Understanding how clinical variables affect stress distribution facilitates optimal prosthesis design and fabrication and may lead to a decrease in mechanical failures as well as improve implant longevity. In this study, the many clinical variations present in an implant-supported prosthesis were analyzed by a 3-dimensional finite-element method. The anterior segment of a human mandible treated with 5 implants supporting a curved beam was created to perform the tests. The variables introduced in the computer model were cantilever length, elastic modulus of cancellous bone, abutment length, implant length, and framework alloy (AgPd or CoCr). The computer was programmed with physical properties of the materials as derived from the literature, and a 100-N vertical load was used to simulate the occlusal force. Images with fringes of stress were obtained, and the maximum stress at each site was plotted in graphs for comparison. Stresses tended to be concentrated at the cortical bone around the neck of the implant closest to the load, whereas stresses in cancellous bone were considered low. In general, the stress distribution was better with stiffer cancellous bone, longer abutments and implants, and shorter cantilevers. The use of a CoCr alloy framework appears to contribute to a better stress distribution.

Since the introduction of the technique of osseointegration by Branemark in 1969,1 the widely accepted and clinically demonstrated prosthetic solution for edentulous patients who cannot adapt to a mandibular complete denture is the construction of a bridge supported by five 10-mm-long implants placed using a 2-stage surgery in the interforaminal region (zone 1).

Stage I surgery aims at implant placement in selected host bone sites that are unloaded during the healing period. After stage I surgery and prosthetic conclusion, the major cause of failure has been attributed to implant overload.2,3 Since the occlusal load will be transferred to the implants and subsequently to the bone, it is believed that the biomechanics of the implant-supported prosthesis play an important role in the longevity of the bone around dental implants.4 Under load, bone tissue undergoes a remodeling process, which ultimately influences the long-term function of a dental implant system.5 The idea is to keep stresses below the failure stress of the bone.4,6,7 

However, the amount of force that can be applied to a dental implant without jeopardizing the surrounding bone is unknown.8 Factors such as the curvature of the mandible, the density of cancellous bone, the length of implants and abutments, the length of cantilever, the number of implants, and the framework stiffness influence stress distribution and have been objects of study.5,9,11 Failure prevention demands testing and stress analysis of the implants and tissues in vitro as well in vivo.9 

Reports indicate that finite element analysis (FEA) is a valuable tool for examining the stress distribution in bone adjacent to dental implants despite its inherent limitations.12,16 

The creation of a computerized model of an implant-supported fixed prosthesis permits an analysis of the possible variation in stress distribution that is likely to occur with diverse prosthetic designs and occlusal load variables. A good understanding of how each component of the implant prosthesis behaves under load could facilitate optimal prosthesis design and fabrication. This could lead to a decrease in likelihood of mechanical failures as well as improve implant prosthesis longevity.

The purpose of this study was to observe stress distribution in bone adjacent to dental implants by means of FEA. Such observation could be used to make comparisons to clinical treatment outcomes.

Model

A computerized 3-dimensional (3-D) finite-element model of the anterior segment of a human mandible provided with an implant-supported bridge was created (Figure 1). The basic model consisted of a curved beam with radius 15.0 mm and dimensions 69.0 mm long, 14.0 mm high, and 6.0 mm wide. This beam was covered with a 1.0-mm-thick layer on the buccal, occlusal, and lingual surfaces and a 3.0-mm layer at the base to simulate cortical bone. The final external dimensions were therefore 71.0 × 18.0 × 8.0 mm, respectively. Five 10.0-mm cylinders 3.75 mm in diameter were placed at the center of the beam, their centers 7.0 mm apart from each other. The cylinders were provided with 3.0-mm-high extensions to simulate the abutments. A second beam (71.0 × 4.0 × 6.0 mm) was added in connection to the abutments to simulate a framework. This resulted in a model with 1714 nodes and 8062 elements. The model was fixed at both ends for the sake of the stress analysis. All materials, bone included, were assumed to be linearly elastic and isotropic. An FEA program (I-DEAS Structural Dynamics Research, Milford, Ohio) installed in a desktop computer was used to analyze the many possible variations in prosthetic design and occlusal load. The model was fed with elastic properties of the materials as derived from the literature (Table 1). Elastic modulus for cortical bone was assumed to be 13.7 GPa, while for cancellous bone it was assumed to be 1.5 GPa based on a density of 25%. A 100-N vertical load was applied at 15.0 mm distally to the terminal abutment to simulate the occlusal force (Table 2). Images with the fringes of stress were obtained for each of the components of the model, and maximum Von Mises stress at each site was plotted in graphs for comparison.

Figures 1–3. Figure 1. Graphic representation of the 3-D model used in the study. Figure 2. Fringes of stress in cortical bone (basic model). Figure 3. Fringes of stress in cancellous bone (basic model)

Figures 1–3. Figure 1. Graphic representation of the 3-D model used in the study. Figure 2. Fringes of stress in cortical bone (basic model). Figure 3. Fringes of stress in cancellous bone (basic model)

Close modal
Table 1.

Elastic properties of materials used for the FEA model

Elastic properties of materials used for the FEA model
Elastic properties of materials used for the FEA model
Table 2.

Clinical variables selected for FEA analysis

Clinical variables selected for FEA analysis
Clinical variables selected for FEA analysis

Variables

Five clinical variables were chosen to be evaluated and are summarized in Table 2 for ease of reference. Each variable was introduced alternately on the basic model, all other conditions being equal.

Cantilever length was defined as the position where the load was applied relative to the center of the terminal abutment. The quality of the cancellous bone was expressed in terms of its modulus of elasticity (E) as derived from the literature. Abutment and implant lengths were based on the Branemark system. Framework alloy was selected according to the modulus of elasticity of the AgPd (E = 95 GPa) and CoCr (E = 218 GPa) alloys.

In the basic model, stress tended to be concentrated at the cortical bone on the disto-lingual aspect of the implant closest to the load (site #1). It gradually decreased as far as the implant is from the load and increased again on the terminal implant on the side opposite to the load (site #5). The same trend was observed on cancellous bone, although the magnitude of stress was considerably lower than on cortical bone. The peak von Mises stress on cortical bone was 9.27 MPa, whereas for cancellous bone it reached 1.23 MPa (Figures 2 and 3).

Cantilever length

Increasing the cantilever length had a remarkable effect on stress concentration (Figure 4). At each increment of 5 mm, the maximum von Mises stress raised by approximately 30% to 37% on the cortical bone around the implants. The same trend was seen on cancellous bone, but on a smaller magnitude.

Figures 4–6. Figure 4. Stress on cortical (a) and cancellous (b) bone when the force was applied at 10, 15, and 20 mm distally to the terminal implant. Figure 5. Stress on cortical (a) and cancellous (b) bone varying the modulus of elasticity of cancellous bone. Figure 6. Stress on cortical (a) and cancellous (b) bone when the abutment length is increased from 3.0 to 5.5 to 7.0 mm

Figures 4–6. Figure 4. Stress on cortical (a) and cancellous (b) bone when the force was applied at 10, 15, and 20 mm distally to the terminal implant. Figure 5. Stress on cortical (a) and cancellous (b) bone varying the modulus of elasticity of cancellous bone. Figure 6. Stress on cortical (a) and cancellous (b) bone when the abutment length is increased from 3.0 to 5.5 to 7.0 mm

Close modal

Quality of cancellous bone

As the elastic modulus of cancellous bone increased, more stress was observed, causing an opposite effect on cortical bone (Figure 5). Stresses increased by 75% in cancellous bone when the elastic modulus was changed from 1.5 to 4.0 GPa and by 45% when changed from from 4.0 to 7.9 GPa. In cortical bone, stresses decreased by 25% and 14% when the elastic modulus was changed from 1.5 to 4.0 GPa and from 4.0 to 7.9 GPa, respectively. Therefore, the stiffer the cancellous bone, the more stress it took, which lessened the stress on the cortical bone.

Abutment length

The stress in cortical and cancellous bone around the implants on sites #1 and #5 remained basically the same when the abutments were changed from 3.0 to 5.5 to 7.0 mm. The stress decreased on site #2 and increased by 50% to 100% on sites #3 and #4, when the abutments were changed from 3.0 mm to 5.5 mm to 7.0 mm (Figure 6).

Implant length

A slight decrease in stress around the implants was observed when the implant length was increased from 10.0 to 13.0 mm (10%) and from 13.0 to 15.0 mm (2%). This observation was valid for cortical and cancellous bone as well (Figure 7).

Framework alloy

Although not remarkably, the use of a CoCr alloy caused a slight decrease in the stress around the implants closest to the load. However, on sites #3 and #4, the stress on cancellous bone increased by up to 90% (Figure 8).

One of the most frequently asked questions about implant prosthesis is how much load implants can withstand without jeopardizing the surrounding bone. Although this question remains unanswered,8,15,19,21 there is a general agreement that a well-planned and executed prosthesis is essential to avoiding excessive forces on bone and implant components. Implant dentistry would greatly profit if it were provided the means to predict how bone and implant components would behave considering each patient's unique jaw anatomy, quality of bone, amount of occlusal force exerted on the prosthesis, etc.

Finite-element analysis, with all its inherent limitations,12,15 is a valuable instrument in pursuing that goal. When associated with clinical findings and accumulation of reliable data on implant loading, bone-implant contact area and other factors could help us understand the problems encountered in daily practice.19 For the reasons above mentioned, the results of this and other FEA studies have to be seen with a critical eye, and the values should not be taken as absolute but should rather be used to compare the possible magnitudes of stress that bone and implant components undergo during function.5,15,22 

In trying to simulate a human mandible, it was found that a 3-D model not of the entire mandible but of the anterior segment (approximately 20 mm posterior to each foramen) would suffice. A previous study concluded that for comparison of the stress distribution around implants it is not necessary to build a model of the entire mandible. By doing so, one can have the advantage of reducing modeling and calculation time.5 

The intention was to build the most versatile model possible to accommodate the largest number of variables with the least modifications. The model seen in Figure 1 was taken as a parameter for comparison. In every situation, the stress tended to be concentrated at the cortical bone on the disto-lingual aspect of the implant closest to the load. This is a common finding in the literature.5,15,21,22 The magnitude of stress found on cortical bone was approximately ten times greater than that seen on cancellous bone. This can be explained by the higher modulus of elasticity of cortical bone, which therefore bears more stress.

Cantilever length

In this study, the effect of the cantilever length was studied by applying the load to 3 points distally to the terminal implant along the framework at 10.0, 15.0, and 20.0 mm. The conventional design of an implant-supported fixed bridge on the edentulous mandible results in bilaterally cantilevered framework extensions, which under load create torque and moment on the implants. Many suggestions have been made in the literature regarding the extension of the cantilever, but in general the various authors agree that according to the quality of bone, a range of 10 to 20 mm of cantilever extension is acceptable.21,23,26 Perhaps one of the most compelling findings of this and other studies17,21,27,28 is the fact that the increase in cantilever length has a remarkable impact upon the stress concentration around the implants. The longer the cantilever arm, the more stress was observed. Fortunately, this is one of the implant-supported prosthesis features most easily controlled by the dentist. The application of the shortened dental arch concept and the express recommendation to the dental technician to keep the length of the cantilever arm to a minimum are procedures that cannot be neglected.

Quality of cancellous bone

Since the elastic modulus of bone is related to its density, cancellous bone was assigned with 3 elastic modulli, 1.5, 4.0, and 7.9 GPa, to verify the influence on stress distribution.

Although the literature shows little variation in cortical bone density, many density ranges for cancellous bone have been found.13,29,30 To guide the surgeon during placement of implants, attempts have been made to classify maxillary and mandibular bone according to the resistance to cut and the cortical:cancellous bone ratio.31,32 The influence of such variation in cancellous bone density on stress distribution has not been the subject of many studies.33 

Long-term clinical studies demonstrated that bone quality strongly correlates to implant success,34,36 whereas micromotion has been regarded as disrupting of the healing process.19 Once osseointegration is achieved and implants are loaded, there should be an effective load transfer from implant to bone. The load transfer is dependent upon the occlusal loads, implant shape and size, biomaterial properties, density of bone, and nature of the interface. Assuming complete contact between a cylindrical implant and bone, which does not necessarily correspond to the clinical situation, this study found that the stress on cortical bone decreased as the elastic modulus of the cancellous bone increased. The presence of a stiffer cancellous bone has the benefit of reducing the stress where it reaches its peak, namely the cervical area around the terminal abutment. At the same time, the stress in cancellous bone somehow increases, balancing the stress distribution. Of course, there is a limit to this benefit, because the more dense the cancellous bone, the more resistant it is to cutting and, therefore, the greater the heat generated during drilling.

Abutment length

Since patients will have different gingival thickness, the height of the abutments will vary accordingly.37 Nevertheless, the higher the abutment, the more unfavorable is the “crown-to-root” ratio. The findings of this study have shown that varying the height of the abutments will have a different effect on bone around each of the 5 implants. Though on sites #1, #2, and #5 only small differences were noted, the stress increased by 50% to 100% on sites #3 and #4. Yet, given the low magnitude of the stress seen in those sites, it is possible that abutment length alone may not be a source of concern.

Implant length

The implant length selection is highly dependent upon the jaw anatomy and the need for implant stabilization. As the presence of micromotion early after implantation has been regarded as disrupting of the bone healing process,19 the selection of an implant that reaches the basal compact layer of bone has been indicated.38 Although there is indication for such procedure, this is seldom accomplished on the mandible zone I where the bone quality usually exempts the need for bicortical stabilization. The longest implant used in this study (15 mm) reached but did not engage the basal compact layer of bone. It has been demonstrated by clinical studies34,39 that implant length does positively influence the outcome of implant therapy: longer implants are associated with higher success rates. This is particularly true when implants are placed in low-density bone, usually in the posterior maxillae (zone II). However, changing the implant length from 10 to 15 to 18 mm did not significantly affect the stress distribution in bone. Only a slight decrease was observed with longer implants. This trend has been observed before in 2-dimensional and 3-D FEA studies,21,40 indicating that some factor other than stress distribution would be responsible for the better results found in clinical studies with longer implants.

Framework alloy

Frameworks for implant-supported prostheses have evolved from soldering a gold alloy framework to gold cylinders, to gold alloy frameworks cast directly onto the cylinders, to silver-palladium alloy frameworks.26,39,41,42 The clinical results showed that the latter option gives consistently good results with respect to accuracy of fit, load-bearing capacity, reduced cost, and design versatility.23 However, the use of an alloy with a higher modulus of elasticity, such as the CoCr alloy, would allow for a more evenly distributed load among implants with a less bulky framework, which would be an advantage if intraoral space is limited.43,45 The structural rigidity of the prosthesis is dependent upon the elastic modulus, shape, and dimensions of the prosthesis, which can affect the force distribution among implants. The lower the elastic modulus, the greater the forces exerted on the abutments closest to the load. Therefore, if a hypothetical rubbery framework was used, the entire load would be concentrated on the implant closest to the loading point. The use of a CoCr alloy framework in this study favored a more even distribution of stress, as a slight increase was observed in cancellous bone around implants #3 and #4, which were opposite to the load. This finding is corroborated by others in the literature4,27,45,46 who found that a stiffer framework, such as one made out of CoCr alloy, provides a more even distribution of forces among the bridge and abutments, decreasing the stress within the retaining screws as a result of the reduced bending of the framework.

Within the limits of the study, the following observations were found:

  1. At each increment of 5 mm in cantilever length, stress increased by approximately 30% to 37% on the cortical bone around implants.

  2. The stiffer the cancellous bone, the more stress it took and the less stress the cortical bone appeared to undergo.

  3. A slight decrease in stress was observed with longer implants and abutments.

  4. The use of a CoCr alloy contributed to a better stress distribution.

Figures 7

and 8. Figure 7. Stress on cortical (a) and cancellous (b) bone when the implant length is increased from 10 to 13 to 15 mm. Figure 8. Stress on cortical (a) and cancellous (b) bone with an AgPd and a CoCr alloy

Figures 7

and 8. Figure 7. Stress on cortical (a) and cancellous (b) bone when the implant length is increased from 10 to 13 to 15 mm. Figure 8. Stress on cortical (a) and cancellous (b) bone with an AgPd and a CoCr alloy

Close modal

The authors wish to thank Professor George Zarb (University of Toronto) for providing the means to conduct this research and to Professor John Curran and his assistant Xiao Yang Lei (Mining Engineering, University of Toronto) for the FEA analysis. This study was supported by Fundação de Amparo a Pesquisa do Estado de São Paulo (FAPESP) and an Implant Prosthodontic Unit Research Grant – Faculty of Dentistry of Toronto.

1
Branemark
,
P. I.
,
U.
Breine
,
R.
Adell
,
B. O.
Hansson
,
J.
Lindström
, and
A.
Ohlsson
.
Intraosseous anchorage of dental prostheses. I. Experimental studies.
Scand J Plast Reconstr Surg
1969
.
3
:
81
100
.
2
Oh
,
T. J.
,
J.
Yoon
,
C. E.
Mish
, and
H. L.
Wang
.
The causes of early implant bone loss: myth or science?
J Periodontol
2002
.
3
:
322
333
.
3
Smith
,
D. C.
Dental implants: materials and design considerations.
Int J Prosthodont
1993
.
6
:
106
117
.
4
Skalak
,
R.
Aspects of biomechanical considerations.
In: PI. Branemark, GA. Zarb, T. Albreksson, eds.
Tissue-Integrated Prostheses—Osseointegration in Clinical Dentistry
Chicago, Ill: Quintessence;
.
1985
.
117
128
.
5
Meijer
,
H. J. A.
,
F. J. M.
Starmans
,
F.
Bosman
, and
W. H. A.
Steen
.
A comparison of finite element models of an edentulous mandible provided with implants.
J Oral Rehabil
1993
.
20
:
147
157
.
6
Frost
,
H. M.
Wolff's Law and bone's structural adaptation to mechanical use: an overview for clinicians.
Angle Orthod
1994
.
64
:
175
188
.
7
Wiskott
,
H. W. A.
and
U. C.
Belser
.
Lack of integration of smooth titanium surfaces: a working hypothesis based on strains generated in the surrounding bone.
Clin Oral Implants Res
1999
.
10
:
429
444
.
8
Clelland
,
N. L.
,
Y. H.
Ismail
,
H. S.
Zaki
, and
D.
Pipko
.
Three-dimensional finite element stress analysis in and around the Screw-Vent implant.
Int J Oral Maxillofac Implants
1991
.
6
:
391
398
.
9
van Oosterwick H
,
Duyck J.
,
J.
Vander Sloten
,
G.
Van der Perre
,
M.
De Cooman
, and
S.
Lievens
.
et al
.
The influence of bone mechanical properties and implant fixation upon bone loading around oral implants.
Clin Oral Implants Res
1998
.
9
:
407
418
.
10
Pierrisnard
,
L.
,
F.
Renouard
,
P.
Renault
, and
M.
Barquins
.
Influence of implant length and bicortical anchorage on stress distribution.
Clin Implant Dent Relat Res
2003
.
5
:
254
262
.
11
Bidez
,
M. W.
and
C. E.
Misch
.
Force transfer in implant dentistry: basic concepts and principles.
J Oral Implantol
1992
.
18
:
264
274
.
12
Brunski
,
J. B.
Biomaterial and biomechanics in dental implant design.
Int J Oral Maxillofac Implants
1988
.
3
:
85
97
.
13
Cook
,
S. D.
,
J. J.
Klawitter
, and
A. M.
Weinstein
.
The influence of implant geometry on the stress distribution around dental implants.
J Biomed Mater Res
1982
.
16
:
369
379
.
14
Davis
,
D. M.
,
G. A.
Zarb
, and
Y. L.
Chao
.
Studies on frameworks for osseointegrated prostheses. Part I: The effect of varying the number of supporting abutments.
Int J Oral Maxillofac Implants
1988
.
3
:
197
201
.
15
Murphy
,
W. M.
,
K. R.
Williams
, and
M. C.
Gregory
.
Stress in bone adjacent to dental implants.
J Oral Rehabil
1995
.
22
:
897
903
.
16
Şahin
,
S.
,
M. C.
Çehreli
, and
E.
Yalçin
.
The influence of functional forces on the biomechanics of implant-supported prostheses—a review.
J Dent
2002
.
30
:
271
282
.
17
Lewinstein
,
I.
,
L.
Banks-Sills
, and
R.
Eliasi
.
Finite element analysis of a new system (IL) for supporting an implant-retained cantilever prosthesis.
Int J Oral Maxillofac Implants
1995
.
10
:
355
366
.
18
Craig
,
R. G.
Restorative Dental Materials. 8th ed
St Louis, Mo: Mosby;
.
1989
.
84.
.
19
Brunski
,
J. B.
Influence of biomechanical factors at the bone-biomaterial interface.
In: JE. Davies, ed.
The Bone-Biomaterial Interface
Toronto, Canada: University of Toronto Press;
.
1991
.
391
404
.
20
Brunski
,
J. B.
Biomechanical factors affecting the bone-dental implant interface.
Clin Mater
1992
.
10
:
153
201
.
21
Sertgöz
,
A.
and
S.
Güvener
.
Finite element analysis of the effect of cantilever and implant length on stress distribution in an implant-supported fixed prosthesis.
J Prosthet Dent
1996
.
76
:
165
169
.
22
Weinberg
,
L. A.
and
B.
Kruger
.
A comparison of implant/prosthesis loading with four clinical variables.
Int J Prosthodont
1995
.
8
:
421
433
.
23
Rangert
,
B.
,
M.
Eng
,
T.
Jemt
, and
L.
Jorneus
.
Forces and moments on Branemark implants.
Int J Oral Maxillofac Implants
1989
.
4
:
241
247
.
24
Taylor
,
T. D.
Fixed implant rehabilitation for the edentulous maxilla.
Int J Oral Maxillofac Implants
1991
.
6
:
329
337
.
25
van Zyl
,
P. P.
,
N.
Grundling
,
C. H.
Jooste
, and
E.
Terblanche
.
Three-dimensional finite element model of a human mandible incorporating six osseointegrated implants for stress analysis of mandibular cantilever prostheses.
Int J Oral Maxillofac Implants
1995
.
10
:
51
57
.
26
Zarb
,
G. A.
and
A.
Schmitt
.
The longitudinal clinical effectiveness of osseointegrated dental implants. The Toronto study. Part II: The prosthetic results.
J Prosthet Dent
1989
.
64
:
53
61
.
27
Benzing
,
U. R.
,
H.
Gall
, and
H.
Weber
.
Biomechanical aspects of two different implant-prosthetic concepts for edentulous maxillae.
Int J Oral Maxillofac Implants
1995
.
10
:
188
198
.
28
White
,
S. N.
,
A. A.
Caputo
, and
T.
Anderkvist
.
Effect of cantilever length on stress transfer by implant-supported prostheses.
J Prosthet Dent
1994
.
71
:
493
499
.
29
Katz
,
J. L.
Mechanics of hard tissue.
In: JD. Brozino, ed.
The Biomedical Engineering Handbook
Boca Raton, Fla: CRC Press;
.
1995
.
273
289
.
30
O'Mahony
,
A. M.
,
J. L.
Williams
,
J. O.
Katz
, and
P.
Spencer
.
Anisotropic elastic properties of cancellous bone from a human edentulous mandible.
Clin Oral Implants Res
2000
.
11
:
415
421
.
31
Cawood
,
J. I.
and
R. A.
Howell
.
A classification of the edentulous jaws.
Int J Oral Maxillofac Surg
1998
.
17
:
232
236
.
32
Lekholm
,
U.
and
G. A.
Zarb
.
Patient selection and preparation.
In: PI. Branemark, GA. Zarb, T. Albreksson, eds.
Tissue-Integrated Prostheses—Osseointegration in Clinical Dentistry
Chicago, Ill: Quintessence;
.
1985
.
199
209
.
33
Holmes
,
D. C.
and
J. T.
Loftus
.
Influence of bone quality on stress distribution for endosseous implants.
J Oral Implantol
1997
.
23
:
104
111
.
34
Friberg
,
B.
,
T.
Jemt
, and
U.
Lekholm
.
Early failures in 4641 consecutively placed Branemark dental implants: a study from stage I surgery to the connection of complete prostheses.
Int J Oral Maxillofac Implants
1991
.
6
:
142
146
.
35
Jemt
,
T.
and
U.
Lekhlom
.
Implant treatment in edentulous maxillae: a 5-year follow-up report on patients with different degrees of jaw resorption.
Int J Oral Maxillofac Implants
1995
.
10
:
303
311
.
36
van Steenberghe
,
D.
,
U.
Lekholm
,
C.
Bolender
,
T.
Folmer
,
P.
Henry
, and
I.
Herrmann
.
et al
.
Applicability of osseointegrated oral implants in the rehabilitation of partial edentulism: a prospective multicenter study on 558 fixtures.
Int J Oral Maxillofac Implants
1990
.
5
:
272
281
.
37
Adell
,
R.
,
U.
Lekholm
, and
P. I.
Branemark
.
Surgical procedures.
In: PI. Branemark, GA. Zarb, T. Albrektsson, eds.
Tissue-Integrated Prostheses—Osseointegration in Clinical Dentistry
Chicago, Ill: Quintessence;
.
1985
.
211
232
.
38
Lekholm
,
U.
,
R.
Adell
, and
P. I.
Branemark
.
Complications.
In: PI. Branemark, GA. Zarb, T. Albreksson, eds.
Tissue-Integrated Prostheses—Osseointegration in Clinical Dentistry
Chicago, Ill: Quintessence;
.
1985
.
233
240
.
39
Jemt
,
T.
Failures and complications in 391 consecutively inserted fixed prosthesis supported by Branemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual check-up.
Int J Oral Maxillofac Implants
1991
.
6
:
270
276
.
40
Meijer
,
H. J. A.
,
J. H.
Kuiper
,
F. J. M.
Starmans
, and
F.
Bosman
.
Stress distribution around dental implants: influence of superstructure, length of implants, and height of mandible.
J Prosthet Dent
1992
.
68
:
96
102
.
41
Cox
,
J.
and
G. A.
Zarb
.
Alternative prosthodontic superstructure designs.
Swedish Dent J Suppl
1985
.
28
:
71
75
.
42
Leung
,
N.
,
G. A.
Zarb
, and
P.
Pilliar
.
Casting of prosthetic superstructures in tissue integrated prostheses.
J Dent Res
1983
.
62
:
293
.
Abstract 1112.
43
Chao
,
Y.
,
D. M.
Davis
,
G. A.
Zarb
, and
H.
Judes
.
A study into the use of chromium-cobalt alloy for constructing the framework for osseointegrated prostheses.
Clin Mater
1988
.
3
:
309
315
.
44
Williams
,
K. R.
,
C. J.
Watson
,
W. M.
Murphy
,
J.
Scott
,
M.
Gregory
, and
D.
Sinobad
.
Finite element analysis of fixed prostheses attached to osseointegrated implants.
Quintessence Int
1990
.
21
:
563
570
.
45
Hulterström
,
M.
and
U.
Nilsson
.
Cobalt-chromium as a framework material in implant-supported fixed prostheses: a 3-year follow-up.
Int J Oral Maxillofac Implants
1994
.
9
:
449
454
.
46
Sertgöz
,
A.
Finite element analysis study of the effect of superstructure material on stress distribution in an implant-supported fixed prosthesis.
Int J Prosthodont
1997
.
10
:
19
27
.

Jose Henrique Rubo, DDS, MSc, PhD, is an associate professor in the Department of Prosthodontics, Bauru School of Dentistry, University of Sao Paulo - USP, Bauru SP, Brazil. Address correspondence to Dr Rubo at Depto. de Prótese FOB – USP, Al. Otávio P. Brisolla 9-75, 17012-901 Bauru SP, Brazil. ([email protected])

Edson Antonio Capello Souza, PhD, is an associate professor in the Department of Mechanical Engineering, Bauru School of Engineering, State University of Sao Paulo - UNESP, Bauru SP, Brazil.