Certain low-frequency magnetic fields cause interference in implantable medical devices. Electromagnetic compatibility (EMC) standards prescribe injecting voltages into a device under evaluation to simplify testing while approximating or simulating real-world exposure situations to low-frequency magnetic fields. The EMC standard ISO 14117:2012, which covers implantable pacemakers and implantable cardioverter defibrillators (ICDs), specifies test levels for the bipolar configuration of sensing leads as being one-tenth of the levels for the unipolar configuration. The committee authoring this standard questioned this testing level difference and its clinical relevance. To evaluate this issue of EMC test levels, we performed both analytical calculations and computational modeling to determine a basis for this difference. Analytical calculations based upon Faraday's law determined the magnetically induced voltage in a 37.6-cm lead. Induced voltages were studied in a bipolar lead configuration with various spacing between a distal tip electrode and a ring electrode. Voltages induced in this bipolar lead configuration were compared with voltages induced in a unipolar lead configuration. Computational modeling of various lead configurations was performed using electromagnetic field simulation software. The two leads that were insulated, except for the distal and proximal tips, were immersed in a saline-conducting media. The leads were parallel and closely spaced to each other along their length. Both analytical calculations and computational modeling support continued use of a one-tenth amplitude reduction for testing pacemakers and ICDs in bipolar mode. The most recent edition of ISO 14117 includes rationale from this study.

Implantable cardiac pacemakers are used in millions of patients to regulate or reproduce normal heart rhythm. Patients are candidates for a pacemaker when the heart's natural rhythm is too slow or if a conduction block is present in the heart's electrical system. An implantable cardioverter defibrillator (ICD) functions like a pacemaker, with the added function of being able to deliver a strong electrical shock to treat life-threatening arrhythmias. Both pacemakers and ICDs have lead(s) that extend from the metallic case under the skin, usually in the pectoral region, to the interior of the heart. Each lead has a distal tip electrode at the far end and one or more ring electrodes spaced a small distance closer to the metallic case (Figure 1). These electrodes are placed in one or more chambers of the heart. The implanted devices detect the heart's intrinsic electrical activity via these electrodes to determine what stimulation to deliver. A voltage can be sensed between the tip electrode and the metallic case (unipolar) or between the tip and a ring electrode (bipolar). The choice of sensing is up to the clinician.

Figure 1.

Tip and ring electrodes are used for a bipolar configuration. For unipolar configuration, only the tip electrode is used.

Figure 1.

Tip and ring electrodes are used for a bipolar configuration. For unipolar configuration, only the tip electrode is used.

Close modal

Pacemakers and ICDs are continuously sensing the heart's electrical activity and are susceptible to electromagnetic interference (EMI), which may be interpreted as cardiac signals. EMI is a disturbance generated by an electrical source, such as a cell phone. EMI to pacemakers and ICDs is well known.1  Because pacemakers and ICDs sense frequencies between 1 and 500 Hz, they are most susceptible to low-frequency magnetic fields. The Food and Drug Administration recognizes ISO 14117:20192  as describing an electromagnetic compatibility (EMC) test method for pacemakers and ICDs.

The Active Implants Joint Working Group 1 (ISO/TC 150 SC 6 JWG 1)3  is the standards group that authors the EMC standard ISO 14117. While writing the second edition of this standard, the group questioned the basis for the material published in the first edition. One topic of discussion was the requirement and lack of rationale in determining the appropriate test levels for bipolar lead configurations. The test levels described for a unipolar configuration are reasonably supported based partly on the reference levels in the European Commission Recommendation 5194  under certain assumptions of magnetic fields inducing a voltage in leads. However, the normative requirements of ISO 14117:2012 simply state, “Bipolar differential mode performance shall be tested using the test signal reduced to one-tenth amplitude” (referring to one-tenth of the test signal specified for devices with unipolar leads).5  The only rationale provided in ISO 14117:2012 is as follows: “Because of the close proximity of tip and ring electrodes, the applicable test signal is reduced to 10% of the common mode test signal amplitude.” No documented scientific basis exists for this 90% reduction for bipolar differential tests.

The objectives of the current study were to determine the appropriate test levels below 10 MHz for the bipolar lead configuration and to provide a clear rationale for those levels. All implantable pacemakers and ICDs are tested to the ISO 14117 standard, and the large majority that are implanted are programmed to a bipolar lead configuration. The current study sought to improve understanding regarding whether implantable pacemakers and ICDs are tested to an adequate level.

Analytical

The voltage induced in the leads of a pacemaker or ICD by a changing magnetic field can be calculated analytically using Faraday's law (equation 1) for frequencies less than 10 MHz. It specifies that the voltage induced in the leads is equal to the magnetic field (flux density) times the area of the inductive loop (times 2 times π times the frequency of the signal). The inductive loop is the area enclosed by the metallic case of the pacemaker or ICD, the lead, and the straight return path through human tissue. The conservative solution that follows is based on fundamental equations and figures presented in EN 50527-2-2:2018,6  which describes how to calculate the voltage induced in the leads by magnetic fields.
formula
where V is the root mean square (RMS) induced voltage (in V), B is the RMS magnetic flux density (in T), A is the area of the inductive loop (in m2), and f is the frequency (in Hz).
As discussed previously, we have confidence that the rationale is valid for test levels in the unipolar configuration. With that assumption, adequate rationale for the bipolar test voltage can be determined with the ratio between the induced voltage in the bipolar configuration versus the induced voltage in the unipolar configuration. This ratio is shown in equation 2.
formula
where the subscript BP is for the bipolar configuration and UP is for the unipolar configuration. The magnetic flux density, B, is constant for this analysis, as it is the reference field that induces the voltage that is being measured. Therefore, the ratio of induced voltage for bipolar to unipolar configurations is entirely dependent on the area enclosed by each lead configuration. In the unipolar configuration, we have the area enclosed by the metallic case of the pacemaker or ICD, the lead, and the straight return path through human tissue (inductive loop). For a lead length of L, the maximum area is a semicircle (Figure 2).
Figure 2.

The region shaded in blue represents the effective induction area for the unipolar configuration.

Figure 2.

The region shaded in blue represents the effective induction area for the unipolar configuration.

Close modal
The area of a semicircle is given by:
formula
where r is the radius of the semicircle.
The lead length, L, is the arc of the semicircle:
formula
The area of 225 cm2 is used as a realistic effective induction area for the unipolar configuration, which provides a substantial safety margin.2,5  Solving for r in equation 4 gives:
formula
Next, replacing r from equation 5 into equation 3 gives:
formula
Solving for L in equation 6:
formula
The bipolar area configuration can be approximated as a triangle (colored gray), as shown in Figure 3. The area of this triangle is given by equation 8:
formula
where b is the base of the triangle and h is the height of the triangle.
Figure 3.

The gray-colored triangle represents the effective induction area for the bipolar differential configuration.

Figure 3.

The gray-colored triangle represents the effective induction area for the bipolar differential configuration.

Close modal
The base of the triangle is approximated as the tip-to-ring spacing and the height of the triangle is the diameter of the unipolar circle. Using the lead length from equation 7 and solving for the radius in equation 5 gives:
formula
Doubling the radius to find the diameter of the semicircle, and using equation 8, gives:
formula
where b is the tip-to-ring spacing
Therefore, the ratio of the induced voltage in a bipolar configuration to the induced voltage in unipolar configuration, in terms of tip-to-ring spacing, b (in cm), is:
formula

Computational Modeling

We performed computational modeling of bipolar and unipolar leads using finite element, three-dimensional electromagnetic simulation software (Ansys HFSS; ANSYS, Canonsburg, PA). Two insulated leads were placed adjacent to one another in a simulated plastic tank (a cylinder with a dielectric constant of 2.0, a diameter of 30 cm, and a height of 30 cm). The cylinder contained a simulated liquid that had electrical properties similar to human soft tissues. The liquid had various conductivities for different simulations. Among the materials used were normal saline with a conductivity of 1.39 S/m and a saline mixture with a conductivity of 0.139 S/m. Saline of various conductivities often is used for simulating human body tissues in simple electromagnetics problems.2,5  A magnetic flux density (B) was generated by a simulated circular Helmholtz coil pair (60 cm radius, 60 cm separation) that was driven by a simulated current generator over various frequencies in the 10 Hz to 10 kHz frequency range. The 1.2-G flux density produced by the Helmholtz coil was highly uniform throughout the volume of the plastic cylinder containing the saline and the leads. The vector of the B field was parallel to the long axis of the cylinder or perpendicular to the surface of the liquid (Figures 4 and 5).

Figure 4.

Simulation configuration to create uniform magnetic field into leads in a saline tank.

Figure 4.

Simulation configuration to create uniform magnetic field into leads in a saline tank.

Close modal
Figure 5.

Computational model of the two leads. Bottom lead = 24 cm diameter.

Figure 5.

Computational model of the two leads. Bottom lead = 24 cm diameter.

Close modal

The leads in this simulation were a pair of adjacent, insulated wires, each forming a semicircular arc. The leads had an inner conducting wire (perfect electrical conductor) with a diameter of 1.0 mm and an insulation thickness surrounding the wire of 0.5 mm (dielectric constant = 2.0). Each wire had a 2-mm bare length at each end. The longer lead, representing the portion of the lead containing the tip, was 37.6 cm long and formed a 180° semicircular arc (24 cm diameter). The shorter lead, representing the portion of the lead containing the ring, was above and parallel to the longer lead. The length of this lead was varied for each simulation performed (Table 1). The two leads were separated by 0.4 mm in the Z direction (Figure 5). The shorter lead formed a semicircular arc, with its angular arc being less than 180° depending on which one of several simulations was performed for different tip-to-ring spacings. The induced voltage on each lead was computed between the distal and the proximal ends of the leads. Among other modeling situations, we performed three separate simulations for the above parameters, with one lead being 10, 20, or 30 mm shorter in length than the longer lead. We used these configurations to determine the ratio of bipolar to unipolar sensing by computing the induced voltages for the longer lead versus the shorter lead, according to the procedure performed for the analytical study.

Table 1.

Computational and analytical results for three tip-to-ring spacings. *The longest known tip-to-ring spacing found via a search of the literature was 16 mm, but the decision was made to display potential large spacings up to 30 mm.

Computational and analytical results for three tip-to-ring spacings. *The longest known tip-to-ring spacing found via a search of the literature was 16 mm, but the decision was made to display potential large spacings up to 30 mm.
Computational and analytical results for three tip-to-ring spacings. *The longest known tip-to-ring spacing found via a search of the literature was 16 mm, but the decision was made to display potential large spacings up to 30 mm.

Table 1 presents data for the computational and analytical results. The table shows the computational results in columns 2 through 4. Columns 5 and 6 present the computational and analytical ratios for the induced voltage in the bipolar configuration versus the induced voltage in the unipolar configuration, respectively. The final column is the ratio between the two methods (computational and analytical). All results are presented as a function of the tip-to-ring spacing.

The last column in Table 1 shows the general agreement between the analytical solution and the computational solution, where a ratio of 1 would be exact agreement. The values between 0.82 and 0.95 give us reasonable confidence in each solution (though not statistically significant due to sample size). Figure 6 shows how the ratio of bipolar to unipolar induced voltages change for each solution based on different tip-to-ring spacing. The longest tip-to-ring spacing found via a search of the literature was 16 mm, but we chose to display potential large spacings up to 30 mm.

Figure 6.

Ratio of bipolar to unipolar induced voltages with respect to tip-to-ring spacing (in millimeters). The analytical solution is represented by the blue line, with several points highlighted beneath the line in blue. Computational results are represented by red dots.

Figure 6.

Ratio of bipolar to unipolar induced voltages with respect to tip-to-ring spacing (in millimeters). The analytical solution is represented by the blue line, with several points highlighted beneath the line in blue. Computational results are represented by red dots.

Close modal

When great differences in conductivities were used in the computational modeling, significant differences in the tip voltage and in the ring voltage were observed. However, the ratio of bipolar to unipolar voltages did not change.

Our test data show that for tip-to-ring spacings less than 19 mm, the induced voltage into a bipolar lead configuration is less than 10% of the induced voltage from a unipolar configuration. This means the normative requirements of ISO 14117:2012 are conservative for tip-to-ring spacings under 19 mm.5  Conservative indicates that pacemakers and ICDs are adequately tested (or appropriately over tested) for leads with tip-to-ring spacing less than 19 mm. The rationale for the ratio of bipolar to unipolar induced voltages has been updated in ISO 14117:2019.2  If future devices use larger tip-to-ring spacing (>19 mm), then higher test levels may be warranted. Currently, we believe that testing to one-tenth of the unipolar test levels, as described in ISO 14117, is appropriate for bipolar differential lead configurations.

Disclaimer

The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health & Human Services.

1.
Napp
A,
Stunder
D,
Maytin
M,
et al.
Are patients with cardiac implants protected against electromagnetic interference in daily life and occupational environment?
Eur Heart J.
2015
;
36
(
28
):
1798
804
.
2.
ISO 14117:2019.
Active implantable medical devices—Electromagnetic compatibility—EMC test protocols for implantable cardiac pacemakers, implantable cardioverter defibrillators and cardiac resynchronization devices.
Geneva, Switzerland
:
International Organization for Standardization
.
3.
International Organization for Standardization.
ISO/TC 150/SC 6: Active implants
.
www.iso.org/committee/53122.html. Accessed June 8, 2021.
4.
European Union.
1999/519/EC: Council Recommendation of 12 July 1999 on the limitation of exposure of the general public to electromagnetic fields (0 Hz to 300 GHz)
.
5.
ISO 14117:2012.
Active implantable medical devices—Electromagnetic compatibility—EMC test protocols for implantable cardiac pacemakers, implantable cardioverter defibrillators and cardiac resynchronization devices.
Geneva, Switzerland
:
International Organization for Standardization
.
6.
EN 50527-2-2:2018.
Procedure for the assessment of the exposure to electromagnetic fields of workers bearing active implantable medical devices—Part 2-2: Specific assessment for workers with cardioverter defibrillators (ICDs).
London, UK
:
BSI Group
.

Author notes

Seth J. Seidman, MS, is the EMC program advisor in the Office of Science and Engineering Laboratories of the Center for Devices and Radiological Health at the Food and Drug Administration in Silver Spring, MD. Email: seth.seidman@fda.hhs.gov

Howard I. Bassen, MS, is a research engineer in the Office of Science and Engineering Laboratories of the Center for Devices and Radiological Health at the Food and Drug Administration in Silver Spring, MD. Email: howard.bassen@fda.hhs.gov