Titanium is the most common material chosen for dental implants because it is highly corrosion resistant because it constantly reforms a protective passive film layer. The formation and composition of the passive film layer is dependent on the environmental conditions. If the stable oxide layer is damaged, the titanium surface underneath can corrode. The purpose of this study was to determine if basic corrosion of commercially pure titanium (CpTi) alloy in artificial saliva was affected by pH and to understand the corrosion kinetics/mechanisms of CpTi as a function of pH. In this study, titanium alloy discs were subjected to corrosion tests. Before the tests, all samples were cleaned and polished using standard metallographic preparation methods. Artificial saliva was used as the testing medium. The following pH values were tested: 3.0, 4.5, 6.0, 6.5, 7.5, and 9.0. Different pH values were achieved by adding lactic acid (acidic) or NaOH (basic) in appropriate amounts. Potentiodynamic curves indicated behavior change at each pH. In addition, the corrosion current density value determined from the potentiodynamic curve exhibited the poorest corrosion resistance for pH 7.5. The Nyquist plot (from the electrochemical impedance spectroscopy results) indicated that pH 7.5 had the poorest resistance. Scanning electron microscopy images indicated that pH levels of 6.5, 7.5, and 9.0 had considerable surface corrosion. The results showed that the media's pH significantly influenced the corrosion behavior of CpTi. The poor corrosion behavior at the neutral pHs invites some concerns and highlights the need for further study.

Titanium (Ti) is the most widespread material used for dental implants because of its mechanical properties, exceptional corrosion resistance, and biocompatibility.13  Although Ti alloys are highly corrosion resistant because of the stability of the TiO2 oxide layer, they are not inert to corrosive attack. When the stable oxide layer is broken down or cannot reform on parts of the surface, Ti can be as corrosive as other base metal alloys.48  In addition, the oral cavity can simulate an electrochemical cell because saliva acts as a weak electrolyte. Electrochemical corrosion of Ti alloys may lead to surface corrosion of dental implants, and ultimately may release the corrosion products into the system over time.9,10  It is not known whether corrosive products are beneficial or harmful to osseointegration.

The corrosion resistance of Ti results from the formation of an extremely stable, continuous, highly adherent, and protective oxide film on the metal's surface.3  The nature, composition, and thickness of the protective surface oxides that form on Ti alloys depend on the environmental conditions.3,7,8  Recent studies have shown that the corrosion resistance of Ti (commercially pure Ti [CpTi] and Yi alloys) becomes weaker with exposure to varying fluoride concentrations and NaF solution with a concentration of 0.5%. In addition, pH has been known to have a strong influence on the corrosion resistance of CpTi and Ti alloys.13,11,12  Understanding the corrosion behavior of Ti in dental implants is very important because it can negatively affect biocompatibility and mechanical integration, thus leading to failure of dental implants over time.9,1315 

In the oral environment, Ti implants are constantly exposed to saliva. Upon placement, the entire surface of the implant can be contaminated with saliva. Once placed, the implant interface is continuously washed with saliva via the gingival sulcus. The presence of electrolytes in saliva can create damage by inducing crevice corrosion, which has been found in many dental implants.16,17 

In addition to the constant contamination, the pH of saliva can vary around dental implants. Many factors can change the pH of saliva. For example, infection can acidify the pH of saliva.11  Food, oral hygiene products, age, periodontitis, oral habits such as smoking, systemic disease, and salivary gland radiation can also contribute to the pH changes in saliva.1722  The normal saliva pH for a healthy adult ranges from 6.3 to 7.23  However, the resting pH of saliva can temporarily decrease to an average of 3.5 if acidic food and beverages are ingested and chronic/systemic disease is present,2426  or can increase to 8.3 if certain alkaline foods are taken in, such as milk, vegetables, and grains.24  The pH of saliva may contribute to the corrosion of dental implants.22,2732 

Many studies have investigated the influence of pH in artificial saliva on the corrosion behavior of Ti.1,3,11,12,18,22,31  However, most of the studies were performed using pH 5, which does not replicate the normal clinical situation. In addition, the interaction between Ti and the oral environment gives rise to electrochemical reactions.22,28  Within the corrosion studies of dental alloys, knowledge of the electrochemical environment of the oral cavity is limited.28  It is hypothesized that a lower saliva pH level induces greater corrosive damage on Ti, and as the pH level increases from 3 to 9, the amount of corrosive damage will decrease. Hence, the aim of this study was to investigate the corrosive behavior of CpTi in artificial saliva at 37°C, as a function of pH values (3–9) and to characterize the Ti surfaces as a result of the corrosion process.

Materials and solutions

In the current study, 25 CpTi alloy discs (American Elements, Los Angeles, Calif) ∅︀12 mm and 7 mm thickness were used. The tests were conducted in artificial saliva with different pH levels: 3, 4.5, 6, 6.5, 7.5, and 9. The artificial saliva and the electrolytes were formed based on studies by Fusayama et al and Meyer.32,33  The composition of this solution, as seen in Table 1, closely resembles that of natural human saliva. To obtain a more acidic artificial saliva solution, lactic acid was added in 1 mL increments, and to obtain a more basic artificial saliva solution, NaOH was added in 1 mL increments.

Table 1

Composition of artificial saliva*

Composition of artificial saliva*
Composition of artificial saliva*

Before the test, all the samples were polished and cleaned using standard metallographic methods. The samples were first mounted in mounting cups (Buehler, Lake Bluff, Ill) with clear orthodontic resin (Dentsply, Milford, Del). The samples were cured under 20 psi. Samples were removed and rough polished using 8″ 320 grit SiC paper to ensure that the sample was flat. The intermediate polishing step was accomplished using a TexMet polishing cloth and 9-μm diamond polishing paste suspended in Meta-Di fluid (Buehler). Once all deep scratches were removed from the surface, the final polishing step was accomplished using a Chemomet polishing cloth and colloidal silica (Buehler). Both the intermediate and final polishing steps were conducted for 10 minutes. Samples were cleaned using 70% Isopropyl alcohol for 15 minutes, then dionized water for 15 minutes in an ultrasonic bath before testing. Samples were dried using a hot air gun.

Electrochemical test methods

Tests were conducted in an electrochemical cell (Figure 1) and performed in triplicate for each group. The corrosion cell held 150 mL of artificial saliva solution. All measurements were made in a standard 3-electrode cell system. A saturated calomel electrode was used as the reference electrode and a graphite rod as a counter electrode. A Gamry potentiostat (Gamry Instruments, Warminster, Pa) was used to carry out the corrosion measurements. After mounting the samples, a period of 30 minutes was allotted for initial stabilization and heating of the solution to 37°C, to mimic the normal body temperature. Initially, the evolution of open circuit potential (OCP) was monitored for 3 600 seconds.

Figure 1.

Electrochemical experimental set-up.

Figure 1.

Electrochemical experimental set-up.

Close modal

The 2-hour electrochemical impedance spectroscopy (EIS) test was conducted after OCP monitoring. The EIS measurements were performed in the frequency range from 100 KHz to 0.01 Hz, with AC sine wave amplitude of 10 mV applied to the electrode at its corrosion potential. These values were used to determine the real (Z') and imaginary (Z”) components of the impedance, which were plotted with the Nyquist plot. The total impedance (|Z|) and phase angle were seen in a Bode plot as a function of the frequency. Impedance data are frequently interpreted in terms of an “equivalent circuit” based on a plausible physical model where the circuit elements represent electrochemical properties of the metal and its oxide film.27  The EIS data were modulated into equivalent electrical circuits using the Zview2 corrosion software (Scribner Association Inc, Southern Pines, NC) to better understand the pH effect on CpTi. Finally, the samples were anodically polarized from −0.8 V to 1.8 V at a scan rate of 5 mV/sec. Potentiodynamic measurements were carried out to determine the initiation and propagation of local corrosion, which is associated with the breakdown of passive (protective) film. The electrochemical corrosion properties of CpTi at each pH were measured in terms of corrosion potential (Ecorr), corrosion current density (Icorr), and passivation current density (Ipass).

Surface characterization

Surface characterization was conducted using scanning electron microscopy (SEM) (Joel JSM-6490 LV, Oxford Instruments, Oxford, UK).

Statistical analysis

The corrosion parameters, including Ecorr, Icorr, double-layer capacitance (Cdl), solution resistance (Rsol), and polarization resistance (Rp) were statistically analyzed using one-way analysis of variance (ANOVA) for analyzing how pH affects the corrosion parameters. Tukey's honestly significant difference test (α = 0.05) was chosen as the multiple-comparison technique when necessary (Statistical Package for the Social Sciences, version 17.0, SPSS Inc, Chicago, Ill).

EIS data

The Nyquist (Figure 2) and Bode plots (Figure 3) provide the corrosion kinetics (impedance and frequency information). The Nyquist plot shows the electrochemical resistance of the Ti surface.

Figures 2–5.

Figure 2. Representative Nyquist plot from electrochemical impedance spectroscopy (EIS) results for commercially pure titanium (CpTi) in artificial saliva as a function of pH. Figure 3. Representative Bode plot from EIS results for CpTi in artificial saliva as a function of pH. Figure 4. Equivalent circuit used for fitting the experimental data. Figure 5. Representative polarization curves at each pH for CpTi in artificial saliva.

Figures 2–5.

Figure 2. Representative Nyquist plot from electrochemical impedance spectroscopy (EIS) results for commercially pure titanium (CpTi) in artificial saliva as a function of pH. Figure 3. Representative Bode plot from EIS results for CpTi in artificial saliva as a function of pH. Figure 4. Equivalent circuit used for fitting the experimental data. Figure 5. Representative polarization curves at each pH for CpTi in artificial saliva.

Close modal

From the Nyquist plot, it can be noted that pH 7.5 had the poorest electrochemical resistance, followed by pH 9.0. For both pH 7.5 and 9.0, there was an increase in the semicircular diameter of the capacitance loop (as indicated by the Nyquist plot). An increase in capacitance leads to a decrease in the overall corrosion resistance. Corrosion resistance appeared to be highest at pH 3.0. Unlike the Bode plot, the Nyquist plot did not account for frequency. The Bode plot also suggested that the resistance of pH 7.5 is the poorest because of the lower Rp value (resistance of the passive film). Only one time constant was observed at each pH, indicating the formation of a homogenous passive film.34 

Figure 4 represents the Randles equivalent circuit used to model the corrosion behavior based on the EIS results. The components of the equivalent circuit are Rs, Rp, and Cdl. Mean values at each pH are compared in Table 2.

Table 2

Corrosion parameters obtained from potentiodynamic polarization curves and from electrochemical impedance spectroscopy modeling of commercially pure titanium in artificial saliva as a function of pH*

Corrosion parameters obtained from potentiodynamic polarization curves and from electrochemical impedance spectroscopy modeling of commercially pure titanium in artificial saliva as a function of pH*
Corrosion parameters obtained from potentiodynamic polarization curves and from electrochemical impedance spectroscopy modeling of commercially pure titanium in artificial saliva as a function of pH*

Potentiodynamic curves

Results of the cyclic potentiodynamic polarization curves for the Ti alloy are presented in Figure 5. All samples revealed an active to passive behavior for each pH. However, the varying potentiodynamic curves indicated a behavior change dependent on pH. This was exhibited by the unique curves represented in Figure 5. A summary of the corrosion data obtained from the cyclic potentiodynamic polarization curves and EIS modeling is provided in Table 2. From the cyclic potentiodynamic polarization curves, the following values were determined: Ecorr, Icorr, and Ipass. Mean values for each pH are compared in Table 2.

Results of one-way ANOVA showed that pH had a significant influence on the Ecorr, Icorr, Cdl, and Rsol values (P < .05). The change in Cdl was significant when comparing pH 3.0 to all other pH values (P < .05). The Ecorr was significant when comparing pH 3.0 to all other pH (P < .05). The Icorr was significant for pH 6.0 and 7.5 when compared to pH 9.0 (P < .05). The Rp showed no significance when comparing pH between tests (P > .05). The Ipass values, although not statistically significant, indicated greater current was required for development of the passive film layer at neutral pH.

Surface examination of corroded surfaces

The SEM images taken after the electrochemical tests suggest surface corrosion at pH 6.5, 7.5, and 9.0 (Figure 6e through g, respectively, arrows). Considerable evidence of surface corrosion represented by surface damage can be seen, particularly at pH 7.5. Negligible surface corrosion existed at pH 3.0, 4.5, and 6.0 with respect to the control sample (Figure 6a through d). The control sample was not subjected to the electrochemical tests.

Figure 6.

Scanning electron microscopy images of commercially pure titanium samples after electrochemical tests were performed taken at ×3000. (a) Control; (b) pH = 3.0; (c) pH = 4.5; (d) pH = 6; (e) pH = 6.5; (f) pH = 7.5; (g) pH = 9.

Figure 6.

Scanning electron microscopy images of commercially pure titanium samples after electrochemical tests were performed taken at ×3000. (a) Control; (b) pH = 3.0; (c) pH = 4.5; (d) pH = 6; (e) pH = 6.5; (f) pH = 7.5; (g) pH = 9.

Close modal

This study showed that the hypothesis was partially accepted. The pH does affect the corrosion behavior of CpTi in artificial saliva; however, Ecorr, Icorr, and Ipass values suggested that artificial saliva at neutral pH (6.0, 6.5, and 7.5) led to greater corrosion of the CpTi discs compared with more acidic saliva. This was indicated by the decreasing Ecorr values, increasing Icorr values, and increasing Ipass values for all three pH values.

Pure Ti and Ti alloys are highly reactive; therefore, when they are exposed to fluid media or air, they quickly develop a layer of titanium dioxide (TiO2). This layer of TiO2, also known as a passive film, acts as a protective interface between the metal structure and biological medium, that is, saliva.4,9  The thickness and composition of the passive film layer is dependent on the environment the Ti is subjected to. Once formed, the passive film layer is constantly broken down and reformed in the oral cavity due to the existence of electrochemical reactions present in the saliva and mechanical actions caused by mastication and micromovements. The combination of chemical reactions and mechanical actions can lead to a complex degradation process of the passive film layer, which can ultimately lead to surface corrosion.

All 3 testing values (Ecorr, Icorr, and Ipass) suggested that an increase in current is required for the development of the passive film layer. This behavior can indicate that although a passive layer is forming, defects in the oxide film may form as well, that is, there is a tendency toward formation of a more irregular or porous oxide layer.2,8,35  The defects can eventually lead to dissolution and subsequent release of metal ions into the system.

The results from the Nyquist and Bode plots indicated that the oxide film formed on the surface of CpTi is thinner at the neutral pH 7.5 (higher corrosion kinetics). If the passive film is thinner, it is easier to penetrate, also leading to an increase in potential pitting. Although the Nyquist and Bode plots agree with potentiodynamic results for pH 7.5, they do not support the results of potentiodynamic tests for pH 6.0 and 6.5. In addition, only the Cdl value for pH 7.5 (capacitance was higher for 7.5) is in agreement with potentiodynamic and SEM results. An increase in capacitance is associated with an decrease in the thickness of the passive layer over time.27,34,36,37  A possible explanation is that the EIS tests performed in this study were conducted over a short time, only 2 hours. Over a prolonged period of time (days, weeks, years) the passive film layer at pH 6.0 and 6.5 may increase in thickness, improving the overall corrosion resistance. However, as noted by the SEM images, during the short time of the potentiodynamic test performed in this study, breakdown of the passive film layer was possible. Earlier studies also reported such particular behavior of heavy surface damage, even pitting, on Ti under the influence of surrounding media, particularly pH level and amount of dissolved oxygen.8,13 

A previous study by Souza et al22  suggested that the protective passive film on Ti alloys is less protective at lower pH. The study evaluated the effect of pH on the electrochemical behavior of Ti alloys in Ringer solution and found that, in contrast to the current study, Ti alloys tend to be more corrosive at a lower pH. In their study, 2 Ti alloys were compared (Ti-13Nb-13Zr and Ti-6Al-4 V), not CpTi, at only 2 pH values, 5.5 and 7.0. In addition, Ringer solution differs greatly in electrolytic composition from artificial saliva.8,13,15,16,34,35 

This study had several limitations. It was limited by the small number of samples tested and the large range of pH values. In addition, acidic or basic components within the artificial saliva solution may have contributed to the corrosion at more neutral pH. Future work will look at close pH intervals of artificial saliva in order to identify the pH value with the greatest electrochemical corrosion. Galvanic studies should be incorporated; in addition, the chemical and mechanical components of corrosion should be combined via tribocorrosion studies of the implant system.37,38 

This study provided a novel and systematic approach to studying corrosion of Ti in the oral environment by using different electrochemical techniques, particularly impedance measurements.35  Understanding the corrosion process of Ti alloys is crucial in predicting the long-term effects of current implant therapy. Corrosion products may have a beneficial or potentially detrimental effect on osseointegration of dental implants.10  Further studies via electrochemical studies and tribocorrosion37,38  will continue the corrosion investigation and ultimately lead to the most effective treatment for patients.

The results from this study showed that the pH of the media had a significant influence on the corrosion behavior of CpTi. The following conclusions were drawn:

  1. The Icorr values indicated that neutral pH 7.5 had inferior corrosion resistance. This finding was supported by the impedance results of the Nyquist plot and Rp value determined via the Bode plot.

  2. The Ecorr, Icorr, and Ipass values suggested a greater corrosion potential at more neutral pH values of 6.0, 6.5, and 7.5. The EIS results did not support greater corrosion kinetics at pH values of 6.0 and 6.5; however, it did support greater corrosion kinetics at pH 7.5.

  3. The SEM images showed considerable surface damage under pH values of 6.5, 7.5, and 9.0.

  4. The weak corrosion behavior at neutral pH invites some concerns and highlights the need for further study.

Abbreviations

Cdl

double-layer capacitance

CpTi

commercially pure titanium

Ecorr

corrosion potential (Tafel method)

EIS

electrochemical impedance spectroscopy

Icorr

corrosion current potential (Tafel method)

Ipass

passivation current density

OCP

open circuit potential

Rp

polarization resistance

Rsol

solution resistance

SEM

scanning electron microscopy

Ti

titanium

This study was partially supported by the American Academy of Implant Dentistry Research Foundation to D.J.L.

1
Nakagawa
M
,
Matsuya
S
,
Shiraishi
T
,
Ohta
M
.
Effect of fluoride concentration and pH on corrosion behavior of titanium for dental use
.
J Dent Res
.
1999
;
78
:
1568
1572
.
2
Assis
SL
,
Wolynec
S
,
Costa
I
.
Corrosion characterization of titanium alloys by electrochemical techniques
.
Electrochemica Acta
.
2006
;
8
:
1815
1819
.
3
Viera
A
,
Rocha
L
,
Celis
J
.
Influence of pH and corrosion inhibitors on the tribocorrosion of titanium in artificial saliva
.
Wear
.
2006
;
261
:
994
1001
.
4
Olmedo
DG
,
Tasat
DR
,
Duffo
G
,
Guglielmotti
MB
,
Cabrini
RL
.
The issue of corrosion in dental implants: a review
.
Acta Odontol Latinoam
.
2009
;
22
:
3
9
.
5
Landolt
D
.
Corrosion and Surface Chemestry of Metals
.
Lausanne, Switzerland
:
CRC Press;
2007
.
6
Williams
DF
,
Williams
RL
.
Degradative effects of the biological environment on metals and ceramics
.
In
:
Ratner
BD
,
ed
.
Biomaterials Science: An Introduction to Material in Medicine
.
Amsterdam
:
Elsevier;
2004
:
430
439
.
7
Williams
DF
.
Physiological and microbiological corrosion
.
Crit Rev Biocompat
.
1985
:
1
:
1
4
.
8
Hanawa
T
.
In vivo metallic biomaterials and surface modification
.
Materials Sci Eng
.
1999
;
A267
:
260
266
.
9
Chaturvedi
TP
.
An overview of the corrosion aspect of dental implants (titanium and its alloys)
.
Indian J Dent Res
.
2009
;
20
:
91
98
.
10
Hallab
NJ
,
Jacobs
JJ
,
Skipor
A
,
Black
J
,
Mikecz
K
,
Galante
JO
.
Systemic metal-protein binding associated with total joint replacement arthroplasty
.
J Biomed Mater Res
.
2000
;
49
:
353
361
.
11
Nakagawa
M
,
Matsuya
S
,
Udoh
K
.
Effects of fluoride and dissolved oxygen concentrations on the corrosion behavior of pure titanium and titanium alloys
.
Dent Mater J
.
2002
;
21
:
83
92
.
12
Ionescu
D
,
Popescu
N
,
Demetrescu
I
.
The kinetic parameters in electrochemical behaviour of titanium in artificial saliva
.
Anale Universitatii Bucuresti
.
2008
;
9
:
77
83
.
13
Cheng
X
,
Roscoe
SG
.
Corrosion behavior of titanium in the presence of calcium phosphate and serum proteins
.
Biomaterials
.
2005
;
26
:
7350
7356
.
14
Slemnik
M
,
Milosev
I
.
An impedance study of two types of stainless steel in Ringer physiological solution containing complexing agents
.
J Mater Sci Mater Med
.
2006
;
17
:
911
918
.
15
Virtanen
S
,
Milosev
I
,
Gomez-Barrena
E
,
Trebse
R
,
Salo
J
,
Konttinen
YT
.
Special modes of corrosion under physiological and simulated physiological conditions
.
Acta Biomater
.
2008
;
4
:
468
476
.
16
Ergun
C
,
Doremus
RH
,
Lanford
WA
.
Interface reaction/diffusion in hydroxylapatite-coated SS316L and CoCrMo alloys
.
Acta Materialia
.
2004
;
52
:
4767
4772
.
17
Nikolopoulou
F
.
Saliva and dental implants
.
Implant Dent
.
2006
;
15
:
372
376
.
18
Schiff
N
,
Grosgogeat
B
,
Lissac
M
,
Dalard
F
.
Influence of fluoride content and pH on the corrosion resistance of titanium and its alloys
.
Biomaterials
.
2002
;
23
:
1995
2002
.
19
Olsen
I
,
Birkeland
JM
.
Assessment of denture plaque pH in subjects with and without denture stomatitis
.
Scand J Dent Res
.
1975
;
83
:
370
374
.
20
Aren
G
,
Sepet
E
,
Ozdemir
D
,
Dinccag
N
,
Guvener
B
,
Firatli
E
.
Periodontal health, salivary status, and metabolic control in children with type 1 diabetes mellitus
.
J Periodontol
.
2003
;
74
:
1789
795
.
21
Edgar
WM
,
Higham
SM
,
Manning
RH
.
Saliva stimulation and caries prevention
.
Adv Dent Res
.
1994
;
8
:
239
245
.
22
Souza
ME
,
Lima
L
,
Lima
CR
,
Zavaglia
CA
,
Freire
CM
.
Effects of pH on the electrochemical behaviour of titanium alloys for implant applications
.
J Mater Sci Mater Med
.
2009
;
20
:
549
552
.
23
Humphrey
SP
,
Williamson
RT
.
A review of saliva: normal composition, flow, and function
.
J Prosthet Dent
.
2001
;
85
:
162
169
.
24
Murrell
S MT
,
Moynihan
PJ
,
Qian
F
,
Wefel
JS
.
Comparison of in vitro erosion potentials between beverages available in the United Kingdom and the United States
.
J Dent
.
2010
;
38
:
284
289
.
25
Dong
YM
,
Pearce
EI
,
Yue
L
,
Larsen
MJ
,
Gao
XJ
,
Wang
JD
.
Plaque pH and associated parameters in relation to caries
.
Caries Res
.
1999
;
33
:
428
436
.
26
Gregory-Head
BL
,
Curtis
DA
,
Kim
L
,
Cello
J
.
Evaluation of dental erosion in patients with gastroesophageal reflux disease
.
J Prosthet Dent
.
2000
;
83
:
675
680
.
27
Grosgogeat
B
,
Boinet
M
.
Electrochemical studies of the corrosion behavior of titanium and the Ti-6Al-4V alloy using electrochemical impedance spectroscopy
.
Biomed Mater Eng
.
2004
;
14
:
323
331
.
28
Kuphasuk
C
,
Oshida
Y
,
Andres
CJ
,
Hovijitra
ST
,
Barco
MT
,
Brown
DT
.
Electrochemical corrosion of titanium and titanium-based alloys
.
J Prosthet Dent
.
2001
;
85
:
195
202
.
29
Cortada
M
,
Giner
L
,
Costa
S
,
Gil
FJ
,
Rodriguez
D
,
Planell
JA
.
Galvanic corrosion behavior of titanium implants coupled to dental alloys
.
J Mater Sci Mater Med
.
2000
;
11
:
287
293
.
30
Huang
HH
.
Effects of fluoride concentration and elastic tensile strain on the corrosion resistance of commercially pure titanium
.
Biomaterials
.
2002
;
23
:
59
63
.
31
Lindholm-Sethson
B
,
Ardlin
BI
.
Effects of pH and fluoride concentration on the corrosion of titanium
.
J Biomed Mater Res A
.
2008
;
86
:
149
159
.
32
Fusayama
T
,
Katayori
T
,
Nomoto
S
.
Corrosion of gold and amalgam placed in contact with each other
.
J Dent Res
.
1963
;
42
:
1183
1197
.
33
Meyer
JM
.
Biocompatibility and corrosion of dental alloys
.
Rev Med Suisse Romande
.
1998
;
118
:
17
25
.
34
Olsson
COA
,
Landolt
D
.
Passive films on stainless steels—chemistry, structure and growth
.
Electrochimica Acta
.
2003
;
48
:
1093
1104
.
35
Valero Vidal AOJ, Igual Munoz A
.
Study of the adsorption mechanisms of albumin on CoCrMo biomedical alloy by means of electrochemical impedance spectroscopy
.
J Electrochem Soc
.
2007
;
154
:
C562
C570
.
36
Milosev
I
,
Strehblow
HH
.
The behavior of stainless steels in physiological solution containing complexing agent studied by X-ray photoelectron spectroscopy
.
J Biomed Mater Res
.
2000
;
52
:
404
412
.
37
Abd-El-Kader
H
,
El-Raghy
SM
.
Wear-corrosion mechanism of stainless steel in chloride media
.
Corrosion Sci. 986;
26
:
647
653
.
38
Mathew
MT
,
Srinivasa
Pai P
,
Pourzal
R
,
Fischer
A
,
Wimmer
MA
,
Significance of tribocorrosion in biomedical applications: overview and current status
.
Adv Tribol
.
2009
;
250986
.