Implant site preparation through drilling procedures may cause bone thermonecrosis. The aim of this in vitro study was to evaluate, using a thermal probe, overheating at implant sites during osteotomies through 2 different drilling methods (continuous drilling technique versus intermittent drilling technique) using irrigation at different temperatures. Five implant sites 13 mm in length were performed on 16 blocks (fresh bovine ribs), for a total of 80 implant sites. The PT-100 thermal probe was positioned 5 mm from each site. Two physiological refrigerant solutions were used: one at 23.7°C and one at 6.0°C. Four experimental groups were considered: group A (continuous drilling with physiological solution at 23.7°C), group B (intermittent drilling with physiological solution at 23.7°C), group C (continuous drilling with physiological solution at 6.0°C), and group D (intermittent drilling with physiological solution at 6.0°C). The Wilcoxon rank-sum test (2-tailed) was used to compare groups. While there was no difference between group A and group B (W = 86; P = .45), statistically significant differences were observed between experimental groups A and C (W = 0; P =.0001), B and D (W = 45; P =.0005), and C and D (W = 41; P = .003). Implant site preparation did not affect the overheating of the bone. Statistically significant differences were found with the refrigerant solutions. Using both irrigating solutions, bone temperature did not exceed 47°C.

Implant site preparation through drilling procedures may not only cause mechanical damage to the bone involved but also a temperature increase in the bone adjacent to the implant site, which can result in thermonecrosis.1  Thermally induced necrosis has been reported as one of the main causes of dental implant failure during osseointegration.25  According to 3 studies,68  a temperature of 47°C for at least 1 minute was identified as a limit threshold to the bone survival. Although intimate contact between the drill and bony wall is mandatory, it is usually considered the main reason for heat generation. Besides, the presence or absence of irrigation during the drilling5,9 ; drill geometry and design10,11 ; drill wear12,13 ; sharpness of the cutting tool14,15 ; bone density16 ; drilling speed, axial force, and pressure applied to the drill1721 ; and internal and external irrigation22  have been investigated by many authors as important factors affecting the development of frictional heat. Nevertheless, few studies have investigated the involvement of the technique in preparing the implant site.2327 

Usually, the preparation of an implant site consists of the use of a series of drills, each of increasing diameter2,6,8 ; the aim is to contain the friction generated during the procedure.28  Nevertheless, Lucchiari et al29  demonstrated that the differences in temperature variations generated by the standard technique versus the single-drill technique are clinically irrelevant.

During the preparation of multiple implant sites, 2 different drilling techniques can be adopted: continuous or intermittent. Many authors observed that the heat generated during implant site preparation is directly proportional to the time of exposure to the friction forces produced during drilling9 ; moreover, higher temperatures are associated with longer preparation times.21,30,31  Conversely, few studies have compared the 2 different methods.

The aim of this in vitro study was to analyze temperature variations generated by continuous and intermittent drilling techniques to prepare multiple implant sites using 2 different refrigerant solutions at different temperatures. The null hypothesis was that no significant difference in temperature variation would be generated by continuous and intermittent drilling techniques with irrigation at different temperatures.

The in vitro experimental study was designed to reproduce the clinical condition of an edentulous mandibular bone by using bone samples dissected from fresh bovine ribs. Bovine ribs represent a well-established bone model for human jaw because of similarities in bone density, thermal conductivity and the ratio between cortical and cancellous bone.1,11,14,18,28  Twenty ribs were dissected through a cutter and a multi-rosette (Meisinger HM 141F023 diameter .3 mm; Hager & Meisinger GmbH, Neuss, Germany) mounted on a handpiece (W&H Dentalwerk Bürmoos GmbH, Bürmoos, Austria) in 16 blocks (40 × 40 × 20 mm). The blocks of bone were prepared as edentulous, and 5 implant sites were prepared on each block for a total of 80 sites. One site/block was set up for the thermal probe (Figure 1). The measurements were carried out using a PT-100 thermal probe (Pico Technology, Tyler, Tex) (temperature range –40°C to +80°C; probe error ± 0; 45°C), a resistance thermometer equipped with a platinum sensor. The center of the hole containing the thermal probe was positioned at 5 mm from each site (Figure 2). The probe was introduced into the bone through the set-up hole and thermally coupled to the inner bone wall through a layer of thermally neutral “thermal grease” (AOS 52029, AOS Thermal Compound, Eatontown, NJ). The thermal insulation was ensured by a coating of 2-component epoxy resin able to thermally isolate a thin layer in the range of –200 to +200°C. The temperature was measured as electrical resistance (ohms [Ω]) by the probe and recorded at 1-second intervals by a digital multimeter (Mod. FLUKE 89 Series IV, Fluke Europe BV, Eindhoven, Netherlands) converted into Celsius degrees according to the equation for linearizing the same software.

Figure 1

One site/block was set up for the thermal probe.

Figure 1

One site/block was set up for the thermal probe.

Close modal
Figure 2

Scheme of the thermal probe positioned 5 mm from each site.

Figure 2

Scheme of the thermal probe positioned 5 mm from each site.

Close modal

After excision from fresh bovine ribs, the blocks were frozen and stored at –20°C up to use. Then they were thawed by immersion in a temperature-controlled saline bath (26.0°C ± 1°C); site preparation began when the internal temperature of the bone was in equilibrium with the bath temperature. All procedures, related to the preparation of implant sites were performed by a single expert operator. The sites were 13 mm in length and 2 mm in diameter; a cutter pilot drill (Biomax 3i, Biomet 3i Implant Innovations, Palm Beach Gardens, Fla) was used. The handpiece was connected to an Implantmed motor (W&H Dentalwerk Bürmoos GmbH) with a speed of 1200 rpm and a torque of 35 N/cm2. The pressure exerted during the implant site preparation was constantly monitored by an electronic precision balance (Excellence Plus XP, Mettler-Toledo, Giesen, Germany) controlled the operator at a range between 20N and 30 N; implant sites prepared with pressures higher than 30N were excluded. Thermal measurements were performed in a climate-controlled room (temperature 23°C to 24°C; relative humidity 5%; absence of direct ventilation); temperature values were measured every 5 seconds. Two different drilling techniques were adopted: method A consisted of continuous drilling, and method B was a drill-stop technique between implant sites. During the preparation of the implant sites, irrigation was done with 2 saline solutions (refAmb or refFri; sodium chloride, Omnia spa, Fidenza, Italy), which had a flow rate of 50 mL/min throughout the drilling. RefAmb was physiological saline maintained in thermal equilibrium with the external environment (23.7°C); refFri was physiological saline refrigerated at 6°C.

Four experimental groups were assessed. Group A consisted of implant sites prepared with the continuous drilling technique and saline solution refAmb; group B consisted of implant sites prepared with the intermittent drilling technique and the refAmb saline solution; Group C consisted of implant sites prepared with the continuous drilling technique and the refFri saline solution; Group D consisted of implant sites prepared with intermittent drilling technique and the refFri saline solution.

The statistical unit considered was temperature (°C). The primary variable was the percentage of temperature variation in the 4 groups. A pilot study was conducted to generate data on the expected effect size and standard deviation to allow for power calculations. Twenty implant sites were used for the calculations, 5 per group. The level of statistical significance was set as α = 0.05 (statistical power: 80%). In group A, the average temperature recorded was 28.07°C (SD 0.32); in group B, the average temperature recorded was 27.71°C (SD 0.39); in group C, the average temperature recorded was 27.37°C (SD 0.72); and in group D, the average temperature recorded was 27.42°C (SD 0.31). Null hypothesis for difference between averages of group A and group B was supposed to be 0.5°C. Fifteen implant sites per group were then estimated after power calculation. The sample size was set at 20 implant sites per group since a 35% bias (5 implant sites per group) was expected. A descriptive statistical analysis was used to calculate the averages and SDs of temperatures in the 4 groups. Wilcoxon matched-pairs signed-ranks test (1-tailed) was used to compare groups (A and B; A and C; B and C; C and D). The level of statistical significance was set as α = 0.05 and statistical power of 80%. Statistical analysis was performed using statistical software SPSS 16.0 (SPSS Inc, Chicago, Ill).

In group A, the average temperature was 29.68°C (range 27.78°C to 31.35°C; SD 1.21); in group B, the average temperature was 29.70°C (range 27.28°C to 31.81°C; SD 1.50), in group C, the average temperature was 27.13°C (range 26.59°C to 28.22°C; SD 0.55), and in group D the average temperature was 28.44°C (range 26.69°C to 31.87°C; SD 1.53). The results are summarized in the Table. There were no statistically significant differences between group A and group B (W = 86; P value = 0.45). Statistically significant differences were found between group A and group C (W = 0; P = .0001); group B and group D (W = 45; P = .0005), and group C and group D (W = 41; P = .003).

Table

Measurement of temperature with continuous and intermittent drilling techniques

Measurement of temperature with continuous and intermittent drilling techniques
Measurement of temperature with continuous and intermittent drilling techniques

Osseointegrated implants have become a viable option in patients affected by partial or total edentulism,32,33  Thus, if osseointegration does not take place, the result is biological failure and consequent implant loss.34  The primary intention of endosseous implant surface modifications is to modulate the implant and host tissue response and accomplish better osseointegration,35  To achieve a predictable osseointegration and to obtain a natural implant restoration, implant site development should be the least traumatic as possible. Tehemar6  tried to identify all the factors that influence the amount of heat generated when bone is drilled. In fact, as reported by Watzek et al,37  the successful healing of implants may be influenced by heat generations during rotary cutting. In particular, Iyer et al30  showed that drilling speed can alter osseointegration as it affects bony tissue if inadequate irrigation is used. Thus, questions remain not only about the optimal design of the drill, the degree of heat generated depending on the bone density and the speed of the drill but also on the best type of irrigation.38 

Increased temperature during drilling has long being identified as critical to avoid a reduction in osteoclast and osteoblast activity and to preserve the surrounding tissues39,40 ; moreover, it could be responsible for the deterioration of the organic component of the bone, which may cause local bone necrosis. The bone itself has a very low thermal conductivity that prevents heat dissipation during the drilling phase.41  Many parameters could influence the increase in temperature during drilling of the bone (ie, characteristics of the drill and the drilling technique as well as characteristics of the cortical bone itself). Hence, to overcome thermal changes, different drilling techniques have been developed and investigated.39  However, few studies have paid attention to the method for preparing the implant site.

The aim of the present in vitro study was to compare heat generation in the bone surrounding the apical region of the implant site during continuous drilling techniques versus intermittent drilling techniques; irrigation solutions at 2 different temperatures were used.

Both internal and external irrigation systems reduced the high heat generated from drilling of implant sites.42  Sener et al9  also demonstrated that external irrigation at room temperature is sufficient for cooling the implant site during drilling. According to the results in this in vitro study, no statistically significant difference between group A and group B (W = 86; P = .45) was found; however, statistically significant differences were found between group A and group C (W = 0; P = .0001); between group B and group D (W = 45; P = .0005), and between group C and group D (W = 41; P = .003). This experimental study showed that the average temperature was 29.68°C in group A, 29.70°C in group B, 27.13°C in group C, and 28.44°C in group D. Hence, the use of a refrigerant solution at a temperature of 6°C reduces the increase in bone temperature during the preparation of implant sites compared with the physiological solution at a temperature of 23.7°C. Moreover, no significant differences were found in the temperature variation generated by continuous and intermittent drilling techniques, except for groups C and D. The temperature increase between group C (continuous drilling techniques) and group D (intermittent drilling techniques) was of 1.31°C. However, a statistical difference of 1.31°C should not be considered clinically relevant, as it does not reduce the risk of thermonecrosis. In parallel, it was observed that the baseline temperature of the bone ribs (26.0 ± 1°C) differed from in vivo temperatures.

Use of a refrigerated solution can provide an advantage during preparation of the implant site; solutions at lower temperature are more effective in cooling the bone. However, concerning the 2 different methods (A and B), there were no statistically significant differences among the 2 strategies in influencing the increased bone temperature.

The method adopted for preparing the implant sites (continuous drilling technique versus intermittent drilling technique) does not affect the overheating of the bone. The use of irrigating solutions is recommended. Even though the temperature of the bone never reached the critical threshold of 47°C with either physiological saline, use of a refrigerated solution instead of an ambient-temperature solution is preferable. Cooler irrigating solutions can confer benefits in the preparation of the implant sites by eliminating several variables that can affect bone overheating (ie, shape of the drills, number of blades, usury).

Abbreviations

Abbreviations
refAmb

physiological solution at 23.7°C.

refFri

physiological solution at 6.0°C.

The authors have no commercial or financial dealings that may pose a conflict of interest or potential conflict of interest.

1
Abouzgia
MB.
James
DF.
Temperature rise during drilling through bone
.
Int J Oral Maxillofac Implants
.
1997
;
12
:
342
353
.
2
Eriksson
RA.
Adell
R.
Temperatures during drilling for the placement of implants using the osseointegration technique
.
J Oral Maxillofacial Surg
.
1986
;
44
:
4
7
.
3
Brisman
DL.
The effect of speed, pressure, and time on bone temperature during the drilling of implant sites
.
Int J Oral Maxillofac Implants
.
1996
;
11
:
35
37
.
4
Bahat
O.
Sullivan
RM.
Parameters for successful implant integration revisited part I: immediate loading considered in light of the original prerequisites for osseointegration
.
Clin Implant Dent Relat Res
.
2010
;
12
:
2
12
.
5
Watanabe
F.
Tawada
Y.
Komatsu
S.
Hata
Y.
Heat distribution in bone during preparation of implant sites: heat analysis by real time thermography
.
Int J Oral Maxillofacial Implants
.
1992
;
7
:
212
219
.
6
Eriksson
AR.
Albrektsson
T.
Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit
.
J Prosthet Dent
.
1983
;
50
:
101
107
.
7
Albrektsson
T.
Brånemark
PI.
Hansson
HA.
Lindström
J.
Osseointegrated titanium implants: requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man
.
Acta Orthop Scand
.
1981
;
52
:
155
170
.
8
Eriksson
AR.
Albrektsson
T.
The effect of heat on bone regeneration: an experimental study in the rabbit using the bone growth chamber
.
J Oral Maxillofac Surg
.
1984
;
42
:
705
711
.
9
Sener
BC.
Dergin
G.
Gursoy
B.
Kelesoglu
E.
Slih
I.
Effects of irrigation temperature on heat control in vitro at different drilling depths
.
Clin Oral Implants Res
.
2009
;
20
:
294
298
.
10
Scarano
A.
Piattelli
A.
Assenza
B.
et al.
Infrared thermographic evaluation of temperature modifications induced during implant site preparation with cylindrical versus conical drills
.
Clin Implant Dent Relat Res
.
2011
;
13
:
319
323
.
11
Cordioli
G.
Majzoub
Z.
Heat generation during implant site preparation: an in vitro study
.
Int J Oral Maxillofacial Implants
.
1997
;
12
:
186
193
.
12
Scarano
A.
Carinci
F.
Quaranta
A.
Di Lorio
D.
Assenza
B.
Piattelli
A.
Effects of bur wear during implant site preparation: an in vitro study
.
Int J Immunopathol Pharmacol
.
2007
;
20
:
23
26
.
13
Oliveira
N.
Alaejos-Algarra
F.
Mareque-Bueno
J.
Ferrés-Padró
E.
Hernández-Alfaro
F.
Thermal changes and drill wear in bovine bone during implant site preparation. A comparative in vitro study: twisted stainless steel and ceramic drills
.
Clin Oral Implants Res
.
2012
;
23
:
963
969
.
14
Chacon
GE.
Bower
DL.
Larsen
PE.
McGlumphy
EA.
Beck
FM.
Heat production by 3 implant drill systems after repeated drilling and sterilization
.
J Oral Maxillofac Surg
.
2006
;
64
:
265
269
.
15
Oh
HJ.
Wikesjö
UM.
Kang
HS.
Ku
Y.
Eom
TG.
Koo
KT.
Effect of implant drill characteristics on heat generation in osteotomy sites: a pilot study
.
Clin Oral Implants Res
.
2011
;
22
:
722
726
.
16
Yacker
MJ.
Klein
M.
The effect of irrigation on osteotomy depth and bur diameter
.
Int J Oral Maxillofac Implants
.
1996
;
11
:
634
638
.
17
Sharawy
M.
Misch
CE.
Weller
N.
Tehemar
S.
Heat generation during implant drilling: the significance of motor speed
.
J Oral Maxillofac Surg
.
2002
;
60
:
1160
1169
.
18
Brisman
DL.
The effect of speed, pressure, and time on bone temperature during the drilling of implant sites
.
Int J Oral Maxillofac Implants
.
1996
;
11
:
35
37
.
19
Augustin
G.
Davila
S.
Udiljak
T.
Vedrina
DS.
Bagatin
D.
Determination of spatial distribution of increase in bone temperature during drilling by infrared thermography: preliminary report
.
Arch Orthop Trauma Surg
.
2009
;
129
:
703
709
.
20
Augustin
G.
Davila
S.
Mihoci
K.
Udiliak
T.
Vedrina
DS.
Antabak
A.
Thermal osteonecrosis and bone drilling parameters revisited
.
Arch Orthop Trauma Surg
.
2008
;
128
:
71
77
.
21
Arbouzgia
MB.
Symington
JM.
Effect of drill speed on bone temperature
.
Int J Oral Maxillofac Surg
.
1996
;
25
:
394
399
.
22
Benington
IC.
Biagioni
PA.
Briggs
J.
Sheridan
S.
Lamey
PJ.
Thermal changes observed at implant sites during internal and external irrigation
.
Clin Oral Implants Res
.
2002
;
13
:
293
297
.
23
Reingewirtz
Y.
Szmukler-Moncler
S.
Senger
B.
Influence of different parameters on bone heating and drilling time in implantology
.
Clin Oral Implants Res
.
1997
;
8
:
189
197
.
24
Sharawy
M.
Misch
CE.
Weller
N.
Tehemar
S.
Heat generation during implant drilling: the significance of motor speed
.
J Oral Maxillofac Surg
.
2002
;
60
:
1160
1169
.
25
Khayat
PG.
Arnal
HM.
Tourbah
BI.
Sennerby
L.
Clinical outcome of dental implants placed with high insertion torques (up to 176 Ncm)
.
Clin Implant Dent Relat Res
.
2013
;
15
:
227
233
.
26
Rashad
A.
Kaiser
A.
Prochnow
N.
Schmitz
I.
Hoffmann
E.
Maurer
P.
Heat production during different ultrasonic and conventional osteotomy preparations for dental implants
.
Clin Oral Implants Res
.
2011
;
22
:
1361
1365
.
27
Wächter
R.
Stoll
P.
Increase of temperature during osteotomy. In vitro and in vivo investigations
.
Int J Oral Maxillofac Surg
.
1991
;
20
:
245
249
.
28
Davidson
SR.
James
DF.
Drilling in bone: modelling heat generation and temperature distribution
.
J Biomech Eng
.
2003
;
125
:
305
314
.
29
Lucchiari
N.
Frigo
AC.
Stellini
E.
Coppe
M.
Berengo
M.
Bacci
C.
In vitro assessment with the infrared thermometer of temperature differences generated during implant site preparation: the traditional technique versus the single-drill technique
.
Clin Implant Dent Relat Res
.
2016
;
18
:
182
191
.
30
Iyer
S.
Weiss
C.
Mehta
A.
Effects of drill speed on heat production and the rate and quality of bone formation in dental implant osteotomies. Part I: relationship between drill speed and heat production
.
Int J Prosthodont
.
1997
;
10
:
411
414
.
31
Iyer
S.
Weiss
C.
Mehta
A.
Effects of drill speed on heat production and the rate and quality of bone formation in dental implant osteotomies. Part II: relationship between drill speed and healing
.
Int J Prosthodont
.
1997
;
10
:
536
540
.
32
Sivolella
S.
Stellini
E.
Testori
T.
Di Fiore
A.
Berengo
M.
Lops
D.
Splinted and unsplinted short implants in mandibles: a retrospective evaluation with 5 to 16 years of follow-up
.
J Periodontol
.
2013
;
84
:
502
512
.
33
Bengazi
F.
Botticelli
D.
Favero
V.
Perini
A.
Urbizo Velez
J.
Lang
NP.
Influence of presence or absence of keratinized mucosa on the alveolar bony crest level as it relates to different buccal marginal bone thicknesses. An experimental study in dogs
.
Clin Oral Implants Res
.
2014
;
25
:
1065
71
.
34
Olmedo-Gaya
MV.
Manzano-Moreno
FJ.
Cañaveral-Cavero
E.
de Dios Luna-del Castillo J, Vallecillo-Capilla M. Risk factors associated with early implant failure: a 5-year retrospective clinical study
.
J Prosthet Dent
.
2016
;
115
:
150
155
.
35
Meng
HW.
Chien
EY.
Chien
HH.
Dental implant bioactive surface modifications and their effects on osseointegration: a review
.
Biomark Res
.
2016
;
14
:
24
26
.
36
Tehemar
SH.
Factors affecting heat generation during implant site preparation: a review of biologic observations and future considerations
.
Int J Oral Maxillofac Implants
.
1999
;
14
:
127
136
.
37
Watzek
G.
Matejka
M.
Lill
W.
Mailath
G.
Matzka
P.
Plenk
H
Jr
Knöchern eingeheilte Implantate (Tübingen, IMZ, Brånemark)—Erfahrungen mit einem Therapiekonzept
.
Z Stomatol
.
1988
;
85
:
207
233
.
38
Möhlhenrich
SC.
Modabber
A.
Steiner
T.
Mitchell
DA.
Hölzle
F.
Heat generation and drill wear during dental implant site preparation: systematic review
.
Br J Oral Maxillofac Surg
.
2015
;
53
:
679
689
.
39
Calvo-Guirado
JL.
Delgado-Peña
J.
Maté-Sánchez
JE.
Mareque Bueno J, Delgado-Ruiz RA, Romanos GE. Novel hybrid drilling protocol: evaluation for the implant healing—thermal changes, crestal bone loss, and bone-to-implant contact
.
Clin Oral Implants Res
.
2015
;
26
:
753
760
.
40
Bettach
R.
Taschieri
S.
Boukhris
G.
Del Fabbro
M.
Implant survival after preparation of the implant site using a single bur: a case series
.
Clin Implant Dent Relat Res
.
2015
;
17
:
13
21
.
41
Gehrke
SA.
Bettach
R.
Taschieri
S.
Boukhris
G.
Corbella
S.
Del Fabbro
M.
Temperature changes in cortical bone after implant site preparation using a single bur versus multiple drilling steps: an in vitro investigation
.
Clin Implant Dent Relat Res
.
2015
;
17
:
700
707
.
42
Benington
IC.
Biagioni
PA.
Crossey
PJ.
Hussey
DL.
Sheridan
S.
Lamey
PJ.
Temperature changes in bovine mandibular bone during implant site preparation: an assessment using infra-red thermography
.
J Dent
.
1996
;
24
:
263
267
.