The drying and storage of flexible endoscopes after liquid chemical disinfection currently is a hot topic of debate and research.1 However, the importance of endoscope drying prior to storage is not new. In 1982, Gerding et al.2 showed that 31% of stored endoscopes were contaminated with bacteria. Introducing forced-air drying prior to storage reduced the contamination rate to 5%. That and many other studies that followed have shown the importance of drying to control microorganism proliferation during storage (see Kovaleva3 and Alfa and Singh4 for comprehensive reviews).
Although the importance of drying flexible endoscopes postprocessing is well recognized, recent articles have shown that:
95% of stored, patient-ready scopes (19 of 20 evaluated) retained residual liquid after automated endoscope processing, which included a programmed alcohol flush and air purge.5
The percentage of stored, patient-ready scopes with residual liquid was 83% (10 of 12) and 85% (11 of 13) at two clinical sites.6 At a third site, endoscopes were consistently dry when a 10-minute forced-air drying procedure was used.
Of 249 U.S. institutions surveyed concerning duodenoscope processing practices, 52.2% did not use forced-air drying after processing.7
Clearly, effective drying and storage of endoscopes is a challenging process.
Why Is Drying so Difficult?
A review of recent flexible endoscope processing manuals by the authors illustrated why drying is so difficult. Explicit channel-drying instruction is lacking. Typically, users are simply told to “dry the channels.” Reference to local standards may be provided, and the manuals may mention that alcohol can be used. However, details about the type of air to use, how to apply the air, and the time required to dry are missing.
ANSI/AAMI ST91:20158 offers guidance for drying of flexible endoscopes. Specifically, the standard states: “Drying can be achieved by flowing air through all endoscope channels for a specified period of time. Drying should be facilitated by using 70–80% ethyl or isopropyl alcohol. When using alcohol, personnel should follow manufacturer's written IFU (instructions for use) on the volume of alcohol and method used for each endoscope lumen and ensure remaining alcohol is removed with medical-grade forced air until no visual signs of moisture remain (or as otherwise recommended by the endoscope manufacturer).” (Note: At the time this article was being written, ST91 was being updated. Once released, readers are encouraged to review the new version of the standard for the most current recommendations for drying and storing flexible endoscopes.)
The standard further instructs to avoid using syringes for channel drying; rather, compressed air at the “force of air pressure to channel size” provided by the manufacturer's IFU should be used. ST91 guides users to the manufacturer's IFU, and as already mentioned, detailed drying instructions are lacking. Therefore, users must determine how to dry on their own.
A lack of simple, sensitive methods for verifying channel dryness further complicates the issue. The instrument channel can be visualized using a borescope. However, most other channels cannot be directly visualized due to their small size (e.g., 1–1.4 mm internal diameter for portions of the air/water channel) and sharp bends that do not allow passage of most borescopes.9 For channels that cannot be directly visualized, the only options are water detection papers6,10–13 or humidity sensors.13 However, these methods are not as sensitive or simple to use compared with visual borescope evaluation performed by trained personnel.
Perumpail et al.10 reported that the detection limit of cobalt (II) chloride paper was greater than 250 μL water for a duodenoscope air/water system, demonstrating that this method is not sensitive enough to confirm all channels in a flexible endoscope are dry. Ofstead et al.,6 using a cobalt chloride-based paper, found that 10 μL water could be detected when using cotton swabs to sample endoscope channel openings. Michels13 showed that both isopropanol wetted copper (II) sulfate paper and a humidity probe were able to detect 5 μL residual water in a simple, simulated endoscope channel. However, the alcohol-wetted paper changed color after five minutes due to the ambient air humidity, potentially making this method difficult to interpret.
Two international standards, DS/EN 16442:201511 and EN ISO 15883-4:2018,12 describe methods for assessing channel and external surface drying efficacy for development of flexible endoscope storage cabinets and washer disinfectors, respectively. Both use copper (II) sulfate paper as the detection method. For internal channels, the distal end of the endoscope is pointed at the paper, which is held 5 to 10 cm away. Next, the channels are flushed with 17.4 psi or greater medical grade air. However, the air flush duration and method used to direct air into the channels are not provided. For external drying verification, the paper is inserted into crevices (e.g., between the control knobs). Of note, drying the control knob crevices is difficult; therefore, verifying dryness of this area using a water-sensitive paper may be beneficial.
Should Alcohol Be Used for Drying Flexible Endoscope Channels?
Some professional organizations recommend using alcohol to enhance drying flexible endoscope channels.8,14 However, use of alcohol is controversial due to its reported fixative properties15,16 and the potential danger and health risks17 associated with its use. Further, Alfa and Sitter18 demonstrated that, with a 10-minute dry time of duodenoscope channels, an alcohol rinse was not required to prevent proliferation of bacteria during storage. The Association of periOperative Registered Nurses (AORN) recommends that a multidisciplinary team conduct a risk assessment to determine whether the channels should be flushed with 70% to 90% alcohol after disinfection.19
These results and directions call into question the necessity of flushing flexible endoscope channels with alcohol to promote drying. Therefore, the decision of whether to use alcohol is another example of facilities not having access to clear guidance.
How Do We Store Flexible Endoscopes?
In the past, many facilities stored patient-ready endoscopes in transport bins, in original endoscope boxes with foam inserts, and upright in cabinets with and without air flow. At that time, storage was viewed as a means to keep the scope out of the patient's reach and to prevent accidental damage. Today's expanded view includes prevention of recontamination of the endoscope prior to patient use, even when stored for days. Users must always follow storage instructions provided by the original equipment manufacturer for each endoscope model. If explicit instructions are not provided, professional societies offer storage recommendations (summarized in Table 1).
Recommendations for proper storage of flexible endoscopes. Abbreviation used: HEPA, high-efficiency particulate air.

The importance of proper storage is highlighted in the recent study by Ofstead et al.,6 which showed microbial growth in 71% (32 of 45) of sampled stored endoscopes. In an Australian study, a contamination rate of 15.5% (of which 0.5% were pathogens) was reported after an average storage period of 37.6 hours (range 5.3–165.3 hours).21 Of important note, having a single organism survive the disinfection process can lead to organism proliferation if that endoscope is not dried and stored properly. According to Thacker and Muthusamy,22 facilities should “consider automated drying steps.” They give examples of cabinets for scope storage with active air circulation, forced-air designs, or both; one used high-efficiency particulate filtered air.
Using automated drying steps is an adjunct for drying of flexible endoscopes. In addition, evidence indicates that drying cabinets with static air do not effectively dry compared with cabinets that circulate filtered air through the endoscope channels for at least an hour.23 For this reason, storage cabinets without filtered circulating air are falling out of favor.
How Long Is Too long for ‘Hang Time'?
How long can an endoscope be stored before it must be processed? Often called “hang time,” the shelf life of a high-level–disinfected or liquid chemically sterilized endoscope depends on the microbiocidal process used, the results of an internal risk assessment, and professional society guideline followed. Table 2 lists professional society recommendations and study results identifying maximum hang times before the endoscope must be processed.
Current state of flexible endoscope hang time. Abbreviations used: ANSI, American National Standards Institute; VHA, Veterans Health Administration.

It is interesting to note that published studies give widely varying results for safe storage times. For example, Schmelzer et al.26 concluded that endoscopes can be stored for seven days (based on their analysis of published studies), but Lacy et al.27 showed that a 12-week hang time is possible. The latter researchers proposed that the key drivers of endoscope contamination are inadequate disinfection and/or poor storage practices.
Considering the current state of understanding, AAMI,8 AORN,19 and the Healthcare Infection Control Practices Advisory Committee20 make it clear that the maximum hang time remains an unresolved issue. Facilities must conduct a risk assessment to determine their own policy, procedures, and practices. Because each facility has a unique inventory of flexible endoscope and storage procedures, as additional evidence-based guidance for endoscope storage is developed, facilities will need to incorporate the new information into their facility-specific risk analysis.
Where Do We Go from Here?
Endoscope manufacturers and researchers must take the lead to provide complete drying and storage instructions so that potentially harmful organisms do not proliferate during storage postprocessing. However, the scientific understanding of “how dry is dry?” is missing. More research is needed to:
Define the benchmark for “dry.”
Develop and validate moisture detection methods that are able to reliably verify whether surfaces and internal channels are “dry.”
Provide practical drying verification tests for use at clinical sites.
If the current situation for drying endoscopes sounds familiar, that is because it echoes medical device cleaning concerns. The question “how clean is clean?” has been debated for two decades, and significant progress has been made. We are just beginning the journey to understanding endoscope drying and storage. For example, the lack of specific drying instructions is in stark contrast to the detailed instructions given for cleaning. Device-specific cleaning adapters direct cleaning fluids and rinse water into the channels, resulting in a consistent and reproducible process. The same diligence used for developing cleaning instructions28 must be applied to drying and storage of flexible endoscopes. Minimally, instructions for manual or automated (e.g., drying cabinet or endoscope processor) processes should include:
Where and how to apply filtered, compressed air, including identifying if adapters are attached to the endoscope and/or the specific type of air nozzle used.
The pressure, quality, and duration of applied filtered compressed air.
The type of cloth used to dry the exterior and, if reusable, the duration of use and how to process.
These drying instructions must be verified to be effective and then validated through human factors testing.28
It is time for action. Manufacturers of flexible endoscopes, endoscope processors, and storage cabinets need to work with each other, as well as with professional societies, regulatory bodies, researchers, processing staff, infection preventionists, and endoscopists, to provide answers to the knowledge gaps highlighted above. The only way to ensure the development of optimal drying—and therefore ensure flexible endoscope safety poststorage—is through a multifunctional team approach.
References
Author notes
Tamara Behm, MSN, RN, CIC, is a clinical education specialist in the Infection Prevention Technologies Division of STERIS Corporation in Mentor, OH. Email:tamara_behm@steris.com
Nancy Robinson, PhD, is a distinguished scientist in the Infection Prevention Technologies Division of STERIS Corporation in Mentor, OH. Email:nancy_robinson@steris.com