Clinical Relevance
Because increase in temperature is related directly to light intensity and exposure time, curing devices with high power density (>1200 mW/cm2) should only be activated for a short time (<15 seconds), even in teeth without cavity preparation.
SUMMARY
This laboratory study examined the effects of curing lights with different light intensities and changing flow rate on the increase in pulpal temperature during the light curing process and the rate of the subsequent decrease in temperature after the termination of light curing.
The tip of a temperature sensor was positioned on the pulpal dentinal wall of the buccal side of the maxillary premolar. Metal tubes were inserted in the palatal and buccal root of the tooth, one for water inflow and the other for water outflow. The tubes were connected to a pump to control the flow rate. The water flow rate was set to 4.2 μl/minute, 28 μl/minute or 70 μl/minute. At each flow rate, the unprepared tooth was light cured from the buccal side 1 mm from the buccal surface, using four different curing lights. The temperature data were recorded and stored on a computer every second for three minutes. The curing lights that were used were: Astralis 10 (QTHhigh, Ivoclar Vivadent), Bluephase 16i (LEDconv, Ivoclar Vivadent) and two experimental LED-curing lights (LEDexp2000, LEDexp3000, Ivoclar Vivadent). The power densities were 1200 mW/cm2, 1600 mW/cm2, 2000 mW/cm2 and 3000 mW/cm2, respectively. The curing lights, LEDconv, LEDexp2000 and LEDexp3000 were activated for 60 seconds, and the QTHhigh was activated for 30 seconds. The maximum intrapulpal temperature (TMAX) and rate of temperature change at 30 seconds after turning off the light (S30LO) were analyzed by two-way ANOVA with a post-hoc Tukey test (p<0.05). The influencing factors were the flow rates and curing lights.
Results: The TMAX ranged from 41.0°C to 53.5°C. There was a difference between the curing lights (p<0.05), with LEDexp3000> LEDexp2000> LEDconv> QTHhigh. There was no difference in TMAX between the different flow rates (p>0.05). Both the curing lights and flow rates affected the S30LO (p<0.05). The S30LO was LEDexp3000> LEDexp2000> LEDconv, QTHhigh (p<0.05). The S30LO at 70 μl/minutes was higher than at 4.2 μl/minutes and 28 μl/minutes (p<0.05).
Clinical Implication: Because the increase in temperature is directly related to the light intensity and exposure time, curing devices with high power density (>1200 mW/cm2) should only be activated for a short period of time (<15 seconds) even in teeth without cavity preparation. The flow rate had only a negligible effect on the temperature increase.
INTRODUCTION
Curing lights have become an indispensable tool in dentistry. Many dental materials are polymerized by them, including bonding materials, direct composite materials, luting materials for inlays and crowns, fissure sealants, bonding materials for orthodontic brackets and more. They are also used to bleach vital teeth, even though it has been reported that these devices do not enhance the bleaching outcome.1
Currently, LED lights have virtually replaced halogen lights and have a number of advantages, such as a longer service life without a decrease in performance, which is a problem frequently encountered with halogen lights, as well as a lower rate of battery discharge. The number of polymerization lights with high light intensities is increasing and units with a light output of 3000 mW/cm2 and higher are now available. However, dentists have become concerned with the possibility of these high-performance lights heating the pulp to such an extent that they could cause pulpal necrosis. The critical threshold value for pulpal necrosis in human teeth is not completely understood. Most studies refer to a study carried out more than 40 years ago. In that trial, the teeth in five Rhesus monkeys were heated using a soldering gun for 5 to 20 seconds; the soldering gun exerted a temperature of 275°C (± 50°C). The results showed that, with a temperature increase of 5.5°C and 11°C, 15% and 60% of the pulpal tissues became necrotic after three months.2 However, it is highly questionable as to whether the values obtained in monkeys are also valid for human beings. A clinical study in volunteers indicated that even a temperature rise of 9°C–15°C did not cause histologically verified pulpal necrosis after three months.3 In that study, heat was applied to the occulsal surface of six premolars and six molars, with individually fitted supports, until the subject complained of a toothache. The contralateral tooth was extracted and the increase in temperature in the pulp was measured using the same parameters as those under in vivo conditions. After three months, the other teeth were extracted and examined histologically. The results indicated that the pulpal tissues could tolerate a temperature rise > 5.5°C without damage. The temperature rise was only for a short-term duration.
There are no reports of pulpal necrosis caused by high-intensity curing lights, even though these devices have been widely used formany years. However, pulpal necrosis due to heat trauma does not involve bacteria and may only be noticed many months or years after the heat exposure has occurred, at which point the dentist would probably relate the pulpal necrosis to other factors, such as possible secondary caries, preparation trauma, fabrication of self-curing tooth temporaries and testing of a pulp reaction using hot gutta purcha. The results of the laboratory tests are difficult to interpret, because the ultimate threshold value for a temperature rise for pulpal necrosis to occur is unknown.
Although the issue of increasing temperature in relation to curing lights has been studied extensively in the laboratory using different models,4–14 the factors influencing the outcome variable have not been investigated systematically. An increase in pulpal temperature can be affected by dentin thickness,8,11,14 duration of light exposure and the type of light-curing device used in the curing process.4,9,12
The pulpal blood flow rate changes in many clinical situations, such as local anaesthesia,15–17 trauma to the tooth,18–21 orthodontic movement22–24 and age [18]. Many composite restorations are placed under these circumstances. It is well known that the blood flow rate affects the thermal response of living tissue. Heat exchange between the living tissue and blood vessels depends on the geometry of the surrounding tissue and the blood flow rate.25 The difference in pulpal blood flow rate can affect the increase in pulpal temperature during the light curing process. The pulp has the highest blood flow per unit weight compared to other oral tissues. Furthermore, capillary blood flow in the coronal portion of pulp is almost double that found in the radicular portion.26 The abundant blood flow in pulpmay also help to disperse heat and help to prevent pulpal damage due to heating.
The current laboratory study examined the effect of curing lights with different light intensities and the changing flow rate on the increase in pulpal tempera ture during the light curing process and the rate of temperature decrease after the termination of light curing.
The null hypotheses in this study were as follows:
There is no difference in intrapulpal peak temperature during the light curing process between curing lights with different power densities.
There is no difference in intrapulpal peak temperature in pulp during the light curing process between groups with different flow rates.
There is no difference in the rate of temperature decrease in pulp after turning off the device between curing lights with different power densities.
There is no difference in the rate of temperature decrease in pulp after turning off the device between groups with different pulpal flow rates.
METHODS AND MATERIALS
The experimental setup used in the current study is similar to one proposed by Daronch and others.5
A. Preparation of Tooth
A maxillary premolar with two separate roots, without caries and/or restorations, was used. No cavity preparation or filling procedures were carried out on the buccal surface, which was exposed to the light curing units. The roots were cut by one-third of their length to expose the canal spaces for metal tube insertion. After checking whether the root canals were free of debris, two metal tubes (diameter 2 mm) were inserted into the apices of both roots about 2 mm and fixed into position using Syntac Classic (Ivoclar Vivadent, Schaan, Liechtenstein) and Heliobond (Ivoclar Vivadent) (Figure 1). The two polyethylene tubes, one for water outflow and one for water inflow, were connected to the metal tubes.
B. Insertion of Thermocouples
On the palatal side of the premolar, a hole (diameter 2 mm) was drilled into the pulp chamber with a cylindrical diamond bur (FG 8614, Intensiv, Grancia, Switzerland). After beveling the orifice, the enamel surrounding the hole was etched with Total Etch (Ivoclar Vivadent) for 30 seconds, then rinsed with water. Heliobond was applied to the etched enamel, gently air-dried, then light-cured for 10 seconds (Astralis 10, 650 mW/cm2). A K-type thermocouple (CHAL-003; OMEGA Engineering, Inc, Stamford, CT, USA) was positioned in the hole, while attention was paid to bringing the tip into contact with the dentin that was opposed to the buccal cavity. Furthermore, care was taken not to insert the thermocouple with too much force in order to avoid bending it. The thermocouple was fixed in position using Variolink II Base material and light-cured for 30 seconds (Astralis 10, 650 mW/cm2). A radiograph was taken to confirm the position of the thermocouple (Figure 2). Other thermocouples for water and air were placed in the water bath and air. All three thermocouples were connected to a computer via a data logger (Agilent 34970A, Agilent Technologies Santa Clara, CA, USA). Software (Agilent BenchLink DataLogger, version 1.4) was used to measure the temperatures at a frequency of 1 Hz.
C. Connection to Water Cycle and Controlling Flow Rates
The polyethylene tube for the water outlet was connected to the pump, and the one for water inflow was placed in the water bath with deionized water. To mimic blood flow in the tooth, the tube for water outflow was connected to the pump. Negative pressure from the pump induced water outflow from the pulp space through the polyethylene tube used for water outflow. At the same time, it also caused water inflow from the water bath into the pulp space through another tube used for water inflow. The flow rate was controlled by a regulator in the pump, and regulator levels 1, 2 and 4 were used. At each level, the amount of water drained from the tooth was collected over a one-minute period and measured. The flow rates in levels 1, 2 and 4, were 4.2 μl/minute(A), 28 μl/minute(B) and 70 μl/minute(C), respectively.
D. Taking the Curing Light's Spectral Radiant Emittance Plot and Power Density
The integration power and its spectral radiant emittance plot of the curing light were measured using an integration sphere and its software (Gigahertz-Optic GmbH, Puchheim, Germany). The curing lights were Astralis 10 (QTHhigh, Ivoclar Vivadent), Bluephase 16i (LEDconv, Ivoclar Vivadent) and two experimental LED curing lights (LEDexp2000, LEDexp3000, Ivoclar Vivadent). AS10 was a halogen type light and the other three devices were LED curing lights. The diameter of the light-curing tip was 8mmin all groups. For LEDexp3000, a cooling system that was specially designed for this unit was installed, which blew a constant stream of cool air in the direction of the light-curing tip. The power density of each curing light was calculated by dividing the integration powers of each curing light by fiber bundle area of each curing light.
E. Measurement of Pulpal Temperature in Relation to the Different Curing Lights and Flow Rates
I. Maximum Intrapulpal Temperature (TMAX)
When the pulp temperature was stabilized between 33°C and 34°C, the tooth was exposed to the four curing lights at each flow rate:
For each curing light, the distance from the light tip to the tooth was set to 1 mm using a metal spacer. The LEDconv, LEDexp2000 and LEDexp3000 curing lights were activated for 60 seconds but QTHhigh was activated for 30 seconds. Subsequently, the decrease in temperature of LEDconv, LEDexp2000 and LEDexp3000 was monitored for an additional 2 minutes and 2.5 minutes for QTHhigh. The temperature data of the pulp, water and air were stored in a computer every second from the start of the light-curing procedure. Five measurements were taken in each group. The highest temperature of the pulp (TMAX) was registered and used for a statistical comparison (Figure 3). The TMAX were compared between groups using two-way ANOVA with a post hoc Tukey test with two fixed variables (flow rate, curing lights) at the 95% confidence level.
II. Calculation of the Rate of Temperature Change at 30 Seconds After Turning Off the Light (S30LO)
In the time-temperature diagram, which shows the results of experiment I (TMAX study), the rate of the temperature change at 30 seconds after turning off the light (S30LO) was calculated. The rate for LEDconv, LEDexp2000 and LEDexp3000 was calculated at 90 seconds and 60 seconds for QTHhigh.
The rate of the temperature change (S) at time x and the temperature Tx were defined using the following equation:

The S30LO were compared between groups using two-way ANOVA with a post hoc Tukey test with two fixed variables (flow rate, curing lights) at the 95% confidence level.
RESULTS
Table 1 summarizes the means and standard deviations of TMAX. The temperature increased to a range between 41.0 °C and 53.5°C, depending on the curing light used. Table 2 lists the results of two-way ANOVA on the effects of the curing light and flow rate on the maximum intrapulpal temperature and their interaction effect. The curing light had a significant statistical effect on the TMAX (p<0.05), whereas the flow rate did not (p>0.05). There was no interaction between the curing light and flow rate (p>0.05). The results of a post hoc multi-comparison Tukey test indicated the following ranking of TMAX: EX2> EX1>BP>AS10 (p<0.05).
Maximum Intrapulpal Temperature (Tmax) Reached During Different Applied Curing Lights and Flow Rates (mean[sd])
![Maximum Intrapulpal Temperature (Tmax) Reached During Different Applied Curing Lights and Flow Rates (mean[sd])](https://allen.silverchair-cdn.com/allen/content_public/journal/operative-dentistry/35/3/10.2341_09-234-l/2/m_i1559-2863-35-3-353-t01.gif?Expires=1683231496&Signature=W~0zlibDFwpBWiNnYy6okDcqMf~VEqc3KB66aCwDmGV44n6o7nnK5RpexD~DOfdqnHhfpMIBac1l53VdUzmOewAQr8QP0HAuC0s4lSqnTy4bp2jje93B7g0MiTXM7vxZlxwtj7Jq0-JYcksIpmEptSapHE3wKWbq6pSmdonxju7Ag4NUvFmvOkOFmfBwlEo2E5~8w9hAF2OdWwOymQ5QVGnkENHixixhDEqV0~A7fIY07AVpBcpjzi8ALiGxZxRiUhK~rLZgpgWVH1BHRt9S~UagAooqv5ICyF7Adz4EPrG2H4yu3a4Q4coTovEY14rLO~Kbz6TqpmLz-bJFkrl5bQ__&Key-Pair-Id=APKAIE5G5CRDK6RD3PGA)
Table 3 shows the means and standard deviations of S30LO. S30LO ranged from 0.11 (A) to 0.27. Table 4 summarizes the results of the two-way ANOVA regarding the effects of the curing light and flow rate on S30LO, along with its correlation. Both the curing light and flow rate had a significant statistical effect on the S30LO (p<0.05). However, Figure 5 suggests that the interaction was minimal, because all lines ran practically parallel. The results of the Tukey test for the curing light factor indicated that the S30LO was LEDexp3000> LEDexp2000> LEDconv, QTHhigh (p<0.05). The results of the Tukey test for the flow rate indicated that the S30LO at 70 μl/minute was higher than that at 4.2 μl/minute and 28 μl/minute (p<0.05).
Temperature Decline Rate (°C/seconds) at 30 Seconds After Light Termination [S30LO] (mean[sd])
![Temperature Decline Rate (°C/seconds) at 30 Seconds After Light Termination [S30LO] (mean[sd])](https://allen.silverchair-cdn.com/allen/content_public/journal/operative-dentistry/35/3/10.2341_09-234-l/2/m_i1559-2863-35-3-353-t03.gif?Expires=1683231496&Signature=XcfOg7CMpOLrGbPiOCE122o9VWpALC-~Xa3jJwom17odTvY6uhqO4Y9M~ZesFoPOnsBkrwdoQTc3QSdHshyFL4-l802KL5ueHKPkIRQETizMYmOE3Up7qjafv3LchMNPiCA5QFuQ-nJLHfMJd69EfilYeUDNaj5XkzRzQIdIXBWMlzDAinj9FNFTu2avDufrFLeJovXrvRvLCSgj91sbQyDKzKtz2Xi4BCYjE8NiM-F6MoKuvh-Ndvt1zLHADNYN6JrpVz4oNEvsKTbJTbh9twQ2R1AOuKtWToyVhxm95b-qDqQ~BJa0KLRk0h~LKWIjwAzt6bYBfFEJ5NlbVLEUvA__&Key-Pair-Id=APKAIE5G5CRDK6RD3PGA)
DISCUSSION
In the current study, the flow rate did not affect the highest temperature in pulp (TMAX) but influenced the decrease in temperature once the device was switched off. The TMAX ranged from 41°C to 53.7°C when it was monitored under a controlled flow rate and curing light exposure. As the initial temperature was approximately 34°C, the amount of increase ranged from approximately 7°C to 19.7°C (Figures 6 and 7; Table 1). In a pilot study, the temperature in pulp was monitored, while changing only the flow rate with the curing light in the turned-off state. The temperature increased with increasing flow rates but the increase was very limited. Considering the results of the current study, the effect of the flow rate appeared to be masked by TMAX, because the temperature changes induced by the curing light were much greater than those induced by the flow rate. If the in vitro results are applied to in vivo conditions, it may be concluded that the flow rate in blood vessels has a negligible effect on the rapid increase in temperature in pulp during the critical phase when the curing light is activated.
When the curing light was turned off, the decrease in pulp temperature was more pronounced in the test group with higher flow rates. It appears that a higher flow rate removed heat more quickly, which affects the S30LO.
The influence of the flow rates on TMAX was quite limited. This phenomenon may be explained by the Womersley number, which is a dimensionless number in biofluid mechanics and describes the heat transfer between a vessel and tissue27: α = R (w/v)1/2, where α is the Womersley number, R is the vessel radius, W is the radial frequency (rad/s) and V is the kinematic viscosity (m2/s). From this equation, the change in flow rate due to pulsating blood, which affects the radial frequency, has a very limited or negligible effect on the heat transfer in small vessels. However, it would be inappropriate to apply this equation directly to the current study, because the canal space and pulp chamber geometry were not circular or consistent. Nevertheless, this equation can explain why the flow rate had such little influence on pulp temperature.
The TMAX data in this study was consistent with that reported in other studies, showing a large increase in temperature with light curing units that have a high power density.3,8,10,12 Considering the initial temperature in pulp (approximately 34°C), the increase in pulp temperature after exposure to QTHhigh (30 seconds), LEDconv (60 seconds), LEDexp2000 (60 seconds) and LEDexp3000 (60 seconds) was 7°C, 11°C, 17°C and 19°C, respectively. According to Zach and Cohen, a 5°C increase in temperature could cause histological changes to the pulp. In the current study, the intrapulpal temperature increased by more than 5°C when the exposure time was extended by more than 20 seconds in QTHhigh and LEDconv, and more than 10 seconds in LEDexp2000 and LEDexp3000 (Figure 6). Although the critical threshold value for pulpal necrosis in human teeth is not completely understood, it would be advisable to limit the exposure time to 20 seconds for a curing device whose power density is between 1200 and 1600mWcm2, and to 10 seconds for a curing device whose power density is between 2000 and 3000 mW/cm2.
Figures 6 and 7 show that the temperature decreased effectively after the curing light was switched off. This effective cooling in the tooth may help to limit the temperature increase in pulp in multiple light curing procedures in clinical situations.
The baseline temperature value of 34°C was reached in QTHhigh only 2.5 minutes after the curing light was turned off. Therefore, the pulpal temperature would be well above 34°C when the second or third exposure to light occurs. Whether multiple exposures lead to a greater increase in pulpal temperature is unknown and should be investigated further.
It should be noted that cavity preparation was not performed prior to light exposure in this study. In some clinical procedures, such as the placement of orthodontic brackets with light-curing materials and the acceleration of bleaching products, the cavity is not prepared. A future study will examine whether a reduced thickness of the remaining hard tissue leads to a greater increase in pulpal temperature.
Another variable is the distance of the light tip to the tooth, which will also be evaluated in another study. Operational measures that may be helpful in reducing the temperature increase include the use of base materials,4 modulation of the light intensity,7 as well as curing tip design and diameter.8
A comparison of QTHhigh and LEDconv in Figure 7 showed that the pattern of the temperature increase was quite similar—from two seconds to 30 seconds—despite the difference in power density of approximately 313mW/cm2 between these two curing lights. This might be because QTHhigh produced a broader spectrum and a slightly greater red shift than LEDconv (Figure 4). Therefore, LEDconv might produce comparatively less heat than QTHhigh, which would eventually result in a similar increase in temperature.10,12
Curing light's spectral radiant mission plot. QTHhigh (Figure 4a), LEDconv (Figure 4b), LEDexp2000 (Figure 4c), LEDexp3000 (Figure 4d).
Curing light's spectral radiant mission plot. QTHhigh (Figure 4a), LEDconv (Figure 4b), LEDexp2000 (Figure 4c), LEDexp3000 (Figure 4d).
Profile plot for S30LO in relation to the curing lights and flow rates. A, B and C represent the flow rates, 4.2 μl/minute, 28 μl/minute and 70 μl/minute, respectively.
Profile plot for S30LO in relation to the curing lights and flow rates. A, B and C represent the flow rates, 4.2 μl/minute, 28 μl/minute and 70 μl/minute, respectively.
Temperature change (°C) in the pulp according to the curing lights used, QTHhigh (6a), LEDconv (6b), LEDexp2000 (6c), LEDexp3000 (6d) and different flow rates A, B and C represent the flow rates, 4.2 μl/minute, 28 μl/minute and 70 μl/minute, respectively.
Temperature change (°C) in the pulp according to the curing lights used, QTHhigh (6a), LEDconv (6b), LEDexp2000 (6c), LEDexp3000 (6d) and different flow rates A, B and C represent the flow rates, 4.2 μl/minute, 28 μl/minute and 70 μl/minute, respectively.
Temperature change (°C) in the pulp according to the flow rates: 4.2 μl/minute (7a), 28 μl/minute (7b) or 70 μl/minute (Figure 7c) and different curing lights.
Temperature change (°C) in the pulp according to the flow rates: 4.2 μl/minute (7a), 28 μl/minute (7b) or 70 μl/minute (Figure 7c) and different curing lights.
Although TMAX was higher in LEDexp3000 than LEDexp2000, the difference was not great, considering that the difference in power density between the two was approximately 1000mW/cm2 (Figure 7, Table 1). The S30LO wasmuch higher in LEDexp3000 than LEDexp2000 (Figure 7, Table 3). This appears to be related to the cooling system in LEDexp3000. This cooling system is effective in reducing the TMAX and increasing the cooling speed, because it is also activated after the light was switched off.
The flow rates (4.2 μl/minute, 28 μl/minute or 70 μl/minute) set in the current study did not represent a real clinical situation, because the real pulpal blood flow rates in humans is unknown. Kim and others17 reported that the pulpal flow in dogs was 33.32 ml/minute in 100g of pulp tissue and was reduced to 7.72 ml/minute after local infiltration anesthesia using 1:100,000 epinephrine with 2% lidocaine. Lustig and others28 reported in their ultrasound/doppler imaging study that the blood flow rate of the sublingual artery that penetrates the bone through the lingual foramen at the midline of the mandible was 0.7–3.7ml/minute. Pitt Ford and others29 reported a significant decrease (31%) in pulpal blood flow after an injection of 1 ml of 2% lidocaine with 1:80,000 adrenaline.
The interaction between curing lights and flow rates was statistically significant (Table 4). The medium flow rate B seems to be responsible for it, as the S30LO of LEDconv was lower for the medium flow rate than for the higher flow rate and the S30LO of QTHhigh was higher for the medium than for the higher flow rate. The reason for that is unclear.
As only one upper premolar was used in this study, the data would vary according to the tooth size and shape. A future study should examine the effect of these factors.
This is the first report in a series of studies that evaluated the effects of the following factors: 1) flow rate, 2) thickness of the overlaying dental hard tissue, 3) distance from the light source to the tooth and 4) presence of composite filling material. All these factors may affect the increase in temperature in pulp during the curing process.
Of the four null hypotheses, the second held and the other three were rejected.
CONCLUSIONS
Within the limits of this study, in which one maxillary premolar was used, the following conclusions were made.
There was a difference in intrapulpal peak temperature during the light-curing process between curing lights with different power densities.
There was no difference in intrapulpal peak temperature in pulp during the light-curing process between groups with different flow rates.
There was a difference in the rate of temperature decrease in pulp after switching off the device between the curing lights with different power densities.
There was a difference in the rate of temperature decrease in pulp after switching off the device between groups with different pulpal flow rates.
Acknowledgments
The authors wish to thank Bruno Senn and Wolfgang Plank from Ivoclar Vivadent for their technical help and support while carrying out the measurements. This work was supported by the Korea Science and Engineering Foundation (KOSEF) grant funded by the Korean government (MEST) (No R13-2003-013-05002-0).
References
Author notes
Sung-Ho Park, professor, Department of Conservative Dentistry, College of Dentistry, Yonsei University, Seoul, Korea
Jean-François Roulet, professor, director, Clinical R&D, Ivoclar Vivadent, Schaan, Liechtenstein
Siegward D Heintze, Dr med dent, PhD, head, In Vitro Laboratory, Research & Development, Ivoclar Vivadent AG, Schaan, Liechtenstein