Context.—

The prothrombin time (PT) and activated partial thromboplastin time (APTT) are screening tests used to detect congenital or acquired bleeding disorders. An unexpected PT and/or APTT prolongation is often evaluated using a mixing test with normal plasma. Failure to correct (“noncorrection”) prolongation upon mixing is attributed to an inhibitor, whereas “correction” points to factor deficiency(ies).

Objective.—

To define an optimal method for determining correction or noncorrection of plasma mixing tests through an international, multisite study that used multiple PT and APTT reagents and well-characterized plasma samples.

Design.—

Each testing site was provided 22 abnormal and 25 normal donor plasma samples, and mixing studies were performed using local PT and APTT reagents. Mixing study results were evaluated using 11 different calculation methods to assess the optimal method based on the expected interpretation for factor deficiencies (correction) and noncorrection (inhibitor effect). Misprediction, which represents the failure of a mixing study interpretation method, was assessed.

Results.—

Percentage correction was the most suitable calculation method for interpreting PT mixing test results for nearly all reagents evaluated. Incubated PT mixing tests should not be performed. For APTT mixing tests, percentage correction should be performed, and if the result indicates a factor deficiency, this should be confirmed with the subtraction III calculation where the normal pooled plasma result (run concurrently) is subtracted from the mixing test result with correction indicated by a result of 0 or less. In general, other calculation methods evaluated that performed well in the identification of factor deficiency tended to have high misprediction rates for inhibitors and vice versa.

Conclusions.—

No single method of mixing test result calculation was consistently successful in accurately distinguishing factor deficiencies from inhibitors, with between-reagent and between-site variability also identified.

Prothrombin time (PT) and activated partial thromboplastin time (APTT) are screening tests used worldwide for the evaluation of patients suspected of congenital or acquired bleeding disorders; they are also used for monitoring purposes (eg, heparin or oral vitamin K antagonists [OVKAs] or replacement therapies) and in the identification of a lupus anticoagulant (LAC).1,2  An unexpected prolongation of the PT and/or APTT can be attributed to a number of causes, including isolated or multiple factor deficiencies or the presence of an inhibitor.3,4  A potentially useful screening mechanism to identify the cause of a prolonged PT or APTT is the mixing test, which is simply repeating the prolonged screening test after mixing the sample with normal pooled plasma (NPP). Failure to correct a prolonged PT or APTT upon mixing is often attributed to an inhibitor, whereas conversely, correction of the prolongation is attributed to single or multiple factor deficiencies.5–7  The ability to differentiate between a factor deficiency and inhibitor will likely direct further evaluation and treatment of the patient.8 

The traditional mixing test is performed by adding patient plasma to NPP at equal volumes/ratio.7  Deviations from this have been described, including using 4 parts patient plasma to 1 part NPP or the addition of an incubation step at 37°C. Although the mixing test has been used for decades, standardization of mixing test performance, as well as methods of calculating results, is lacking.

Plasma mixing tests should be considered in the investigation of select patients with unexplained prolongation of PT and/or APTT. Guidance has been provided for mixing test use in patients suspected of having a LAC as part of the diagnostic algorithm.9–11  Where available, patient history, including recent drug exposure, must be reviewed prior to initiating testing, as this information could be useful for result interpretation. Mixing tests should not be performed on samples with normal baseline PT and/or APTT results, nor should they be performed in patients with known exposure to anticoagulant agents.

A recent international study of 990 laboratories revealed large variations in mixing test methodologies, with only 49% of responding participants providing accurate interpretation of the mixing test results.12  This high percentage of inaccurate result reporting likely reflects the lack of standardization for both determining what constitutes a corrected mixing test result and the method for determining correction. As such, in 2019 the International Council for Standardization in Haematology (ICSH) agreed to provide funding for an international, multisite study to evaluate PT and APTT mixing tests using well-characterized plasma samples with the purpose of determining whether there is an optimal method for determining correction or noncorrection. In this field study, 11 different, previously described mixing test calculation methods to ascertain correction or noncorrection using a variety of PT and APTT reagent platforms were compared.

This study was funded and supported by ICSH such that the test samples and NPP were provided at no cost to participating clinical laboratory sites. Buffered citrated plasma samples used in this study were prepared, stored, and shipped by a single-source commercial vendor (Precision BioLogic, Halifax, Nova Scotia, Canada). The ICSH mixing test kit consisted of frozen plasma sets including 22 abnormal samples, 25 normal donor samples, and NPP. Kits were shipped on dry ice for overnight delivery to each selected laboratory site. The characteristics of each abnormal sample are listed in Table 1. Normal donors consisted of 12 women (median age, 37 years; range, 22–50 years) and 13 men (median age, 40 years; range, 23–64 years). The NPP was collected from a minimum of 20 ostensibly healthy individuals with no evidence of hemostatic disease, then buffered with hydroxyethylpiperazine ethane sulfonic acid buffer. Single-factor–deficient plasmas were prepared by immunoadsorption, with each preparation quality tested to confirm normal and decreased factor activity levels as appropriate (Supplemental Table 1; see supplemental digital content containing 3 tables at https://meridian.allenpress.com/aplm in the August 2024 table of contents). Combined-factor–deficient plasma samples were prepared using either adsorbed plasma or plasma from patients on OVKA. Individual inhibitor-type anticoagulants were added to pooled normal plasma to achieve concentrations consistent with therapeutic or “on-therapy” concentrations.13,14  Factor VIII inhibitor and LAC plasma samples were from well-characterized human donor sources (Supplemental Table 2).

Table 1.

Abnormal Sample Characteristics and Expected Mixing Study Results

Abnormal Sample Characteristics and Expected Mixing Study Results
Abnormal Sample Characteristics and Expected Mixing Study Results

All prepared plasma samples were quality tested for accuracy of sample characterization, aliquoted, rapidly frozen, and stored at −40°C to −80°C. Confirmation of kit receipt and sample status (residual dry ice present, samples maintained in frozen state) was made to the shipper and ICSH study investigators.

Twenty clinical laboratories that used the most common instrument/reagent platforms based on external quality assurance programs were sought for participation. The number of clinical laboratories using the same PT or APTT reagent platform was limited to 6. Testing sites had the option to use more than one reagent platform; however, a limiting factor was the provided test sample volume (1.0 mL). Laboratories using in-house or laboratory-developed PT or APTT methods were excluded from participation. In vitro diagnostic companies that provide PT and APTT reagents were separately invited to participate, although their participation required the ICSH mixing study kits to be purchased.

Each testing site was provided specific instructions for thawing the samples and performing PT and APTT mixing tests and was encouraged to perform such studies during a series of days using 2 or more technicians. For each mixing study run, laboratories were instructed to test the NPP as a stand-alone sample, as this was necessary for select calculations (Table 2). A clotting time that exceeded the upper limit of the reference interval (RI) provided by the testing site was considered an abnormal baseline result. After completion of baseline values, each sample (abnormal samples and normal donor samples) was mixed with NPP at a 1:1 ratio (immediate phase) and retested for PT and APTT. Mixing tests were performed on all samples, including those from normal donors, regardless of initial (baseline) values.

Table 2.

Calculation Methods Used

Calculation Methods Used
Calculation Methods Used

After completion of the immediate phase testing, capped mixed samples were incubated in a 37°C water bath for 1 hour (incubated phase) and screening PT and APTT were performed.

Each testing site provided its PT and APTT reagent source, manufacturer, local RIs, and local mechanism for interpreting mixing test results (Supplemental Table 3). Each testing site was provided a spreadsheet for result recording, and a single investigator performed the mixing study calculations (Table 2). Each testing site was blinded to other sites’ participation and data.

Although for the purposes of this study all abnormal and normal samples underwent mixing studies, only baseline samples that were outside the upper end of the local RI were analyzed for interpretations. The RIs used in determining correction and noncorrection were based on 2 different determinations: (1) locally provided RI and (2) RI calculated using the mean + 2 SDs derived from the provided 25 normal donor plasmas mixed with the NPP. Three different subtraction methods were used. Subtraction I evaluated the mixing study result to the local upper RI, where mixing results that fell below were considered correction and above the upper RI noncorrection. Subtraction II compared mixing study results with upper RI, but the upper RI was determined using the study donor set and NPP result (mean + 2 SDs). Subtraction III compared the mixing study result with the NPP result from the run. Three mixing ratio determinations were also calculated: mixing study result divided by local RI mean, mixing study result divided by study donors + NPP mean, and mixing study result divided by NPP value obtained during the run. Samples that demonstrated any obvious operational or instrument errors were excluded from analysis. Samples that exceeded the upper limit for clotting time detection (eg, >90 seconds) were included in analysis and given a 1-second addition to threshold (eg, 91 seconds for >90 seconds reported value).

Test samples consisting of multiple or single factor deficiencies or OVKA-induced factor deficiencies were expected to demonstrate correction following NPP mix (Table 1). The expected mixing test result interpretation was determined to be noncorrected for those samples with baseline prolongation due to inhibitory anticoagulants and specific or nonspecific factor inhibitors.

To determine whether a reagent performed as expected, the concept of misprediction was used to analyze the pooled data set. Misprediction occurs when the reported mixing test interpretation or baseline PT and/or APTT result does not match the expected baseline result or mixing test interpretation. In addition, analysis was grouped by specific reagent if 3 or more participants used the same reagent(s). Misprediction was calculated for all baseline, immediate, and incubated mixing results. Multiple rationales for determining misprediction according to the parameters listed in Table 1 were used. First was that isolated or combined factor deficiencies are expected to correct after mixing with NPP. A second rationale for unfractionated heparin and low-molecular-weight heparin therapies is that the PT is expected to be normal, as commercial reagents contain heparin neutralizers. The APTT is expected to be prolonged in a sample containing therapeutic levels of unfractionated heparin, but with LWMH the APTT may be normal or elevated, depending on the reagent’s sensitivity to low-molecular-weight heparin. A prolonged baseline APTT due to heparin effect is not expected to correct when performing mixing tests, but correction may occur in samples containing lower drug concentrations. A third rationale is that although it is well recognized that PT and/or APTT reagent sensitivity to direct oral anticoagulants (DOACs) (rivaroxaban, apixaban, edoxaban, and dabigatran) varies widely between platforms,13,15,16  the desired or expected effect on the baseline PT and APTT for “on-therapy” drug concentrations would be a prolonged clotting time that does not correct with mixing studies.13,14  A fourth rationale used is that factor VIII inhibitor samples will have a prolonged baseline APTT (and normal PT), which may or may not correct in immediate mixing test results but will prolong with incubated mix given that these antibodies tend to demonstrate time and temperature dependence.17  Finally, as with DOACs and heparins, there is known between-APTT reagent variation in sensitivity to the presence of LAC.18,19  For this reason, it has been suggested that LAC-insensitive reagents should be used for general (ie, not LAC) screening purposes.18,19  The LAC interpretation data were assessed only on those samples with a reported baseline APTT prolongation, and the diagnostic criterion for LAC determination is noncorrection of a mixing study indicating an inhibitor.19,20  The PT is rarely prolonged in patients with LAC, so a normal PT is expected in these samples.11  As only prolonged baseline results were evaluated, mixing study misprediction in LAC samples would be for those methods that indicated correction.

There were 20 clinical laboratories and 4 reagent manufacturers (Diagnostica Stago, Roche Diagnostics, Siemens Healthcare Diagnostics, and TCoag) that agreed to participate. Eight clinical sites were in the United States, 4 in Australia, 3 in the United Kingdom, 2 in Spain, and 1 each in Ireland, Switzerland, and Malta. Because of logistical issues with shipping, the clinical site in Malta was not able to complete the sample testing. One Australian clinical laboratory, because of staffing limitations secondary to the SARS-CoV-2 pandemic, was unable to perform sample testing.

Two clinical testing sites provided more than one reagent for APTT, and one provided more than one reagent for PT, whereas all industry participants provided more than one reagent for APTT and 2 of 4 industry participants provided more than 1 PT reagent for testing. Combining both clinical laboratory and industry data, there were 28 PT reagent results sets and 34 APTT reagent result sets used for evaluation. Different reagent lots were used between testing sites.

Of the clinical laboratory participants providing local guidance for mixing test correction, 7 of 18 (39%) indicated a mixing study correction if results returned to the RI, 2 of 18 (11%) if results returned to within 2 to 5 seconds of the upper limit of the RI, 2 of 18 (11%) using index of correction (Rosner index),20  2 of 18 (11%) using a percentage correction (or Chang score),21  and 5 of 18 (28%) using other strategies, including ratios.22 

PT Baseline and Mixing Test Results

For PT testing, baseline misprediction was based on an unexpected prolongation of the clotting time and occurred in 9 of 28 factor VIII–deficient and low-titer factor VIII inhibitor samples (32%) and 11 of 28 for the LAC samples (39%) (Table 3). For DOACs, baseline misprediction was based on a normal PT for 16 of 28 apixaban samples (57%), 1 of 27 edoxaban samples (4%), and 1 of 27 dabigatran samples (4%).

Table 3.

Prothrombin Time (PT) Baseline and Immediate Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa

Prothrombin Time (PT) Baseline and Immediate Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa
Prothrombin Time (PT) Baseline and Immediate Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa

For immediate or incubated phase mixing test results, there was no interpretation method that correctly classified all samples (Tables 3 and 4). However, for percentage correction in the immediate PT mixing interpretation, there were single-laboratory mispredictions for one combined deficiency sample, the apixaban and edoxaban sample (Table 3). Subtraction method I, the most commonly performed method of interpretation by study participants, performed poorly for both deficient and inhibitor samples (Tables 3 and 4). The other 2 subtraction methods had decreased misprediction rates for inhibitors, but yielded nearly 100% misprediction for single or multiple factor deficiencies. The 3 ratio methods performed similarly to one another. In general, using a ratio greater than 1.2 improved the performance for factor deficiency interpretations but caused performance in the interpretation of factor inhibitors to worsen (Tables 3 and 4). The percentage correction interpretation demonstrated the most optimal method for interpreting immediate-phase PT mixing tests results. This was also apparent when 5 different PT reagents were evaluated individually, as there were no mispredictions in immediate-phase mixing results for both deficient and inhibitor samples, except for one reagent in the presence of edoxaban (Table 5). The Rosner index performed well in the correct interpretation of factor deficiencies, but there were frequent misinterpretations in the DOAC inhibitor samples. This was also apparent when 5 different PT reagents were evaluated individually, with the exception of one OVKA sample and one reagent where the Rosner index had frequent mispredictions in the DOAC samples. Misprediction rates markedly increased for all inhibitor sample results following incubation (Tables 4 through 6).

Table 4

Prothrombin Time Incubated Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa

Prothrombin Time Incubated Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa
Prothrombin Time Incubated Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa
Table 5.

Interpretation Misprediction Values for Respective Prothrombin Time Reagents Using Percentage Correction Method

Interpretation Misprediction Values for Respective Prothrombin Time Reagents Using Percentage Correction Method
Interpretation Misprediction Values for Respective Prothrombin Time Reagents Using Percentage Correction Method
Table 6.

Interpretation Misprediction Values for Respective Prothrombin Time Reagents Using Rosner Index Method

Interpretation Misprediction Values for Respective Prothrombin Time Reagents Using Rosner Index Method
Interpretation Misprediction Values for Respective Prothrombin Time Reagents Using Rosner Index Method

APTT Baseline and Mixing Test Results

For APTT results, baseline misprediction was based on an unexpected prolongation of the clotting time and occurred in 7 of 34 factor VII–deficient plasma samples (20%) and 17 of 34 apixaban samples (50%) (Table 7). For samples with expected prolongation of the baseline APTT, there were normal clotting times for 5 of 34 edoxaban samples (15%) and 4 of 34 (12%) and 9 of 34 (26%) strong and weak LAC samples, respectively.

Table 7.

Activated Partial Thromboplastin Time Baseline and Immediate Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa

Activated Partial Thromboplastin Time Baseline and Immediate Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa
Activated Partial Thromboplastin Time Baseline and Immediate Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa

For immediate- or incubated-phase mixing tests there was no interpretation method that classified all samples correctly. Results for immediate APTT mixing mirror PT mixing results. Subtraction method I, the most commonly performed method of interpretation by study participants, performed poorly for both deficient and inhibitor samples (Tables 7 and 8). The other 2 subtraction methods had decreased misprediction rates for inhibitors but yielded nearly 100% misprediction for single or multiple factor deficiencies. The 3 ratio methods performed similarly. In general, using a ratio greater than 1.2 improved performance for factor deficiency interpretations but caused performance in the interpretation of factor inhibitors to worsen. All methods except subtraction II and III failed to detect a weak factor VIII inhibitor in the immediate mix, with some improvement following incubation (Tables 7 and 8). The percentage correction and Rosner index had nearly 100% misprediction for the weak FVIII inhibitor in both immediate and incubated mix results. Although the percentage correction and Rosner index performed well in correctly identifying factor deficient plasmas, there were frequent mispredictions in the inhibitor samples. Only subtraction method III correctly classified all inhibitor test samples, but it had nearly 100% misprediction for factor deficiency samples. Heparinized LAC and DOAC samples yielded misprediction interpretations for mixing study results for various APTT reagents using both percentage correction and the Rosner index. This is not surprising given the varied reagent responsiveness to heparin, DOACs, and LACs (Tables 9 and 10), but it should be noted that mispredictions for percentage correction on the incubated phase, with the exception of the weak factor VIII inhibitor, were mostly confined to a single laboratory (Table 9) for any given reagent.

Table 8.

Activated Partial Thromboplastin Time Incubated Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa

Activated Partial Thromboplastin Time Incubated Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa
Activated Partial Thromboplastin Time Incubated Phase Mixing Study Results: Misprediction Values for Abnormal Samplesa
Table 9.

Interpretation Misprediction Values for Respective Activated Partial Thromboplastin Time Reagents Using Percentage Correction Methoda

Interpretation Misprediction Values for Respective Activated Partial Thromboplastin Time Reagents Using Percentage Correction Methoda
Interpretation Misprediction Values for Respective Activated Partial Thromboplastin Time Reagents Using Percentage Correction Methoda
Table 10.

Interpretation Misprediction Values for Respective Activated Partial Thromboplastin Time Reagents Using Rosner Index Method

Interpretation Misprediction Values for Respective Activated Partial Thromboplastin Time Reagents Using Rosner Index Method
Interpretation Misprediction Values for Respective Activated Partial Thromboplastin Time Reagents Using Rosner Index Method

Although it is generally appreciated that PT and APTT reagents vary in their sensitivities to factor levels, heparins, DOACs, and LACs, it requires emphasis that in this study, misprediction was based on the inability for a mixing study interpretation to correctly identify a deficiency or inhibitor in a sample that demonstrated a prolonged baseline clotting time. Samples that did not demonstrate a baseline prolonged clotting time were not subject to mixing study evaluations. Factors that contributed to mixing study misprediction included reagent sensitivity, local RI determination, and inappropriate reagent use (ie, using a LAC-sensitive reagent for screening rather than a LAC-insensitive reagent, which would be more appropriate), further emphasizing the need for appropriate RI determination and local laboratory validation of its reagent sensitivity and mixing study interpretive methods.

This study revealed that no single mixing test interpretation method, regardless of reagent used, demonstrated zero misprediction in all samples. Calculation methods tended to perform well in the correct identification of factor deficiency or inhibitor, but not both. This suggests that a staggered approach in performing calculations may be optimal, especially for the more common APTT mixing study. Although the percentage correction method demonstrated the most accurate interpretation for PT mixing tests, for APTT mixing tests this interpretation method would be optimized using the subtraction III method when the percentage correction interpretation suggests a factor deficiency. This algorithm rules out a factor inhibitor. The calculation method reportedly used by the majority of laboratories participating in this study, specifically correction of the mix into the RI, performed poorly in correctly categorizing both PT and APTT factor-deficient and inhibitor samples (Table 7).

This study encompasses multiple reagent platforms using a wide selection of abnormal samples, in contrast to other publications that focused on mixing studies for LAC assessment,11,23  only addressed APTT and not PT prolongations,22  or used only contrived samples that may mimic coagulation abnormalities.8  In a large-single site retrospective study of APTT mixing studies using a single platform, Liu and colleagues24  evaluated modified thresholds for different interpretative methods, but it is unclear whether these same thresholds could be applicable to other platforms. In the present study, these same thresholds were evaluated using multiple platforms. Results from this study provide additional clinical and laboratory guidance as to the use of PT mixing studies, which are often neglected given the rarity of inhibitors to extrinsic and common pathway factors.25–27  With increased use of anti–factor Xa DOACs, there may be increased demand for PT mixing studies in acute settings in patients with unknown drug exposure.7,28 

A surprising number of mispredictions were reported for baseline sample results. Factor VIII–deficient and low-titer factor VIII inhibitor (0.6 Bethesda units) samples should yield a normal PT, yet 32% (9 of 28) reported a prolonged PT for those sample types. Similar to the expectation that NPPs used in mixing tests should contain more than 80 IU/dL of each coagulation factor,6,8  manufacturer-provided information for the deficient plasmas used in this study indicated that all levels other than intended deficiencies were 50 IU/dL activity or more, and thus a deficiency or inhibitor of a factor other than factor VIII cannot explain this misprediction. The basis of this high misprediction rate requires further investigation. The high degree of baseline misprediction with DOAC was not unexpected, as it is well known that APTT and PT reagents vary in their responsiveness to these agents.13,15,16,29  For APTT reagents, there were also varying degrees of baseline misprediction between reagent platforms, as some reagents were identified as LAC sensitive and others not. The misprediction in baseline results for weak and strong LAC of 12% (4 of 34) and 29% (9 of 34), respectively, emphasizes the need to use APTT reagents that are sensitive to LAC when screening for the presence of these antibodies or laboratories should seek other testing schemes.9,10  An unexpected prolongation of the PT and APTT in these otherwise normal samples could be due to inappropriate methodologies used for local reference range generation, yet these seemingly abnormal results may lead to erroneous diagnosis, unnecessary laboratory investigation, unnecessary delays in treatment, or patient mismanagement.30  Previous studies have also demonstrated such differences between reagent platforms and the varied factor sensitivities of both PT and APTT reagents.31,32 

The best calculation method for PT mixing tests was the percentage correction followed by the Rosner index. The Rosner index (also known as the index of circulating anticoagulant), performed poorly in the identification of inhibitor samples, for which it was originally designed. Misprediction rates dramatically increased when incubated-phase calculations were used; possible contributors could be partial loss of labile factors and pH drift. These data suggest that only immediate-phase and not incubated PT mixing tests should be performed, as incubation tends to result in misprediction of inhibitor samples.33–35  The authors are unaware of the existence of time- and temperature-dependent antibodies requiring incubated PT mixing studies.

This study demonstrates that for APTT-based mixing tests, mixing study subtraction calculation methods that correctly identified inhibitors had high misprediction rates in identifying factor deficiencies and vice versa. Subtraction method III correctly identified all inhibitor samples but performed poorly on samples with single or multiple factor deficiencies. Calculations using the percentage correction or Rosner index showed better performance in factor-deficient samples than any of the subtraction- or ratio-based calculations evaluated. Both percentage correction and Rosner index, however, did not perform correctly for all samples, and surprisingly, both methods performed poorly in identifying the weak FVIII inhibitor sample as an inhibitor in both the immediate and incubated mixing studies. When evaluating individual APTT reagents (Tables 9 and 10), the percentage correction method demonstrated fewer mispredictions than the Rosner index method. Further investigation is needed to determine if using locally derived cutoffs instead of standard 1.1 and 1.2 cutoffs for the ratio methods could improve assay performance, particularly if these cutoffs are derived from normal donors mixed with the same NPP in local use.11 

In this study, multiple laboratories, regardless of the APTT reagent used, had difficulty identifying the weak factor VIII inhibitor, even though all sites demonstrated a prolonged baseline APTT in this sample. This could be because a 1-hour incubation may be insufficient for detecting weak factor VIII inhibitors. For other factor inhibitors studied, the incubated mixing test (versus the immediate) substantially improved inhibitor detection using the subtraction and ratio methods. In patients with suspected factor VIII inhibitors, because of its potential serious consequences, additional studies, such as performance of factor VIII activity levels in addition to the performance of an incubated mixing study, should be considered.

Given the variability in misprediction rates between calculation methods demonstrated in this study, validation, or, at a minimum, verification by individual laboratories of the accuracy of the mixing calculation used is paramount. This should be conducted using a variety of deficient and inhibitor sample types. Furthermore, in those situations where a complete patient history including medications is not available, more selective laboratory methods to screen for anticoagulant presence are recommended, as mixing studies should not be performed when inhibitor-type anticoagulants are present.

The strength of this study is the multinational clinical laboratory and industry participation, which represents multiple reagent testing platforms, variable RIs (even when using the same reagent/instrument system), and differing lots within the same reagent platforms. Study weaknesses include that some samples used were considered surrogate or contrived samples, which is a commonly used method for assessing reagent performance but may not have the same testing characteristics as patient samples. This study used a 1-hour incubation period, which may not be sufficient to detect weak factor VIII inhibitors, and further studies are warranted to elucidate the ideal incubation time for detecting these weak titer inhibitors. Additionally, only a single source and type (buffered) of NPP was used; it may differ from the NPP source and characteristics used in many laboratories, which may influence misprediction rates. Hence the importance of laboratories validating their mix test interpretations under local test conditions.

The APTT and to a lesser degree the PT mixing tests have long been a diagnostic staple for clinicians as a guide to suggest whether prolongation of the clotting time is due to factor deficiency versus an inhibitor. Standardization of mixing test performance as well as limited evidence for accurate interpretative methods is lacking. Although limited result variation between testing sites reportedly using the same reagent platforms and interpretative methods was evident, the following conclusions are notable. Testing of samples with normal baseline PT or APTT results and in patients with known inhibitory anticoagulants, such as DOACs or heparins, should be avoided. The PT mixing test is optimal for the immediate phase only, as incubated PT mixing studies dramatically increased misprediction. This study suggests that no single mixing study interpretation method, regardless of reagents used, demonstrates zero misprediction in all samples tested. For the PT and APTT mixing tests, the percentage correction was the most suitable interpretation method, but APTT studies may be further optimized when used in conjunction with a subtraction method to the NPP value used in the percentage correction calculation. As differences in misprediction between laboratories using the same PT and APTT reagents was evident, each laboratory should validate or at a minimum verify the performance of their mixing study interpretive method(s) using well-characterized samples expected to yield corrected and noncorrected interpretations, before the results of mixing tests are reported clinically. Given that no single approach for mixing study interpretations had zero misprediction, clinical correlation with laboratory findings is required.

1.
Johnsen
JM,
Konkle
BA.
Differential diagnosis of the bleeding patient. In:
Marder
VJ,
Aird
WC,
Bennett
JS,
Schulman
S,
White
GC
, eds.
Haemostasis and Thrombosis: Basic Principles and Clinical Practice
. 6th ed.
Philadelphia, PA
:
Lippincott Williams & Wilkins
;
2014
:
648
658
.
2.
Bethel
M,
Adcock
DM.
Laboratory evaluation of prolonged APTT and PT
.
Lab Med
.
2004
;
35
(
5
):
285
291
.
3.
Adcock
DM,
Moore
GW,
Montalvão
SL,
Kershaw
G,
Gosselin
RC.
Activated partial thromboplastin time and prothrombin time mixing studies: current state of the art [published online September 2, 2022]
.
Semin Thromb Hemost
.
4.
Adcock
DM,
Poirier
BF.
Screening coagulation assays, factor XIII and d-dimer. In:
Teruya
J
, eds.
Management of Bleeding Patients
.
Cham, Switzerland
:
Springer International Publishing
;
2016
:
3
16
.
5.
National Committee for Clinical Laboratory Standards.
One-stage prothrombin time (PT) and activated partial thromboplastin time (APTT) test; approved guidelines
.
Wayne, PA
:
NCCLS
;
1996
.
NCCLS document H47-A
.
6.
Choi
SH,
Rambally
S,
Shen
YM.
Mixing study for the evaluation of abnormal coagulation test
.
JAMA
.
2016
;
316
(
20
);
2146
2147
.
7.
Favaloro
EJ.
Coagulation mixing studies: Utility, algorithmic strategies and limitations for lupus anticoagulant testing or follow up of abnormal coagulation tests
.
Am J Hematol
.
2020
;
95
(
1
):
117
128
.
8.
Favaloro
EJ,
Bonar
R,
Duncan
E,
et al.
Mis-identification of factor inhibitors by diagnostic hemostasis laboratories: recognition of pitfalls and elucidation of strategies: a follow up to a large multicenter evaluation
.
Pathology
.
2007
;
39
:
504
511
.
9.
Devreese
KMJ,
de Groot
PG,
de Laat
B,
et al.
Guidance from the Scientific and Standardization Committee for lupus anticoagulant/antiphospholipid antibodies of the International Society on Thrombosis and Haemostasis: update of the guidelines for lupus anticoagulant detection and interpretation
.
J Thromb Haemost
.
2020
;
18
(
11
):
2828
2839
.
10.
Keeling
D,
Mackie
I,
Moore
GW,
Greer
IA,
Greaves
M;
British Committee for Standardization in Haematology
.
Guidelines on the investigation and management of antiphospholipid syndrome
.
Br J Haematol
.
2012
:
157
:
47
58
.
11.
Moore
GW,
Rangarajan
S,
Holland
LJ,
Henley
A,
Savidge
GF.
Low frequency of elevated prothrombin times in patients with lupus anticoagulants when using a recombinant thromboplastin reagent: implications for dosing and monitoring of oral anticoagulant therapy
.
Br J Biomed Sci
.
2005
;
62
:
15
18
.
12.
Azner
E,
Rogic
D,
Meijer
P,
et al.
;
for joint Working Group on Postanalytical Phase of the European Federation of Clinical Chemistry and Laboratory Medicine and European Organisation for External Quality Assurance Providers in Laboratory Medicine. An international study of how laboratories handle and evaluate patient samples after detecting an unexpected APTT prolongation
.
Clin Chem Lab Med
.
2015
;
53
(
1
):
1593
1603
.
13.
Gosselin
RC,
Adcock
DM,
Bates
SM,
et al.
International Council for Standardization in Haematology (ICSH) recommendations for laboratory measurement of direct oral anticoagulants
.
Thromb Haemost
.
2018
;
118
(
3
):
437
450
.
14.
Douxfils
J,
Gosselin
RC.
Laboratory assessment of direct oral anticoagulants
.
Semin Thromb Hemost
.
2017
;
43
(
3
):
277
290
.
15.
Adcock
DM,
Gosselin
R.
Direct oral anticoagulants (DOACs) in the laboratory: 2015 review
.
Thromb Res
.
2015
;
136
(
1
):
7
12
.
16.
Douxfils
J,
Ageno
W,
Samama
CM,
et al.
Laboratory testing in patients treated with direct oral anticoagulants: a practical guide for clinicians
.
J Thromb Haemost
.
2018
;
16
(
2
):
209
219
.
17.
Adcock
DM,
Favaloro
EJ.
Pearls and pitfalls in factor inhibitor assays
.
Int J Lab Hematol
.
2015
;
37
(
suppl 1
):
52
60
.
18.
Fritsma
GA,
Dembitzer
FR,
Randhawa
A,
et al.
Recommendations for appropriate activated partial thromboplastin time reagent selection and utilization
.
Am J Clin Pathol
.
2012
;
137
(
6
):
904
908
.
19.
Clinical and Laboratory Standards Institute (CLSI)
.
Laboratory testing for the lupus anticoagulant; approved guideline
.
Wayne, PA
:
CLSI
;
2014
.
CLSI H-60
.
20.
Rosner
E,
Pauzner
R,
Lusky
A,
Modan
M,
Many
A.
Detection and quantitative evaluation of lupus circulating anticoagulant activity
.
Thromb Haemost
.
1987
;
57
(
2
):
144
147
.
21.
Chang
SH,
Tillema
V,
Scherr
D.
A “percent correction” formula for evaluation of mixing studies
.
Am J Clin Pathol
.
2002
;
117
(
1
):
62
73
. .
22.
Kershaw
G,
Orellana
D.
Mixing tests: diagnostic aides in the investigation of prolonged prothrombin times and activated partial thromboplastin times
.
Semin Thromb Hemost
.
2013
;
39
(
3
):
283
290
. .
23.
Depreter
B,
Devreese
KM.
Differences in lupus anticoagulant final conclusion through clotting time or Rosner index for mixing test interpretation
.
Clin Chem Lab Med
.
2016
;
54
(
9
):
1511
1516
.
24.
Liu
C,
Ling
L,
Huang
X,
et al.
Evaluation of activated partial thromboplastin time mixing studies using several methods
.
Arch Pathol Lab Med
.
2022
;
146
(
11
):
1387
1394
.
25.
Goulenok
T,
Vasco
C,
Faille
D,
et al.;
RAVI study group
.
Acquired factor V inhibitor: a nation-wide study of 38 patients
.
Br J Haematol
.
2021
;
192
(
5
):
892
899
. .
26.
Mulliez
SM,
Devreese
KM.
Isolated acquired factor VII deficiency: review of the literature
.
Acta Clin Belg
.
2016
;
71
(
2
):
63
70
. .
27.
Hazim
AZ,
Ruan
GJ,
Khodadadi
RB,
Slusser
JP,
Marshall
AL,
Pruthi
RK.
A single-institution retrospective study of causes of prolonged prothrombin time and activated partial thromboplastin time in the outpatient setting
.
Int J Lab Hematol
.
2022
;
44
(
1
):
209
215
.
28.
Funk
DM.
Coagulation assays and anticoagulant monitoring
.
Hematology Am Soc Hematol Educ Program
.
2012
;
2012
:
460
465
.
29.
Douxfils
J,
Adcock
DM,
Bates
SM
, et al.
2021 Update of the International Council for Standardization in Haematology recommendations for laboratory measurement of direct oral anticoagulants
.
Thromb Haemost
.
2021
;
121
(
8
):
1008
1020
.
30.
Barbosa
CAN,
Montalvao
SAL,
Barbosa
KGN,
et al.
Prolonged APTT of unknown etiology: a systematic evaluation of causes and laboratory resource use in an outpatient hemostasis academic unit
.
Res Pract Thromb Haemost
.
2019
;
3
(
4
):
749
757
.
31.
Bowyer
A,
Kitchen
S,
Makris
M.
The responsiveness of different APTT reagents to mild factor VIII, IX and XI deficiencies
.
Int J Lab Hematol
.
2011
;
33
(
2
):
154
158
.
32.
Toulon
P,
Eloit
Y,
Smahi
M,
Sigard
C,
Jambou
D,
Fischer
F,
Appert-Flory
A.
In vitro sensitivity of different activated partial thromboplastin time reagents to mild clotting factor deficiencies
.
Int J Lab Hematol
.
2016
;
38
(
4
):
389
396
.
33.
Adcock
DM,
Gosselin
RC.
The danger of relying on the APTT and PT in patients on DOAC therapy, a potential patient safety issue
.
Int J Lab Hematol
.
2017
;
39
(
suppl 1
):
37
40
. .
34.
Pernod
G,
Albaladejo
P,
Godier
A,
et al.
Working Group on Perioperative Haemostasis. Management of major bleeding complications and emergency surgery in patients on long-term treatment with direct oral anticoagulants, thrombin or factor-Xa inhibitors: proposals of the working group on perioperative haemostasis (GIHP)—March 2013
.
Arch Cardiovasc Dis
.
2013
;
106
(
6–7
):
382
393
. .
35.
Thomas
S,
Makris
M.
The reversal of anticoagulation in clinical practice 
Clin Med (Lond)
.
2018
;
18
(
4
):
314
319
. .

Author notes

Supplemental digital content is available for this article at https://meridian.allenpress.com/aplm in the August 2024 table of contents.

The authors have no relevant financial interest in the products or companies described in this article.

Supplementary data