BACKGROUND:

Foot drop in people with multiple sclerosis (MS) commonly leads to decreased mobility and quality of life (QOL). Functional electrical stimulation (FES) of the peroneal nerve can improve the gait of people with foot drop, yet various barriers restrict widespread use. The purpose of this case series was to examine the feasibility of a telerehabilitation-monitored FES device and report changes in functional mobility and QOL in people with moderate MS-related disability.

METHODS:

FES use was progressed over 8 weeks via 3 telerehabilitation sessions. Feasibility of telerehabilitation was assessed by percentage of telerehabilitation visits completed and participant-reported satisfaction. At baseline and study completion, functional mobility with and without FES were assessed by the Timed 25-Foot Walk (T25FW), Timed Up and Go (TUG), and 2-Minute Walk Test (2MWT), Multiple Sclerosis Impact Scale (MSIS-29), and the 12-item Multiple Sclerosis Walking Scale (MSWS-12). Fatigue was assessed via the Modified Fatigue Impact Scale (MFIS) before and after the intervention.

RESULTS:

Eleven participants (mean age = 50.4 years [SD 10.8]; 2 males) completed the study. All (33/33) telerehabilitation visits were completed and participants attained high levels of satisfaction with no adverse events. At 8 weeks, compared to baseline, there were clinically meaningful improvements on the T25FW, 2MWT, and TUG for 45%, 55%, and 82% of participants, respectively. Clinically meaningful improvements on the MSIS-29 and MSWS-12 were also recorded for 64% and 36% of participants, respectively.

CONCLUSIONS:

Telerehabilitation was safe and feasible for FES intervention, and improvements in functional mobility and QOL were observed. Telerehabilitation to monitor FES may improve access and reduce patient burden; therefore, studying its efficacy is warranted.

Multiple sclerosis (MS) is an autoimmune, neurodegenerative disorder characterized by demyelination of axons in the central nervous system (CNS) affecting approximately 3 million people worldwide.1  Difficulty walking is highly prevalent in people with MS and is a defining feature of disease progression.2,3  Improving walking ability is also one of the top priorities for people with MS.4,5  One important cause of walking difficulty is foot drop, or the inability to sufficiently dorsiflex the ankle during the swing phase of gait. Individuals with foot drop report an increased incidence of falls, worse participation in mobility-related activities, and reduced quality of life.6,7 

Functional electrical stimulation (FES) of the peroneal nerve is a common first-line treatment to address foot drop in people with MS.8  Surface electrodes placed over the common peroneal nerve provide electrical stimulation to elicit ankle dorsiflexion as the swing limb advances.9  FES can improve gait speed, decrease energy cost of walking, and reduce falls when used by people with MS who have foot drop.10,11  In spite of the acknowledged advantages provided by FES for enhancing mobility, there are various barriers to extensive adoption of this assistive technology.12  First, both in terms of the cost of the device itself and the multiple, in-person clinic visits required to optimize the device settings, high initial costs limit access to wearable FES.13  Second, challenges with precise electrode placement12  and time to set up the device14  affect long-term adherence and contribute to high abandonment rates of the assistive technology.15  Finally, a recent qualitative study13  and clinical practice guideline16  found that FES users value “access to professional help” when initially learning to use FES, and that physical therapy (PT) intervention is imperative to promote optimal results. However, to date, many of the studies investigating the use of FES in people with MS offer no or limited ongoing support and training for participants, ranging from 0 to 8 PT visits,10  and thus, the optimal amount and type of PT intervention is not clearly defined.

Telerehabilitation may mitigate several of the aforementioned barriers to FES delivery and use. This virtual mode of health care has emerged as a viable rehabilitation option for people with MS,17  reducing travel time for in-person appointments18  and providing an ecologically valid environment for clinicians to evaluate and treat without the burden of an in-person visit. Recent evidence supports the use of telerehabilitation as a safe and effective method of delivering care for people with MS, which can lead to functional improvements in gait and balance outcomes.19  However, to the best of our knowledge, no studies to date have investigated the use of telerehabilitation to facilitate delivery of FES intervention for people with MS.

Therefore, the primary aim of this case series was to examine the feasibility of telerehabilitation to remotely monitor and advance an FES intervention in people with MS. The secondary aim was to report changes in functional mobility and quality of life (QOL) after 8 weeks of FES and telerehabilitation.

Design and Participants

This prospective case series enrolled adults (> 18 years of age) with either progressive or relapsing MS from the departments of Physical Therapy and Neurology at the University of California San Francisco. Participants were included in the study if they met the following criteria: diagnosis of MS (by 2010 International Panel criteria)20 ; ability to walk with or without an assistive device (Expanded Disability Status Scale score of ≤ 6.5) for at least 10 meters; presence of unilateral foot drop as determined by a physical therapist; access to a smartphone and the internet; and adequate cognitive and communicative function to understand training instructions and use the FES device. Participants in this study had never used an FES device previously. Participants were excluded if they were unable to communicate in English (the software used in the study was only available in English); had comorbid neurologic, orthopedic, or cardiovascular diagnoses that would impact the participant's mobility or treatment response; had an active implanted device or cancer in proximity to the area of electrode placement; were pregnant; had a fixed plantarflexion contracture; or had anatomical contraindications (eg, swollen/infected skin, edema). All subjects provided signed informed consent approved by the University of California San Francisco Institutional Review Board.

Study Procedures

Participants received an initial in-person assessment from a licensed physical therapist. During the baseline assessment, each participant performed functional mobility testing first without FES then with FES following a 5-minute acclimation period with FES in which participants walked overground with the FES device turned on. Participants were provided a rest break between the acclimation period and the formal functional mobility testing. Initial device tolerability and changes in gait parameters were assessed with and without FES activated. If FES was well-tolerated, participants were instructed on how to don and doff the device and prescribed an individualized wear-time for the first week of use. Participants then completed a total of 3 telehealth visits at weeks 2, 4, and 6 with the same physical therapist during the 8-week intervention period (FIGURE S1). The 30-minute telehealth visits assessed participants’ self-reported experience with the FES device, screened for adverse events, and allowed the physical therapist to assist remotely with electrode placement and FES progress on an individual basis in the participant's environment. All participants completed an in-person final assessment at study completion. The same acclimation period to FES was not required during the final study visit; however, participants were again provided a rest break prior to performing functional mobility testing.

FIGURE S1

Overview of Study Activities

FIGURE S1

Overview of Study Activities

Close modal

FES Device

The EvoWalk FES device (Evolution Devices, Inc) was used to administer FES. The EvoWalk FES device uses an artificial intelligence–based algorithm and a single sensor to detect when the user initiates the swing phase of gait and appropriately delivers stimulation to the common peroneal nerve to elicit dorsiflexion. The EvoWalk is paired via Bluetooth to a smartphone-based application so participants and clinicians can adjust stimulation intensity as needed. During the initial study visit, the study physical therapist applied 2 surface electrodes over the common peroneal nerve as it passes over the head of the fibula, and the other over the tibialis anterior muscle belly. The level of stimulation was selected individually, based on the participants’ motor response and tolerance.

Feasibility

Feasibility was measured by retention (ie, percentage of participants who completed the study), acceptability (ie, percentage of the telerehabilitation visits completed), adherence (ie, use of the FES device), participant-reported satisfaction with the telerehabilitation visits, and safety (via recording of adverse events). At the end of each telerehabilitation visit, participants completed the Telerehabilitation Satisfaction Survey, a 10-item self-report questionnaire (TABLE S1). Each item is scored on a 5-point Likert scale from 1 (not at all satisfied) to 5 (very satisfied). Scores range from 10 to 50, with a higher score indicating more satisfaction.21 

TABLE S1

Participant Baseline Characteristics, Acceptability, Satisfaction

Participant Baseline Characteristics, Acceptability, Satisfaction
Participant Baseline Characteristics, Acceptability, Satisfaction

Functional Mobility Tests

Functional mobility was assessed by the Timed 25-Foot Walk Test (T25FW), the 2-Minute Walk Test (2MWT), and the Timed Up and Go (TUG). For all functional mobility tests, participants were allowed to use their preferred assistive device, if needed. The T25FW test, the most common measure of walking speed in people with MS, was obtained. Participants were asked to walk 25 feet in a straight hallway as quickly as possible, safely. The time taken to walk 25 feet was recorded by a stopwatch. This test was repeated 3 times and the average was taken and converted to meters per second for analysis. An improvement of 20% has been suggested as a minimal clinically important difference (MCID).6  The 2MWT was used as a measure of walking endurance. Participants were asked to walk for 2 minutes at maximal effort around cones placed in a 50-foot hallway. The 2MWT has high intrarater reliability for people with MS and can detect meaningful changes in walking endurance.22  An improvement of 11.4 meters has been suggested as a minimally important change (MIC).23  The TUG was used as a measure of balance and mobility. Participants began while seated in a chair with armrests. Participants were instructed to stand up and walk around a cone placed 3 meters away and return to a seated position as quickly as possible, safely. The time taken to complete this task was recorded by a stopwatch. The TUG has good construct validity for people with MS24  but lacks sensitivity in detecting clinically relevant changes.25 

Step count was measured by the remote capture of steps during FES device use. The EvoWalk device collects step count through an accelerometer and gyroscope with a sampling frequency of 66.7 Hz. The step count feature has yet to be validated against accelerometry in people with MS.26  Mean FES device wear-time (ie, wearing and using the FES device) and weekly step count (ie, a continuous 7-day period) were derived from the EvoWalk software and averaged for each participant for their first and last week of FES use.

Self-Reported Measures

The Multiple Sclerosis Impact Scale is a 29-item questionnaire (MSIS-29); it assesses the impact that MS has on a person's physical and psychological QOL. Items related to physical and psychological function are rated on a 1-to-5 response scale as being affected by MS or bothersome from “not at all” to “extremely.” Scores range from 29 to 145, with a score of 29 indicating that physical and psychological functions have not been affected at all and a score of 145 indicating that these domains have been extremely affected by the individual's MS. The MSIS-29 is a highly recommended outcome measure with strong psychometrics.27  An improvement of 8 points has been suggested as the MCID.28 

The Multiple Sclerosis Walking Scale (MSWS-12) is a 12-item questionnaire; it assesses the impact that MS has on a person's walking function over the previous 2 weeks. Items related to physical and psychological function are rated on a 1-to-5 response scale as being affected by MS from “not at all” to “extremely.” Individual item scores are summed, and the total score is standardized to a scale ranging from 0 to 100, with a lower score indicating that MS has not affected walking function and a higher score indicating that walking function has been extremely affected by MS. The MSWS-12 is a valid and reliable measure of walking ability in people with MS.29  An improvement of 8.6 points has been suggested as a MIC.25 

The Modified Fatigue Impact Scale (MFIS-5) is a 5-item questionnaire that quantifies fatigue symptoms in people with MS. This is a shortened version of the MFIS and includes questions pertaining to the physical, cognitive, and psychosocial subscales of the longer questionnaire. The MFIS-5 scores each item on a 0 (never) to 4 (almost always) rating scale. Scores range from 0 to 20, with a higher score indicating more fatigue.30  An improvement of 4 points has been suggested as the MCID.31 

Data Analysis

Descriptive statistics for demographic data and feasibility (ie, retention, acceptability, adherence) are presented as mean plus or minus SD or medians, as appropriate. The impact of FES was measured by the initial and combined orthotic effect. The initial orthotic effect was defined as the difference in walking without FES and with FES at baseline; the combined orthotic effect was defined as the difference between walking without FES at baseline and with FES after the 8-week intervention period.16  The initial and combined orthotic effects are presented using mean change and 95% CI. Due to the exploratory nature of this case series, only participants who completed both assessments were considered for analysis. All analyses were performed in SPSS Statistics 28 (IBM Corp).

Twelve individuals were enrolled, and 11 completed the study (9 women and 2 men); 1 participant withdrew after 4 weeks of the study because of an unrelated health problem. Participants ranged in age from 33 to 63 years and were 50.6 ± 10.8 years old. Baseline characteristics of the participants in this study are further outlined in TABLE 1.

TABLE 1

Baseline Demographics, Acceptability, and Device Wear-Time (N = 11)

Baseline Demographics, Acceptability, and Device Wear-Time (N = 11)
Baseline Demographics, Acceptability, and Device Wear-Time (N = 11)

Feasibility

The use of telerehabilitation to remotely monitor and progress FES intervention was feasible, acceptable, and safe: 92% of participants completed the study (11/12) and 91% (10/11) of participants who completed the 10-item Telerehabilitation

Satisfaction Survey21  reported being “more than satisfied” with the telerehabilitation sessions (Table 1). Adherence to the biweekly telerehabilitation sessions during the study was 100% (33/33 telerehabilitation sessions) for the 11 participants who completed the study, and there was an average increase in FES device wear-time of 129.4 ± 124.3 minutes from week 1 to week 8 of FES use. There was 1 noninjurious fall reported and it was determined to be unrelated to any study activities. No other adverse events were reported.

Functional Mobility

There was no initial orthotic effect on the T25FW (TABLE 2). There was an initial orthotic effect on the 2MWT (Table 2) and TUG exceeding the MIC for 18% (2/11) and 55% (6/11) of participants, respectively. Five out of 11 (45%) participants demonstrated a combined orthotic effect with FES on the T25FW exceeding the MCID (Table 2). Fifty-five percent (6/11) and 82% (9/11) of participants demonstrated a combined orthotic effect with FES exceeding the MIC on the 2MWT (Table 2) and TUG, respectively. Eighteen percent (2/11) of participants required assistive devices (ie, 4-wheeled walker) during functional mobility testing and maintained the same level of assistive device use throughout the duration of the study. Individual participant data for the T25FW and 2MWT data can be found in TABLE S2 and TABLE S3, respectively.

TABLE 2

Results of Walking Tests and Patient-Reported Outcomes

Results of Walking Tests and Patient-Reported Outcomes
Results of Walking Tests and Patient-Reported Outcomes
TABLE S2

Participant Results of T25FW

Participant Results of T25FW
Participant Results of T25FW
TABLE S3

Participant Results of the 2MWT

Participant Results of the 2MWT
Participant Results of the 2MWT

Step Count

Remote step count data were available for 10 participants during FES use (TABLE S4). There was an average increase in weekly step count of 1537.7 ± 1185.1 steps (Table 2) from week 1 to week 8 of FES use. A comparison to the MCID for daily step count was not possible given the limited amount of time that step data were recorded (ie, during FES device use).32 

TABLE S4

Participant FES Wear-Time and Step Count

Participant FES Wear-Time and Step Count
Participant FES Wear-Time and Step Count

Self-Reported Measures

There were improvements at study completion exceeding the MCID for 64% (7/11) on the MSIS-29 and 36% (4/11) on the MSWS-12 (Table 2). There were no clinically meaningful improvements observed in fatigue (Table 2). Individual participant data for all self-reported measures can be found in TABLE S5.

TABLE S5

Participant PROM Results

Participant PROM Results
Participant PROM Results

Eight weeks of FES training monitored and advanced remotely through telerehabilitation was feasible, acceptable, and showed an overall positive impact on functional mobility and QOL in this case series of 11 participants with MS.

To our knowledge, this is the first study utilizing telerehabilitation-assisted delivery of FES. This novel use of telerehabilitation was feasible, safe, and practical as highlighted by high adherence and participant satisfaction. The majority of participants in the present study demonstrated a mean increase in wear-time of the FES device from week 1 to week 8, indicating the relative usability of the device. This is an important finding, as other researchers have highlighted several barriers to FES use in people with MS.12,13  Squires et al33  found that factors related to a supportive environment played a significant role in predicting long-term use of assistive technology. Building upon this, Miller Renfrew et al13  identified that access to professional help and the support of caregivers were crucial in facilitating use of FES. Telerehabilitation offers physical therapists a means to address these barriers and promote a positive experience with FES, without requiring patients and caregivers to engage in additional in-person visits.34  By mitigating some of the commonly encountered obstacles to FES, telerehabilitation may encourage increased usage and offset the higher cost of FES over a longer period of time.35 

Due to the exploratory design of this case series, conclusions cannot be drawn regarding the efficacy of telerehabilitation-delivered FES on functional mobility and self-reported outcomes. Numerous studies36,37  have found that the provision of behavioral interventions delivered remotely increased physical activity and reduced self-reported walking impairment in people with MS. Knowing this, it is possible that the improvements observed in the present study were attributable to the increased interaction between participants and the study physical therapist, which may have been an impetus to be more physically active. Regardless of the mechanism, the growing body of evidence supporting the remote delivery of behavioral interventions to increase physical activity in people with MS may be translatable to FES delivery by providing patients with necessary skills to promote a specific behavior change (eg, FES use).

The current study did not demonstrate initial orthotic effects on the T25FW. While there are mixed results in the literature on the initial orthotic effects of FES in people with MS,11,35,38  9 of 15 studies included in a recent systematic review and meta-analysis found significant initial orthotic effects on short walking tests.10  One explanation for the variability across these results could be the higher average baseline walking speed of the participants in this case series (1.03 ± 0.30 m/s) compared to those included (0.69 ± 0.03 m/s) in the aforementioned review.10  It is possible that individuals who walk faster may take longer to see benefit from the use of FES, as these individuals may have less marked gait abnormalities and may not experience an immediate impact compared with those with more significant disability. Barr et al39  found similar results in an MS cohort, where participants with an average gait speed of 1.18 m/s did not demonstrate initial orthotic effects, but demonstrated significant combined orthotic effects after 8 weeks of FES use. Our results extend these findings by demonstrating that individuals with a faster baseline gait speed (> 0.8 m/s) and mild disability can also benefit from FES and highlight the importance of a training period to facilitate improvements in gait performance. The improvements observed in the present study were comparable to those reported in previous literature supporting the positive effects of FES on short10  and long walking tests40  in people with MS.

An important finding in this case series was the changes on patient-reported outcomes. The scores on the MSIS-29 indicate the intervention may have had positive effects on QOL. Previous studies using FES in people with MS demonstrate conflicting results regarding the impact on QOL (using the MSIS-29).40,41  Downing et al41  studied a 2-week FES intervention which yielded significant improvements in QOL, whereas a 12-week FES intervention in a study by van der Linden et al40  did not. Our findings show improved QOL over an 8-week period of FES use, contrasting previous results from longer FES intervention periods,40  during which people with MS may be more likely to experience symptom fluctuations. There are a few possible explanations for our results. First, the use of FES may have provided a physical and psychological benefit, in which participants could walk faster with less concern about their foot drop. This may have led to increased motivation and confidence to participate in meaningful activities that center around ambulation. Indeed, we saw some preliminary evidence of this as participants modestly increased their weekly step count from week 1 to week 8 of FES use. It is also plausible that the telerehabilitation component of this study contributed to the positive QOL improvements. Selfmanagement interventions are increasingly being used and have demonstrated a positive impact on managing MS-related symptoms.42  The telerehabilitation sessions used in this study may have fostered the development of self-management strategies with the FES device, which could have led to a sense of empowerment and increased confidence in its usage. If participants feel competent in navigating the common challenges to FES use, they may be more likely to utilize the device and garner the benefits associated with a training period.

A small group of participants (36%) demonstrated clinically meaningful improvements on the MSWS-12, which suggests these individuals experienced a positive effect from the intervention on self-perceived effects of MS on walking ability. The lack of improvement in the other participants is in line with previous literature with FES intervention periods greater than 2 weeks, which failed to show improvements on the MSWS-12.40  It is possible the MSWS-12 is not as responsive as other outcomes to detect changes in walking-related function across participants with a broad range of disability levels.43 

While reductions in energy cost of walking and perceived exertion with FES use have been previously reported,40,44  we observed no changes in self-reported fatigue. A limitation of the current study was the measurement of fatigue using solely the MFIS-5 before and after the intervention, which provided narrow insight into the multidimensional nature of fatigue and did not capture changes in performance fatigue or fatigability.45  Future studies may consider increasing the frequency of self-reported fatigue measures and ecologically valid activity monitoring using wearable devices to further elucidate the impact of FES use on fatigue and fatigability.46 

This study has several limitations. Due to the feasibility nature of this case series and small sample size, the results should be interpreted cautiously. There was no comparison group and so it is not known if telerehabilitation-delivered FES intervention was the cause of the change or if it is superior to a traditional, in-person approach. Potential bias may have been introduced as no blinding was performed; the same member of the study team conducted the telerehabilitation sessions as well as the study outcome tests. Furthermore, the FES intervention period in this study was 8 weeks, which varies from other studies investigating FES in MS, and this may limit the ability to compare the effects of FES dosing needed to produce meaningful change in functional mobility and QOL measures. Lastly, while the virtual mode of FES delivery was feasible with high adherence, future studies with qualitative thematic analysis and a longer follow-up period are needed to determine which aspects of the telerehabilitation sessions are most useful, and if this can reduce the high levels of assistive device abandonment in people with MS.

The results of this study demonstrate the feasibility of using telerehabilitation to monitor and advance an FES intervention for people with MS. Eight weeks of FES training with 3 telerehabilitation visits had positive effects on functional mobility and QOL. Further research is needed to compare FES training delivered in-clinic versus remotely via telerehabilitation to determine which method is superior in improving mobility outcomes and adherence to FES use, and which is preferable to individuals with MS. Investigation of optimal dosing for FES training is also needed to guide clinicians in prescription of this intervention.

PRACTICE POINTS
  • The use of telerehabilitation to monitor a functional electrical stimulation (FES) intervention for people with foot drop due to multiple sclerosis (MS) was safe, feasible, and practical.

  • The results of this case series suggest that telerehabilitation-delivered FES may have a positive impact on walking and quality of life for people with MS and may mitigate commonly encountered obstacles to continuous FES use.

For their contributions to this study, we thank Alexander Dien, Erica Pitsch, and Natalie Anzures.

1.
Walton
C
,
King
R
,
Rechtman
L
, et al.
Rising prevalence of multiple sclerosis worldwide: insights from the Atlas of MS, third edition
.
Mult Scler
.
2020
;
26
(
14
):
1816
1821
.
doi:
10.1177/1352458520970841
2.
Larocca
NG
.
Impact of walking impairment in multiple sclerosis: perspectives of patients and care partners
.
Patient
.
2011
;
4
(
3
):
189
201
.
doi:
10.2165/11591150-000000000-00000
3.
Heesen
C
,
Böhm
J
,
Reich
C
,
Kasper
J
,
Goebel
M
,
Gold
SM
.
Patient perception of bodily functions in multiple sclerosis: gait and visual function are the most valuable
.
Mult Scler
.
2008
;
14
(
7
):
988
991
.
doi:
10.1177/1352458508088916
4.
Motl
RW
,
Pilutti
LA
,
Learmonth
YC
,
Goldman
MD
,
Brown
T
.
Clinical importance of steps taken per day among persons with multiple sclerosis
.
PLoS One
.
2013
;
8
(
9
):
e73247
.
doi:
10.1371/journal.pone.0073247
5.
Goldman
MD
,
Motl
RW
,
Rudick
RA
.
Possible clinical outcome measures for clinical trials in patients with multiple sclerosis
.
Ther Adv Neurol Disord
.
2010
;
3
(
4
):
229
239
.
doi:
10.1177/1756285610374117
6.
Hobart
J
,
Blight
AR
,
Goodman
A
,
Lynn
F
,
Putzki
N
.
Timed 25-foot walk: direct evidence that improving 20% or greater is clinically meaningful in MS
.
Neurology
.
2013
;
80
(
16
):
1509
1517
.
doi:
10.1212/WNL.0b013e31828cf7f3
7.
Gunn
H
,
Creanor
S
,
Haas
B
,
Marsden
J
,
Freeman
J
.
Frequency, characteristics, and consequences of falls in multiple sclerosis: findings from a cohort study
.
Arch Phys Med Rehabil
.
2014
;
95
(
3
):
538
545
.
doi:
10.1016/j.apmr.2013.08.244
8.
Alnajjar
F
,
Zaier
R
,
Khalid
S
,
Gochoo
M
.
Trends and technologies in rehabilitation of foot drop: a systematic review
.
Expert Rev Med Devices
.
2021
;
18
(
1
):
31
46
.
doi:
10.1080/17434440.2021.1857729
9.
Smith
AD
,
Prokopiusova
T
,
Jones
R
,
Burge
T
,
Rasova
K
.
Functional electrical stimulation for foot drop in people with multiple sclerosis: the relevance and importance of addressing quality of movement
.
Mult Scler
.
2021
;
27
(
5
):
653
660
.
doi:
10.1177/1352458520923958
10.
Miller
L
,
McFadyen
A
,
Lord
AC
, et al.
Functional electrical stimulation for foot drop in multiple sclerosis: a systematic review and meta-analysis of the effect on gait speed
.
Arch Phys Med Rehabil
.
2017
;
98
(
7
):
1435
1452
.
doi:
10.1016/j.apmr.2016.12.007
11.
Gervasoni
E
,
Parelli
R
,
Uszynski
M
, et al.
Effects of functional electrical stimulation on reducing falls and improving gait parameters in multiple sclerosis and stroke
.
PM R
.
2017
;
9
(
4
):
339
347.e1
.
doi:
10.1016/j.pmrj.2016.10.019
12.
Bulley
C
,
Mercer
TH
,
Hooper
JE
,
Cowan
P
,
Scott
S
,
van der Linden
ML
.
Experiences of functional electrical stimulation (FES) and ankle foot orthoses (AFOs) for foot-drop in people with multiple sclerosis
.
Disabil Rehabil Assist Technol
.
2015
;
10
(
6
):
458
467
.
doi:
10.3109/17483107.2014.913713
13.
Miller Renfrew
L
,
Flowers
P
,
Lord
AC
, et al.
An exploration of the experiences and utility of functional electrical stimulation for foot drop in people with multiple sclerosis
.
Disabil Rehabil
.
2020
;
42
(
4
):
510
518
.
doi:
10.1080/09638288.2018.1501100
14.
Auchstaetter
N
,
Luc
J
,
Lukye
S
, et al.
Physical therapists’ use of functional electrical stimulation for clients with stroke: frequency, barriers, and facilitators
.
Phys Ther
.
2016
;
96
(
7
):
995
1005
.
doi:
10.2522/ptj.20150464
15.
Sugawara
AT
,
Ramos
VD
,
Alfieri
FM
,
Battistella
LR
.
Abandonment of assistive products: assessing abandonment levels and factors that impact on it
.
Disabil Rehabil Assist Technol
.
2018
;
13
(
7
):
716
723
.
doi:
10.1080/17483107.2018.1425748
16.
Johnston
TE
,
Keller
S
,
Denzer-Weiler
C
,
Brown
L
.
A clinical practice guideline for the use of ankle-foot orthoses and functional electrical stimulation post-stroke
.
J Neurol Phys Ther
.
2021
;
45
(
2
):
112
196
.
doi:
10.1097/NPT.0000000000000347
17.
Bove
R
,
Bevan
C
,
Crabtree
E
, et al.
Toward a low-cost, in-home, telemedicine-enabled assessment of disability in multiple sclerosis
.
Mult Scler
.
2019
;
25
(
11
):
1526
1534
.
doi:
10.1177/1352458518793527
18.
Landi
D
,
Ponzano
M
,
Nicoletti
CG
, et al.
Patient's point of view on the use of telemedicine in multiple sclerosis: a web-based survey
.
Neurol Sci
.
2022
;
43
(
2
):
1197
1205
.
doi:
10.1007/s10072-021-05398-6
19.
Di Tella
S
,
Pagliari
C
,
Blasi
V
,
Mendozzi
L
,
Rovaris
M
,
Baglio
F.
Integrated telerehabilitation approach in multiple sclerosis: a systematic review and meta-analysis
.
J Telemed Telecare
.
2020
;
26
(
7-8
):
385
399
.
doi:
10.1177/1357633×19850381
20.
Thompson
AJ
,
Banwell
BL
,
Barkhof
F
, et al.
Diagnosis of multiple sclerosis: 2017 revisions of the McDonald criteria
.
Lancet Neurol
.
2018
;
17
(
2
):
162
173
.
doi:
10.1016/S1474-4422(17)30470-2
21.
Miller
MJ
,
Pak
SS
,
Keller
DR
,
Barnes
DE
.
Evaluation of pragmatic telehealth physical therapy implementation during the COVID-19 pandemic
.
Phys Ther
.
2021
;
101
(
1
):
pzaa193
.
doi:
10.1093/ptj/pzaa193
22.
Gijbels
D
,
Eijnde
BO
,
Feys
P
.
Comparison of the 2- and 6-minute walk test in multiple sclerosis
.
Mult Scler
.
2011
;
17
(
10
):
1269
1272
.
doi:
10.1177/1352458511408475
23.
Baert
I
,
Freeman
J
,
Smedal
T
, et al.
Responsiveness and clinically meaningful improvement, according to disability level, of five walking measures after rehabilitation in multiple sclerosis: a European multicenter study
.
Neurorehabil Neural Repair
.
2014
;
28
(
7
):
621
631
.
doi:
10.1177/1545968314521010
24.
Kalron
A
,
Dolev
M
,
Givon
U
.
Further construct validity of the Timed Up-and-Go Test as a measure of ambulation in multiple sclerosis patients
.
Eur J Phys Rehabil Med
.
2017
;
53
(
6
):
841
847
.
doi:
10.23736/S1973-9087.17.04599-3
25.
Baert
I
,
Smedal
T
,
Kalron
A
, et al.
Responsiveness and meaningful improvement of mobility measures following MS rehabilitation
.
Neurology
.
2018
;
91
(
20
):
e1880
e1892
.
doi:
10.1212/WNL.0000000000006532
26.
Fang
J
,
Kumar
DP
,
Celnikier
E
,
Snyder
M
,
Potts
E
,
Sosnoff
J
.
Enhancing mobility in persons with footdrop in multiple sclerosis through functional electrical stimulation
.
Mult Scler Relat Disord
.
2023
;
73
:
104651
.
doi:
10.1016/j.msard.2023.104651
27.
Ramp
M
,
Khan
F
,
Misajon
RA
,
Pallant
JF
.
Rasch analysis of the Multiple Sclerosis Impact Scale MSIS-29
.
Health Qual Life Outcomes
.
2009
;
7
:
58
.
doi:
10.1186/1477-7525-7-58
28.
Costelloe
L
,
O'Rourke
K
,
Kearney
H
, et al.
The patient knows best: significant change in the physical component of the Multiple Sclerosis Impact Scale (MSIS-29 physical)
.
J Neurol Neurosurg Psychiatry
.
2007
;
78
(
8
):
841
844
.
doi:
10.1136/jnnp.2006.105759
29.
McGuigan
C
,
Hutchinson
M
.
Confirming the validity and responsiveness of the Multiple Sclerosis Walking Scale-12 (MSWS-12)
.
Neurology
.
2004
;
62
(
11
):
2103
2105
.
doi:
10.1212/01.wnl.0000127604.84575.0d
30.
Ritvo
PG
,
Fischer
JS
,
Miller
DM
,
Andrews
H
,
Paty
DW
,
LaRocca
,
NG
;
National Multiple Sclerosis Society
.
Multiple Sclerosis Quality of Life Inventory: a user's manual
.
1997
. Accessed July 17, 2023. http://walkcoc.nationalmssociety.org/docs/HOM/MSQLI_Manual_and_Forms.pdf
31.
Meca-Lallana
V
,
Brañas-Pampillón
M
,
Higueras
Y
, et al.
Assessing fatigue in multiple sclerosis: psychometric properties of the five-item Modified Fatigue Impact Scale (MFIS-5)
.
Mult Scler J Exp Transl Clin
.
2019
;
5
(
4
):
2055217319887987
.
doi:
10.1177/2055217319887987
32.
Motl
RW
,
Sandroff
BM
,
Sosnoff
JJ
.
Commercially available accelerometry as an ecologically valid measure of ambulation in individuals with multiple sclerosis
.
Expert Rev Neurother
.
2012
;
12
(
9
):
1079
1088
.
doi:
10.1586/ern.12.74
33.
Squires
LA
,
Williams
N
,
Morrison
VL
.
Matching and accepting assistive technology in multiple sclerosis: a focus group study with people with multiple sclerosis, carers and occupational therapists
.
J Health Psychol
.
2019
;
24
(
4
):
480
494
.
doi:
10.1177/1359105316677293
34.
Bove
R
,
Garcha
P
,
Bevan
CJ
,
Crabtree-Hartman
E
,
Green
AJ
,
Gelfand
JM
.
Clinic to in-home telemedicine reduces barriers to care for patients with MS or other neuroimmunologic conditions
.
Neurol Neuroimmunol Neuroinflamm
.
2018
;
5
(
6
):
e505
.
doi:
10.1212/NXI.0000000000000505
35.
Renfrew
LM
,
Paul
L
,
McFadyen
A
, et al.
The clinical- and cost-effectiveness of functional electrical stimulation and ankle-foot orthoses for foot drop in multiple sclerosis: a multicentre randomized trial
.
Clin Rehabil
.
2019
;
33
(
7
):
1150
1162
.
doi:
10.1177/0269215519842254
36.
Motl
RW
,
Hubbard
EA
,
Bollaert
RE
, et al.
Randomized controlled trial of an e-learning designed behavioral intervention for increasing physical activity behavior in multiple sclerosis
.
Mult Scler J Exp Transl Clin
.
2017
;
3
(
4
):
2055217317734886
.
doi:
10.1177/2055217317734886
37.
Rintala
A
,
Hakala
S
,
Paltamaa
J
,
Heinonen
A
,
Karvanen
J
,
Sjögren
T
.
Effectiveness of technology-based distance physical rehabilitation interventions on physical activity and walking in multiple sclerosis: a systematic review and meta-analysis of randomized controlled trials
.
Disabil Rehabil
.
2018
;
40
(
4
):
373
387
.
doi:
10.1080/09638288.2016.1260649
38.
Barrett
C
,
Taylor
P
.
The effects of the odstock drop foot stimulator on perceived quality of life for people with stroke and multiple sclerosis
.
Neuromodulation
.
2010
;
13
(
1
):
58
64
.
doi:
10.1111/j.1525-1403.2009.00250.x
39.
Barr
CJ
,
Patritti
BL
,
Bowes
R
,
Crotty
M
,
McLoughlin
JV
.
Orthotic and therapeutic effect of functional electrical stimulation on fatigue induced gait patterns in people with multiple sclerosis
.
Disabil Rehabil Assist Technol
.
2017
;
12
(
6
):
560
572
.
doi:
10.3109/17483107.2015.1136702
40.
van der Linden
ML
,
Hooper
JE
,
Cowan
P
,
Weller
BB
,
Mercer
TH
.
Habitual functional electrical stimulation therapy improves gait kinematics and walking performance, but not patient-reported functional outcomes, of people with multiple sclerosis who present with foot-drop
.
PLoS One
.
2014
;
9
(
8
):e103368.
doi:
10.1371/journal.pone.0103368
41.
Downing
A
,
Van Ryn
D
,
Fecko
A
, et al.
Effect of a 2-week trial of functional electrical stimulation on gait function and quality of life in people with multiple sclerosis
.
Int J MS Care
.
2014
;
16
(
3
):
146
152
.
doi:
10.7224/1537-2073.2013-032
42.
Kratz
AL
,
Alschuler
KN
,
Williams
DA
,
Ehde
DM
.
Development and pilot testing of a web-based symptom management program for multiple sclerosis: My MS Toolkit
.
Rehabil Psychol
.
2021
;
66
(
2
):
224
232
.
doi:
10.1037/rep0000375
43.
Mañago
MM
,
Cohen
ET
,
Cameron
MH
,
Christiansen
CL
,
Bade
M
.
Reliability, validity, and responsiveness of the Patient-Specific Functional Scale for measuring mobility-related goals in people with multiple sclerosis
.
J Neurol Phys Ther
.
2023
;
47
(
3
):
139
145
.
doi:
10.1097/NPT.0000000000000439
44.
Paul
L
,
Rafferty
D
,
Young
S
,
Miller
L
,
Mattison
P
,
McFadyen
A
.
The effect of functional electrical stimulation on the physiological cost of gait in people with multiple sclerosis
.
Mult Scler
.
2008
;
14
(
7
):
954
961
.
doi:
10.1177/1352458508090667
45.
Kluger
BM
,
Krupp
LB
,
Enoka
RM
.
Fatigue and fatigability in neurologic illnesses: proposal for a unified taxonomy
.
Neurology
.
2013
;
80
(
4
):
409
416
.
doi:
10.1212/WNL.0b013e31827f07be
46.
Kratz
AL
,
Murphy
SL
,
Braley
TJ
.
Ecological momentary assessment of pain, fatigue, depressive, and cognitive symptoms reveals significant daily variability in multiple sclerosis
.
Arch Phys Med Rehabil
.
2017
;
98
(
11
):
2142
2150
.
doi:
10.1016/j.apmr.2017.07.002

FINANCIAL DISCLOSURES:

Elie Celnikier is the chief science officer for Evolution Devices, Inc. He was responsible for deriving the step-count and wear-time data from the EvoWalk FES device.

FUNDING/SUPPORT:

This study was partially supported by a seed grant from the University of California at San Francisco Department of Physical Therapy and Rehabilitation Science.

PRIOR PRESENTATION:

This material was presented in abstract form at the International Symposium of Gait and Balance in Multiple Sclerosis; April 2023; Aurora, CO.