The aim of this study was to evaluate the efficacy of osteotomes in enhancing bone quality as compared to conventional implant bed preparation using burrs. Polyurethane foam blocks differing in density (10 pcf, 20 pcf) and structure (cellular, solid) were used. Ten implant sockets were prepared in each of the materials by means of drilling and by using osteotomes. Bone quality was assessed by measuring implant insertion torque and primary implant stability (resonance frequency analysis). Additionally, a newly designed device (BoneProbe) for bone quality testing during dental implant surgery based on intraoperative compressive tests was applied. Multivariate analysis of variance with Pillai's trace was used as test statistic (level of significance: α = 0.05) and Pearson correlation coefficients were calculated for all combinations of parameters. Whereas a significant influence of bone type on all measurement results (P = 0) could be found, the factor preparation technique only showed a significant effect on BoneProbe measurements in the cortical area (P = 0) and on implant insertion torque (P = 0). The interaction of bone type and preparation technique showed a significant effect on BoneProbe measurements in the trabecular area (P = .002) and on implant insertion torque (P = 0). Significant correlations between all parameters were found. The application of osteotomes leads to higher values for compressive testing of bone in the apical part of an osteotomy depending on the initial density of the bone. Intraoperative compressive testing appears to be sensitive enough for quantifying localized changes in bone quality.

Local alveolar bone quality and quantity, geometry of the implant as well as the placement technique used have been shown to affect primary implant stability.1,2  As this purely mechanical phenomenon of primary stability is difficult to achieve in low quality bone, modifications of the regular surgical protocol for implant placement such as undersized drilling,3  the use of tapered implants,2  and the use of osteotomes46  have been suggested.

For nonablative implant bed preparation by means of osteotomes, it has been claimed that bone would be condensed and relocated both laterally and apically, thereby enhancing local bone density and consequently primary implant stability.4,7 

However, the literature available reports nonuniform findings with the use of osteotomes.8  Positive effects of bone condensing with respect to bone architecture and early healing of implants have been described in human cadaver8,9  and animal studies,10,11  respectively.

Applying this technique in a clinical study, Gulsahi et al12  found no significant improvements in bone related parameters such as bone mineral density. Even on the contrary, in two cadaver studies, greater implant stability could be achieved using conventional methods as compared to using osteotomes.13,14  Based on these contradictory findings, it was the goal of this investigation to compare conventional and osteotome-based implant preparation techniques under well-controlled, in vitro conditions. A recently introduced diagnostic tool based on compressive testing of bone surrounding an implant site was used to directly evaluate possible effects on the material properties of bone in different areas of the osteotomies.

BoneProbe—Measuring device for the intraoperative determination of alveolar bone quality

A novel measurement system named BoneProbe for the direct evaluation of bone quality during dental implant surgery was developped.15,16  The sensing element of the device consists of a metal cylinder with a diameter of 3.50 mm, which is divided into 6 segments and fits into a readily prepared implant socket for a regular sized implant (Straumann Standard Implant; Institut Straumann AG, Basel, Switzerland). For determining bone quality the sensor is placed in the cortical and trabecular part of an osteotomy and expanded to 3.57 mm and 3.52 mm respectively while the force needed is recorded (Figure 1).

Figure 1.

Sensing element of the BoneProbe consisting of a metal cylinder with a diameter of 3.5 mm, which is split into 6 segments. The sensor can be positioned in an implant socket and expanded while the actual force needed is measured.

Figure 1.

Sensing element of the BoneProbe consisting of a metal cylinder with a diameter of 3.5 mm, which is split into 6 segments. The sensor can be positioned in an implant socket and expanded while the actual force needed is measured.

Close modal

Bone surrogate materials

Commercially available polyurethane foam blocks (Figure 2) differing in structure and density were used as bone surrogate materials17  (Sawbones Europe AB, Malmö, Sweden):

Figure 2.

Polyurethane foam materials used as bone surrogate. The foam blocks either had a cellular or solid structure and densities of 10 pcf or 20 pcf.

Figure 2.

Polyurethane foam materials used as bone surrogate. The foam blocks either had a cellular or solid structure and densities of 10 pcf or 20 pcf.

Close modal
  • Blocks with a solid, homogeneous structure (Solid Rigid Polyurethane Foam 10 pcf, 20 pcf)

  • Blocks with a cellular structure (Cellular Rigid Polyurethane Foam 10 pcf, 20 pcf)

Conventional implant bed preparation

Ten sockets 3.5 mm in diameter and 11 mm in length were prepared for the placement of screw-shaped cylindrical implants with a diameter of 4.1 mm and 10 mm length (Straumann Standard Implant; Institut Straumann AG) in each of the materials. The implant positions were marked with a round burr and a set of twist drills 2.2 mm, 2.8 mm and 3.5 mm in diameter was used in combination with a surgical motor (KaVo INTRAsurg 1000; KaVo Dental GmbH, Biberach, Germany) to create standardized implant beds (Figure 3).

Figures 3 and 4.

Set of drills used for conventional implant bed preparation. A round burr is used to mark the implant position prior to the creation of an osteotomy using twist drills with diameters ranging from 2.2 mm to 3.5 mm. Figure 4. (a) Set of bone condensing osteotomes used for nonablative implant bed preparation following the creation of an initial osteotomy with a diameter of 2.2 mm. (b) Sinus lift osteotome with concave end, which was used for final shaping of the implant sockets.

Figures 3 and 4.

Set of drills used for conventional implant bed preparation. A round burr is used to mark the implant position prior to the creation of an osteotomy using twist drills with diameters ranging from 2.2 mm to 3.5 mm. Figure 4. (a) Set of bone condensing osteotomes used for nonablative implant bed preparation following the creation of an initial osteotomy with a diameter of 2.2 mm. (b) Sinus lift osteotome with concave end, which was used for final shaping of the implant sockets.

Close modal

Implant bed preparation applying the osteotome technique

For implant placement with the osteotomes technique, the implant positions (10 implants per material) were marked with a round burr, and a twist drill 2.2 mm in diameter was used to create an initial osteotomy of 11 mm in length. Bone condensing osteotomes 2.2 mm, 2.8 mm, and 3.5 mm in diameter with tapered, convex ends (Institut Straumann AG, Basel, Switzerland) were used to further prepare the implant sockets (Figure 4a). Final shaping of the osteotomies was done using a sinus lift osteotome 3.5 mm in diameter (Institut Straumann AG, Basel, Switzerland) with a parallel walled, concave end (Figure 4b).

Measurements of bone quality and primary implant stability

Upon completion of implant bed preparation, the sensor of the BoneProbe was positioned in the socket at depths of 2 mm (cortical area) and 10 mm (trabecular area). The sensor was opened to 3.57 mm in the cortical area and 3.52 mm in the trabecular area. The forces needed to open the sensor were recorded in N. Implants were installed using a surgical motor (KaVo INTRAsurg 1000, KaVo Dental GmbH) measuring the maximum torque values needed to seat the implants.18  Primary implant stability was determined by means of resonance frequency measurements (Osstell mentor, Osstell AB, Gothenburg, Sweden) and recorded as implant stability quotients (ISQ).19 

Statistical analysis

Multivariate analysis of variance (MANOVA, SPSS 14.0 for Windows, SPSS Inc, Chicago, IL) was applied with preparation technique and bone type as fixed factors. Since the Box test of homogeneity of covariances revealed significant differences among groups (P = 0), Pillai's trace, which is robust to heterogeneous variances,20,21  was used as test statistic. The level of significance was set at α = 0.05. In addition, Pearson correlation coefficients were calculated for all combinations of parameters.

The mean values and standard deviations for all parameters obtained are given in Table 1. In general, the use of osteotomes led to higher measurement values in a specific material for all parameters studied with the following exceptions. In Cellular 20 material, lower values were recorded with the BoneProbe in the trabecular area as well as with the Osstell mentor device (Osstell aB) when implants were placed using osteotomes. Similarly, the use of osteotomes led to lower values with the BoneProbe in the trabecular area in Solid 20 material.

Table 1

Mean values and standard deviations for all parameters measured in the different polyurethane foam materials following conventional and osteotome based implant site preparation.*

Mean values and standard deviations for all parameters measured in the different polyurethane foam materials following conventional and osteotome based implant site preparation.*
Mean values and standard deviations for all parameters measured in the different polyurethane foam materials following conventional and osteotome based implant site preparation.*

Global MANOVA with Pillai's trace revealed a significant influence of both, preparation technique and bone type on all measurement results (P = 0).

Tests of between subject effects (Table 2) revealed a significant influence of bone type on all measurement results (P = 0). The factor preparation technique, however, only showed a significant effect on BoneProbe measurements in the cortical area (P = 0) and implant insertion torque (P = 0). The interaction of bone type and preparation technique showed a significant effect on BoneProbe measurements in the trabecular area (P = .002) and on implant insertion torque (P = 0).

Significant correlations between all parameters were found based on Pearson correlation coefficients (Table 3).

Table 3

Pearson correlation coefficients describing possible correlations between all parameters studied; all correlations are significant at a level of α = 0.05.*

Pearson correlation coefficients describing possible correlations between all parameters studied; all correlations are significant at a level of α = 0.05.*
Pearson correlation coefficients describing possible correlations between all parameters studied; all correlations are significant at a level of α = 0.05.*

It could be shown that particularly in low-density bone the use of osteotomes may lead to an enhancement of bone quality and subsequently primary implant stability as assessed by resonance frequency analysis.22  In materials with greater density, such an effect could not be shown consistently. On the contrary, in Cellular and Solid materials with a density of 20 pcf, even lower values have been recorded with the BoneProbe in trabecular bone. This may partially be explained by deviations in the insertion path of the osteotomes as well as in their repeated application. Supporting this interpretation, no significant effect of the preparation technique on BoneProbe readings in the trabecular area could be found. The interaction preparation technique and bone type, however, showed a significant effect for this parameter. This may be indicative for osteotomes condensing bone predominantly in the apical part of an osteotomy depending on the initial density of the bone. This was further supported by the fact that the interaction preparation technique and bone type did not show a significant effect on BoneProbe readings in the cortical part of an osteotomy whereas the factor preparation technique alone had a significant influence on this parameter.

No effect of preparation technique and the interaction preparation technique and bone type on Osstell measurements1,19  could be revealed. It therefore appears that resonance frequency measurements of implants placed in sites prepared with either of the techniques is not sensitive enough for detecting slight differences in bone density as occurring as a result of using osteotomes.

The parameter implant insertion torque18  appears to be sensitive enough to detect slight differences in bone quality as all factors showed significant effects on this parameter. However, it is not possible to differentiate between effects of osteotomes in the cervical and apical part of an osteotomy as can be done using the BoneProbe as diagnostic device.

Specific limitations have to be taken into account when interpreting the results of this study. Although it has been shown that human bone has a structure which is comparable to polymeric foams,17  the use of polyurethane foam as bone surrogate material may have affected the results presented. However, these materials guarantee for standardized experimental conditions as well as simple handling characteristics.

It has been shown that measurements done using the BoneProbe correlate well with other parameters of bone quality and primary implant stability.15,16  Nevertheless, this device has to be considered experimental in the current status as clinical trials proving its diagnostic value are missing. Having completed experiments in human cadaver bone,15  the BoneProbe should be tested extensively in animal models and human trials in order to clarify the diagnostic value of BoneProbe measurements.

Due to deviations in the diameter of the osteotomies, especially in the cervical area, the sensing element of the BoneProbe had to be opened to 3.57 mm, whereas in the apical part consistent measurements could be performed by opening the sensor to 3.52 mm. With the BoneProbe not being able to provide absolute results on the material parameters of the bone type investigated,23  the force needed for opening the sensor from minimum to maximum diameter in the range of 7.5 N was neglected in the final analysis.

Within the limitations of this in vitro study, it can be concluded that the BoneProbe is more sensitive for determining local differences in bone quality as compared to conventional techniques such as resonance frequency measurements and measurements of implant insertion torque. As could be shown based on the BoneProbe measurements, the use of osteotomes for implant bed preparation predominantly increases bone quality in the apical part of an osteotomy.

Table 2

Statistical analysis of measurement results based on tests of between subject effects; significant influences (α = 0.05) of the fixed factors preparation technique and bone type on measurement results (BoneProbe, Osstell mentor and insertion torque) are written in bold.

Statistical analysis of measurement results based on tests of between subject effects; significant influences (α = 0.05) of the fixed factors preparation technique and bone type on measurement results (BoneProbe, Osstell mentor and insertion torque) are written in bold.
Statistical analysis of measurement results based on tests of between subject effects; significant influences (α = 0.05) of the fixed factors preparation technique and bone type on measurement results (BoneProbe, Osstell mentor and insertion torque) are written in bold.

This study has been generously supported by a grant from the American Academy of Implant Dentistry.

ISQ

implant stability quotient

MANOVA

multivariate analysis of variance

1
Meredith
N
.
Assessment of implant stability as a prognostic determinant
.
Int J Prosthodont
.
1998
;
11
:
491
501
.
2
Ostman
PO
,
Hellman
M
,
Wendelhag
I
,
Sennerby
L
.
Resonance frequency analysis measurements of implants at placement surgery
.
Int J Prosthodont
.
2006
;
19
:
77
83
;
discussion 84
.
3
Beer
A
,
Gahleitner
A
,
Holm
A
,
Birkfellner
W
,
Homolka
P
.
Adapted preparation technique for screw-type implants: explorative in vitro pilot study in a porcine bone model
.
Clin Oral Implants Res
.
2007
;
18
:
103
107
.
4
Strietzel
FP
,
Nowak
M
,
Küchler
I
,
Friedmann
A
.
Peri-implant alveolar bone loss with respect to bone quality after use of the osteotome technique: results of a retrospective study
.
Clin Oral Implants Res
.
2002
;
13
:
508
513
.
5
Siddiqui
AA
,
Sosovicka
M
.
Lateral bone condensing and expansion for placement of endosseous dental implants: a new technique
.
J Oral Implantol
.
2006
;
32
:
87
94
.
6
Peñarrocha-Diago
M
,
Rambla-Ferrer
J
,
Perez
V
,
Pérez-Garrigues
H
.
Benign paroxysmal vertigo secondary to placement of maxillary implants using the alveolar expansion technique with osteotomes: a study of 4 cases
.
Int J Oral Maxillofac Implants
.
2008
;
23
:
129
132
.
7
Summers
RB
.
A new concept in maxillary implant surgery: the osteotome technique
.
Compendium
.
1994
;
15
:
152
,
154
156
,
158 passim; quiz 162
.
8
Proff
P
,
Bayerlein
T
,
Rottner
K
,
Mai
R
,
Fanghänel
J
,
Gedrange
T
.
Effect of bone conditioning on primary stability of FRIALIT-2 implants
.
Clin Oral Implants Res
.
2008
;
19
:
42
47
.
9
Fanuscu
MI
,
Chang
TL
,
Akça
K
.
Effect of surgical techniques on primary implant stability and peri-implant bone
.
J Oral Maxillofac Surg
.
2007
;
65
:
2487
2491
.
10
Nkenke
E
,
Kloss
F
,
Wiltfang
J
,
et al
.
Histomorphometric and fluorescence microscopic analysis of bone remodelling after installation of implants using an osteotome technique
.
Clin Oral Implants Res
.
2002
;
13
:
595
602
.
11
Schlegel
KA
,
Kloss
FR
,
Kessler
P
,
Schultze-Mosgau
S
,
Nkenke
E
,
Wiltfang
J
.
Bone conditioning to enhance implant osseointegration: an experimental study in pigs
.
Int J Oral Maxillofac Implants
.
2003
;
18
:
505
511
.
12
Gulsahi
A
,
Paksoy
CS
,
Yazıcıoglu
N
,
Arpak
N
,
Kucuk
NO
,
Terzioglu
H
.
Assessment of bone density differences between conventional and bone-condensing techniques using dual energy x-ray absorptiometry and radiography
.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod
.
2007
;
104
:
692
698
.
13
Büchter
A
,
Kleinheinz
J
,
Joos
U
,
Meyer
U
.
Primary implant stability with different bone surgery techniques. An in vitro study of the mandible of the minipig [in German]
.
Mund Kiefer Gesichtschir
.
2003
;
7
:
351
355
.
14
Cehreli
MC
,
Kökat
AM
,
Comert
A
,
Akkocaoğlu
M
,
Tekdemir
I
,
Akça
K
.
Implant stability and bone density: assessment of correlation in fresh cadavers using conventional and osteotome implant sockets
.
Clin Oral Implants Res
.
2009
;
20
:
1163
1169
.
15
Karl
M
,
Winter
W
,
Graef
F
,
Wichmann
MG
,
Krafft
T
.
Determination of alveolar bone quality during dental implant surgery by means of compressive testing
.
In
:
Lim
CT
,
Goh
,
JCH
,
eds
.
IFMBE Proceedings 31 - 6th World Congress of Biomechanics
.
Berlin, Heidelberg, New York
:
Springer;
2010
:
946
949
.
16
Winter
W
,
Steinmann
P
,
Wichmann
MG
,
Krafft
T
,
Karl
M
.
Quality of alveolar bone—Structure dependant material properties and design of a novel measurement technique
.
In
:
Lim
CT
,
Goh
JCH
,
eds
.
IFMBE Proceedings 31 - 6th World Congress of Biomechanics
.
Berlin, Heidelberg, New York
:
Springer;
2010
:
950
953
.
17
Gibson
LJ
.
The mechanical behaviour of cancellous bone
.
J Biomech
.
1985
;
18
:
317
328
.
18
Johansson
B
,
Back
T
,
Hirsch
JM
.
Cutting torque measurements in conjunction with implant placement in grafted and nongrafted maxillas as an objective evaluation of bone density: a possible method for identifying early implant failures?
Clin Implant Dent Relat Res
.
2004
;
6
:
9
15
.
19
Friberg
B
,
Sennerby
L
,
Meredith
N
,
Lekholm
U
.
A comparison between cutting torque and resonance frequency measurements of maxillary implants. A 20-month clinical study
.
Int J Oral Maxillofac Surg
.
1999
;
28
:
297
303
.
20
Olson
CL
.
Comparative robustness of six tests in multivariate analysis of variance
.
J Am Stat Assoc
.
1974
;
69
:
894
908
.
21
Olson
CL
.
On choosing a test statistic in multivariate analysis of variance
.
Psychol Bull
.
1976
;
83
:
579
586
.
22
Aparicio
C
,
Lang
NP
,
Rangert
B
.
Validity and clinical significance of biomechanical testing of implant/bone interface
.
Clin Oral Implants Res
.
2006
;
17
Suppl 2
:
2
7
.
23
Linde
F
.
Elastic and viscoelastic properties of trabecular bone by a compression testing approach
.
Dan Med Bull
.
1994
;
41
:
119
138
.