The objective of this study was to evaluate the primary stability of short and conventional dental implants with different platform types at different site densities in vitro. One hundred twenty implants were placed in polyurethane blocks that simulate different bone densities (bone types I and IV). The implants were divided into 10 groups, with 12 implants each according to the type of prosthetic connections (external hexagon, EH; morse taper, MT) and size of the implants (conventional: 4 × 10 mm; short: 5 × 5, 5.5 × 5, 5 × 6, and 5.5 × 6 mm). Insertion torque and resonance frequency analyses were performed to evaluate the primary stability. The Kruskal-Wallis test complemented by Dunn's test and the Mann-Whitney test were used for statistical analysis. These tests were applied at the confidence level of 95% (P < .05). The implants installed in blocks with density type IV exhibited reduced insertion torque compared with implants placed in blocks with density type I. Short implants with EH exhibited increased insertion torque compared with short implants with MT in blocks with bone density type I. In general, implants installed in blocks with density type I exhibited greater primary stability. The short implants with EH with a 5.5-mm diameter and the short implants with MT with a 5-mm diameter exhibited reduced primary stability. No differences between short and conventional implants were noted. Short implants have primary stability and insertion torque at least equivalent to conventional implants irrespective of the platform type and density of the site.

The use of implants in the rehabilitation of edentulous and partially edentulous patients promotes a more conservative alternative in prosthetic treatment because the missing teeth can be replaced with great predictability, with no wear of adjacent teeth during the prosthetic procedures,1  thus promoting greater patient satisfaction.2  However, the installation of implants is dependent on bone tissue availability, which allows obtaining a suitable positioning of the dental implants without causing injury to noble structures present in the jaws.3 

Some procedures have been indicated for the resolution of bone viability, such as the elevation of the maxillary sinus,4  the use of onlay graft blocks,5  the lateralization of the inferior alveolar nerve,6  and osteodistraction.7  However, these surgical procedures are complex and increase patient discomfort, the possibility of injuries of noble structures, and the time in the beginning of the prosthetic procedures.8,9 

Short implants have been cited as a less traumatic alternative to the rehabilitation of areas with poor bone availability.8,10,11  Studies have demonstrated that the success rate of short implants is similar to that of conventional implants.8,11,12  However, the success of implants is highly dependent on obtaining primary stability,1315  and factors such as the implant design,16  bone density,17  and insertion torque18  influence this parameter.

Given that short implants can be used in sites with low bone density, such as the posterior region of the maxillary, to avoid the sinus floor augmentation surgeries, it is important that the short implants present an appropriate design, which can improve the primary stability even in regions with low bone density. Thus, the objective of the study was to assess the primary stability through resonance frequency analysis and the insertion torque of conventional and short size implants with external hexagon or morse taper placed in polyurethane blocks with similar densities of bone types I and IV to answer the following questions. (1) Can the short implants provide primary stability at the same level of the conventional implants? (2) Can the type of platform and the different densities modify the primary stability of the short and conventional implants?

Implants

One hundred twenty tapered implants were used in this study (Implacil de Bortoli, São Paulo, SP, Brazil). The implants were divided into 10 groups according to the type of prosthetic connection (external hexagon [EH] and morse taper [MT]) and implant size (conventional: 4 × 10 mm; short: 5 × 5, 5.5 × 5, 5 × 6, and 5.5 × 6 mm). Six implants of each type were installed in polyurethane blocks of different densities (type I or type IV).

Implants placement

The implants were inserted into the specimens by a single operator following the sequence of surgical drills recommended by the manufacturer. To reproduce in vitro the evaluation of the primary stability of the implant in similar conditions to the bone tissue, blocks of polyurethane (Nacional Ossos, Jaú, Brazil) of 2 different densities simulating bone type I (40 per cubic foot [PCF]/0.64 g/cm3) and type IV (15 PCF or 0.24 g/cm3) were used. The blocks were generated in such a manner to have a length of 9.9 cm, width of 2.55 cm, and height of 2.02 cm. Each block received 5 implants. In total, 24 blocks were produced. Twelve of these blocks had a density similar to type I human bone, whereas the other 12 blocks had a density similar to type IV human bone (the Figure). The surgical drills were replaced every 5 perforations for implant placement.

Figure

(a) The polyurethane block with a density similar to type I human bone. (b) The scanning electron microscopy (SEM) image of the block with density similar to type I human bone. Note that this block has a more massive structure. (c) The polyurethane block with a density similar to type IV human bone. (d) The SEM image of the block with density similar to type IV human bone. Note that this block has a more porosity structure. (e) Implants in the final position into the polyurethane blocks. The external hexagon implants were placed at the level of the blocks, whereas the morse taper implants were installed 2 mm below the end of the blocks.

Figure

(a) The polyurethane block with a density similar to type I human bone. (b) The scanning electron microscopy (SEM) image of the block with density similar to type I human bone. Note that this block has a more massive structure. (c) The polyurethane block with a density similar to type IV human bone. (d) The SEM image of the block with density similar to type IV human bone. Note that this block has a more porosity structure. (e) Implants in the final position into the polyurethane blocks. The external hexagon implants were placed at the level of the blocks, whereas the morse taper implants were installed 2 mm below the end of the blocks.

Close modal

To facilitate the standardization of implant placement in the blocks, 2 surgical guides were prepared with resin acrylic. These surgical guides had the same width and length of the blocks, allowing the placement of the implants at the same position and equidistantly. A distance of 1 cm between the implants and the edge of the blocks was maintained. The EH implants were placed at the level of the blocks, whereas the MT implants were installed 2 mm below the end of the blocks (the Figure).

Evaluation of primary stability: insertion torque and resonance frequency analysis

The primary stability of each implant was assessed via insertion torque and resonance frequency analysis. Insertion torque was measured at the time of implant installation in the block until the implant reached the ideal position using the torque wrench (Implacil de Bortolli, São Paulo, Brazil). Resonance frequency analysis was performed with the Osstell (Osstell, Integration Diagnostic, Göteborg, Sweden). This machine uses a piezoelectric effect to produce a deflection of the implant on the transducer (smart peg), which was adapted directly over the implant and was stimulated to vibrate by means of sinusoidal waves. The stability measured by this device was measured on a scale from 1 (minimum) to 100 (maximum). The torque wrench and the Osstell were calibrated before and after the measurements of each implant. The measurements were repeated 3 times in each implant, and the average of this value was considered the value of the sample to eliminate the measurement bias.

Statistical analysis

The software GraphPad Prism 5 (San Diego, Calif) was used to perform statistical analyses of the study. The data obtained from insertion torque and resonance frequency analyses were assessed for a normal distribution using the Kolgomorov-Smirnov, D'Agostino-Pearson, and Shapiro-Wilk tests. These tests revealed that data were not distributed according to the theorem of the central data distribution. The comparison of the subgroups was performed as follows: (1) the same implants were compared variating the block density (type I vs type IV) by the Mann-Whitney test; (2) the same implants placed in the same variating the implants plataform (EH vs MT) by the Mann-Whitney test; (3) the comparison of 5 types of implants (conventional: 4 × 10 mm; short: 5 × 5, 5.5 × 5, 5 × 6, and 5.5 × 6 mm) in the same block density and with the same type of platform was performed by the Kruskal-Wallis test complemented by the Dunn's test. The comparison of similar implant types installed in blocks with different densities was performed by the Mann-Whitney test. All tests were applied with a significance α level of 95% (P < .05).

The insertion torque of the implants installed in the blocks with a density similar to bone type I was consistently increased compared with the insertion torque of the implants installed in the blocks with a density similar to bone type IV (P < .05). The conventional EH implants presented lower insertion torque compared with short EH implants when the implants were placed in blocks with a bone type I density. In addition, the short MT implants presented lower insertion torque compared with short EH implants when these implants where placed in blocks with a bone type I density. No differences were noted between the implants placed in blocks with densities similar to bone type IV. Table 1 presents the mean (median) ± SD of the insertion torque values in all the groups.

Table 1

Median (Q1, Q3) of the insertion torque values (N/cm2) in all groups*

Median (Q1, Q3) of the insertion torque values (N/cm2) in all groups*
Median (Q1, Q3) of the insertion torque values (N/cm2) in all groups*

*Note the different variables tested in this study (implant size, prosthetic connection, and block density). NS indicates not significant.

Lower insertion torque than the short implants placed in blocks with density similar to type I bone based on the Kruskal-Wallis test with Dunn's test.

Lower insertion torque of the short implants with external hexagon placed in blocks with a density similar to type I bone based on the Mann-Whitney test.

Regarding the data of resonance frequency analysis, reduced primary stability values were noted in implants installed in blocks with a density similar to bone type IV compared with implants placed in blocks with a density similar to bone type I, except the short EH implants (5 × 5 and 5 × 6 mm) and short MT implants (5.5 × 5 mm), which exhibited similar stability when installed in blocks with different densities. Regarding the implants placed in blocks with a density similar to bone type I, short implants with EH (5 × 6 mm) and MT (5.5 × 5 mm) exhibited increased primary stability compared with implants with MT (5 × 6 mm) and EH (5.5 × 5 mm), respectively. Regarding the implants placed in blocks with a density similar to bone type IV, short implants with EH and a 5-mm diameter presented increased primary stability compared with MT implants with the same diameter, whereas the short implants with MT with a 5.5-mm diameter exhibited increased primary stability compared with EH implants with the same diameter. Table 2 presents the mean (median) ± SD of the resonance frequency analysis for all groups.

Table 2

Median (Q1, Q3) of the resonance frequency analysis (RFA) in all groups*

Median (Q1, Q3) of the resonance frequency analysis (RFA) in all groups*
Median (Q1, Q3) of the resonance frequency analysis (RFA) in all groups*

*Note the different variables tested in this study (implant size, prosthetic connection, and block density). Different letters represent significant differences between the implants with the same prosthetic connections placed in blocks with the same density based on the Kruskal-Wallis test with Dunn's test. NS indicates not significant.

Lower RFA than implants with similar dimensions but different prosthetic connections placed in blocks with a density similar to the type I bone using the Mann-Whitney test.

Lower RFA than implants with similar dimensions but different prosthetic connections placed in blocks with a density similar to type IV bone based on the Mann-Whitney test.

Obtaining primary stability has been reported as an important factor for the osseointegration of implants,17,19,20  and this parameter has been used as a reference for determining the best time for application of occlusal loads on the dental implant.21,22  The density of the surgical sites influences the obtained primary stability.17,20,23  This statement was confirmed in this study, given that we verified increased insertion torque and resonance frequency values on implants installed in blocks with a density similar to bone type I compared with implants placed in blocks that simulated the density of bone type IV.

Another important finding from this study was that conventional implants with external hexagons had a reduced insertion torque compared with short implants placed in blocks with a density similar to bone type I. The diameters of the short implants in this study were 5 or 5.5 mm, whereas conventional implants were 4 mm in diameter. The increased diameter of the short implants may explain the increased primary stability of the short implants. It was previously reported that the diameter of the implants has more influence on the primary stability than the length of the implant.16,24 

EH implants exhibited increased insertion torque compared with MT implants in blocks with a density similar to bone type I. Finite element studies revealed that EH implants exhibit increased concentrated stress in the cervical portion of the implant when subjected to the occlusal loads compared with MT implants.25  It is possible that the distribution of forces during implant placement is similar to that observed during occlusal loads,25  and this fact may explain the increased insertion torque of HE implants compared with MT implants. It is worth noting that despite this significant difference, the MT implants presented insertion torque values within the range of 28.33–30 N in the blocks that simulated the density of bone type I, and these values represent good primary stability of the implant during the clinical procedures.22 

An important finding of this study was that resonance frequency analysis data for short implants were generally similar to the data recorded for conventional implants. Changes in the short implants design (eg, increase of the diameter, modification of the implants surfaces)16,26  provided an improvement in clinical outcomes for these implants, inducing increased primary stability12  and promoting clinical outcomes similar to the conventional implants.1012 

In our study, the statistically significant differences that occurred in the insertion torque and resonance frequency analyses were not coincidental. Both tests analyze the primary stability, and their values are influenced by similar parameters (eg, bone density). A positive correlation was noted between these data; however, the result is not statistically significant.27,28  This effect occurs because the insertion torque measures the compression induced by implant placement into the surgical site, whereas the resonance frequency analysis evaluates the micro-movements of the implant in the final position into the jaws.29 

This in vitro study has limitations that should be taken into consideration when interpreting the results. Although the blocks exhibited density characteristics similar to bone, these blocks exhibited homogeneous density, which is not observed in bone tissue. The implants were also installed under optimum conditions, with no conditions that hinder access to the surgical site. Thus, although the preliminary data are important for the knowledge regarding the effect of the short implant design on primary stability, clinical studies assessing the primary stability of short implants are needed to confirm the findings of this study.

In conclusion, the short implants tested in this study exhibited primary stability and insertion torque at least equivalent to conventional implants irrespective of the platform type and density of the site.

Abbreviations

Abbreviations
EH

external hexagon

MT

morse taper

PCF

per cubic foot

1
Al-Quran
FA,
Al-Ghalayini
RF,
Al-Zu'bi
BN.
Single-tooth replacement: factors affecting different prosthetic treatment modalities
.
BMC Oral Health
.
2011
;
21
:
11
34
.
2
Zembic
A,
Wismeijer
D.
Patient-reported outcomes of maxillary implant-supported overdentures compared with conventional dentures
.
Clin Oral Implants Res
.
2014
;
25
:
441
450
.
3
Arlin
ML.
Short dental implants as a treatment option: results from an observational study in a single private practice
.
Int J Oral Maxillofac Implants
.
2006
;
21
:
769
776
.
4
Boyne
PJ,
James
RA.
Grafting of the maxillary sinus floor with autogenous marrow and bone
.
J Oral Surg
.
1980
;
38
:
613
616
.
5
Misch
CM.
Ridge augmentation using mandibular ramus bone grafts for the placement of dental implants: presentation of a technique
.
Pract Periodontics Aesthet Dent
.
1996
;
8
:
127
135
.
6
Friberg
B,
Ivanoff
CJ,
Lekholm
U.
Inferior alveolar nerve transposition in combination with Brånemark implant treatment
.
Int J Periodontics Restorative Dent
.
1992
;
12
:
440
449
.
7
Chiapasco
M,
Zaniboni
M,
Rimondini
L.
Autogenous onlay bone grafts vs. alveolar distraction osteogenesis for the correction of vertically deficient edentulous ridges: a 2-4-year prospective study on humans
.
Clin Oral Implants Res
.
2007
;
4
:
432
440
.
8
Felice
P,
Pistilli
R,
Piattelli
M,
Soardi
E,
Corvino
V,
Esposito
M.
Posterior atrophic jaws rehabilitated with prostheses supported by 5 × 5 mm implants with a novel nanostructured calcium-incorporated titanium surface or by longer implants in augmented bone. Preliminary results from a randomised controlled trial
.
Eur J Oral Implantol
.
2012
;
5
:
149
161
.
9
Vasco
MA,
Hecke
MB,
Bezzon
OL.
Analysis of short implants and lateralization of the inferior alveolar nerve with 2-stage dental implants by finite element method
.
J Craniofac Surg
.
2011
;
22
:
2064
2071
.
10
Anitua
E,
Orive
G,
Aguirre
JJ,
Andía
I.
Five-year clinical evaluation of short dental implants placed in posterior areas: a retrospective study
.
J Periodontol
.
2008
;
79
:
42
48
.
11
Schincaglia
GP,
Thoma
DS,
Haas
R,
Tutak
M,
Garcia
A,
Taylor
TD,
Hämmerle
CH.
Randomized controlled multicenter study comparing short dental implants (6 mm) versus longer dental implants (11-15 mm) in combination with sinus floor elevation procedures. Part 2: clinical and radiographic outcomes at 1 year of loading
.
J Clin Periodontol
.
2015
;
42
:
1042
1051
.
12
Calvo-Guirado
JL,
López Torres
JA,
Dard
M,
Javed
F,
Pérez-Albacete Martínez C, Maté Sánchez de Val JE. Evaluation of extrashort 4-mm implants in mandibular edentulous patients with reduced bone height in comparison with standard implants: a 12-month results
.
Clin Oral Implants Res
.
2016
;
27
:
867
874
.
13
Javed
F,
Romanos
GE.
Role of implant diameter on long-term survival of dental implants placed in posterior maxilla: a systematic review
.
Clin Oral Invest
.
2015
;
19
:
1
10
.
14
Moraschini
V,
Velloso
G,
Luz
D,
Barboza
EP.
Implant survival rates, marginal bone level changes, and complications in full-mouth rehabilitation with flapless computer-guided surgery: a systematic review and meta-analysis
.
Int J Oral Maxillofac Surg
.
2015
;
44
:
892
901
.
15
Queiroz
TP,
Aguiar
SC,
Margonar
R,
de Souza Faloni
AP,
Gruber
R,
Luvizuto
ER.
Clinical study on survival rate of short implants placed in the posterior mandibular region: resonance frequency analysis
.
Clin Oral Implants Res
.
2015
;
26
:
1036
1042
.
16
Möhlhenrich
SC,
Heussen
N,
Elvers
D,
Steiner
T,
Hölzle
F,
Modabber
A.
Compensating for poor primary implant stability in different bone densities by varying implant geometry: a laboratory study
.
Int J Oral Maxillofac Surg
.
2015
;
44
:
1514
1520
.
17
Martinez
H,
Davarpanah
M,
Missika
P,
Celletti
R,
Lazzara
R.
Optimal implant stabilization in low density bone
.
Clin Oral Implants Res
.
2001
;
12
:
423
432
.
18
Trisi
P,
Perfetti
Q,
Baldoni
E,
Berardi
D,
Colagiovanni
M,
Scogna
G.
Implant micromotion is related to peak insertion torque and bone density
.
Clin Oral implants Res
.
2009
;
20
:
467
471
.
19
Albrektsson
T,
Brånemark
PI,
Hansson
HA,
Lindström
J.
Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man
.
Acta Orthop Scand
.
1981
;
52
:
155
170
.
20
Bilhan
H,
Bilmenoglu
C,
Urgun
AC,
Ates
G,
Bural
C,
Cilingir
A,
Geckili
O.
Comparison of the primary stability of two implant designs in two different bone types: an in vitro study
.
Int J Oral Maxillofac Implants
.
2015
;
30
:
1036
1040
.
21
Bischof
M,
Nedir
R,
Szmukler-Moncler
S,
Bernard
JP,
Samson
J.
Implant stability measurement of delayed and immediately loaded implants during healing
.
Clin Oral Implants Res
.
2004
;
15
:
529
539
.
22
Barewal
RM,
Stanford
C,
Weesner
TC.
A randomized controlled clinical trial comparing the effects of three loading protocols on dental implant stability
.
Int J Oral Maxillofac Implants
.
2012
;
27
:
945
956
.
23
Oliscovicz
NF,
Shimano
AC,
Marcantonio
Junior
E,
Lepri
CP,
Dos Reis
AC.
Analysis of primary stability of dental implants inserted in different substrates using the pullout test and insertion torque
.
Int J Dent
.
2013
;
2013
:
194987
.
24
Maiorana
C,
Farronato
D,
Pieroni
S,
Cicciu
M,
Andreoni
D,
Santoro
F.
A four-year survival rate multicenter prospective clinical study on 377 implants: correlations between implant insertion torque, diameter, and bone quality
.
J Oral Implantol
.
2015
;
41
:
e60
e65
.
25
Dayrell
AC,
Noritomi
PY,
Takahashi
JM,
Consani
RL,
Mesquita
MF,
dos Santos
MB.
Biomechanical analysis of implant-supported prostheses with different implant-abutment connections
.
Int J Prosthodont
.
2015
;
28
:
621
623
.
26
Urdaneta
RA,
Daher
S,
Leary
J,
Emanuel
KM,
Chuang
SK.
The survival of ultrashort locking-taper implants
.
Int J Oral Maxillofac Implants
.
2012
;
27
:
644
654
.
27
da Cunha
HA,
Francischone
CE,
Filho
HN,
de Oliveira
RC.
A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading
.
Int J Oral Maxillofac Implants
.
2004
;
19
:
578
585
.
28
Valente
ML,
de Castro
DT,
Shimano
AC,
Lepri
CP,
dos Reis
AC.
Analysis of the influence of implant shape on primary stability using the correlation of multiple methods
.
Clin Oral Invest
.
2015
;
19
:
1861
1866
.
29
Herekar
M,
Sethi
M,
Ahmad
T,
Fernandes
AS,
Patil
V,
Kulkarni
H.
A correlation between bone (B), insertion torque (IT), and implant stability (S): BITS score
.
J Prosthet Dent
.
2014
;
112
:
805
810
.