Different types of internal implant–abutment connections, namely hexagon and conical, have been used for implant restoration. However, data regarding the benefits of these internal connections in terms of clinical outcomes are scarce. Accordingly, the aim of this study was to compare radiographic marginal bone loss (RMBL) and associated implant complications between implants with internal hexagon (IH) connections and those with internal conical (IC) connections. Forty-nine patients with 98 implants (2 per patient) placed in the posterior mandible were recruited. All implants were inserted in pairs into solid D2 bone according to a randomized sequence; the first patient received an IH connection implant on the mesial side, while the second patient received an IC connection implant on the mesial side. Each patient received 1 implant with an IH connection and 1 with an IC connection, placed side by side. Four months after placement, all implants were loaded with single screw-retained ceramic restorations with IH or IC connections. RMBL and complications, including implant/prosthesis failure, were recorded at the time of implant loading (baseline) and at 6, 12, and 36 months after loading. The results revealed no significant between-group differences in RMBL (P = .74), gingival bleeding on probing (P = .29), and complications (P = .32). Thus, the type of internal implant–abutment connection did not affect clinical outcomes, including RMBL and implant/prosthesis failure. Future studies should additionally evaluate long-term prosthesis-related complications, such as screw loosening and fracture, between the 2 types of internal connections.

Oral implantology is an evolving field with continuously developing designs, concepts, and materials. Every year, millions of dental implants are inserted and loaded with prosthetic superstructures, and, as observed in any therapeutic technique, biological and biomechanical complications may occur.1  Currently, most dental implants have different abutment connections to minimize the risk of abutment movement and loosening of the inner screws. With screw-retained restorations (SCRs), the abutment can be removed if necessary. Sherif et al2  showed that there was no significant difference between cement-retained restorations and SCRs in terms of clinical outcomes. This suggests that SCRs are as reliable as cement-retained restorations, although the latter are more popular in current clinical practice.

Another abutment modification is the so-called platform switching, which involves placing an abutment of smaller diameter into an implant of larger diameter. However, research has demonstrated that platform switching alone does not prevent crestal bone loss, regardless of mucosal thickness.3,4  Nonetheless, one study reported a small positive effect of platform switching in terms of minimizing marginal bone loss (approximately 0.3 mm) 1 year after loading.5  Similar results were reported in another study.6 

The implant–abutment microgap and its relationship with crestal bone loss have also been evaluated. The microgap traps bacteria and can function as a reservoir for them; this often triggers inflammatory responses and infection around implants.6  Interestingly, one study compared implants with screw-retained internal hexagon (IH) abutments and implants with a cone-morse internal connection and found no detectable separation or microgap between the implants and the abutments around the conical connection.7  In contrast, numerous gaps and voids were present between the implant and the screw-retained IH abutment.7 

In a study on the biomechanical behavior of multiple implant-supported prostheses, Borie et al8  showed minimum stress in the bone surrounding implants with internal conical (IC) connections. In another study that evaluated fracture resistance under oblique compressive forces for IH and IC connection implants, the maximum deformation force for IC connection implants was significantly higher than that for IH connection implants.9  The friction-locking mechanics and the solid design of the IC connection have been shown to provide greater resistance to deformation and to fracture under oblique compressive loading.

To our knowledge, data on the benefits of IH and IC connections in terms of clinical outcomes are scarce. Therefore, the aim of this study is to compare radiographic marginal bone loss (RMBL) and associated implant complications between implants with identical surface treatments but with different internal connections, namely, IH and IC connections.

Subjects

This clinical study was approved by the appropriate ethical committee, and written informed consent was obtained from all human participants. The study was conducted at the Department of Implant Supported Restorations from March 1, 2013, to May 31, 2017. Patients aged ≥21 years who could understand and sign a written informed consent form, who required a 2-implant–supported prostheses (excluding overdentures) in a free-end zone of the posterior mandible, and who exhibited adequate bone in the treatment area (width ≥ 6.75 mm; height ≥ 12 mm) were considered eligible. Additional inclusion criteria were as follows: presence of D2 bone, good oral health, no smoking history, and presence of a thick gingival biotype (ie, ≥2 mm of tissue thickness measured at the implant insertion site using cone-beam computed tomography [CBCT] and a periodontal probe). Bone density was assessed on the basis of computed tomography–derived Hounsfield units.10  Preoperative radiographs and CBCT scans, as well as clinical examinations, were used to determine soft tissue and bone volumes. No hard or soft tissue grafting procedures were permitted for the included patients.

The exclusion criteria were as follows: medical conditions requiring prolonged steroid treatment and/or medications that could interfere with bone metabolism, irradiation of the head and neck area, autoimmune diseases requiring medical treatment, intravenous bisphosphonate or oral bisphosphonate for >3 years, local inflammation and untreated periodontitis, acute or purulent infection in the area of implant placement, poor oral hygiene and motivation (O'Leary plaque index ≥ 30%),11  uncontrolled diabetes (HbA1c > 8%), pregnancy or nursing, psychiatric problems, absence of an antagonistic occlusal surface for the implant-supported prosthesis at the time of implant loading, any type of bone augmentation performed at the implant insertion site with a healing period of <6 months, bruxism, alcoholism or chronic drug abuse, and inability to commit to the follow-up schedule.

The exclusion criteria during surgery were as follows: lack of primary stability of at least 25 N-cm, need for bone augmentation procedures during implant insertion, and inability to place the implant in a correct position from a prosthetic point of view.

Implants with IH (SEVEN, MIS) or with IC (C1, MIS) connections were used (Figure 1). For the IC connection, the type of conical connection was a 6-degree, which ensures a secure fit between abutment and implant with minimal micromovements. A single operator performed all the surgeries to minimize bias. All implants were inserted in pairs into solid D2 bone according to a randomized sequence; the first patient received the IH connection implant on the mesial side, while the second patient received the IC connection implant on the mesial side. Each patient received 1 implant with an IH connection and 1 with an IC connection, placed side by side. The diameter and the length of the inserted implants were 3.75 or 4.20 mm and 10 or 11.50 mm, respectively. The diameter and the length of the 2 implants were the same for each patient.

Figures 1 and 2.

Figure 1. The 2 types of internal implant–abutment connections used in this study. (a) Internal hexagon connection. (b) Internal conical connection.

Figure 2. Illustration of implant insertion and uncovering procedures for implants with internal hexagon or conical connections. (a) Site after completion of the short-drilling protocol. (b) Implants with internal hexagon or conical connections are inserted to manage the free-end situation in the posterior mandible. (c) Clinical appearance of the crest during the 3-month osteointegration period. (d) Uncovered implant after the osteointegration period. (e) Gingival tissue in the tissue punch. (f) Healing screws.

Figures 1 and 2.

Figure 1. The 2 types of internal implant–abutment connections used in this study. (a) Internal hexagon connection. (b) Internal conical connection.

Figure 2. Illustration of implant insertion and uncovering procedures for implants with internal hexagon or conical connections. (a) Site after completion of the short-drilling protocol. (b) Implants with internal hexagon or conical connections are inserted to manage the free-end situation in the posterior mandible. (c) Clinical appearance of the crest during the 3-month osteointegration period. (d) Uncovered implant after the osteointegration period. (e) Gingival tissue in the tissue punch. (f) Healing screws.

Close modal

Measurement

Peri-implant marginal bone remodeling was analyzed on intraoral radiographs obtained using the parallel cone technique at the time of implant loading (baseline) and at 6, 12, and 36 months after loading (Figure 4). To standardize the intraoral periapical radiographs, a silicone holder was used to maintain parallelism and minimize errors, and the X-ray (ProX, Planmeca) was directed perpendicular to the implant. The holder included a minimum of 3–4 teeth adjacent to the edentulous space. The radiographic sensor surface was retained in the mount to ensure reproducibility.

Figures 3 and 4.

Figure 3. Illustration of implant restoration procedures. (a) Prefabricated titanium abutment connections. (b) All-ceramic screw-retained restorations. (c and d) Adhesive cementation of inlays for secure sealing of access holes. (e) Clinical aspects of the final fixed restorations.

Figure 4. Radiographs of implants with internal hexagon or conical connections at (a) the time of insertion and at (b) 6, (c) 12, and (d) 36 months after loading, with the final restorations.

Figures 3 and 4.

Figure 3. Illustration of implant restoration procedures. (a) Prefabricated titanium abutment connections. (b) All-ceramic screw-retained restorations. (c and d) Adhesive cementation of inlays for secure sealing of access holes. (e) Clinical aspects of the final fixed restorations.

Figure 4. Radiographs of implants with internal hexagon or conical connections at (a) the time of insertion and at (b) 6, (c) 12, and (d) 36 months after loading, with the final restorations.

Close modal

The peri-implant marginal bone level was evaluated by measuring the difference between implant–bone contact and implant–abutment junction using ImageJ 1.50e software (National Institutes of Health). All radiographs were registered on a 1:1 scale (Figure 5) according to the length (in mm) of the implant from the neck to the apex. The most common and reliable method used to assess marginal bone loss at the patient level involves determination of mesial/distal bone loss followed by calculation of the average of these measurements. Data were obtained by measuring the parallel axis of the implant, and results were determined by using 2 points: a reference point on the neck of the implant and another on the bone–implant contact.

Figures 5 and 6.

Figure 5. Radiographic analysis of bone level changes after the insertion of implants with internal hexagon or conical connections. (a) Calibration (in mm) with the known implant length using Image J software. (b) Measurement of bone level changes on the mesial aspect of the implant, from the platform to the implant bone level.

Figure 6. Clinical aspects of prostheses fitted on implants with internal hexagon or conical connections at (a) 6 months, (b) 12 months, and (c) 36 months.

Figures 5 and 6.

Figure 5. Radiographic analysis of bone level changes after the insertion of implants with internal hexagon or conical connections. (a) Calibration (in mm) with the known implant length using Image J software. (b) Measurement of bone level changes on the mesial aspect of the implant, from the platform to the implant bone level.

Figure 6. Clinical aspects of prostheses fitted on implants with internal hexagon or conical connections at (a) 6 months, (b) 12 months, and (c) 36 months.

Close modal

Clinical preference regarding connection type was recorded at 6, 12, and 36 months after implant loading. Any complications, including implant/prosthesis failure, were recorded for both connection types.

Materials/Procedures

All patients received a prophylactic oral antibiotic 1 hour before surgery (2 g amoxicillin or 600 mg clindamycin in case of allergy to penicillin) and were instructed to rinse their oral cavities with a 0.2% chlorhexidine mouth rinse for 1 minute. All patients were treated under local anesthesia.

A gingival flap was elevated for implant insertion. The implant site was prepared using a short-drilling protocol,12  and all implants were inserted with an insertion torque of <50 N-cm. During implant site preparation, the operator also subjectively assessed bone quality. Once the implant site was prepared, the operator inserted the 2 implants as per the randomization sequence. Saline irrigation was applied during the entire site preparation and implant insertion procedure. At least 3 mm of bone was left between each implant. The torque was measured using an insertion ratchet for the last 3–4 rotations. After insertion, all implants embedded at the crestal bone level received cover screws provided by the manufacturer of each type of implant. All implants were submerged for the healing process.

After surgery, the patients were prescribed 200 mg Ketoprofen (1–3 times a day as needed) and an antibiotic (1 g amoxicillin twice a day for 5 days or, in case of penicillin allergy, 300 mg clindamycin twice a day for 5 days). Follow-up assessments were performed within 1 week, and sutures were removed after 10 days.

Three months after implant insertion, a second surgery was performed to uncover the implants and place healing screws. The implant insertion and uncovering procedures are depicted in Figure 2.

Four weeks after implant uncovering, impressions were recorded using the direct transfer technique, and prosthetic restorations were delivered 1 week later. All patients received single screw-retained implant-supported monolithic lithium disilicate (e.max Press, Ivoclar Vivadent, Schaan, Liechtenstein) crowns. Either IH (Ti Base Kit, Single Tooth Restoration, MIS) or IC (Ti Base Kit, Single Tooth Restoration, MIS) prefabricated titanium implant–abutment connections were selected. All single-crown restorations were extraorally cemented on the titanium abutments using an adhesive cement (Multilink Hybrid Abutment, Ivoclar Vivadent). All SCRs with IH or IC connections were fixed to the implants using a torque of 25 N-cm. The screw access holes were closed with ceramic inlays.13  A rubber dam was used during inlay cementation to minimize humidity from the oral cavity and peri-implant sulcus. The restorative procedures are depicted in Figure 3.

In this study, determination of a sample size was not attempted. Forty-nine patients were included, and each received 2 implants. Two Excel spreadsheets were generated: 1 for the IH connection and 1 for the IC connection. An independent statistician analyzed all data.

All statistical analyses were performed using OS X El Capitan version 10.11.3 software (Apple Inc). The significance of the differences observed was evaluated using 1-way analysis of variance. A P value of ≤.05 was considered statistically significant. All statistical calculations are described in the tables, and the results are presented as mean ± SD with minimum and maximum values. The rate of complication, including implant/prosthesis failure, was compared between groups using the Fisher exact test. The measurements taken at each follow-up visit (6, 12, and 36 months) were compared to the baseline measurements for evaluation of remodeling at the marginal bone level. The Scheffé method was used to analyze the mean radiographic values at 6, 12, and 36 months after loading, with baseline values as covariates.

In total, 80 patients were screened for eligibility; from these, 17 who were unable to commit to the 3-year follow-up schedule, 3 with bruxism, 8 who required bone augmentation because of insufficient bone <6 months before implant insertion, 1 who had received irradiation, and 2 with poor oral hygiene were excluded (n = 31). Finally, 49 patients (28 men, 21 women) were included. Each patient received 2 implants, yielding a total of 98 implants analyzed. The patient and intervention characteristics according to type of connection are presented in Table 1. Two of the 49 patients (a total of 4 implants) were excluded from the analysis due to not completing the 36-month follow-up (1 died during the first year in a car accident and the other could not be contacted after 12 months). The 47 remaining patients completed all follow-up evaluations, and their implants exhibited stable osseointegration.

Table 1

Patient characteristics and implant specifications

Patient characteristics and implant specifications
Patient characteristics and implant specifications

No implant presented with clinically or radiographically evident complications during the 3-year follow-up, and there was no statistically significant difference with regard to prosthesis failure between the IH and the IC connection groups (P = .32; difference = −0.08; 95% CI, −0.25–0.08). In addition, the 2 groups showed no statistically significant differences in terms of gingival bleeding or pathologies involving the peri-implant tissue (P = .29; difference = 0.12; 95% CI, −10–0.35).

With regard to the radiographic bone level changes, there was no significant difference between groups from baseline to the 3-year follow-up (P = .73; difference = −0.03; 95% CI, −0.25–0.17). The average RMBL in the IH and the IC connection groups was 0.42 and 0.35 mm, respectively, at 6 months (Table 2); 0.58 and 0.70 mm, respectively, at 1 year; and 0.68 and 0.64 mm, respectively, at 3 years (Table 2).

Table 2

Changes in the peri-implant marginal bone level for implants with internal hexagon or conical connections*

Changes in the peri-implant marginal bone level for implants with internal hexagon or conical connections*
Changes in the peri-implant marginal bone level for implants with internal hexagon or conical connections*

The Fisher least significant difference test confirmed that bone level changes did not differ significantly between the 2 groups at any time point (P > .5; Table 3). However, there was a significant difference between baseline and 36 months in the IH connection group (P < .05) and between baseline and 12 (P < .001) and 36 months (P < .01) in the IC connection group.

Table 3

Comparison of P values of bone level changes between internal hexagon connections and internal conical connections in implants (Fisher least significant difference test)

Comparison of P values of bone level changes between internal hexagon connections and internal conical connections in implants (Fisher least significant difference test)
Comparison of P values of bone level changes between internal hexagon connections and internal conical connections in implants (Fisher least significant difference test)

Several factors, including but not limited to abutment type, mucosa biotype, alveolar bone thickness, and oral hygiene, have been shown to influence the peri-implant marginal bone level. The results of this study showed that 3 years after loading, there were no statistically significant differences between IH and IC implant–abutment connections in terms of RMBL and complications (including implant/prosthesis failure). This is consistent with the outcome reported in a systematic review that compared conical and nonconical implant–abutment connections in vitro and in vivo.4  The authors of that review concluded that both abutment connections exhibited sufficient resistance to maximal bending forces and to fatigue loading.4 

For the analysis of marginal bone loss, most studies14,15  recommend mesial and distal measures obtained from a conventional periapical radiograph. From a statistical point of view, there are no differences between mesial and distal radiographic measurements. One in vitro study examined the accuracy of interproximal peri-implant bone defect detection with titanium (Ti) and zirconium dioxide (ZrO2) implants using digital periapical radiography and CBCT. The results of the above-mentioned study indicate that the assessment of interproximal peri-implant defect width in both Ti and ZrO2 implants was more accurate with periapical radiography compared to CBCT. CBCT measurements always led to an overestimation of the defect width, which reached clinical relevance in ZrO2 implants.16 

Meetal et al,17  in a study comparing CBCT and conventional periapical radiography, found that digital long-cone periapical radiographs are a reliable and valid method for the detection of circumferential peri-implant bone defects and that conventional periapical radiography performs significantly better than CBCT. Conversely, another study18  reported that CBCT and conventional periapical radiography produce similar results in the evaluation of radicular changes. However, it indicates the use of conventional periapical radiography because CBCT implies a higher radiation dose.

Periapical radiography is a reliable method19,20  and in our study it was also more accurate than CBCT-3D evaluation, although we did use a silicone holder for reproducibility on the first technique.

In the present study, RMBL was similar between the 2 connection types, even though the IC connection implants were designed to allow platform switching. Thus, there appeared to be no added benefit of platform switching, consistent with the findings of a study by Esposito et al,3  who found no statistically significant differences in clinical outcomes between internal and external connections. Our results are also consistent with those of another systematic review that found similarities between platform-switched and regular-platform implants placed in the molar region.21  In another study, however, conical abutments did show some superiority in terms of seal performance, microgap formation, torque maintenance, and abutment stability.22  Other in vivo studies (human and animal) have also indicated that conical and nonconical systems are comparable in terms of implant success and survival rates. However, some have argued that a conical implant–abutment connection provides better abutment fit, stability, and seal performance than a nonconical abutment connection.22 

Regarding bacterial microleakage from the human saliva to the implant–abutment connection, some studies have demonstrated that Morse taper connections and conical implant–abutment connections lead to significantly less bacterial species than IH and external hexagon connections.23,24  In the current study, the results regarding peri-implant soft tissue loss and bone loss indicated a slight bone loss around the IH implant–abutment connections.

In the present study, a delayed loading approach was applied to minimize any potential influencing variables and promote uneventful healing. Because no bone or tissue grafting procedures were performed, the conditions were uniform and permitted a fair evaluation of the influence of connection type.

Another factor that should be taken into consideration is the initial gingival tissue thickness at the crest level. Generally, thick bony plates are associated with thick tissue biotypes, while thin bony plates are associated with thin tissue biotypes that are more prone to fenestration and dehiscence after dental extractions and implant placement. In the present study, patients with thick mucosa (≥2 mm) generally exhibited a normal architecture and a stable contour at the follow-up examinations. During the 3-year follow-up period, the peri-implant soft tissues showed normal physiological characteristics (Figure 6).

Future studies should consider the use of optical coherence tomography (OCT) for the analysis of bone-implant contact. One article25  has reported that the implant-bone interface can be successfully investigated through a noninvasive method. Compared to conventional dental imaging, OCT is safer and can achieve the best depth resolution of all known methods. In a pilot study, Sanda et al26  demonstrated that OCT is an effective diagnostic imaging tool only if the mucosa covering the implant body is thinner than 1 mm. On the other hand, some reports indicate that OCT imaging of implants is useful up to mucosa thicknesses of 3 mm. In another study,27  OCT was used for the visualization and identification of peri-implant bone defects, and quantitative measurements of defect depth were obtained. It was concluded that OCT can be successfully used to determine peri-implant bone levels and to identify bone defects and that it produces quantitative noninvasive measurements of bone loss.

This study found no statistically significant differences in marginal bone remodeling or complications (including implant/prosthesis failure) between implants with IH connections and those with IC connections 3 years after loading. Thus, the type of internal implant–abutment connection did not influence clinical outcomes, including RMBL and implant/prosthesis failure. Future studies should address long-term prosthesis-related complications, such as screw loosening and fracture.

Abbreviations

Abbreviations
CBCT:

cone-beam computed tomography

IC:

internal conical

IH:

internal hexagon

OCT:

optical coherence tomography

RMBL:

radiographic marginal bone loss

SCR:

screw-retained restorations

This work was partly supported by a project grant (no. P III-C5-TC-2017-03-ERPOCSI). We would like to acknowledge the participation of Anghel Cernescu (Mechanics and Strength of Materials, Politehnica University, Timisoara, Romania) as an independent statistician who performed all statistical analyses.

The authors do not have any financial interests, either direct or indirect, in the products or information listed in the manuscript.

1. 
Geckili
 
O,
Bilhan
 
H,
Geckili
 
E,
Cilingir
 
A,
Mumcu
 
E,
Bural
 
C.
Evaluation of possible prognostic factors for the success, survival, and failure of dental implants
.
Implant Dent
.
2014
;
23
:
44
50
.
2. 
Sherif
 
S,
Susarla
 
HK,
Kapos
 
T,
Munoz
 
D,
Chang
 
BM,
Wright
 
RF.
A systematic review of screw- versus cement-retained implant-supported fixed restorations
.
J Prosthodont
.
2014
;
23
:
1
9
.
3. 
Esposito
 
M,
Maghaireh
 
H,
Pistilli
 
R,
et al
Dental implants with internal versus external connections: 5-year post-loading results from a pragmatic multicenter randomised controlled trial
.
Eur J Oral Implantol
.
2016
;
9
:
129
141
.
4. 
Meloni
 
SM,
Jovanovic
 
SA,
Lolli
 
FM,
et al
Platform switching vs regular platform implants: nine-month post-loading results from a randomised controlled trial
.
Eur J Oral Implantol
.
2014
;
7
:
257
265
.
5. 
Guerra
 
F,
Wagner
 
W,
Wiltfang
 
J,
et al
Platform switch versus platform match in the posterior mandible—1-year results of a multicentre randomized clinical trial
.
J Clin Periodontol
.
2014
;
41
:
521
529
.
6. 
Del Fabbro
 
M,
Bianchessi
 
C,
Del Lupo
 
R,
Landi
 
L,
Taschieri
 
S,
Corbella
 
S.
Platform switching vs standard implants in partially edentulous patients using the Dental Tech Implant System: clinical and radiological results from a prospective multicenter study
.
Clin Oral Invest
.
2015
;
19
:
2233
2244
.
7. 
Scarano
 
A,
Mortellaro
 
C,
Mavriqi
 
L,
Pecci
 
R,
Valbonetti
 
L.
Evaluation of microgap with three-dimensional X-ray microtomography: internal hexagon versus cone morse
.
J Craniofac Surg
.
2016
;
27
:
682
685
.
8. 
Borie
 
E,
Orsi
 
IA,
Noritomi
 
PY,
Kemmoku
 
DT.
Three-dimensional finite element analysis of the biomechanical behaviors of implants with different connections, lengths, and diameters placed in the maxillary anterior region
.
Int J Oral Maxillofac Implants
.
2016
;
31
:
101
110
.
9. 
Coppedê
 
AR,
Bersani
 
E,
de Mattos Mda
 
G,
Rodrigues
 
RC,
Sartori
 
IA,
Ribeiro
 
RF.
Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: an in vitro study
.
Int J Prosthodont
.
2009
;
22
:
283
286
.
10. 
Misch
 
CE.
Density of bone: effect on treatment plans, surgical approach, healing, and progressive bone loading
.
Int J Oral Implantol
.
1990
;
6
:
23
31
.
11. 
Silberman
 
SL,
Le Jeune
 
RC,
Serio
 
FG,
Devidas
 
M,
Davidson
 
L,
Vernon
 
K.
A method for determining patient oral care skills: the University of Mississippi Oral Hygiene Index
.
J Periodontol
.
1998
;
69
:
1176
1180
.
12. 
Bratu
 
E,
Mihali
 
S,
Shapira
 
L,
Bratu
 
DC,
Wang
 
HL.
Crestal bone remodeling around implants placed using a short drilling protocol
.
Int J Oral Maxillofac Implants
.
2015
;
30
:
435
440
.
13. 
Mihali
 
S,
Canjau
 
S,
Bratu
 
E,
Wang
 
HL.
Utilization of ceramic inlays for sealing implant prostheses screw access holes: a case-control study
.
Int J Oral Maxillofac Implants
.
2016
;
31
:
1142
1149
.
14. 
Misch
 
CE,
Perel
 
ML,
Wang
 
HL,
et al
Implant success, survival, and failure: the International Congress of Oral Implantologists (ICOI) Pisa Consensus Conference
.
Implant Dent
.
2008
;
17
:
5
15
.
15. 
Linkevicius
 
T,
Apse
 
P,
Grybauskas
 
S,
Puisys
 
A.
Influence of thin mucosal tissues on crestal bone stability around implants with platform switching: a 1-year pilot study
.
J Oral Maxillofac Surg
.
2010
;
68
:
2272
2277
.
16. 
Steiger-Ronay
 
V,
Krcmaric
 
Z,
Schmidlin
 
PR,
Sahrmann
 
P,
Wiedemeier
 
DB,
Benic
 
GI.
Assessment of peri-implant defects at titanium and zirconium dioxide implants by means of periapical radiographs and cone beam computed tomography: an in-vitro examination
.
Clin Oral Implants Res
.
2018
;
29
:
1195
1201
.
17. 
Dave
 
M,
Davies
 
J,
Wilson
 
R,
Palmer
 
R.
A comparison of cone beam computed tomography and conventional periapical radiography at detecting peri-implant bone defects
.
Oral Implants Res
.
2013
;
24
:
671
8
.
18. 
Takeshita
 
WM,
Chicarelli
 
M,
Iwaki
 
LCV.
Comparison of diagnostic accuracy of root perforation, external resorption and fractures using cone-beam computed tomography, panoramic radiography and conventional and digital periapical radiography
.
Indian J Dent Res
.
2015
;
26
:
619
626
.
19. 
Hilgenfeld
 
T,
Juerchott
 
A,
Deisenhofer
 
UK,
et al
Accuracy of cone-beam computed tomography, dental magnetic resonance imaging, and intraoral radiography for detecting peri-implant bone defects at single zirconia implants—an in vitro study
.
Clin Oral Implants Res
.
2018
;
29
:
922
930
.
20. 
Saberi
 
BV,
Khosravifard
 
N,
Ghandari
 
F,
Hadinezhad
 
A.
Detection of peri-implant bone defects using cone-beam computed tomography and digital periapical radiography with parallel and oblique projection
.
Imaging Sci Dent
.
2019
;
49
:
265
272
.
21. 
Enkling
 
N,
Jöhren
 
P,
Klimberg
 
V,
Bayer
 
S,
Mericske-Stern
 
R,
Jepsen
 
S.
Effect of platform switching on peri-implant bone levels: a randomized clinical trial
.
Clin Oral Implants Res
.
2011
;
22
:
1185
1192
.
22. 
Schmitt
 
CM,
Nogueira-Filho
 
G,
Tenenbaum
 
HC,
et al
Performance of conical abutment (Morse Taper) connection implants: a systematic review
.
J Biomed Mater Res A
.
2014
;
102
:
552
574
.
23. 
do Nascimento
 
C,
Miani
 
PK,
Pedrazzi
 
V,
et al
Leakage of saliva through the implant-abutment interface: in vitro evaluation of three different implant connections under unloaded and loaded conditions
.
Int J Oral Maxillofac Implants
.
2012
;
27
:
551
560
.
24. 
Gherlone
 
EF,
Capparé
 
P,
Pasciuta
 
R,
Grusovin
 
MG,
Mancini
 
N,
Burioni
 
R.
Evaluation of resistance against bacterial microleakage of a new conical implant-abutment connection versus conventional connections: an in vitro study
.
New Microbiol
.
2016
;
39
:
49
56
.
25. 
Sinescu
 
C,
Hughes
 
M,
Bradu
 
A,
et al
Implant bone interface investigated with a non-invasive method: optical coherence tomography
.
Proceedings of SPIE—The International Society for Optical Engineering
2008
;
6991
:
69911L
2
.
26. 
Sanda
 
M,
Shiota
 
M,
Imakita
 
C,
Sakuyama
 
A,
Kasugai
 
S,
Sumi
 
Y.
The effectiveness of optical coherence tomography for evaluating peri-implant tissue: a pilot study
.
Imaging Sci Dent
.
2016
;
46
:
173
178
.
27. 
Kim
 
S,
Kang
 
SR,
Park
 
HJ,
Kim
 
B,
Kim
 
TI,
Yi
 
WJ.
Quantitative measurement of peri-implant bone defects using optical coherence tomography
.
J Periodontal Implant Sci
.
2018
;
48
:
84
91
.