Dental implants are widely used for replacing the missing teeth of partially or completely edentulous patients.1,2  Many studies have confirmed that dental implants show predictable results with a high percentage of success.35  However, clinicians must be aware of the mechanical complications, such as screw loosening, screw or implant fracture, and prosthetic framework fracture, which may result in implant failures.611  Among these, the incidence of implant fracture is low (0.08–0.74%),12  but it poses great difficulties for the clinician who must solve this problem. According to Balshi,8  implant fractures may be the result of (1) defects in implant design or material, (2) nonpassive fit of the prosthetic framework, and (3) physiological or biomechanical overload.

The aim of this case letter was to analyze a dental implant fracture and define how to effectively manage this challenging situation.

In August 2008, a 56-year-old nonsmoking male patient was referred to our clinic (Baskent University, Faculty of Dentistry, Oral and Maxillofacial Surgery Department, Ankara, Turkey). The patient complained of pain, bleeding at the gingival margin, and loosening of a 4-unit implant supported prosthesis in the mandibular left premolar–molar region. The fixed prosthesis was mobile; however, the implant mobility was not present. The panoramic radiograph revealed no evidence of fracture or marginal bone loss around the implant (Figure 1).

Figure 1.

The panoramic radiograph of the patient's first visit.

Figure 1.

The panoramic radiograph of the patient's first visit.

Close modal

The patient's anamnesis revealed the placement of two titanium implants (Astra Tech, Mölndal, Sweden) in the mandibular left region (3.5 × 11 mm for tooth #21 and 4 × 13 mm for tooth #18) 4 years previously. An implant-supported fixed partial denture (FPD) with 2 pontics was performed. Two years after the FPD was completed, the patient complained about the mobilization of this unit and went to their own clinician. The fixed denture was removed and the gold screws of the crowns were retightened. Following this retightening the patient had no problems except for interim pain and spontaneous bleeding.

When the patient presented to our clinic, the patient's implant-supported bridge was loose and, therefore, it was removed. The abutments were also mobile and had to be removed. To evaluate the situation, a mucoperiosteal flap was raised and the implants were exposed. Unfortunately, it was not feasible to retain the same denture and implants because the buccal coronal portions of the implants were fractured and there was buccal bone loss (Figure 2). The treatment plan was to remove the implants, rectify the defect with autogenous bone grafting, and then insert the new implants. The implants were retrieved with a trephine bur under copious irrigation to reduce heat from friction (Figure 3). The defects were filled with synthetic bone graft (Osteobiol Gen Os, 250–1000 micron, 1 gr, Italy) and membranes (Osteobiol Evaluation, 30 × 30 mm, Italy). Following a 6-month post-graft healing period, 2 implants (4 × 11 mm each) were inserted in the related region. However, tooth position #18's grafted site was not suitable for inserting an implant; thus, the second implant was inserted in tooth position #19. After 3 months, a 4-unit bridge was placed with a single pontic and cantilever. The patient's re-evaluation appointments after 6 months to 3 years showed no evidence of problems related either to the prosthesis or to the implants. (Figure 4)

Figure 2.

Fractured buccal portion of the osseointegrated implants.

Figure 2.

Fractured buccal portion of the osseointegrated implants.

Close modal
Figure 3.

Extracted dental implants with trephine bur.

Figure 3.

Extracted dental implants with trephine bur.

Close modal
Figure 4.

The panoramic radiograph of the patient after 3 years.

Figure 4.

The panoramic radiograph of the patient after 3 years.

Close modal

Mechanical problems after implant insertion occur principally due to framework misfit, excessive occlusal force, poor prosthetic component design, and unfavorable leverage or parafunctional activities.13  Sanchez-Perez et al14  have provided a most comprehensive classification of the fracture risk factors. They categorized the fracture cases according to patient factors, implant-related factors, and prosthetic factors.

Patient factors are related to bone loss, bruxism, or pocket depth (PD). According to Rangert et al,9  most of the patients (56%) with implant fracture had bruxism or heavy occlusal forces, which might lead to bone loss around the cervical area of loaded implants. Coronal bone resorption produces higher bending stress around the implant and also causes PD, bleeding, or an inflammatory response. In the present case, the panoramic radiographs did not reveal any bone resorption around the implant area. However, the existing resorption on the buccal part of the implant was easily observable when the mucoperiosteal flap was raised. This unusual event was quite confusing for properly diagnosing the problem. The patient was also evaluated for bruxism and parafunctional habits, but no evidence was found. Bleeding and interim pain of the patient were assumed to be caused by the edges of the existing prosthesis.

Implant-related factors include diameter, crown implant ratio, and implants design. A 4-mm implant has a 30% higher fatigue resistance compared to the 3.75-mm implant.9  In the present case, the diameters of the implants were 3.5 and 4 mm. Implants of at least 4 mm diameter are suitable for the molar region. However, a 4-mm diameter implant was also fractured, which might be due to the buccolingual and mesiodistal dimensions of the 4-unit bridge used for its restoration.

Prosthetic factors include cantilevers or loosening of the screws. Eckert et al11  stated that insufficient or excessive torque to the retaining screws may also lead to improper prosthesis connection. Rangert et al9  stated that most of the fracture cases (82%) had distal or mesial cantilevers. Conversely in our case, the 4-unit bridge with 2 pontics was fractured while the 4-unit bridge with 1 pontic and 1 distal cantilever had no problems even after 3 years. However, loosening of the screws occurred 2 times prior to the implant fracture. When screw loosening occurs, the clinician must immediately take precautionary measures so that the problem does not become more complicated.

Our treatment included complete removal of the fractured implants, bone grafting, and placement of 2 new implants with larger diameters. This treatment was chosen because of the chronic screw loosening and bone resorption on the buccal part of the implants. Wider and multiple implants may help prevent this problem. This treatment option is based upon previous reports concerning dental implant fractures.

Abbreviations

FPD

fixed partial denture

PD

pocket depth

1
Brånemark
PI
,
Hansson
BO
,
Adell
R
,
et al
.
Osseointegrated implants in the treatment of the edentulous jaw: experience from a 10-year period
.
Scand J Plast Reconstr Surg Suppl
.
1977
;
16
:
1
132
.
2
Adell
R
,
Lekholm
U
,
Rockler
B
,
Brånemark
PI. A
15-year study of osseointegrated implants in the treatment of the edentulous jaw
.
Int J Oral Surg
.
1981
;
10
:
387
416
.
3
Lekholm
U
,
Gunne
J
,
Henry
P
,
Higuchi
K
,
Lindén
U
,
Bergström
C
,
et al
.
Survival of the Brånemark implant in partially edentulous jaws: a 10-year prospective multicenter study
.
Int J Oral Maxillofac Implants
.
1999
;
14
:
639
645
.
4
Charyeva
O
,
Altynbekov
K
,
Zhartybaev
R
,
Sabdanaliev
A
.
Long-term dental implant success and survival–a clinical study after an observation period up to 6 years
.
Swed Dent J
.
2012
;
36
:
1
6
.
5
Higuchi
KW
,
Folmer
T
,
Kultje
C
.
Implant survival rates in partially edentulous patients: a 3-year prospective multicenter study
.
J Oral Maxillofac Surg
.
1995
;
53
:
264
268
.
6
Velásquez-Plata
D
,
Lutonsky
J
,
Oshida
Y
,
Jones
R
.
A close-up look at an implant fracture: a case report
.
Int J Periodontics Restorative Dent
.
2002
;
22
:
483
491
.
7
Mendonça
G
,
Mendonça
DB
,
Fernandes-Neto
AJ
,
Neves
FD
.
Management of fractured dental implants: a case report
.
Implant Dent
.
2009
;
18
:
10
16
.
8
Balshi
TJ
.
An analysis and management of fractured implants: a clinical report
.
Int J Oral Maxillofac Implants
.
1996
;
11
:
660
666
.
9
Rangert
B
,
Krogh
PH
,
Langer
B
,
Van Roekel
N
.
Bending overload and implant fracture: a retrospective clinical analysis
.
Int J Oral Maxillofac Implants
.
1995
;
10
:
326
334
.
10
Weinberg
LA
.
The biomechanics of force distribution in implant-supported prostheses
.
Int J Oral Maxillofac Implants
.
1993
;
8
:
19
31
.
11
Eckert
SE
,
Meraw
SJ
,
Cal
E
,
Ow
RK
.
Analysis of incidence and associated factors with fractured implants: a retrospective study
.
Int J Oral Maxillofac Implants
.
2000
;
15
:
662
667
.
12
Berglundh
T
,
Persson
L
,
Klinge
B
.
A systematic review of the incidence of biological and technical complications in implant dentistry reported in prospective longitudinal studies of at least 5 years
.
J Clin Periodontol
.
2002
;
29
Suppl 3
:
197
212
;
discussion 232–233. Review
.
13
McGlumphy
EA
,
Larsen
PE
,
Peterson
LJ
.
Etiology of implant complications: anecdotal reports vs prospective clinical trials
.
Compend Suppl
.
1993
;
15
:
S544
548
;
quiz S565–566
.
14
Sánchez-Pérez
A
,
Moya-Villaescusa
MJ
,
Jornet-Garcia
A
,
Gomez
S
.
Etiology, risk factors and management of implant fractures
.
Med Oral Patol Oral Cir Bucal
.
2010
;
1
;
15
:
e504
e508
.
Review
.