The aim of this study was to assess the effect of a diamond-like carbon (DLC) coating on the removal torque of prefabricated implant screws after cyclic loading. Four groups with two crowns supported by two implants (n = 5) were obtained according to splinted and nonsplinted prosthesis, using titanium or DLC screws (splinted crowns with titanium screw [STi], splinted crowns with DLC screw [SC], nonsplinted crowns and titanium screw [NSTi], and nonsplinted crowns and DLC screw [NSC]). The prosthetic screws were tightened at 32 Ncm and retightened, and the specimens were submitted to 106 mechanical cycles (4 Hz/98 N). After cyclic loading, loosening torque was evaluated, and the final measurements were performed. Data were analyzed by two-way analysis of variance and Tukey's test (α = .005). There was statistically significance in the interaction of screw × splinting (P = .003). For the group that used titanium screws, NSTi showed smaller removal torque compared with STi. It was concluded that the use of the DLC coating screws in nonsplinted prosthesis maintain the torque after cyclic loading.

Complications with single-tooth implant restorations frequently involve the integrity of the implant-abutment screw joint, with screw loosening as a common implication.14  Loss or decrease on the retention torque of screws, with consequent abutment screw loosening, may be caused by different factors and/or their interaction.57  Occlusal forces seem to play an important role in screw loosening of implants with hexagonal connections, in which preloading is the only force that resists these forces, avoiding the abutment to separate from the implant. The axial load generated in the screw on tightening, called preload, induces the clamping force that acts at the interface between the abutment and the implant-bearing surfaces, which holds the joint closed and also counteracts any load applied on the joint. Thus, if the preload is exceeded by the occlusal force, the screw will loosen.8 

The tension generated in the screw as a result of the applied torque depends on friction and some characteristics of screw material, such as yield strength, elasticity modulus, and fatigue life.9,10  Clearly, the coefficient of friction is a major factor influencing the preload achieved at a given torque force.11  The coefficient of friction is influenced by the hardness of the threads, surface finishing, speed of torquing, surface treatments, presence and quality of lubricants, and fit and machining tolerances.1214  Therefore, implant screw manufacturers have used dry lubricants such as gold coating, diamond-like carbon (DLC), and nitride coating to reduce friction, increase fastening, and allow more turning of the screw for a given torque.1517  Among the many favorable properties of DLC films, the most valuable for dentistry is the compatibility to titanium and 316 stainless steel.17  However, the effects of the use of DLC film on screw removal torque (RT) have not been clearly reported in the literature.

Splinting implants with prosthetic crowns may provide better stress and strain distribution to the bone to avoid overloading,18,19  especially for those placed in poor-quality bone, such as the posterior region of the jaws.20  Nevertheless, compared with nonsplinted single crowns, there are some disadvantages in the splinting approach, including difficulties to fit the framework and to provide adequate emergence profile, besides interproximal hygiene issues.21 

Thus, the objective of this study was to evaluate the effect of DLC coating on the RT of the prefabricated implant screw after mechanical cycling (MC) in splinted and nonsplinted prostheses. The hypothesis tested was that the DLC coating and splinting of the prostheses are factors that influence the screw loosening.

Specimen manufacturing

Forty implants analogues of the Branemark system (Neodent, Curitiba, Brazil), with platform 4.1 (external hexagon interface) were used. Forty metallic crowns were made, including 20 premolars and 20 molars. The specimens were set in 10 splinted and 10 nonsplinted, each sample having a premolar and one molar, and divided into 4 groups (n = 5): splinted or nonsplinted crowns using titanium screw or titanium screw with DLC coating surface (Neotorque, Neodent, Curitiba, PR, Brazil). The specimens consisted of two metallic structures (Ni-Cr) simulating fixed partial dentures over two adjacent implants corresponding to teeth #29 and #30 as follows: STi, splinted crowns with titanium screw; SC, splinted crowns with DLC screw; NSTi, nonsplinted crowns and titanium screw; NSC, nonsplinted crowns and DLC screw. The wax patterns for the crowns were produced with the CAD/CAM system (Milling Unit M5, Zirkonzahn, Zirkonzahn, Germany) from the milling of wax discs (Grew Wax 95H10, Zirkonzahn). Optical impressions (Scanner S600 ARTI, Zirkonzahn) to analogs were made. The crowns were designed (Zirkonzahn Software) with two types of interproximal contact: splinted or broad contact surface (3 mm2) (Figures 1 and 2).

Figures 1 and 2.

Figure 1. Virtual planning crowns in CAD/CAM, simulating region of the posterior mandible. Figure 2. (a) Splinted crowns, (b) nonsplinted crowns, (c) titanium screws, and (d) neotorque screw (diamond-like carbon cover).

Figures 1 and 2.

Figure 1. Virtual planning crowns in CAD/CAM, simulating region of the posterior mandible. Figure 2. (a) Splinted crowns, (b) nonsplinted crowns, (c) titanium screws, and (d) neotorque screw (diamond-like carbon cover).

Close modal

The wax patterns were then invested in a rapid cycle, carbon-free, phosphate-bonded investment (Micro Fine 1700, Talmax, Curitiba, PR, Brasil) and cast using a nickel-chromium alloy (Fit CAST-SB Plus, Talmax). Castings were divested with no further polishing or finishing (Figure 3).

Figure 3

Casting process. (a) Crowns at the end of the closure process, ready to be included. (b) Crowns attached to the sprues. (c) Crowns with inclusion of the coating material. (d) Crowns after casting and polishing.

Figure 3

Casting process. (a) Crowns at the end of the closure process, ready to be included. (b) Crowns attached to the sprues. (c) Crowns with inclusion of the coating material. (d) Crowns after casting and polishing.

Close modal

Implant analogues and metal crowns were placed in polyurethane molds (10 mm height and 18 mm diameter) and embedded in autopolymerizing acrylic resin (Jet, Classic Dental Products, São Paulo, SP, Brazil), 1 mm distant from the platform, to standardize the positioning with no vertical axis inclination. First, the crowns were slightly screwed to the implants by hand.22  The metal crowns were attached to each implant analog according to the screws of each group. Then, the set was placed in a socket at the device base to avoid any rotational movement. A torque of 32 Ncm was applied using a precision digital torque wrench (TQ-680, Instrutherm, São Paulo, SP, Brazil), according to the manufacturer's recommendation. After 3 minutes, the screw was retightened to the same torque to minimize embedment relaxation23  (Figures 4 and 5).

Figure 4

(a) Crowns in position with its analogs, ready for inclusion. (b) Samples fixed on a solid base and ready to receive the resin. (c) Device of each sample after adding the resin. (d) Contact point of the assessment of nonsplinted samples.

Figure 4

(a) Crowns in position with its analogs, ready for inclusion. (b) Samples fixed on a solid base and ready to receive the resin. (c) Device of each sample after adding the resin. (d) Contact point of the assessment of nonsplinted samples.

Close modal
Figures 5 and 6.

Figure 5. Evaluation of device removal torque. (a) Adjustable device for the digital switch setting. (b) Adjustable platform to hold the specimens. (c) Torquemeter connected to a data link receiver and software. (d) Capturing in real time any movement of screw loosening. Figure 6 . Samples under cyclic load with two loading tips focusing in the center of each crown.

Figures 5 and 6.

Figure 5. Evaluation of device removal torque. (a) Adjustable device for the digital switch setting. (b) Adjustable platform to hold the specimens. (c) Torquemeter connected to a data link receiver and software. (d) Capturing in real time any movement of screw loosening. Figure 6 . Samples under cyclic load with two loading tips focusing in the center of each crown.

Close modal

Cyclic loading

All specimens were submitted to cyclic fatigue machine (Biocycle, Biopdi, São Carlos, SP, Brazil) and submitted to 1 × 106 cycles at 4 Hz and 2 bar (50-N force in each tooth) in distilled water at 37°C. The eccentric load was applied on the occlusal surface of the crowns. Screw joint failure was defined as abutment mobility resulting from screw loosening (Figure 6).

Removal torque

After cyclic loading, loosening values of each specimen were measured with a digital torque meter (TQ-680, Instrutherm) connected to a data acquisition system (Lynx Technology, San Diego, Calif). To remove the screw, torque was applied in a counterclockwise direction, using a hexagonal wrench attached to the digital torque meter. The digital torque was recorded immediately after releasing the screw. All procedures were conducted by the same operator that was previously calibrated.

Statistical analysis

Two-way analysis of variance (ANOVA) and the Tukey post hoc test (α = .05) were made considering the screw type and splinting for each group of teeth (premolars and molars).

The mean and standard deviation of loosening values are shown in the Table. The analyses were done for each type of tooth (molars and premolars separately), and there was no comparison between tooth types. For the group of molar teeth, there was no statistical difference for any of the evaluated factors.

Table

Mean and standard deviation of detorque for each individual teeth group*

Mean and standard deviation of detorque for each individual teeth group*
Mean and standard deviation of detorque for each individual teeth group*

For the premolar teeth, statistical analysis showed that for the interaction of splinting × screw type, there were statistically significant differences (P = .003). For the splinted group, the titanium screw showed a higher value (P = .039). Otherwise, for the nonsplinted group, it was the DLC screw that had higher values (P = .014). When the DLC screw was used, there was no difference between the splinted and nonsplinted groups (P = .381). However, with the use of a titanium screw, the splinted group had higher values (P < .001).

The hypothesis was accepted. The factors studied (screw type and splinting) were found to be important for screw loosening. The cyclic loading was used to evaluate the torque maintenance on the retention of abutment screws. The mechanical cycling used represents 5 years of intraoral clinical service for the implant-supported restoration.24 

This study showed that all screws had reduced torque values after mechanical cycling. This can be due to low loosening values (the Table) compared with the initial torque of 32 Ncm. Each manufacturer establishes the correct value of torque that the screw needs to receive to be tightened and to achieve the required preload value to be resistant to loosening.25  The preload force is a compression force generated between the screw and the thread of the implant at the time the screw is tightened for the first time.7,12  The loosening values represented the amount of tightening torque retained in the screw, which is an valuable indirect estimation for preload maintenance.25  Thus, it shows that mechanical cycling induced a loss in the initial torque for all groups.

The results of this study show that the titanium screw splinted prostheses demonstrated satisfactory performance, with higher values of torque at the time of loosening after mechanical cycling compared with the DLC screw. The coated screws may present lower reductions in preload after cyclic loading, depending on the clinical situation. When using the DLC screw, there was no significant difference between splinted and nonsplinted prostheses. Possibly, abutment screw coating provided greater preload and thus more stable joints, thereby resulting in less screw vibration and micromovements during cyclic loading.26 

The DLC coating is an engineering process that reduces the coefficient of friction and makes a smooth surface.15  This effect may be responsible for the decreased torque to remove the DLC screw relative to the titanium screw splinted prostheses. This result shows the importance of choosing the screw according to the type of prosthesis. For splinted prostheses, titanium screws showed significantly greater values compared with loosening DLC screws. However, for the nonsplinted prostheses, the DLC screw presented greater loosening values. This result corroborates with another study, which showed a similar result for single crowns after cyclic loading.27 

Implant failure depends on two distinct types of factors: biological and mechanical. Biological causes are essentially peri-implantitis, affecting the soft and hard tissues surrounding dental implants, whereas mechanical causes involve implant-prosthetic components in general. Mechanical complications are as follows: implant fracture, abutment fracture, screw loosening, loss, and overstructure (ceramic and/or metal) fracture.28,29 

The separation of the crowns may be beneficial to avoid peri-implant diseases, improving the interproximal contour and avoiding concentration of stress during jaw flexure, as well as increasing patient satisfaction.30  However, splinted crowns present biomechanical advantages compared with nonsplinted crowns, and this is because stress distribution is better and more homogeneous.31  Regarding the dissipation of stresses on the implant/prosthesis system, some studies have used the finite element method (FEM), which reports the importance of the direction of application of the forces in the vertical/axial direction for the long-term success of prosthetic rehabilitation.3234  These factors are important and should guide the dental surgeon in choosing the type of prosthesis.32  Factors such as the occlusal load must be considered during the treatment planning process to ensure that prosthetic design and the number and location of dental implants are not subject to overloading.35  In addition to the concern with implant planning, occlusal factors also influence prosthetic planning. Therefore, in areas of great masticatory effort, the option to splint the crowns is well indicated for a better distribution of stresses on the crown/implant system. Therefore, in this type of clinical situation, the titanium screw is indicated. In an analysis of stress distribution by FEM, Hasan et al36  showed that the splinting of implant causes a noticeable drop in the stress in the implant system and surrounding bone.

Considering that the value of the removal torque is a measure of the remaining preload in the abutment screw, the reduction in removal torque after cyclic loading in the present study conforms to the joint failure mechanism described by Bickford.14  The external forces progressively erode the preload resulting from screw vibration, wear of the mating surfaces, and settling (embedment relaxation).9  Occlusal forces are complex in vector and magnitude; however, clinically, the occlusal forces are distributed over the prosthesis and remaining teeth, and cyclical forces vary in intensity.37  Hence, it is possible to find different results in the literature.

The clinical relevance of this is that, although screw loosening was detected, the significant decrease in preload indicates that patient follow-up is needed to ensure the integrity of abutment on single-implant restorations. This shows that the screwed prosthesis for single external hexagon implants is the best option because of the reversibility that ensures the case if there is screw loosening. Cemented prostheses for these cases are not indicated as well, because in the case of screw loosening, the crowns are lost.26  A study that used FEM reported that the cemented prosthesis presented significant physical and mechanical advantages because it did not present an occlusal orifice; this type of prosthesis presented greater resistance compared with the screwed one.38  Nevertheless, in the current study, we used screwed prostheses that had holes, but we believe that it was not a factor that could alter the result, because the application of load was always performed in the same place in all samples, and no load was applied to the areas of the holes. Currently, Morse taper implants guarantee a good indication for unitary prosthesis due to its esthetic and biomechanical characteristics related mainly to the advent of the use of the switching platform, which means that there is a decreased microgap abutment/implant joint, reducing the biofilm accumulation and consequently the bone loss and soft tissue around the implant.39 

Caution must be exerted in extrapolating laboratory data to clinical situations. Loading conditions in a clinical situation are not the same as in an in vitro investigation. Loading can vary in different patients. Each in vitro test may represent only one approach to a clinical situation, because many variables are usually excluded. Further laboratory and clinical studies should be conducted to confirm the effectiveness of coatings in preventing screw loosening and ensuring screw joint stability of abutments. Additionally, in vitro and in vivo studies should be done with Morse implants for comparison of real safety for the loss of torque and to promote a better planning scenario.

In conclusion, with the limitations of an in vitro study, the choice of the screw should be made according to the type of prosthesis. The use of the DLC coating screw in nonsplinted prostheses is important to maintain the torque after cyclic loading.

Abbreviations

Abbreviations
DLC

diamond-like carbon

MC

mechanical cycling

NSC

nonsplinted crowns and DLC screw

NSTi

nonsplinted crowns and titanium screw

RT

screw removal torque

SC

splinted crowns with DLC screw

STi

splinted crowns with titanium screw

The authors thank The Mechanical Projects Laboratory Prof. Henner A. Gomide (LPM)/UFU, Dental Research Center Biomechanics, Biomaterials, Cell Biology (CPBio)/UFU and Neodent for research incentive. The authors thank CNPq and FAPEMIG for financial support.

The authors declare no conflicts of interest.

1
Goodacre
CJ,
Kan
JY,
Rungcharassaeng
K.
Clinical complications of osseointegrated implants
.
J Prosthet Dent
.
1999
;
8
:
537
552
.
2
Jung
RE,
Pjetursson
BE,
Glauser
R,
Zembic
A,
Zwahlen
M.
A systematic review of the 5-year survival and complication rates of implant supported single crowns
.
Clin Oral Impl Res
.
2008
;
19
:
119
130
.
3
Kreissl
ME,
Gerds
T,
Muche
R,
Strub
JR.
Technical complications of implant supported fixed partial dentures in partially edentulous cases after an average observation period of 5 years
.
Clin Oral Impl Res
.
2007
;
18
:
720
726
.
4
Guda
T,
Ross
TA,
Lang
LA,
Millwater
HR.
Probabilistic analysis of preload in the abutment screw of a dental implant complex
.
J Prosthet Dent
.
2008
;
100
:
183
193
.
5
Theoharidou
A,
Petridis
HP,
Tzannas
K,
Garefis
P.
Abutment screw loosening in single-implant restorations: a systematic review
.
Int J Oral Maxillofac Implants
.
2008
;
23
:
681
690
.
6
Sakaguchi
RL,
Borgersen
SE.
Nonlinear contact analysis of preload in dental implant screws
.
Int J Oral Maxillofac Implants
.
1995
;
10
:
295
302
.
7
Binon
P,
Sutter
F,
Beaty
K,
Brumski
J,
Gulbransen
H,
Wei Ner
R.
The role of screws in implant systems
.
Int J Oral Maxillofac Implants
.
1994
;
9
:
48
63
.
8
Schwarz
MS.
Mechanical complications of dental implants
.
Clin Oral Implants Res
.
2000
;
11
:
156
158
.
9
Bickford
JH.
Self-loosening
.
In
:
Bickford
JH,
ed
.
An Introduction to the Design and Behavior of Bolted Joints. 3rd ed
.
New York, NY
:
Marcel Dekker;
1995
:
528
542
.
10
Yousef
H,
Luke
A,
Ricci
J,
Weiner
S.
Analysis of changes in implant screws subject to occlusal loading: a preliminary analysis
.
Implant Dent
.
2005
;
14
:
378
382
.
11
Bozkaya
D,
Müftü
S.
Mechanics of the taper integrated screwed-in (TIS) abutments used in dental implants
.
J Biomech
.
2005
;
38
:
87
97
.
12
Weiss
EI,
Kozak
D,
Gross
MD.
Effect of repeated closures on opening torque values in seven abutment-implant systems
.
J Prosthet Dent
.
2000
;
84
:
194
199
.
13
Lang
LA,
Kang
B,
Wang
RF,
Lang
BR.
Finite element analysis to determine implant pre-load
.
J Prosthet Dent
.
2003
;
90
:
539
46
.
14
Tzenakis
GK,
Nagy
WW,
Fournelle
RA,
Dhuru
VB.
The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws
.
J Prosthet Dent
.
2002
;
88
:
183
191
.
15
Martin
WC,
Woody
RD,
Miller
BH,
Miller
AW.
Implant abutment screw rotations and preloads for four different screw materials and surfaces
.
J Prosthet Dent
.
2001
;
86
:
24
32
.
16
Byrne
D,
Jacobs
S,
O'Connel
B,
Houston
F,
Claffey
N.
Preloads generated with repeated tightening in three types of screws used in dental implant assemblies
.
J Prosthodont
.
2006
;
15
:
164
171
.
17
Kim
SK,
Lee
JB,
Koak
JY,
et al.
An abutment screw loosening study of a diamond like carbon-coated CP titanium implant
.
J Oral Rehabil
.
2005
;
32
:
346
350
.
18
Yilmaz
B,
Seidt
JD,
McGlumphy
EA,
Clelland
NL.
Comparison of strains for splinted and non-splinted screw-retained prostheses on short implants
.
Int J Oral Maxillofac Implants
.
2011
;
26
:
1176
1182
.
19
Guichet
DL,
Yoshinobu
D,
Caputo
AA.
Effect of splinting and interproximal contact tightness on load transfer by implant restorations
.
J Prosthet Dent
.
2002
;
87
:
528
535
.
20
Wang
TM,
Leu
LJ,
Wang
J,
Lin
LD.
Effects of prosthesis materials and prosthesis splinting on peri-implant bone stress around implants in poor-quality bone: a numeric analysis
.
Int J Oral Maxillofac Implants
.
2002
;
17
:
231
237
.
21
Faucher
RR,
Bryant
RA.
Bilateral fixed splints
.
Int J Periodontics Restorative Dent
.
1983
;
3
:
8
37
.
22
Cardoso
M,
Torres
MF,
Lourenço
EJ,
de Moraes Telles D, Rodrigues RC, Ribeiro RF. Torque removal evaluation of prosthetic screws after tightening and loosening cycles: an in vitro study
.
Clin Oral Implants Res
.
2012
;
23
:
475
480
.
23
Jorge
JR,
Barão
VA,
Delben
JA,
Assunção
WG.
The role of implant abutment system on torque maintenance of retention screws and vertical misfit implant-supported crowns before and after mechanical cycling
.
Int J Oral Maxillofac Implants
.
2013
;
28
:
415
422
.
24
Lee
J,
Kim
YS,
Kim
CW,
Han
JS.
Wave analysis of implant screw loosening using an air cylindrical cyclic loading device
.
J Prosthet Dent
.
2002
;
88
:
402
408
.
25
Assunção
WG,
Delben
JA,
Tabata
LF,
Barão
VA,
Gomes
EA,
Garcia
IR
Jr.
Preload evaluation of different screws in external hexagon joint
.
Implant Dent
.
2012
;
21
:
46
50
.
26
Prado
CJ,
Neves
FD,
Soares
CJ,
Dantas
KA,
Dantas
TS,
Naves
LZ.
Influence of abutment screw design and surface coating on the bending flexural strength of the implant set
.
J Oral Implantol
.
2014
;
40
:
123
128
.
27
Bacchi
A,
Regalin
A,
Bhering
CL,
Alessandretti
R,
Spazzin
AO.
Loosening torque of Universal Abutment screws after cyclic loading: influence of tightening technique and screw coating
.
J Adv Prosthodont
.
2015
;
7
:
375
379
.
28
Albrektsson
TA.
Multicenter report on osseointegrated oral implants
.
J Prosthet Dent
.
1988
;
60
:
75
84
.
29
Oh
TJ,
Yoon
J,
Misch
CE,
Wang
HL.
The causes of early implant bone loss: myth or science
.
J Periodontol
.
2002
;
73
:
322
333
.
30
van Steenberghe
D,
Lekholm
U,
Bolender
C,
et al.
Applicability of osseointegrated oral implants in the rehabilitation of partial edentulism: a prospective multicenter study on 558 fixtures
.
Int J Oral Maxillofac Implants
.
1990
;
5
:
272
281
.
31
Naves
MD,
Menezes
HH,
de Magalhaes
D,
de Araújo
CA,
Simamoto Júnior P. The influence of interproximal contact on implant-supported fixed partial dentures in the posterior jaw: a photoelastic analysis
.
J Oral Implantol
.
2015
;
41
:
3
9
.
32
Cicciù
M,
Cervino
G,
Bramanti
E,
et al.
FEM analysis of mandibular prosthetic overdenture supported by dental implants: evaluation of different retention methods
.
Comput Meth Methods Med
.
2015
;
1
16
.
33
Lauritano
F,
Runci
M,
Cervino
G,
Fiorillo
L,
Bramanti
E,
Cicciù
M.
Three-dimensional evaluation of different prosthesis retention systems using finite element analysis and the Von Mises stress test
.
Minerva Stomatol
.
2016
;
65
:
353
367
.
34
Cicciù
M,
Risitano
G,
Maiorana
C,
Franceschini
G.
Parametric analysis of the strength in the “Toronto” osseous-prosthesis system
.
Minerva Stomatol
.
2009
;
58
:
9
23
.
35
Rangert
B,
Krogh
PH,
Langer
B,
Van Roekel
N.
Bending overload and implant fracture: a retrospective clinical analysis
.
Int J Oral Maxillofac Implants
.
1995
;
10
:
326
334
.
36
Hasan
I,
Bourauel
C,
Keilig
L,
Stark
H,
Lückerath
W.
The effect implant splinting on the load distribution in bone bed around implant-supported fixed prosthesis with different framework materials: a finite element study
.
Ann Anat
.
2015
;
199
:
43
51
.
37
Khraisat
A,
Hashimoto
A,
Nomura
S,
Miyakawa
O.
Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system
.
J Prosthet Dent
.
2004
;
91
:
326
334
.
38
Cicciù
M,
Bramanti
E,
Matacena
G,
Guglielmino
E,
Risitano
G.
FEM evaluation of cemented-retained versus screw-retained dental implants single-tooth crown prosthesis
.
Int J Clin Exp Med
.
2014
;
7
:
817
825
.
39
Macedo
JP,
Pereira
J,
Vahey
BR,
et al.
Morse taper dental implants and platform switching: The new paradigm in oral implantology
.
Eur J Dent
.
2016
;
10
:
148
154
.