The present study focused on investigating whether universal screwdriver kits cause less friction between the screwdriver and the abutment screw than original screwdrivers. For this purpose, two original screwdrivers (Straumann and BEGO) and a universal screwdriver kit (bredent) were investigated. On one implant per screwdriver, 26 abutments were properly attached one after the other with the corresponding abutment screws. After tightening the abutment screw, the force required to pull the screwdriver off the screw head was determined with a spring balance.For both manufacturers, greater pull-off forces were measured when using the original screwdrivers than when using the universal screwdriver. The pull-off force required for the Straumann original screwdriver was 3.7 N ± 1.4, while that required for the universal screwdriver was 0.1 N ± 0.1 (p < 0.001).The mean pull-off force was 1.5 N ± 1.5 for the BEGO original screwdriver and 0.7 N ±0.9 for the universal screwdriver (p = 0.19). Using original manufacturer-supplied screwdrivers could thus minimize the risk of the screwdriver slipping out of the screw head during dental treatment and being swallowed or aspirated by the patient.
Skip Nav Destination
Article navigation
Research Article|
March 09 2023
Comparing original and universal screwdrivers and how they affect friction in the screw
Marcus Schiller
Marcus Schiller
Medical High School Hannover
Oral and Maxillofacial Surgery
Carl-Neuberg-Str. 1
GERMANY
Hannover
Lower Saxony
30625
Search for other works by this author on:
J Oral Implantol (2023)
Article history
Received:
August 03 2020
Revision Received:
November 11 2020
Revision Received:
March 07 2021
Revision Received:
August 05 2021
Accepted:
September 21 2022
Citation
Vasileios Alevizakos, Nadine Achilles, Björn Rahlf, Philipp Jehn, Constantin von See, Marcus Schiller; Comparing original and universal screwdrivers and how they affect friction in the screw. J Oral Implantol 2023; doi: https://doi.org/10.1563/aaid-joi-D-20-00249
Download citation file:
Sign in
Don't already have an account? Register
Client Account
You could not be signed in. Please check your email address / username and password and try again.
Could not validate captcha. Please try again.
Sign in via your Institution
Sign in via your Institution
1
Views
Citing articles via
Enhanced Lip-Contour During Facial Scan for 3D DSD and Implant Planning: A Blue Screen Approach
Jing-Wen Yang, DDS, PhD, Yu-Jia Zhu, DDS, Kelvin I. Afrashtehfar, DDS, MSc Dr med dent, FDSRCS, FRCDC, Yi-Jiao Zhao, SE, Yong Wang, MS
Immediate Implant and Customized Healing Abutment Promotes Tissues Regeneration: A 5-Year Clinical Report
Francesco Corrado, DDS, Simone Marconcini, DDS, PhD, Saverio Cosola, DDS, Enrica Giammarinaro, DDS, Ugo Covani, MD
Marijuana – Clinical Concerns for Implant Dentistry – Part 2
James L. Rutkowski, DMD, PhD
Does the Prosthetic Phase of Dental Implants Justify the Prescription of Preventive Antibiotics in Healthy Patients? A Systematic Review
Angel-Orión Salgado-Peralvo, DDS, MSc, MSc, PhD, Andrea Uribarri, DDS, MSc, MSc, Juan-Francisco Peña-Cardelles, DDS, MSc, PhD, Naresh Kewalramani, DDS, MSc, Jorge Luis Garnier Rodríguez, DDS, MSc, Eugenio Velasco-Ortega, MD, PhD
Comparing original and universal screwdrivers and how they affect friction in the screw
Vasileios Alevizakos, Nadine Achilles, Björn Rahlf, Philipp Jehn, Constantin von See, Marcus Schiller