.
2017
;
63
(1)
:
73
81
.
57.
Peacock
WF,
Diercks
D,
Birkhahn
R,
et al
Can a point-of-care Troponin I Assay be as good as a central laboratory assay?: a MIDAS investigation
.
Ann Lab Med
.
2016
;
36
(5)
:
405
412
.
58.
Collinson
P,
Goodacre
S,
Gaze
D,
Gray
A.
Very early diagnosis of chest pain by point-of-care testing: comparison of the diagnostic efficiency of a panel of cardiac biomarkers compared with troponin measurement alone in the RATPAC trial
.
Heart
.
2012
;
98
(4)
:
312
318
.
59.
Vafaie
M,
Biener
M,
Mueller
M,
et al
Addition of copeptin improves diagnostic performance of point-of-care testing (POCT) for cardiac troponin T in early rule-out of myocardial infarction—a pilot study
.
Int J Cardiol
.
2015
;
198
:
26
30
.
60.
Bingisser
R,
Cairns
CB,
Christ
M,
et al
Measurement of natriuretic peptides at the point of care in the emergency and ambulatory setting: current status and future perspectives
.
Am Heart J
.
2013
;
166
(4)
:
614
621.e611
.
61.
Iwaz
JA,
Maisel
AS.
Recent advances in point-of-care testing for natriuretic peptides: potential impact on heart failure diagnosis and management
.
Expert Rev Mol Diagn
.
2016
;
16
(6)
:
641
650
.
62.
Bogaert
L,
Schiemsky
T,
Van Hover
P,
De Schrijver
P,
Van Hoovels
L.
Analytical and diagnostic performance evaluation of five creatinine POCT devices in the identification of patients at risk for post-contrast acute kidney injury (PCAKI)
.
Clin Chem Lab Med
.
2019
;
57
(9)
:
e214
e217
.
63.
Snaith
B,
Harris
MA,
Shinkins
B,
Jordaan
M,
Messenger
M,
Lewington
A.
Point-of-care creatinine testing for kidney function measurement prior to contrast-enhanced diagnostic imaging: evaluation of the performance of three systems for clinical utility
.
Clin Chem Lab Med
.
2018
;
56
(8)
:
1269
1276
.
64.
van der Heijden
C,
Roosens
L,
Cluckers
H,
Van Craenenbroeck
AH,
Peeters
B.
Analytical and clinical performance of three hand-held point-of-care creatinine analyzers for renal function measurements prior to contrast-enhanced imaging
.
Clin Chim Acta
.
2019
;
497
:
13
19
.
65.
Niska
R,
Bhuiya
F,
Xu
J.
National Hospital Ambulatory Medical Care Survey: 2007 emergency department summary
.
Natl Health Stat Report
.
2010;(26):1–31.
66.
Braunstein
GD.
The long gestation of the modern home pregnancy test
.
Clin Chem
.
2014
;
60
(1)
:
18
21
.
67.
Nerenz
RD,
Song
H,
Gronowski
AM.
Screening method to evaluate point-of-care human chorionic gonadotropin (hCG) devices for susceptibility to the hook effect by hCG beta core fragment: evaluation of 11 devices
.
Clin Chem
.
2014
;
60
(4)
:
667
674
.
68.
Nerenz
RD,
Gronowski
AM.
Point-of-care and over-the-counter qualitative human chorionic gonadotropin (hCG) devices remain susceptible to false-negative results caused by excess hCG beta core fragment
.
Clin Chem
.
2013
;
59
(11)
:
1672
1674
.
69.
Gronowski
AM,
Powers
M,
Stenman
UH,
Ashby
L,
Scott
MG.
False-negative results from point-of-care qualitative human chorionic gonadotropin (hCG) devices caused by excess hCGbeta core fragment vary with device lot number
.
Clin Chem
.
2009
;
55
(10)
:
1885
1886
.
70.
Gronowski
AM,
Cervinski
M,
Stenman
UH,
Woodworth
A,
Ashby
L,
Scott
MG.
False-negative results in point-of-care qualitative human chorionic gonadotropin (hCG) devices due to excess hCGbeta core fragment
.
Clin Chem
.
2009
;
55
(7)
:
1389
1394
.
71.
Griffey
RT,
Trent
CJ,
Bavolek
RA,
Keeperman
JB,
Sampson
C,
Poirier
RF.
“Hook-like effect” causes false-negative point-of-care urine pregnancy testing in emergency patients
.
J Emerg Med
.
2013
;
44
(1)
:
155
160
.
72.
Nickmans
S,
Vermeersch
P,
Van Eldere
J,
Billen
J.
Performance of qualitative urinary hCG assays
.
Acta Clin Belg
.
2014
;
69
(4)
:
277
279
.
73.
Greene
DN,
Schmidt
RL,
Kamer
SM,
Grenache
DG,
Hoke
C,
Lorey
TS.
Limitations in qualitative point of care hCG tests for detecting early pregnancy
.
Clin Chim Acta
.
2013
;
415
:
317
321
.
74.
Terwijn
M,
van Schie
A,
Blankenstein
MA,
Heijboer
AC.
Pregnancy detection by quantitative urine hCG analysis: the need for a lower cut-off
.
Clin Chim Acta
.
2013
;
424
:
174
.
75.
Kamer
SM,
Foley
KF,
Schmidt
RL,
Greene
DN.
Analytical sensitivity of four commonly used hCG point of care devices
.
Clin Biochem
.
2015
;
48
(6)
:
448
452
.
76.
Tenenbein
M.
Do you really need that emergency drug screen?
Clin Toxicol (Phila)
.
2009
;
47
(4)
:
286
291
.
77.
Mastrovitch
TA,
Bithoney
WG,
DeBari
VA,
Nina
AG.
Point-of-care testing for drugs of abuse in an urban emergency department
.
Ann Clin Lab Sci
.
2002
;
32
(4)
:
383
386
.
78.
Attema-de Jonge
ME,
Peeters
SY,
Franssen
EJ.
Performance of three point-of-care urinalysis test devices for drugs of abuse and therapeutic drugs applied in the emergency department
.
J Emerg Med
.
2012
;
42
(6)
:
682
691
.
79.
Tomaszewski
C,
Runge
J,
Gibbs
M,
Colucciello
S,
Price
M.
Evaluation of a rapid bedside toxicology screen in patients suspected of drug toxicity
.
J Emerg Med
.
2005
;
28
(4)
:
389
394
.
80.
Zhang
Y,
Kwong
TC.
Utilization management in toxicology
.
Clin Chim Acta
.
2014
;
427
:
158
166
.
81.
Wiencek
JR,
Colby
JM,
Nichols
JH.
Rapid assessment of drugs of abuse
.
Adv Clin Chem
.
2017
;
80
:
193
225
.
82.
Wang
LL,
Ren
XX,
He
Y,
Cui
GF,
Wei
ZW,
Yun
KM.
An experimental pharmacokinetics study of diazepam and its metabolites in oral fluid of Chinese population
[published online
January
21,
2020]
.
J Anal Toxicol. doi:10.1093/jat/bkz101
83.
Scherer
JN,
Fiorentin
TR,
Borille
BT,
et al
Reliability of point-of-collection testing devices for drugs of abuse in oral fluid: a systematic review and meta-analysis
.
J Pharm Biomed Anal
.
2017
;
143
:
77
85
.
84.
Wagmann
L,
Maurer
HH.
Bioanalytical methods for new psychoactive substances
.
Handb Exp Pharmacol
.
2018
;
252
:
413
439
.
85.
Nichols
JH,
Christenson
RH,
Clarke
W,
et al
Executive summary: the National Academy of Clinical Biochemistry Laboratory Medicine Practice Guideline: evidence-based practice for point-of-care testing
.
Clin Chim Acta
.
2007
;
379
(1–2)
:
14
28
;
discussion 29–30.
86.
Zeserson
E,
Goodgame
B,
Hess
JD,
et al
Correlation of venous blood gas and pulse oximetry with arterial blood gas in the undifferentiated critically ill patient
.
J Intensive Care Med
.
2018
;
33
(3)
:
176
181
.
87.
Schrauben
SJ,
Negoianu
D,
Costa
C,
et al
Accuracy of acid-base diagnoses using the central venous blood gas in the medical intensive care unit
.
Nephron
.
2018
;
139
(4)
:
293
298
.
88.
Ekstrom
M,
Engblom
A,
Ilic
A,
Holthius
N,
Nordstrom
P,
Vaara
I.
Calculated arterial blood gas values from a venous sample and pulse oximetry: clinical validation
.
PLoS One
.
2019
;
14
(4)
:
e0215413
.
89.
Mirzazadeh
M,
Morovat
A,
James
T,
Smith
I,
Kirby
J,
Shine
B.
Point-of-care testing of electrolytes and calcium using blood gas analysers: it is time we trusted the results
.
Emerg Med J
.
2016
;
33
(3)
:
181
186
.
90.
Triplett
KE,
Wibrow
BA,
Norman
R,
et al
Can the blood gas analyser results be believed? A prospective multicentre study comparing haemoglobin, sodium and potassium measurements by blood gas analysers and laboratory auto-analysers
.
Anaesth Intensive Care
.
2019
;
47
(2)
:
120
127
.
91.
Gibbons
M,
Klim
S,
Mantzaris
A,
Dillon
O,
Kelly
AM.
How closely do blood gas electrolytes and haemoglobin agree with serum values in adult emergency department patients: an observational study
.
Emerg Med Australas
.
2019
;
31
(2)
:
241
246
.
92.
Kankaanpaa
M,
Holma-Eriksson
M,
Kapanen
S,
et al
Comparison of the use of comprehensive point-of-care test panel to conventional laboratory process in emergency department
.
BMC Emerg Med
.
2018
;
18
(1)
:
43
.
93.
Van den Bruel
A,
Thompson
MJ,
Haj-Hassan
T,
et al
Diagnostic value of laboratory tests in identifying serious infections in febrile children: systematic review
.
BMJ
.
2011
;
342:d3082.
94.
Chaisson
LH,
Semitala
FC,
Asege
L,
et al
Point-of-care C-reactive protein and risk of early mortality among adults initiating antiretroviral therapy
.
AIDS
.
2019
;
33
(5)
:
895
902
.
95.
Brouwer
N,
van Pelt
J.
Validation and evaluation of eight commercially available point of care CRP methods
.
Clin Chim Acta
.
2015
;
439
:
195
201
.
96.
Na
JY,
Park
JT.
Evaluation of point-of-care testing for C-reactive protein in forensic medicine
.
J Korean Med Sci
.
2019
;
34
(38)
:
e246
.
97.
Thaler
M,
Luppa
PB.
Highly sensitive immunodiagnostics at the point of care employing alternative recognition elements and smartphones: hype, trend, or revolution?
Anal Bioanal Chem
.
2019
;
411
(29)
:
7623
7635
.
98.
Bulterys
MA,
Wagner
B,
Redard-Jacot
M,
et al
Point-of-care urine LAM tests for tuberculosis diagnosis: a status update [published online December 31, 2019]
.
J Clin Med
.
2020
;
9
(1)
:
111
.
99.
Kim
H,
Chung
DR,
Kang
M.
A new point-of-care test for the diagnosis of infectious diseases based on multiplex lateral flow immunoassays
.
Analyst
.
2019
;
144
(8)
:
2460
2466
.
100.
Qin
Q,
Wang
K,
Yang
J,
et al
Algorithms for immunochromatographic assay: review and impact on future application
.
Analyst
.
2019
;
144
(19)
:
5659
5676
.
101.
Egilmezer
E,
Walker
GJ,
Bakthavathsalam
P,
et al
Systematic review of the impact of point-of-care testing for influenza on the outcomes of patients with acute respiratory tract infection
.
Rev Med Virol
.
2018
;
28
(5)
:
e1995
.
102.
Popowitch
EB,
Miller
MB.
Performance characteristics of Xpert Flu/RSV XC assay
.
J Clin Microbiol
.
2015
;
53
(8)
:
2720
2721
.
103.
Wahrenbrock
MG,
Matushek
S,
Boonlayangoor
S,
Tesic
V,
Beavis
KG,
Charnot-Katsikas
A.
Comparison of Cepheid Xpert Flu/RSV XC and BioFire FilmArray for detection of influenza A, influenza B, and respiratory syncytial virus
.
J Clin Microbiol
.
2016
;
54
(7)
:
1902
1903
.
104.
Nitsch-Osuch
A,
Kuchar
E,
Golebiak
I,
Kanecki
K,
Tarka
P,
Brydak
LB.
Rapid influenza diagnostic tests improve suitability of antiviral treatment in hospitalized children
.
Adv Exp Med Biol
.
2017
;
968
:
1
6
.
105.
Brendish
NJ,
Malachira
AK,
Armstrong
L,
et al
Routine molecular point-of-care testing for respiratory viruses in adults presenting to hospital with acute respiratory illness (ResPOC): a pragmatic, open-label, randomised controlled trial
.
Lancet Respir Med
.
2017
;
5
(5)
:
401
411
.
106.
Li-Kim-Moy
J,
Dastouri
F,
Rashid
H,
et al
Utility of early influenza diagnosis through point-of-care testing in children presenting to an emergency department
.
J Paediatr Child Health
.
2016
;
52
(4)
:
422
429
.
107.
Brendish
NJ,
Schiff
HF,
Clark
TW.
Point-of-care testing for respiratory viruses in adults: the current landscape and future potential
.
J Infect
.
2015
;
71
(5)
:
501
510
.
108.
Hagen
A,
Eichinger
A,
Meyer-Buehn
M,
Schober
T,
Huebner
J.
Comparison of antibiotic and acyclovir usage before and after the implementation of an on-site FilmArray meningitis/encephalitis panel in an academic tertiary pediatric hospital: a retrospective observational study
.
BMC Pediatr
.
2020
;
20
(1)
:
56
.
109.
Brendish
NJ,
Mills
S,
Ewings
S,
Clark
TW.
Impact of point-of-care testing for respiratory viruses on antibiotic use in adults with exacerbation of airways disease
.
J Infect
.
2019
;
79
(4)
:
357
362
.
110.
Chavez
PR,
Bradley
HM,
Wesolowski
LG,
et al
Performance evaluation of four point-of-care HIV tests using unprocessed specimens
.
J Clin Virol
.
2020
;
124
:
104282
.
111.
Pai
NP,
Tulsky
JP,
Cohan
D,
Colford
JM
Jr,
Reingold
AL.
Rapid point-of-care HIV testing in pregnant women: a systematic review and meta-analysis
.
Trop Med Int Health
.
2007
;
12
(2)
:
162
173
.
112.
Dalcin
D,
Bogoch II. Point-of-care testing for HIV
.
CMAJ
.
2018
;
190
(18)
:
E569
.
113.
Gliddon
HD,
Peeling
RW,
Kamb
ML,
Toskin
I,
Wi
TE,
Taylor
MM.
A systematic review and meta-analysis of studies evaluating the performance and operational characteristics of dual point-of-care tests for HIV and syphilis
.
Sex Transm Infect
.
2017
;
93
(S4)
:
S3
S15
.
114.
Negedu-Momoh
OR,
Jegede
FE,
Yakubu
A,
et al
Performance evaluation of BD FACSPresto point of care CD4 analyzer to enumerate CD4 counts for monitoring HIV infected individuals in Nigeria
.
PLoS One
.
2017
;
12
(5)
:
e0178037
.
115.
Moran
Z,
Sacks
JA,
Frimpong
FK,
Frimpong
AB,
Ben Amor
Y.
Performance of the BD-FACS Presto for CD4 count and hemoglobin measurement in a district hospital and rural laboratory in Ghana
.
PLoS One
.
2019
;
14
(2)
:
e0212684
.
116.
Bwana
P,
Vojnov
L,
Adhiambo
M,
et al
The BD FACSPresto point of care CD4 test accurately enumerates CD4+ T cell counts
.
PLoS One
.
2015
;
10
(12)
:
e0145586
.
117.
Hong
KH,
Lee
SW,
Kim
TS,
et al
Guidelines for laboratory diagnosis of coronavirus disease 2019 (COVID-19) in Korea
.
Ann Lab Med
.
2020
;
40
(5)
:
351
360
.
118.
Long
QX,
Liu
BZ,
Deng
HJ,
et al
Antibody responses to SARS-CoV-2 in patients with COVID-19
.
Nat Med
.
2020
;
26
(6)
:
845
848
.
119.
Imai
K,
Tabata
S,
Ikeda
M,
et al
Clinical evaluation of an immunochromatographic IgM/IgG antibody assay and chest computed tomography for the diagnosis of COVID-19
.
J Clin Virol
.
2020
;
128
:
104393
.
120.
Brendish
NJ,
Malachira
AK,
Clark
TW.
Molecular point-of-care testing for respiratory viruses versus routine clinical care in adults with acute respiratory illness presenting to secondary care: a pragmatic randomised controlled trial protocol (ResPOC)
.
BMC Infect Dis
.
2017
;
17
(1)
:
128
.

Competing Interests

The author has no relevant financial interest in the products or companies described in this article.