Abstract
Validating a thermal disinfection process for the processing of medical devices using moist heat via direct temperature monitoring is a conservative approach and has been established as the A0 method. Traditional use of disinfection challenge microorganisms and testing techniques, although widely used and applicable for chemical disinfection studies, do not provide as robust a challenge for testing the efficacy of a thermal disinfection process. Considerable research has been established in the literature to demonstrate the relationship between the thermal resistance of microorganisms to inactivation and the A0 method formula. The A0 method, therefore, should be used as the preferred method for validating a thermal disinfection process using moist heat.
Disinfection, which is defined as reducing the number of viable microorganisms on a product to a level previously specified as appropriate for its intended further handling or use, can be achieved thermally by the action of moist heat.1 Thermal disinfection during the processing of medical devices, typically performed in a washer-disinfector, is widely used for two purposes. The first is for reducing product bioburden (disinfection) either as a terminal step (e.g., for noncritical or semicritical devices) or prior to packaging and sterilization (e.g., for critical devices) in preparation for patient use. The second is to render the devices safe for handling for central service professionals during inspection and packaging.2,3 Thermal disinfection requirements therefore should consider the potential levels of microbial contamination on reusable devices after use, the desired level of reduction to render those devices safe for handling and for their intended purpose, and the reliability of the disinfection process to consistently achieve that endpoint.
The microbial load on device types after patient use has been established in the literature and can vary depending on the typical clinical use of the device. For example, critical (surgical) devices, on average, have demonstrated relatively low levels of viable microorganisms (bioburden level <102 colony-forming units [CFU]/cm2).4 However, these same studies have shown the concentration of other testing analytes (e.g., protein, total organic carbon, hemoglobin) to be more noteworthy. Although the data indicate that residual clinical soil (e.g., human secretions, blood, tissue) can harbor microorganisms, the incoming product bioburden levels are far below the microbial populations challenged during an overkill sterilization process (e.g., moist heat or gaseous processes).
Conservative sterilization processes have been demonstrated to achieve at least a 12-log10 reduction of microorganisms with a known higher resistance versus typical bioburden.3,5 Cleaning, which is defined as the removal of contamination from an item to the extent necessary for its further processing and its intended subsequent use, is an important step to render the device ready for sterilization and will further reduce the levels of microorganisms prior to sterilization. Therefore, with critical devices, adequate cleaning followed by sterilization is the minimum requirement to ensure the device is safe for patient use.
It is not likely that, for the intended use of the device, a disinfection process is strictly necessary as an intermediate step prior to sterilization. A benefit may exist to having an interim disinfection step to render the device safe for handling during inspection and packaging for sterilization. For example, the expectation in the Occupational Safety and Health Administration's Bloodborne Pathogens standard 29 CFR 1910.1030 is that an employer will minimize the occupational exposure to bloodborne pathogens.
Thermal disinfection has been used by sterile processing departments as a universal precaution to reduce the risk of exposure to processing personnel postcleaning. Although routine thermal disinfection at less than 100°C (212°F) may not be effective in deactivating all types of microorganisms (e.g., certain types of bacteria spores), it is a reliable and consistent disinfection process. As the temperature increases above a certain point (typically ≥70°C or 158°F), so does the activity against microorganisms, with variable intrinsic and acquired resistance mechanisms to heat.3 Thermal disinfection therefore will provide processing personnel with a minimized risk of bloodborne pathogens exposure.
In other situations, the microbiological load can be much higher (e.g., with flexible endoscopes used in the gastrointestinal system6 ) or more variable (e.g., with noncritical devices or surfaces depending on their use7 ). Where practical, thermal disinfection is still viewed as the preferred and more reliable method to render these devices safe for use due to its known efficacy against microbial pathogens.5 Chemical disinfection generally is only considered if thermal disinfection cannot be applied (e.g., due to thermo-sensitivity of device or surface materials).
Disinfection Efficacy
When thermal disinfection is specified in the device manufacturer's instructions for use (IFU), the disinfection process must be validated to demonstrate the effectiveness of the cycle to inactivate microorganisms for that specific device.8 Chemical disinfection claims and validations traditionally are performed by inoculating the most difficult-to-disinfect locations with a known titer of challenge vegetative microorganisms (e.g., Escherichia coli, Staphylococcus aureus, Pseudomonas aeruginosa, Klebsiella pneumoniae, Mycobacterium species) in the presence of soil and calculating the log reduction postdisinfection. The levels and types of microorganism inactivation are the basis for disinfection claims (e.g., in many countries defined as low-, intermediate-, and high-level disinfection) to describe the effectiveness of the process).9,10 Unfortunately, using a similar strategy for thermal disinfection validation may not appropriately or reliably challenge the device or load.
The disinfection levels established for chemical disinfection leverage the chemical resistance profile of microorganisms such as Mycobacterium species ( e.g., M. terrae), which are known to represent some of the more challenging vegetative microorganism to inactivate.11 It may not be appropriate to apply the same microorganism resistance profile to a thermal disinfection method using moist heat, even when a thermoresistant strain is used. This concept was confirmed in studies evaluating the resistance of microorganisms using thermal disinfection, suggesting that from the 19 microorganisms tested, only Micrococcus luteus showed any resistance (equivalent to a low-level disinfection claim).12 This study showed that a bioburden reduction of a heat resistant Mycobacterium species may not be the most resistant challenge for a device during thermal disinfection validation studies. Other reports suggested the use of other bacteria due to their notable resistance profiles to thermal inactivation (e.g., Enterococcus species13 ).
The procedure for performing a bioburden reduction study in a liquid environment (e.g., washer-disinfector) also presents challenges. The most obvious is that the process of exposure within the washer-disinfector is designed for the physical removal of microorganisms (and other soil components) during the full washer-disinfector process. The method of direct inoculation of the challenge microorganism to a device surface used in chemical disinfection validations is impractical because a high population of the inoculum would wash off the device during the exposure cycle (even if just the disinfection phase). To circumvent this problem, glass ampoules are prepared with the inoculation titer and placed at worst-case locations within the load (e.g., load, load carrier) in the washer-disinfector. These ampoules are secured on the outside or inside feature of a device. Due to their size (Figure 1), they may only represent portions outside the device surfaces or loads. They also depend on heat transfer into the vials, which does not directly simulate exposure directly on a device surface. Such ampoules therefore are not a representative challenge.
The challenges of working with microorganisms during thermal disinfection are addressed by using thermometric testing, as described in ISO 15883-1:2006.1 The reliability of thermometric testing is based on the known microbial resistance profiles of different types of microorganisms to thermal inactivation, which has been established in various industrial microbiology applications since the early 1900s.5,11 Essentially, most bacteria, fungi, viruses, and protozoa are inactivated at temperatures in excess of 70°C or 158°F. The much higher temperature conditions defined for steam sterilization applications are to accommodate the specific inactivation of thermoresistant bacterial spores, which are outside the scope of this discussion on disinfection.5,14
As the temperature increases above levels of 70°C, the time to inactivate microorganisms decreases considerably—a concept similar to the well-established microbial inactivation profile used for steam sterilization and the basis for lethality modelling in heat inactivation studies (the F-value or F0 for moist heat processes).14,15 Similar to steam sterilization processes, the thermometric test for thermal disinfection is performed by challenging the disinfection phase of the washer-disinfector cycle with each load carrier, including a representative load to determine the thermal distribution of the chamber, load, and carrier load. The relationship of the recorded temperature to microorganism inactivation can be expressed using the A0 calculation.1 Unlike the size challenges with the ampoules, thermocouples may be placed at multiple locations to identify the greatest challenge for thermal disinfection within the chamber, chamber loads, and individual devices (Figure 2).
Examples of the positioning of thermocouples at various positions within the load and washer-disinfector for temperature distribution validation testing. Thermocouples are colored orange or brown.
Examples of the positioning of thermocouples at various positions within the load and washer-disinfector for temperature distribution validation testing. Thermocouples are colored orange or brown.
A0 Calculation and Microbial Reduction
Similar to other methods used to evaluate microbiological inactivation rates (e.g., F0 method for steam sterilization), the A0 method incorporates safety factors to conservatively measure the actual disinfection value. The z value of microorganisms increases as the resistance of the microorganism increases to the disinfection method. The z value of 10°C selected for the A0 equation is representative of bacterial spores, which are the most resistant of all microorganisms and not the goal of the thermal disinfection cycle for reduction.16 Although the A0 value can consist of the sum of many subvalues (the combination of any time/temperature during the disinfection phase), additional safety is added to the calculation by only including the time associated with a 70°C or greater temperature range in the calculation, even when it has been established that microorganism inactivation occurs at temperatures below this value.16 The A0 method therefore can be considered a conservative process based on the effectiveness of microbial inactivation.12
An example of a washer-disinfector cycle with the A0 value calculated during the performance of the cycle, as depicted in Figure 3, is as follows
Intermediate Rinse: 70°C/0.5 min(conservative estimate using minimum phase temperature)
A0 =10(70–80)/10 × 30 = 3
Rinse: 90°C/5min
A0 =10(90–80)/10 × 300 = 3,000
Disinfection: 90°C/1 min
A0 =10(90–80)/10 × 60 = 600
Dry: 90°C/9min
A0 =10(90–80)/10 × 540 = 5,400
Sum of A0 for entire washer-disinfector cycle:
A0 =∑(10(70–80)/10×30) + (10(90–80)/10× 300) + (10(90× 80)/10 x 60) +(10(90×80)/10× 540) =9,003
Although the ISO 15883-1 standard only applies the A0 concept to the disinfection phase of the washer-disinfector cycle, any temperature above 70°C in the cycle will achieve disinfection. Therefore, when looking at the total disinfection efficacy in the rinse, disinfection, and drying phases of the cycle, the conservative estimate for the total cycle A0 value is 9,003.
Of note, during the drying phase, moist heat (steam) is created during the process; therefore, this phase of the process can also be considered cumulative to the A0. Figure 3 shows the temperature and duration of exposure for devices during the entirety of the washer-disinfector cycle. If the accumulative kill is calculated using all subparts of the cycle, then the A0 method demonstrates a very conservative approach for thermal disinfection.
Data from peer-reviewed literature over many years initially were used to establish the relationship between moist heat and microbial reduction using thermal disinfection, especially from the use of heat for the pasteurization of foods. During the past 20 years, further research has confirmed the thermoresistant profiles of microorganism in device disinfection studies and applied these to the A0 method (Table 1).
Testing for statistical significance has been performed to demonstrate the correlation and highly predictable behavior between microbial reduction and the A0 value. In fact, the minimum A0 value (i.e., A0 = 60) was achievable even when microorganisms were tested in the presence of blood or biofilms.12,17 The data within these literature references clearly show that using the A0 method to demonstrate microbial inactivation of microorganisms is far superior to the traditional microbiological reduction method adopted from chemical disinfection efficacy studies.
Conclusion
The application of microbial reduction for the thermal disinfection process does not accurately account for the efficiency of the cycle. The A0 method is a far more conservative method and allows for greater scientific understanding of the true distribution of temperature in a typical washer-disinfector (and associated load). The method also applies a greater safety factor to ensure the safe handling or use (when applicable) of devices after cleaning using a washer-disinfector validated under the requirements of the ISO 15883 standard series. The relationship between A0 and microbial reduction has been well documented in the literature. Using the A0 method with thermometric profiling is a practical way to establish the relationship between heat and microbial inactivation during validation testing and should be considered the preferred method of validating thermal disinfection processes.
References
Author notes
Terra A. Kremer, BS, is a senior program manager in Microbiological Quality & Sterility Assurance at Johnson & Johnson in Raritan, NJ. Email: tkremer@its.jnj.com
Gerald McDonnell, BSc, PhD, is a senior director in Microbiological Quality & Sterility Assurance at Johnson & Johnson in Raritan, NJ. Email: gmcdonne@its.jnj.com
Emily Mitzel, MS, is a director global cleaning and disinfection at GE Healthcare in Park City, UT. Email: emily.mitzel@ge.com
Nupur Jain, BS, MBA, is a director of device reprocessing & sterilization at Intuitive Surgical in Sunnyvale, CA. Email: nupur.jain@intugurg.com
Henri Hubert is the head of department research and development at SMP GmbH in Tübingen, Germany. Email: hhubert@smpgmbh.com
Klaus Roth is the owner of SMP GmbH in Tübingen, Germany. Email: kroth@smpgmbh.com
Philippe Labrie, BSc, PhD, is a scientific director of the research and validation team and Medisafe R&D team at STERIS Canada ULC in Québec, Canada. Email: philippe_labrie@steris.com
Alex Villella, BA, is a senior scientist of device reprocessing and disinfection at Cenorin Products for Healthcare in Kent, WA. Email: avillella@cenorin.com