Initial breakdown of the implant-tissue interface generally begins at the crestal region in successfully osseointegrated implants. The purpose of this study was to evaluate the effect on crestal bone loss (CBL) around implants specially developed for immediate loading with a unique crestal drill. After 8 weeks postextraction, 6 young male mongrel dogs received 48 implants (XiVE) in the region corresponding to the 4 mandibular premolars. The implant sites were prepared according to the manufacturer's protocol with conventional standard drills. Before implant placement, the crestal drill was used in the experimental group but not in the control group. After a healing period of 12 weeks, the dogs were sedated and euthanized. Through linear measurements, from the top of the implant to the first bone-implant contact, the amount of CBL was determined. The histomorphometric results of CBL (mean ± SEM) were 0.88 ± 0.13 mm (range 0.0–3.0 mm) in the experimental group and 1.69 ± 0.17 mm (range 0.0–4.2 mm) in the control group. The difference was statistically significant (P < .05) when the implants were used as the experimental units. The statistical analysis also revealed significance when the dogs were used as the experimental units (P < .05). When the median was used for analyses, the CBL was 0.44 mm for the experimental group and 1.91 mm for the control group. Crestal bone loss was minimized when the crestal preparation protocol was carefully followed by using the osseocondensating XiVE implant system.

The success of dental implants is highly dependent on integration between the implant components and the hard and soft tissues. Initial breakdown of the implant-tissue interface generally begins at the crestal region in successfully osseointegrated dental implants.1–3 

Possible etiological factors of early implant bone loss (from implant placement to 1-year postloading) include surgical trauma, formation of biologic width, implant crestal module, and other factors.4 However, surgical trauma has been regarded as one of the most commonly suspected etiologies for early crestal bone loss (CBL).5 Implants that fail because of surgical trauma are often surrounded by a fibrous connective tissue or have an apical extension of the junctional epithelium.6 Heat generated at the time of drilling, elevation of periosteal flap, and excessive pressure at the crestal region during implant placement may contribute to implant bone loss during the healing period.

Eriksson and Albrektsson7 reported that the critical temperature for implant site preparation is 47°C for 1 minute or 40°C for 7 minutes. When the bone is overheated, the risk for implant failure is significantly increased. Periosteal elevation has also been speculated as one of the possible contributing factors for CBL. Wilderman and coworkers8 reported that mean horizontal bone loss after osseous surgery around teeth with periosteal elevation is approximately 0.8 mm and that the reparative potential is highly dependent on the amount of cancellous bone existing underneath the cortical bone.

Implant surgical techniques must be precise and atraumatic. This ensures congruent implant site preparation, which provides firm seating of the implant9 and avoids injury to the bone by excessive heat build-up.10 Structure and vascularization play an important role in the reaction of bone to the effect of surgical trauma. Well supplied with blood vessels, spongy bone dissipates the stress faster and has a greater capacity for repair than compact bone, which has a limited blood supply by comparison.11 

A new implant system has been developed that focuses on implant site preparation in areas with different bone qualities. The macroscopic design of the implant and threads promotes considerable bone condensation during its placement. Specific site preparation is performed according to the different bone densities to ensure a gentle and atraumatic implant insertion. A unique crestal drill, 0.2 mm wider than the final drill in sequence, is used to control the level of bony condensation during implant placement. The drill matches the shape of the implant and the collar flare. The use of the crestal drill influences the degree of internal bony condensation achieved by the implant during insertion. According to the manufacturer, depending on the bone quality, a varying insertion depth between 2 mm (for bone type IV) and 6 mm (for bone type I or II) is recommended for the crestal drill. The surgeon assesses bone density during initial site preparation with the pilot drill.12,13 If there is little or no resistance to the drill, bone density can be assumed to be low (bone type IV). This is taken into consideration during the use of the crestal drill. In sites with low bone quality, the preparation depth would be 2 mm. High resistance in the bone corresponds to bone type I, II, or III. In sites with dense bone, preparation is made to a depth of up to 6 mm.

The success of this immediate-loading implant system is highly dependent on the use of this crestal drill preparation. However, little is known regarding its effect on crestal resorption during healing. Therefore, the purpose of this study was to evaluate early CBL around XiVE implants when the crestal preparation protocol was followed.

The Animal Research Committee of the School of Dentistry of Ribeirão Preto, University of São Paulo, approved the experimental protocol. Six young male mongrel dogs, weighing approximately 15 kg each, were used in this study. The dogs had intact maxillae, had no mucosal lesions, and were in good systemic health as determined by a veterinarian after clinical examination.

The night before the surgery, the dogs received a combination of 20 000 IU penicillin and 1.0 g streptomycin per 10 kg body weight. Because each dose provides antibiotic coverage for 4 days, another dose was injected 4 days later, totaling 8 days of antibiotic coverage.14,15 In the first stage of the study, the dogs were sedated and anesthetized with 1 mL/kg thiopental (20 mg/kg thiopental diluted in 50 mL saline). After local anesthesia, full-thickness flaps were elevated bilaterally in the area of the first to fourth mandibular premolars. The teeth were sectioned in a bucco-lingual direction at the bifurcation so that the roots could be individually extracted without damaging the alveolus. After repositioning of the periodontal flaps, the wound was closed with resorbable sutures. The dogs were maintained on a soft diet for 14 days. Healing was evaluated periodically, and the teeth were cleaned monthly with ultrasonic scalers. To try to achieve a bone with intermediate quality, a healing period of 8 weeks was allowed before the next surgical intervention. The dogs were anesthetized in the same manner as described before. The night before the second surgery, the dogs received another dose of antibiotics in the same manner as described before. Full-thickness flaps were elevated bilaterally in the area corresponding to the first to fourth mandibular premolars (Figure 1). Four blasted and acid-etched surface, self-tapping threaded implants (XiVE, Dentsply Friadent, Mannheim, Germany), 4.5 mm in diameter and 9.5 mm in length, were placed bilaterally in each dog. The crestal drill was used in the experimental group but not in the control group. The experimental and control sites were randomly selected by a coin toss. The implant sites were sequentially enlarged to 4.5 mm in diameter with the pilot and spiral drills according to the manufacturer's standard protocol. After completion of the implant site preparation, crestal preparation of the bone in a vertical direction was performed in the experimental sites only. The crestal drill (Figure 2) with a cutting depth of 6 mm was used for this purpose. In this study, crestal preparation of 6 mm was chosen for producing a more representative field for histomorphometric analysis. The implants were placed at the bone crest according to the manufacturer's instruction (Figure 3). The dogs were maintained on a soft diet for 14 days. Healing was evaluated periodically and the teeth were cleaned monthly with ultrasonic points. The dogs were sedated and then euthanized with an overdose of thiopental 12 weeks after implant placement. Hemimandibles were harvested, dissected, and fixed in 4% phosphate-buffered formalin pH 7 for 10 days and transferred to a solution of 70% ethanol until processing. The specimens were dehydrated in increasing concentrations of alcohol up to 100%, infiltrated, embedded in methylmethacrylate resin, and then hard sectioned by the technique described by Donath and Breuner.16 The sections were prepared for histomorphometry and stained with Stevenell's blue and Alizarin red S for optic microscopic analysis.

Figures 1–3. Figure 1. Full-thickness flap and clinical photograph of the alveolar crest before second surgery. Figure 2. The crestal drill. Figure 3. Implants in position, occlusal view

Figures 1–3. Figure 1. Full-thickness flap and clinical photograph of the alveolar crest before second surgery. Figure 2. The crestal drill. Figure 3. Implants in position, occlusal view

Close modal

Histomorphometric analysis

Two 20- to 30-μm longitudinal histological sections from each implant were captured with a video camera Leica DC 300F (Leica Microsystems GmbH, Nussloch, Germany) joined to a stereomicroscope Leica MZFL III (Leica Microsystems GmbH). The images were analyzed through the Image J program (National Institutes of Health, Bethesda, Md), where CBL around the implants was determined. Through linear measurements, from the top of the implant to the first bone-implant contact, the amount of bone loss was determined (Figure 4). A blinded single examiner, with no knowledge if the sections were from experimental or control groups, made the measurements.

Figures 4–6. Figure 4. Histologic section depicting where the crestal bone loss was measured (original magnification, ×3.2). Figure 5. Crestal bone loss in the control group (original magnification, ×2.0) showing more crestal bone loss. The arrows indicate the first bone-implant contact. Figure 6. Crestal bone loss in the experimental group (original magnification, ×2.0) showing less crestal bone loss. The arrows indicate the first bone-implant contact

Figures 4–6. Figure 4. Histologic section depicting where the crestal bone loss was measured (original magnification, ×3.2). Figure 5. Crestal bone loss in the control group (original magnification, ×2.0) showing more crestal bone loss. The arrows indicate the first bone-implant contact. Figure 6. Crestal bone loss in the experimental group (original magnification, ×2.0) showing less crestal bone loss. The arrows indicate the first bone-implant contact

Close modal

Statistical analysis

Mean values and SEM were calculated. The data were grouped by using the implants and dogs as units for analysis. The mean differences of CBL between the groups were verified through the Mann-Whitney nonparametric test when the implants were used as the experimental units and through the Kruskal-Wallis test when the dogs were used as the experimental units, with a significance level of P ≤ .05. All calculations were performed by SPSS for Windows (SSP Inc, Chicago, Ill).

Clinical findings

Postextraction healing was uneventful in all dogs. At implant surgery 8 weeks later, the extraction sites appeared to have healed clinically. The alveolar ridge showed evidence of remodeling (Figure 1). During site preparation, resistance to the pilot drill resembled type II or III bone qualities. Healing was also uneventful after implant placement, without complications throughout the experimental period.

Histologic observations

The bone-implant interface had mineralized bone matrix in intimate contact with the implant surfaces in both the experimental and the control groups. Crestal bone loss was more pronounced in the control group (Figure 5) than in the experimental group (Figure 6).

Histomorphometric findings

The histomorphometric measurements CBL (mean ± SEM) around the implants are shown in Table 1. The data revealed that the CBL was 0.88 ± 0.13 mm (range 0–3.0 mm) in the experimental group and 1.69 ± 0.17 mm (range 0–4.2 mm) in the control group. The differences were statistically significant (P < .05) when implants were used as the experimental units. The statistical analysis also revealed statistical significance when the dogs were used as experimental units (P < .05), as seen in Table 2. Figure 7 presents the implant-frequency distribution in relation to the CBL for the experimental and control groups. The experimental group showed 5 implants without bone loss (0 mm), 9 implants with CBL of 1 mm, 6 implants with CBL of 2 mm, 3 implants with CBL of 3 mm, and only 1 implant with CBL of more than 3 mm. The control group showed 3 implants without bone loss (0 mm), 5 implants with CBL of 1 mm, 9 implants with CBL of 3 mm, and 3 implants with CBL of more than 3 mm. When the median was used for analyses, eliminating the best and worst results, the CBL was 0.44 mm for the experimental group and 1.91 mm for the control group.

Table 1

Crestal bone loss

Crestal bone loss
Crestal bone loss
Figure 7.

Frequency distribution of the number of implants in relation to the crestal bone loss values in millimeters

Figure 7.

Frequency distribution of the number of implants in relation to the crestal bone loss values in millimeters

Close modal

The current longevity of dental implants is strongly dependent on the interaction between implant components and oral tissues as a result of contemporary improvements in implant materials, geometry and surfacing, and the perfection of surgical techniques. Early CBL is often observed after the first year of function, followed by minimal bone loss annually thereafter. There is a lack of agreement as to why greater bone loss occurs during initial healing and the first year of implant function compared with the following years. Several possible etiologies of early implant bone loss (from placement to 1-year postloading) are suggested. Surgical trauma has been proposed as one of the most commonly suspected etiologies for early implant failure.5,17,18 Implants that fail because of surgical trauma are often surrounded by a fibrous connective tissue capsule or have an apical extension of the junctional epithelium.6 Heat generated at the time of osteotomy preparation,19 elevation of a periosteal flap,20,21 and excessive pressure at the crestal region during implant placement may contribute to implant bone resorption during the healing period. Preparation of the bone and subsequent implant placement damages blood vessels that provide nutrition to the surrounding bone, which in turn may compromise the adjacent bone. Another explanation for early bone loss may be related to stresses that are placed on the bone immediately surrounding the implant. These stresses may cause microfractures in the bone and stimulate bone turnover, which, without contiguous blood supply as a result of the implant body, may result in resorption.22 The marginal bone loss that has been reported in vivo may also be caused by high stress concentrations in the crestal region.23,24 When bone is overheated, the risk of implant failure is significantly increased. Overheating may be generated by excessive pressure at the crestal region during implant surgery. Wiskott and Belser25 also attributed CBL to increased pressure on the osseous bed during implant placement, establishment of a physiologic “biologic width,” stress shielding, and lack of adequate biomechanical coupling between the load-bearing implant surface and the surrounding bone.

Implant countersinking below the alveolar crest was recommended in the Branemark surgical protocol to minimize the risk of implant exposure during healing and to accommodate the platform to enhance the emergence profile for implant prostheses at the expense of crestal bone.26 To reduce the stress on crestal bone during implant placement, a new approach with a self-tapping implant and a crestal preparation protocol was developed.

Analysis between the experimental and the control groups showed statistically significant differences in CBL in favor of the experimental group (0.88 mm for experimental group and 1.69 mm for control group) measured from the top of the implant to the first bone-implant contact. Additional observations revealed that 14 of the 24 implants in the experimental group had CBL between 0 and 1 mm and only 3 implants had CBL of 3 mm, whereas 8 implants in the control group had CBL between 0 and 1 mm and 12 implants had CBL between 3 and 4 mm. The better distribution achieved in the experimental group reveals advantages in favor of the crestal preparation protocol. When median values were used, the CBL was 0.44 mm for the experimental group and 1.91 mm for the control group, thus increasing the difference in favor of the experimental group. The crestal preparation protocol led to less traumatic implant placement reducing the CBL, which is a common finding in dogs.27–30 

The macrodesign of the XiVE implant with the core and thread pattern is intended to achieve high primary stability during placement. Very high torque may occur particularly in the mandible because of the high proportion of cortical bone. Compression may compromise the in vivo periosteal blood supply,31 lead to necrosis,32 and increase the risk of bone resorption.33 The crestal drill controls the degree of internal compression; therefore, the correct use of the crestal drill must be observed to avoid excessive CBL. The depth of crestal preparation should vary depending on the bone quality. A varied insertion depth of 2 mm in bone type IV, 3 mm in bone type III, and 6 mm in bone types I and II is recommended. The internal condensation effect induced by the thread design is particularly desirable in spongy bone. In contrast, internal condensation is not required to increase the primary stability in more cortical bone.

Crestal bone loss was minimized when the crestal preparation protocol was carefully followed in conjunction with the placement of the osseocondensating XiVE implant system.

Table 2

Crestal bone loss with dogs as experimental units

Crestal bone loss with dogs as experimental units
Crestal bone loss with dogs as experimental units

This study was partially supported by Dentsply Friadent and Fundacao de Amparo a Pesquisa do Estado de São Paulo (grant 03/01815-6).

1
Adell
,
R.
,
U.
Lekholm
,
B.
Rockler
, and
P. I.
Brånemark
.
A 15 year study of osseointegrated implants in the treatment of edentulous jaw.
Int J Oral Surg
1981
.
10
:
387
416
.
2
Buser
,
D.
,
H. P.
Weber
, and
N. P.
Lang
.
Tissue integration of non-submerged implants. One-year results of a prospective study with 100 ITI hollow-cylinder and hollow-screw implants.
Clin Oral Implant Res
1990
.
1
:
33
40
.
3
Jemt
,
T.
,
U.
Lekholm
, and
K.
Grondahl
.
A 3-year follow up study of early single implant restoration ad modum Brånemark.
Int J Periodont Restor Dent
1990
.
10
:
341
349
.
4
Oh
,
T-J.
,
J.
Yoon
,
C. E.
Misch
, and
H-L.
Wang
.
The causes of early implant bone loss. Myth or science?
J Periodontol
2002
.
73
:
322
333
.
5
Esposito
,
M.
,
J. M.
Hirsch
,
U.
Lekholm
, and
P.
Thomsen
.
Biological factors contributing to failures of osseointegrated oral implants. (I) Success criteria and epidemiology.
Eur J Oral Sci
1998
.
106
:
527
531
.
6
Piatelli
,
A.
,
A.
Scarano
, and
M.
Piatelli
.
Microscopical aspects of failure in osseointegrated dental implants: a report of five cases.
Biomaterials
1996
.
17
:
917
920
.
7
Eriksson
,
A. R.
and
T.
Albrektsson
.
The effect of heat on bone regeneration: an experimental study in the rabbit using the bone growth chamber.
Int J Oral Maxillofac Surg
1984
.
42
:
705
711
.
8
Wilderman
,
M. N.
,
F. M.
Wentz
, and
B. J.
Orban
.
Histogenesis of repair after osseous surgery.
J Periodontol
1970
.
41
:
551
565
.
9
Roberts
,
W. E.
,
R. K.
Smith
,
Y.
Zilberman
,
P. G.
Mozsary
, and
R. S.
Smith
.
Osseous adaptation to continuous loading of rigid endosseous implants.
Am J Orthod
1984
.
86
:
95
111
.
10
Eriksson
,
A. R.
and
T.
Albrektsson
.
Temperature threshold levels for the heat-induced bone tissue injury: a vital microscopic study in the rabbit.
J Prosthet Dent
1983
.
50
:
101
108
.
11
Huiskes
,
R.
Some fundamental aspects of human joint replacement. Analyses of stress and heat conduction in bone-prosthesis structures.
Acta Orthop Scand
1980
.
185
:
1
208
.
12
Engquist
,
B.
,
T.
Bergendal
,
T.
Kallus
, and
U.
Linden
.
A retrospective multicenter evaluation of osseointegrated implants supporting overdentures.
Int J Oral Maxillofac Implants
1998
.
3
:
129
134
.
13
Jaffin
,
R.
and
C.
Berman
.
The excessive loss of Brånemark fixtures in type IV bone: a 5-year analysis.
J Periodontol
1991
.
62
:
2
4
.
14
Novaes
Jr,
A. B.
,
G. M.
Vidigal
Jr
,
A. B.
Novaes
,
M. F. M.
Grisi
,
S.
Poloni
, and
A.
Rosa
.
Immediate implants placed into infected sites: a histomorphometric study in dogs.
Int J Oral Maxillofac Implants
1998
.
13
:
113
118
.
15
Novaes
Jr,
A. B.
,
V.
Papalexiou
,
M. F. M.
Grisi
,
S. L. S.
Souza
,
M.
Taba
Jr
, and
J. K.
Kajiwara
.
Influence of implant microstructure on the osseointegration of immediate implants placed into periodontally infected sites. A histomorphometric study in dogs.
Clin Oral Implant Res
2004
.
15
:
34
43
.
16
Donath
,
K.
and
G.
Breuner
.
A method for the study of undecalcified bones and teeth with attached soft tissue.
J Oral Pathol
1982
.
11
:
318
326
.
17
Brånemark
,
P-I.
,
U.
Breine
,
R.
Adell
,
B. O.
Hansson
,
J.
Lindstrom
, and
A.
Ohisson
.
Intraosseous anchorage of dental prostheses.
Scand J Plast Reconstr Surg
1969
.
3
:
81
100
.
18
Albrektsson
,
T.
Bone tissue response.
In: Brånemark P-I, Zarb GA, Albrektsson T, eds. Tissue-Integrated Prostheses. Chicago, Ill: Quintessence Publishing Co; 1985:129–143
.
19
Haider
,
R.
,
G.
Watzek
, and
H.
Plenk
Jr.
.
Effects of drill cooling and bone structure on IMZ implant fixation.
Int J Oral Maxillofac Implants
1993
.
8
:
83
91
.
20
Hillerup
,
S.
Profile changes of bone and soft tissues following vestibular sulcus extension by the operation of Edlan and Mejchar. A 2-year follow-up study-II.
Int J Oral Surg
1979
.
8
:
340
346
.
21
Hillerup
,
S.
,
E.
Eriksen
, and
B.
Solow
.
Reduction of mandibular residual ridge after vestibuloplasty. A two-year follow-up study comparing the Edlan flap, mucosal and skin graft operations.
Int J Oral Maxillofac Surg
1989
.
18
:
271
276
.
22
Cochran
,
D. L.
,
P. V.
Nummikoski
,
F. L.
Higginbotton
,
J. S.
Hermann
,
S. R.
Makins
, and
D.
Buser
.
Evaluation of an endosseous titanium implant with sandblasted and acid-etched surface in the canine mandible: radiographic results.
Clin Oral Implant Res
1996
.
7
:
240
252
.
23
Quirynen
,
M.
,
I.
Naert
, and
D.
van Steenberghe
.
Fixture design and overload influence marginal bone loss and fixture success in the Brånemark system.
Clin Oral Implant Res
1992
.
3
:
104
111
.
24
Rangert
,
B.
,
P.
Krogh
,
B.
Langer
, and
N.
van Hoekel
.
Bending and overload and implant fracture.
Int J Oral Maxillofac Implants
1995
.
10
:
326
334
.
25
Wiskott
,
H. W.
and
U. C.
Belser
.
Lack of integration of smooth titanium surface: a working hypothesis based on strains generated in the surrounding bone.
Clin Oral Implant Res
1999
.
10
:
429
444
.
26
Adell
,
R.
,
U.
Lekholm
, and
P-I.
Brånemark
.
Surgical procedures.
In: Brånemark P-I, Zarb GA, Albrektsson T, eds. Tissue-Integrated Prostheses. Chicago, Ill: Quintessence Publishing Co; 1985:211–240
.
27
Block
,
M. S.
,
J. N.
Kent
, and
J. F.
Kay
.
Evaluation of hydroxyapatite-coated titanium dental implants in dogs.
Int J Oral Maxillofac Surg
1987
.
45
:
601
607
.
28
Evans
,
G. H.
,
A. J.
Mendez
, and
R. F.
Caudill
.
Loaded and nonleaded titanium versus hydroxyapatite-coated threaded implants in the canine mandible.
Int J Oral Maxillofac Implants
1996
.
11
:
360
371
.
29
Gammage
,
D. D.
,
A. E.
Bowman
,
R. M.
Meffert
,
R. J.
Cassingham
, and
W. A.
Dawenport
.
A histologic and scanning electron micrographic comparison of the osseous interface in loaded IMZ and integral implants.
Int J Periodont Restor Dent
1990
.
10
:
125
135
.
30
Weber
,
H. P.
,
D.
Buser
, and
K.
Donath
.
et al
.
Comparison of healed tissues adjacent to submerged and non-submerged unloaded titanium dental implants.
Clin Oral Implant Res
1996
.
7
:
11
19
.
31
Roberts
,
W. E.
and
L. P.
Garetto
.
Bone physiology and metabolism.
In: Misch CE, ed. Implant Dentistry. 2nd ed. St Louis, Mo: Mosby; 1999:225–237
.
32
Egger
,
G. W. N.
,
T. O.
Shindler
, and
C. M.
Pomerat
.
The influence of the contact-compression factor on osteogenesis in surgical fractures.
J Bone Joint Surg
1949
.
31
:
693
716
.
33
Trehame
,
R. W.
Review of Wolff's law and its proposed means of operation.
Orthop Rev
1981
.
10
:
35
47
.

Author notes

Arthur B. Novaes Jr, DDS, is professor and chairman, Rafael R. de Oliveira is a graduate student, and Mário Taba Jr, DDS, Sérgio L. S. de Souza, DDS, Daniela B. Palioto, DDS, and Márcio F. M. Grisi, DDS, are professors in the Department of Bucco Maxillo Facial Surgery and Periodontology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, São Paulo, Brazil. Address correspondence to Dr Arthur B. Novaes Jr, DDS, Faculdade de Odontologia de Ribeirão Preto, Universidade de São Paulo, Av do Café s/n, 14040-904, Ribeirão Preto, São Paulo, Brasil ([email protected]).