An in vitro experimental model was designed and tested to determine the influence that peri-implant strain may have on the overall crestal bone. Strain gages were attached to polymethylmethacrylate (PMMA) models containing a screw-type root form implant at sites 1 mm from the resin-implant interface. Three different types of crown superstructures (cemented, 1-screw [UCLA] and 2-screw abutment types) were tested. Loading (1 Hz, 200 N load) was performed using a MTS Mechanical Test System. The strain gage data were stored and organized in a computer for statistical treatment. Strains for all abutment types did not exceed the physiological range for modeling and remodeling of cancellous bone, 200–2500 μɛ (microstrain). For approximately one-quarter of the trials, the strain values were less than 200 μɛ the zone for bone atrophy. The mean microstrain obtained was 517.7 μɛ. In conclusion, the peri-implant strain in this in vitro model did not exceed the physiologic range of bone remodeling under axial occlusal loading.
Introduction
While the use of endosseous implants has achieved success rates that are generally greater than 95%,1 the issue of crestal bone resorption continues to receive significant interest from both researchers and clinicians. There are a number of theories to explain this phenomenon, including the following: the trauma associated with second stage surgery,2 plaque-associated peri-implantitis,3 excessive force transfer associated with occlusal loads during tooth contact,4 the inability of the rigid bone-implant interface to absorb physiologic forces without a periodontal ligament—particularly when the bone is thin, and stress shielding.5,6,7,8 These conflicting hypotheses suggest that a basic understanding of the mechanisms underlying crestal bone changes is still lacking, since occlusal forces for patients with implant-supported restorations appear to be within the range of the masticatory forces observed in the natural dentition.6,9
Both in vitro photoelastic studies and finite element computer modeling suggest that the crestal bone receives significant shear stress, particularly in cortical areas. Nonetheless, Pilliar and others10,11 claim that, at least under some circumstances, the crestal bone is stress shielded and receives an inadequate stimulus to maintain bone volume.
Martin and Burr12 and Frost13 categorize the effect of compressive strain on bone as demonstrating a physiological modeling-remodeling zone between 200 and 2500 μɛ, an overload zone between 2500 and 4000 μɛ (where damage and microcracking of the bone occurs with some repair), and finally a pathological overload zone of strains greater than 4000 μɛ where little or no repair of the bone is observed. Furthermore, Mori and Burr14 show that microdamage can be identified in areas of active remodeling.
Application of these observations to the dental implant-bone interface suggests this as a mechanism to explain the repair of the crestal interface from fatigue-induced microdamage. For example, Hoshaw et al15 observed an increase in both the modeling and remodeling responses with a net bone loss near the coronal portion of the loaded implants. Another mechanism suggested is that abutment movement on the implant creates strain in the peri-implant crestal bone by distortion of the collar of the implant.16 However, other investigators suggest stress shielding—that is, a lack of adequate strain in the crestal bone—as an explanation for the resorption seen in the crestal bone.17,18
The purpose of these experiments was to examine the magnitude of the strains in the crestal bone adjacent to the machined collar of a dental implant in an in vitro model system.
Methods
The overall experimental strategy was to develop an in vitro system in which a strain gage was located proximal to the crestal area of the implant to measure the strain from occlusal loading.
Model fabrication
We constructed a wax pattern of a box 20 mm long, 10 mm wide, and 15 mm high with an open top and containing a hollow wax cylinder with inner diameter of 5 mm and a wall thickness of 1 mm (Figure 1). It was invested in an elastic polyvinylsiloxane material (Reprosil, Dentsply International Inc, Milford, Del) to obtain a mold. Polymethylmethacrylate (PMMA) was utilized as the material in which the implants were embedded because it has a modulus of elasticity similar to that of cancellous bone,19–22 1.3–2.0 GPa;23 therefore, it is analogous to clinical situations in which implants are placed into Type 2 and Type 3 osseous beds.24 A mixture of PMMA resin monomer liquid and polymer powder (Lang Dental Mfg Co, Wheeling, Ill) in a ratio of 1:1.5 was poured into the mold and allowed to cure under 20 psi in warm water for 15 minutes. Five samples were made.
Cementation of implants and strain gages
Branemark Mark II, 4 mm diameter × 10 mm length, implants (Nobel Biocare Inc, Yorba Linda, Calif) were cemented into the cylinder of each experimental sample flush with the top using a jig from a surveyor (Ney Dental International, Bloomfield, Conn), such that the implant was placed perpendicular to the floor of the sample using a 1:1.5 mixture of acrylic resin monomer and polymer. Following polymerization, the PMMA coating around the implant surface was shaved with a milling machine (Cenders & Netaux SA, Bienne, Switzerland) to a 1 mm thickness to allow the positioning of the strain gages proximal to the implant in the crestal area (Figure 2).
Strain gages, type EA-06-062AQ-350 (Measurements Group, Inc, Raleigh, NC) were cemented on the buccal, mesial, and distal surfaces of each resin cylinder using cyanoacrylate cement (Measurements Group, Inc, Raleigh, NC) following the manufacturer's standard protocols. Following this, a 3 mm–thick layer of PMMA was applied uniformly over the strain gages to simulate the clinical situation where there is generally a minimum of 3 mm of bone between an implant and the neighboring tooth or implant. Embedding has no effect on the data recorded from the strain gages25,26 (Figure 3).
Abutments and crowns
Three types of abutment systems were utilized:
- 1.
Standard abutment with a cemented crown;
- 2.
UCLA abutment crown cast as one unit;
- 3.
Screw-retained crown with a separate abutment (EsthetiCone abutment and cast-to coping, NobelBiocare, Yorba Linda, Calif).
Standard laboratory procedures were used for fabrication of each crown. The superstructures for all 3 abutment systems had the same occluso-gingival, buccolingual, and mesiodistal dimensions, simulating a premolar. The height of the occlusal surface of the crowns measured 9 mm from the implant-abutment interface. One set of crowns was fabricated and utilized for all 5 samples to eliminate variability in the crown superstructure.
Occlusal loading protocol
Each of the samples was mounted onto a Dental Materials Testing System (Model #810, MTS System Corp, Minneapolis, Minn), a computer-guided simulated oral environment where occlusal loading (Figure 4) against a natural premolar loading was cyclic, 1 Hz, at 200 N, along the long axis of the implant to simulate vertical chewing27 (Figure 4).
Data collection
The signals from the strain gages were passed through a Model 2120A Strain Gage Conditioner/Amplifier (Measurements Group, Raleigh, NC) to a GraphTec WR7700 chart recorder (Schindler Elevator Corp, Randolph, NJ) at an acquisition rate of 32 Hz. The distal, mesial, and buccal strain gage outputs and the load sensor output from the MTS machine were recorded.
The load signal was sinusoidal with constant amplitude for each set of nine cycles with a frequency of 1 Hz. The data used for analysis included the peak values of the strain gage signal on the chart recorder for each sample run. A conversion factor was used to transform the chart peak values from millimeters into strain units using a calibration resistor in the 2120 signal conditioning unit (31 mm / 951 Micro-Strain). The calibration resistor is a fixed value resistor switched into either arm of a Wheatstone bridge to balance the circuit, giving a 951 μɛ value with an excitation voltage of 2 volts and a gage factor of 2.125. The strain signal was then routed through a data acquisition board interface (A/D converter) connected to a personal computer, and transferred into a file for analysis.
Statistical analysis
- 1.
The data was first subjected to a chi-square analysis by dichotomizing it into 2 groups. The first included strains between 0 and +200 μɛ. The second included strains ≥200 μɛ. The analysis was cross-tabulated with abutment type.
- 2.
A general linear model (GLM) procedure ANOVA for unbalanced data was used to compare the mean strains between abutment types by isolating the large experimental error due to trial-to-trial variations and placement of the gages among the 5 samples. Posthoc pair-wise comparisons were made using Tukey's protected t-test.28
Results
The mean microstrain values for the trials with the 3 different abutment types were as follows: 596.7 μɛ for the1-screw (UCLA) abutment type, 462.4 μɛ for the 2-screw abutment type, and 386.4 μɛ for the cemented abutment type.
When the data were dichotomized by microstrain values, 73.33% of the trials were within the physiologic range for the 1-screw (UCLA) abutment type, 76.3% for the 2-screw abutment type, and 68.89% for the cemented abutment type. The remaining trials for each of the abutment types were within the disuse atrophy range. No trials with microstrain values greater than 2500 μɛ were observed (Table 1).
Chi-square analysis
Comparison of the dichotomized data for the 3 different abutment types using Fisher exact test showed that the percentage of trials where the measured strains 1 mm from the implant-resin interface were within the physiological range were statistically the same for all three groups (Table 2).
Analysis of variance
Statistically significant differences were found between mean peri-implant strains observed with the 3 abutment types (P < .05). The 1-screw (UCLA) abutment type demonstrated the highest peri-implant strains. The 2-screw abutment type had an intermediate level of strains 1 mm from the implant-resin interface, while the cemented abutment type had the lowest strains. The ANOVA is shown in Table 3. No statistical differences were found in the data recorded from the buccal, mesial, and distal gages; thus, the data from the 3 sites were combined. Because of the large variation among samples, an additional nonparametric analysis—the Wilcoxan signed-rank test—was performed and demonstrated similar results (Table 4).
Discussion
These experiments demonstrate that, in this in vitro system, normal masticatory loads resulted in strains within the physiological capacities of peri-implant bone to remodel and maintain its integrity. In these experiments, 73% of the recorded microstrain values were categorized as within the physiological range, 200–2500 μɛ, while the remainder fell in the subphysiological range of 0–200 μɛ. Loads in the latter range could result in resorption associated with disuse atrophy.29 These results are based on masticatory loads of 200 N. Many individuals, however, function with masticatory loading that are less than 200 N,30–34 which could lead to a reduced strain that is associated with disuse atrophy. That being said, loading of the samples in these experiments was centered over the long axis of the fixture. Eccentric loading might result in magnification of the microstrain. This would be especially true in situations where the implant superstructure includes cantilevered units that would magnify the effects of the applied load, particularly in the crestal bone adjacent to the cantilever.35
There is considerable uncertainty regarding the etiology of the 1–1.5 mm of crestal bone loss commonly seen in the first year after implants are loaded. A number of photoelastic studies36–38 claim that relatively high levels of strain occur in the peri-implant crestal area. In support of these studies, finite element modeling studies by Borchers et al,39 Cook et a1,40 and Misch et al41 make similar predictions. Another explanation to support the hypothesis of high stress levels in the peri-implant area is based on the observation that there is a very significant discrepancy—10 times order of magnitude—between the modulus of elasticity of bone and that of the titanium fixture.42 As a result, relative motion is generated between the implant (made of titanium) and the bone, which flexes because of its lower modulus. Thus, high levels of strain may be generated in the bone. No experimental evidence, however, is currently available to support this theory.
A second contrasting theory presented by Pillar et al,31,43,16 with both experimental histological evidence and finite element modeling, claims that stress shielding occurs in the peri-implant crestal bone resulting in disuse atrophy. This does not exclude the possibility that in some circumstances with long cantilevers or short implants, excessive strain may lead to microfracture and destruction of the peri-implant bone.35
Vaillancourt et al,18 who proposed that the loss of crestal peri-implant bone observed in the first year of loading is a result of disuse atrophy, estimate on the basis of finite element analysis that a minimum load of 153 N on the implant is needed to produce an adequate strain in the crestal peri-implant bone to stimulate physiologic remodeling and prevent disuse atrophy.
Loading of the 3 different abutment systems resulted in statistical differences in the peri-implant crestal strains. The 1-screw (UCLA) system showed the highest strains at 596.7 μɛ; the 2-screw measured an intermediate strain at 462.4 μɛ; and the lowest stress transmission was observed around the cemented abutment at 386.4 μɛ. It has been suggested that the gold screw of the 2-screw system served to reduce the load transferred to the implant-resin interface.44 With these microstrain values, however, it is possible that both the gold screw and the cement layer reduced the stress transmission.
While the large standard deviations of the sample mean is disappointing, the strain records obtained during cyclic loading of each implant sample were consistent in each group of trials. The large intertrial variance was controlled by use of the GLM procedure, ANOVA, and the application of a nonparametric analysis. Most importantly, however, the large variability in the standard deviation did not affect the major findings of the study: that is, there were no trials in which the mean values exceeded the 2500 microstrain limit of physiologic modeling and remodeling activity. Furthermore, Brennan et al45 Wang and Stohler,46 and Kaukinen et al47 all demonstrate a wide inter-sample variation in their strain gage studies , similar to those observed in this research. This points to the sensitivity of the strain gage measuring systems and the difficulties associated with standardization of the samples.
The use of PMMA as a bone analogue has significant limitations. While Young's modulus for cancellous bone and PMMA are similar,18–22,25 bone is an heterogenous substance composed of cortical bone, as well as trabecular bone with anisotropic properties.48 PMMA, however, is a homogenous substance. Thus, extrapolation of the results of these experiments using PMMA to biologic systems must be approached with caution. It is more realistic to view these series of experiments as an initial study that should serve as the basis for more elaborate experimental methodologies that employ bone specimens and in vivo models.
There are a variety of surface treatments for the collar of the implant. These include machined, roughened (by etching, blasting, or oxidizing), and grooving. The stress transfer to the bone probably differs among them. This study examined the stress transfer from the machined collar. Further studies are needed to evaluate the effects of a more intimate interface between the collar of the implant and the bone, as occurs with a roughened or grooved surface. This does not, however, invalidate the results of the study; in fact, the differences—if indeed they exist—may be a partial explanation for the phenomenon of crestal bone loss. Another issue that requires further investigation is that of the effects of eccentric loading. This was a preliminary experiment to present an approach other than FEA to examine crestal bone stress and strain.
Conclusions
For this in vitro model system,
- 1.
All abutment types and locations around the implant showed axially loaded generated strains in the peri-implant region that fell within or below the physiologic zone for cancellous bone remodeling, which is 200–2500 μɛ.
- 2.
With the UCLA abutment, the highest strains occurred in the peri-implant region (596 μɛ), while the cemented abutment type showed the lowest strain distribution (386.41 μɛ), and the 2-screw abutment type showed strains between the other abutment types (462.4 μɛ).
- 3.
With axial loading, little difference was observed in the strain distribution in the peri-implant region on the facial, mesial, and distal sides of the implant.
Abbreviations